Gas Chromatography
•General Design of a Gas Chromatograph
•Separation Processes in Gas Chromatography
•GC Columns
•GC Injectors
•GC Detectors
General Design of a GC
Some of the design“details”
• Gas supplies usually have either in-line or instrument
mounted traps to remove any water, oxygen,
hydrocarbons or other “contaminants” from compressed
gases
• Gas flows can be controlled using either needle valves or
mass-flow controllers (electronic sensors)
• Instruments can have multiple injectors, detectors or
columns
• Injectors and detectors usually have their own
temperature controlled zones (small heaters)
• The GC oven has a large fan and a vent door to help with
rapid cooling of the oven
• Data collection (and integration) can be done using a chart
recorder, integrator or a computerized data system
Separation Processes in GC
• Gas Chromatography as it is usually performed is correctly called gas-
liquid chromatography
• the analyte is in the gas phase in the GC and partitions between the mobile
phase (carrier gas) and the liquid stationary phase that is coated on the
inside of an open-tubular capillary column or on particles inside a packed
column
• Some packed-column GC uses non-coated solid stationary phases, in
which case one is performing gas-solid adsorption chromatography
• Capillary, open-tubular (WCOT specifically) column GC is the primary
type of GC used in quantitative analysis:
• higher resolution = greater ability to discriminate between components
• smaller capacity of the column is not important as long as sufficient analyte is
available for detection
• pg/mL (ppt) to g/mL (ppm) concentration range for liquid analytes
The Objective in
Chromatography (all types)
• Separate your analytes (resolution of 1.5 or better) in
the shortest amount of time possible and detect
them….
• How can we do this in GC?
• Use different columns for different analyte types
• stationary phase
• diameter of column, stationary phase thickness
• column length
• Use different injection types/temperatures to optimize the
process of loading the sample on the column
• Use different temperature (or pressure) programs for the
column
• Select and use a detector that is suitable for the analyte(s) of
interest
GC Columns (concentrating on open-tubular capillary
columns)
• Column “frame” constructed of fused silica tubing
• Polyamide coating on the outside gives it strength
• Liquid stationary phases coated or bonded to the inside of the
tubing
• 0.1 - 0.53 mm + ID, 5-100 meters in length, stationary phases
usually 0.10 to 1.5 m in thickness
• Mounted on a wire cage to make them easier to handle
• 5-150 meters long.
Capillary
Column
Stationary
Phases
Choosing a GC Column…
• Is the column compatible with your analytes
• polar analytes require polar stationary phases so they will
spend some of their “time” in the stationary phase
• non-polar analytes require non-polar stationary phases
• You usually have to compromise on the stationary phase to
get a good column for your analytes (which are probably a
mix of polar and non-polar)
• DB-5, HP-5, EC-5, RTX-5 (5% dimethyl, 95% diphenyl
polysiloxane) most common general use column.
• Temperature range, solvent and carrier gas
compatibility
• Sample capacity versus resolution
• usually determines packed vs.. capillary
• GC’s usually setup for either packed or capillary
• Let’s say you choose a capillary column, there’s more
to think about!
For capillary GC columns….
 Increased length = greater N, therefore a greater R
 expense is possible band broadening if analytes are on the
column too long!
 Increased length leads to longer separations. Do you have
the time?
 Increased stationary phase thickness and column
diameter provides increased sample capacity and
can provide increased resolution
 tradeoffs are a longer analysis time and more column bleed
with thicker stationary phases
 Is the column compatible with the detector?
 Thick stationary phases bleed more and will contaminate a
mass spectrometer.
 For most analytical work, a best “compromise”
column is chosen and other variables (temp, etc.) are
altered to optimize the separation.
Capillary vs. Packed Columns
 Capillary Columns:
 Higher resolution (R)
 Greater HETP and N
 Shorter analysis time
 Greater sensitivity
 Most common in analytical
laboratory GC instruments
 Smaller sample capacity
 Higher cost/column
 Columns more susceptible
to damage
 Packed Columns
 Greater sample capacity
 Lower cost (can make your own)
 More rugged
 Most common in process labs or
separating/determining major
components in a sample (prep
GC)
 Limited lengths reduces R and N
 Not compatible with some GC
detectors
Temperature Programming in GC
 The “simplest” way to alter the separation in GC is to
alter the temperature program in the oven. You can
also alter the pressure of the carrier gas, but this is
less common (much).
 Isothermal = constant temperature
 Gradient = varied temperature
 By altering the temperature, you vary the rate of the
reaction for any analyte:
 they spend more or less time in the stationary phase
 the greater the difference in the times between analytes, the
better the separation!
Analyte
Analyte phase
stationary
phase
mobile 
The traps of temperature…
If your temperature at a given time is too high, you
can cause the peaks to co-elute
poor resolution vs but a faster separation
If your temperature at a given time is too low, you
can get still get a good separation
adequate resolution, but a separation that takes very
long
You have to choose a compromise temperature
program
GC Carrier Gases (the mobile phase)
 Usually “inert” gases (don’t react with analytes except
sometimes in the detector)
 Purpose:
 sweep sample through the column
 protect column from oxygen exposure at temperature
 assist with function of the detector
 Most common:
 Helium (available relatively pure without extensive purification after
it leaves a compressed gas cylinder)
 Nitrogen (usually requires an oxygen and water trap)
 Hydrogen
normally used only with flame ionization detectors (FID) since the FID
needs it as fuel for the flame
still rarely used due to safety concerns (and chromatographic ones)
GC Injection….
 Samples are injected through a septum:
 keeps oxygen out of the column
 provides a seal to keep the carrier gas pressure up at the
head of the column
carrier gas flow rate is determined by the pressure or the gas at
the opening of the column
 Many different (mostly proprietary) materials
red rubber (bleeds at about 250 C)
Thermogreen (good up to about 300 C)
High-temperature blue (good a little over 300 C)
 The injector is usually lined with a de-activated glass
liner
 prevents metal injector-sample reactions that would alter
analytes or damage the metal of the injector
 Can be cleaned/replaced regularly
Injection types
 On-Column Injection:
 used widely in packed-column GC, less in capillary GC
 sample is deposited directly on the column
Good for thermally unstable compounds
Good for quantitative analysis at low concentrations
 all sample is available to travel to the detector
BUT, you can inject only a relatively small amount of sample in
capillary GC anyhow.
 Splitless Injection:
 Sample is vaporized in the injector itself and ALL of the
sample is swept onto the column by the carrier gas
 Again, relatively small samples are injected (10  L or less in
capillary GC)
 Sample spends a large amount of time in the injector
 Best for trace (1 -100 ppm range) concentrations of high
boiling point analytes in low boiling point solvents
extra time in the injector helps volatilize the analytes.
Split Injection:
the injection is split, with only a portion of the
sample (usually 1% - 20%) actually making it to
the column
the most common method of injecting samples
onto small diameter, open-tubular columns.
Even if you inject 20 L, only a fraction (adjustable)
makes it on to the column
Not good for analytes with a wide range of
boiling points
some may be swept out the split vent before they are
volatilized
Modern capillary GCs come with a
Split/Splitless injectors standard
you switch between modes by changing the split
vent gas flow and using a different injection liner.
Don’t Forget SPME (Solid Phase Microextraction)
GC Detectors
 A dozen or more varieties (some obscure)
 Must be:
 sensitive to the analytes of interest
 compatible with the column, carrier gas, solvent, etc.
 rugged enough to withstand general unattended used
I’ve run our new GC for 36 hours straight without touching it!
 Should have a known linear range
 if the detector response is very linear, you can use a response
factor instead of a calibration curve for quantitation!
 Usually require separate gas supplies (other than the
carrier gas), have their own temperature control.
 Measure nothing more than a voltage or a current.
FID
FPD
Thermal Conductivity
(TCD)
 The carrier gas has a known
thermal conductivity.
 As the thermal conductivity of
the column eluent (gas flow in)
changes, the resistance of the
filament changes.
 The presence of analyte
molecules in the carrier gas
alter the thermal conductivity of
the gas (usually He)
 There is normally a second
filament to act as a reference
(the carrier gas is split)
 Increased sensitivity with
decreasing temperature
(detector), flow rate and applied
current.
 Filaments will burn out
(oxidized) in the presence of
oxygen if hot!
Non-destructive
FID
 Destructive, sample lost.
 Analytes containing C burn
in a hydrogen-oxygen flame
and produce ions
 CHO+ ions are collected on a
cathode and the current they
produce results in the signal
 WILL NOT detect non-C
containing compounds!
 Requires H2 supply (tank or
generator) and O2 supply
(compressed air)
 H2 carrier gas can be used,
eliminating the need for a
supply for the detector
 A makeup gas can also be
required!
-
Flame
O
,
H
e
CHO
O
CH
2
2


 
ECD
 Particularly sensitive to halogens nitriles,
carbonyls, nitro compounds
 Analytes pass through a cell, in which
electrons are traveling between a 63Ni
electrode and a collector electrode
 As analytes with “electron capturing
ability” pass through the cell, the flow of
electrons is interrupted.
 The change in current, due to reduced
flow of electrons, is recorded.
 EXTREMELY SENSITIVE TO
HALOGENS
 could ruin detector with 1 ppm
hexachlorocyclohexane by contaminating it
with excess analyte
 Widely used for the determination of
pesticides, herbicides and PCBs in
environmental samples.
 Non-destructive

Weitere ähnliche Inhalte

Ähnlich wie GC.ppt

CHROMATOGRAPHY learn chemistry enjoy this
CHROMATOGRAPHY learn chemistry enjoy thisCHROMATOGRAPHY learn chemistry enjoy this
CHROMATOGRAPHY learn chemistry enjoy thisgiridharrao67
 
Chromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.pptChromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.pptMajdolenAhrki
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentationArgha Sen
 
Cromatografía de gases principios y fundamentos
Cromatografía de gases principios y fundamentosCromatografía de gases principios y fundamentos
Cromatografía de gases principios y fundamentosAngela Mora
 
lecture-8-gc c.pptx
lecture-8-gc c.pptxlecture-8-gc c.pptx
lecture-8-gc c.pptxSahilSwe
 
GC and GC-MS slide.ppt
GC and GC-MS slide.pptGC and GC-MS slide.ppt
GC and GC-MS slide.pptAhnaf maznun
 
Ph.D Course_Gas chromatography.ppt
Ph.D Course_Gas chromatography.pptPh.D Course_Gas chromatography.ppt
Ph.D Course_Gas chromatography.pptKartikThakkar24
 
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...DipeshGamare
 
Gas chromatography instrumentation
Gas chromatography instrumentationGas chromatography instrumentation
Gas chromatography instrumentationMaryamAfzal19
 
GC PPT...pptx
GC PPT...pptxGC PPT...pptx
GC PPT...pptxDipu77
 
Gas chromatography by Dr.Rashmi Agarwal
Gas chromatography by Dr.Rashmi AgarwalGas chromatography by Dr.Rashmi Agarwal
Gas chromatography by Dr.Rashmi AgarwalRashmiAgarwal60
 

Ähnlich wie GC.ppt (20)

CHROMATOGRAPHY learn chemistry enjoy this
CHROMATOGRAPHY learn chemistry enjoy thisCHROMATOGRAPHY learn chemistry enjoy this
CHROMATOGRAPHY learn chemistry enjoy this
 
Chromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.pptChromatography الشيت الثالث.ppt
Chromatography الشيت الثالث.ppt
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 
gas chromatography
gas chromatographygas chromatography
gas chromatography
 
Rajeshri patil
Rajeshri patilRajeshri patil
Rajeshri patil
 
Cromatografía de gases principios y fundamentos
Cromatografía de gases principios y fundamentosCromatografía de gases principios y fundamentos
Cromatografía de gases principios y fundamentos
 
lecture-8-gc c.pptx
lecture-8-gc c.pptxlecture-8-gc c.pptx
lecture-8-gc c.pptx
 
GC and GC-MS slide.ppt
GC and GC-MS slide.pptGC and GC-MS slide.ppt
GC and GC-MS slide.ppt
 
Ph.D Course_Gas chromatography.ppt
Ph.D Course_Gas chromatography.pptPh.D Course_Gas chromatography.ppt
Ph.D Course_Gas chromatography.ppt
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
GAS CHROMATOGRAPHY-Principle, Instrumentation Advantage and disadvantage appl...
 
Gas Chromatography.pptx
Gas Chromatography.pptxGas Chromatography.pptx
Gas Chromatography.pptx
 
Gas chromatography instrumentation
Gas chromatography instrumentationGas chromatography instrumentation
Gas chromatography instrumentation
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
GC PPT...pptx
GC PPT...pptxGC PPT...pptx
GC PPT...pptx
 
GC .pdf
GC .pdfGC .pdf
GC .pdf
 
GC Power Point.pptx
GC Power Point.pptxGC Power Point.pptx
GC Power Point.pptx
 
Gas chromatography by Dr.Rashmi Agarwal
Gas chromatography by Dr.Rashmi AgarwalGas chromatography by Dr.Rashmi Agarwal
Gas chromatography by Dr.Rashmi Agarwal
 

Mehr von Priyanka Yadav

pharma quiz for pharmaceutical field.pptx
pharma quiz for pharmaceutical field.pptxpharma quiz for pharmaceutical field.pptx
pharma quiz for pharmaceutical field.pptxPriyanka Yadav
 
WHO_ppt_final - Copy.ppt
WHO_ppt_final - Copy.pptWHO_ppt_final - Copy.ppt
WHO_ppt_final - Copy.pptPriyanka Yadav
 
hyphenatedtechniques-200429145545.pdf
hyphenatedtechniques-200429145545.pdfhyphenatedtechniques-200429145545.pdf
hyphenatedtechniques-200429145545.pdfPriyanka Yadav
 
extractionprocess-190307055449.pdf
extractionprocess-190307055449.pdfextractionprocess-190307055449.pdf
extractionprocess-190307055449.pdfPriyanka Yadav
 
potentiometry-180713090143.pdf
potentiometry-180713090143.pdfpotentiometry-180713090143.pdf
potentiometry-180713090143.pdfPriyanka Yadav
 
Chromatography HPLC.ppt
Chromatography HPLC.pptChromatography HPLC.ppt
Chromatography HPLC.pptPriyanka Yadav
 
Antimalerial agent.pdf
Antimalerial agent.pdfAntimalerial agent.pdf
Antimalerial agent.pdfPriyanka Yadav
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.pptPriyanka Yadav
 
Standardisation of herbal drg
Standardisation of herbal drgStandardisation of herbal drg
Standardisation of herbal drgPriyanka Yadav
 

Mehr von Priyanka Yadav (15)

pharma quiz for pharmaceutical field.pptx
pharma quiz for pharmaceutical field.pptxpharma quiz for pharmaceutical field.pptx
pharma quiz for pharmaceutical field.pptx
 
IR-.ppt
IR-.pptIR-.ppt
IR-.ppt
 
WHO_ppt_final - Copy.ppt
WHO_ppt_final - Copy.pptWHO_ppt_final - Copy.ppt
WHO_ppt_final - Copy.ppt
 
hyphenatedtechniques-200429145545.pdf
hyphenatedtechniques-200429145545.pdfhyphenatedtechniques-200429145545.pdf
hyphenatedtechniques-200429145545.pdf
 
extractionprocess-190307055449.pdf
extractionprocess-190307055449.pdfextractionprocess-190307055449.pdf
extractionprocess-190307055449.pdf
 
IPQC Seminar.pptx
IPQC Seminar.pptxIPQC Seminar.pptx
IPQC Seminar.pptx
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
potentiometry-180713090143.pdf
potentiometry-180713090143.pdfpotentiometry-180713090143.pdf
potentiometry-180713090143.pdf
 
Chromatography HPLC.ppt
Chromatography HPLC.pptChromatography HPLC.ppt
Chromatography HPLC.ppt
 
Antimalerial agent.pdf
Antimalerial agent.pdfAntimalerial agent.pdf
Antimalerial agent.pdf
 
NMR B.Ph 8th.pdf
NMR B.Ph 8th.pdfNMR B.Ph 8th.pdf
NMR B.Ph 8th.pdf
 
mass.ppt
mass.pptmass.ppt
mass.ppt
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.ppt
 
NMR SPECTROSCOPY .ppt
NMR SPECTROSCOPY .pptNMR SPECTROSCOPY .ppt
NMR SPECTROSCOPY .ppt
 
Standardisation of herbal drg
Standardisation of herbal drgStandardisation of herbal drg
Standardisation of herbal drg
 

Kürzlich hochgeladen

Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsMedicoseAcademics
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfSasikiranMarri
 
Report Back from SGO: What’s New in Uterine Cancer?.pptx
Report Back from SGO: What’s New in Uterine Cancer?.pptxReport Back from SGO: What’s New in Uterine Cancer?.pptx
Report Back from SGO: What’s New in Uterine Cancer?.pptxbkling
 
Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?bkling
 
Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!ibtesaam huma
 
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranMusic Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranTara Rajendran
 
Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Mohamed Rizk Khodair
 
epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptxMohamed Rizk Khodair
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptMumux Mirani
 
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxPERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxdrashraf369
 
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATROApril 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATROKanhu Charan
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt downloadAnkitKumar311566
 
Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxDr. Dheeraj Kumar
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfSreeja Cherukuru
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdfDolisha Warbi
 
POST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxPOST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxvirengeeta
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
The next social challenge to public health: the information environment.pptx
The next social challenge to public health:  the information environment.pptxThe next social challenge to public health:  the information environment.pptx
The next social challenge to public health: the information environment.pptxTina Purnat
 

Kürzlich hochgeladen (20)

Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes Functions
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdf
 
Report Back from SGO: What’s New in Uterine Cancer?.pptx
Report Back from SGO: What’s New in Uterine Cancer?.pptxReport Back from SGO: What’s New in Uterine Cancer?.pptx
Report Back from SGO: What’s New in Uterine Cancer?.pptx
 
Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?
 
Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!
 
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranMusic Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
 
Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)
 
epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptx
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.ppt
 
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptxPERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
PERFECT BUT PAINFUL TKR -ROLE OF SYNOVECTOMY.pptx
 
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATROApril 2024 ONCOLOGY CARTOON by  DR KANHU CHARAN PATRO
April 2024 ONCOLOGY CARTOON by DR KANHU CHARAN PATRO
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt download
 
Culture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptxCulture and Health Disorders Social change.pptx
Culture and Health Disorders Social change.pptx
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
 
POST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxPOST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptx
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
The next social challenge to public health: the information environment.pptx
The next social challenge to public health:  the information environment.pptxThe next social challenge to public health:  the information environment.pptx
The next social challenge to public health: the information environment.pptx
 

GC.ppt

  • 1. Gas Chromatography •General Design of a Gas Chromatograph •Separation Processes in Gas Chromatography •GC Columns •GC Injectors •GC Detectors
  • 3. Some of the design“details” • Gas supplies usually have either in-line or instrument mounted traps to remove any water, oxygen, hydrocarbons or other “contaminants” from compressed gases • Gas flows can be controlled using either needle valves or mass-flow controllers (electronic sensors) • Instruments can have multiple injectors, detectors or columns • Injectors and detectors usually have their own temperature controlled zones (small heaters) • The GC oven has a large fan and a vent door to help with rapid cooling of the oven • Data collection (and integration) can be done using a chart recorder, integrator or a computerized data system
  • 4. Separation Processes in GC • Gas Chromatography as it is usually performed is correctly called gas- liquid chromatography • the analyte is in the gas phase in the GC and partitions between the mobile phase (carrier gas) and the liquid stationary phase that is coated on the inside of an open-tubular capillary column or on particles inside a packed column • Some packed-column GC uses non-coated solid stationary phases, in which case one is performing gas-solid adsorption chromatography • Capillary, open-tubular (WCOT specifically) column GC is the primary type of GC used in quantitative analysis: • higher resolution = greater ability to discriminate between components • smaller capacity of the column is not important as long as sufficient analyte is available for detection • pg/mL (ppt) to g/mL (ppm) concentration range for liquid analytes
  • 5.
  • 6. The Objective in Chromatography (all types) • Separate your analytes (resolution of 1.5 or better) in the shortest amount of time possible and detect them…. • How can we do this in GC? • Use different columns for different analyte types • stationary phase • diameter of column, stationary phase thickness • column length • Use different injection types/temperatures to optimize the process of loading the sample on the column • Use different temperature (or pressure) programs for the column • Select and use a detector that is suitable for the analyte(s) of interest
  • 7. GC Columns (concentrating on open-tubular capillary columns) • Column “frame” constructed of fused silica tubing • Polyamide coating on the outside gives it strength • Liquid stationary phases coated or bonded to the inside of the tubing • 0.1 - 0.53 mm + ID, 5-100 meters in length, stationary phases usually 0.10 to 1.5 m in thickness • Mounted on a wire cage to make them easier to handle • 5-150 meters long.
  • 9.
  • 10.
  • 11. Choosing a GC Column… • Is the column compatible with your analytes • polar analytes require polar stationary phases so they will spend some of their “time” in the stationary phase • non-polar analytes require non-polar stationary phases • You usually have to compromise on the stationary phase to get a good column for your analytes (which are probably a mix of polar and non-polar) • DB-5, HP-5, EC-5, RTX-5 (5% dimethyl, 95% diphenyl polysiloxane) most common general use column. • Temperature range, solvent and carrier gas compatibility • Sample capacity versus resolution • usually determines packed vs.. capillary • GC’s usually setup for either packed or capillary • Let’s say you choose a capillary column, there’s more to think about!
  • 12.
  • 13.
  • 14.
  • 15.
  • 16. For capillary GC columns….  Increased length = greater N, therefore a greater R  expense is possible band broadening if analytes are on the column too long!  Increased length leads to longer separations. Do you have the time?  Increased stationary phase thickness and column diameter provides increased sample capacity and can provide increased resolution  tradeoffs are a longer analysis time and more column bleed with thicker stationary phases  Is the column compatible with the detector?  Thick stationary phases bleed more and will contaminate a mass spectrometer.  For most analytical work, a best “compromise” column is chosen and other variables (temp, etc.) are altered to optimize the separation.
  • 17.
  • 18.
  • 19. Capillary vs. Packed Columns  Capillary Columns:  Higher resolution (R)  Greater HETP and N  Shorter analysis time  Greater sensitivity  Most common in analytical laboratory GC instruments  Smaller sample capacity  Higher cost/column  Columns more susceptible to damage  Packed Columns  Greater sample capacity  Lower cost (can make your own)  More rugged  Most common in process labs or separating/determining major components in a sample (prep GC)  Limited lengths reduces R and N  Not compatible with some GC detectors
  • 20. Temperature Programming in GC  The “simplest” way to alter the separation in GC is to alter the temperature program in the oven. You can also alter the pressure of the carrier gas, but this is less common (much).  Isothermal = constant temperature  Gradient = varied temperature  By altering the temperature, you vary the rate of the reaction for any analyte:  they spend more or less time in the stationary phase  the greater the difference in the times between analytes, the better the separation! Analyte Analyte phase stationary phase mobile 
  • 21.
  • 22. The traps of temperature… If your temperature at a given time is too high, you can cause the peaks to co-elute poor resolution vs but a faster separation If your temperature at a given time is too low, you can get still get a good separation adequate resolution, but a separation that takes very long You have to choose a compromise temperature program
  • 23.
  • 24. GC Carrier Gases (the mobile phase)  Usually “inert” gases (don’t react with analytes except sometimes in the detector)  Purpose:  sweep sample through the column  protect column from oxygen exposure at temperature  assist with function of the detector  Most common:  Helium (available relatively pure without extensive purification after it leaves a compressed gas cylinder)  Nitrogen (usually requires an oxygen and water trap)  Hydrogen normally used only with flame ionization detectors (FID) since the FID needs it as fuel for the flame still rarely used due to safety concerns (and chromatographic ones)
  • 25.
  • 26. GC Injection….  Samples are injected through a septum:  keeps oxygen out of the column  provides a seal to keep the carrier gas pressure up at the head of the column carrier gas flow rate is determined by the pressure or the gas at the opening of the column  Many different (mostly proprietary) materials red rubber (bleeds at about 250 C) Thermogreen (good up to about 300 C) High-temperature blue (good a little over 300 C)  The injector is usually lined with a de-activated glass liner  prevents metal injector-sample reactions that would alter analytes or damage the metal of the injector  Can be cleaned/replaced regularly
  • 27.
  • 28.
  • 30.  On-Column Injection:  used widely in packed-column GC, less in capillary GC  sample is deposited directly on the column Good for thermally unstable compounds Good for quantitative analysis at low concentrations  all sample is available to travel to the detector BUT, you can inject only a relatively small amount of sample in capillary GC anyhow.  Splitless Injection:  Sample is vaporized in the injector itself and ALL of the sample is swept onto the column by the carrier gas  Again, relatively small samples are injected (10  L or less in capillary GC)  Sample spends a large amount of time in the injector  Best for trace (1 -100 ppm range) concentrations of high boiling point analytes in low boiling point solvents extra time in the injector helps volatilize the analytes.
  • 31. Split Injection: the injection is split, with only a portion of the sample (usually 1% - 20%) actually making it to the column the most common method of injecting samples onto small diameter, open-tubular columns. Even if you inject 20 L, only a fraction (adjustable) makes it on to the column Not good for analytes with a wide range of boiling points some may be swept out the split vent before they are volatilized Modern capillary GCs come with a Split/Splitless injectors standard you switch between modes by changing the split vent gas flow and using a different injection liner.
  • 32. Don’t Forget SPME (Solid Phase Microextraction)
  • 33. GC Detectors  A dozen or more varieties (some obscure)  Must be:  sensitive to the analytes of interest  compatible with the column, carrier gas, solvent, etc.  rugged enough to withstand general unattended used I’ve run our new GC for 36 hours straight without touching it!  Should have a known linear range  if the detector response is very linear, you can use a response factor instead of a calibration curve for quantitation!  Usually require separate gas supplies (other than the carrier gas), have their own temperature control.  Measure nothing more than a voltage or a current.
  • 34.
  • 35.
  • 36.
  • 37.
  • 39. Thermal Conductivity (TCD)  The carrier gas has a known thermal conductivity.  As the thermal conductivity of the column eluent (gas flow in) changes, the resistance of the filament changes.  The presence of analyte molecules in the carrier gas alter the thermal conductivity of the gas (usually He)  There is normally a second filament to act as a reference (the carrier gas is split)  Increased sensitivity with decreasing temperature (detector), flow rate and applied current.  Filaments will burn out (oxidized) in the presence of oxygen if hot! Non-destructive
  • 40.
  • 41. FID  Destructive, sample lost.  Analytes containing C burn in a hydrogen-oxygen flame and produce ions  CHO+ ions are collected on a cathode and the current they produce results in the signal  WILL NOT detect non-C containing compounds!  Requires H2 supply (tank or generator) and O2 supply (compressed air)  H2 carrier gas can be used, eliminating the need for a supply for the detector  A makeup gas can also be required! - Flame O , H e CHO O CH 2 2    
  • 42.
  • 43. ECD  Particularly sensitive to halogens nitriles, carbonyls, nitro compounds  Analytes pass through a cell, in which electrons are traveling between a 63Ni electrode and a collector electrode  As analytes with “electron capturing ability” pass through the cell, the flow of electrons is interrupted.  The change in current, due to reduced flow of electrons, is recorded.  EXTREMELY SENSITIVE TO HALOGENS  could ruin detector with 1 ppm hexachlorocyclohexane by contaminating it with excess analyte  Widely used for the determination of pesticides, herbicides and PCBs in environmental samples.  Non-destructive