The document presents an overview of high performance liquid chromatography (HPLC). It discusses the key components of an HPLC system including the reservoir, pump, injector, separation column, and detector. It explains that compounds are separated on the column based on differences in how they partition between the mobile and stationary phases. The document also reviews different modes of HPLC, common applications, and advantages over gas chromatography.
2. Outlook:
History
Introduction
The components of the high performance liquid chromatograph
What affects system
The separation process
The chromatogram
The most common modes of HPLC
HPLC Applications
The differences between HPLC and GC
Advantages of High Performance Liquid Chromatography
Mehdi Soleymani 2
Medical Nanotechnology Department
3. 3
Ether
Chlorophyll
CaCO3
Chromatography
Colors
Mehdi Soleymani
Medical Nanotechnology Department
4. 4
Water flow
Base
Light leaf
Heavy stone
Mehdi Soleymani
Medical Nanotechnology Department
5. 5
Chromatography in which the mobile phase is a liquid.
The liquid used as the mobile phase is called the “eluent”.
The stationary phase is usually a solid or a liquid.
In general, it is possible to analyze any substance that
can be stably dissolved in the mobile phase.
Mehdi Soleymani
Medical Nanotechnology Department
6. 6
Chromatography: Analytical technique
Chromatograph: Instrument
Chromatogram: Obtained “picture”
Chromatographer: Person
Mehdi Soleymani
Medical Nanotechnology Department
7. 7
Higher degree of separation!
Refinement of packing material (3 to 10 μm)
Reduction of analysis time!
Delivery of eluent by pump
Demand for special equipment that can withstand high pressures
The arrival of high performance liquid chromatography!
Mehdi Soleymani
Medical Nanotechnology Department
8. HPLC is a form of liquid chromatography used to separate compounds
that are dissolved in solution. HPLC instruments consist of a reservoir of
mobile phase, a pump, an injector, a separation column, and a detector.
Compounds are separated by injecting a sample mixture onto the
column. The different component in the mixture pass through the
column at differentiates due to differences in their partition behavior
between the mobile phase and the stationary phase. The mobile phase
must be degassed to eliminate the formation of air bubbles.
Mehdi Soleymani 8
Medical Nanotechnology Department
10. 10
Isocratic system
Constant eluent composition
Gradient system
Varying eluent composition
HPGE (High Pressure Gradient)
LPGE (Low Pressure Gradient)
Mehdi Soleymani
Medical Nanotechnology Department
11. 11
Mixer
High-pressure gradient
Low-pressure
gradient unit
Mixer
Low-pressure gradient
Mehdi Soleymani
Medical Nanotechnology Department
12. 12
Principle Pattern An Example
Solvent Cabinet Vacuum Degasser
Binary Pump
Autosampler
Thermostatted
Column Compartment
Detector
Solvent Reservoirs
Controller
Mehdi Soleymani
Medical Nanotechnology Department
13. The column is one of the most
important components of the HPLC
chromatograph because the separation
of the sample components is achieved
when those components pass through
the column. The High performance
liquid chromatography apparatus is
made out of stainless steel tubes with a
diameter of 3 to 5mm and a length
ranging from 10 to 30cm.
Normally, columns are filled with silica
gel because its particle shape, surface
properties, and pore structure help to get
a good separation. Silica is wetted by
nearly every potential mobile phase, is
inert to most compounds and has a high
surface activity which can be modified
easily with water and other agents. Silica
can be used to separate a wide variety of
chemical compounds, and its
chromatographic behavior is generally
predictable and reproducible.
Mehdi Soleymani
Medical Nanotechnology Department
13
15. 15
Problems caused by dissolved air in the eluent
Unstable delivery by pump
More noise and large baseline drift in detector cell
In order to avoid these problems, the eluent
must be degassed.
Mehdi Soleymani
Medical Nanotechnology Department
16. The function of the injector is to place the sample into the high-pressure flow
in as narrow volume as possible so that the sample enters the column as a
homogeneous, low-volume plug. To minimize spreading of the injected volume
during transport to the column, the shortest possible length of tubing should be
used from the injector to the column.
When an injection is started, an air actuator rotates the valve: solvent goes
directly to the column; and the injector needle is connected to the syringe. The
air pressure lifts the needle and the vial is moved into position beneath the
needle. Then, the needle is lowered to the vial.
Mehdi Soleymani
Medical Nanotechnology Department
16
17. 17
Front View
Inject
Rear View
Load - Inject
Sample Loop
Mehdi Soleymani
Medical Nanotechnology Department
18. Absorbance (UV with
Filters, UV with
Monochromators)
IR Absorbance
Fluorescence
Refractive-Index
Evaporative Light
Scattering Detector (ELSD)
Electrochemical
Mass-Spectrometric
Photo-Diode Array
Mehdi Soleymani 18
Medical Nanotechnology Department
19. 19
Carbohydrates
1. fructose
2. Glucose
3. Saccharose
4. Palatinose
5. Trehalulose
6. isomaltose
Zorbax NH2 (4.6 x 250 mm)
70/30 Acetonitrile/Water
1 mL/min Detect=Refractive Index
1
2
3
4
5
mAU
time
6
Mehdi Soleymani
Medical Nanotechnology Department
20. Waste
Mehdi Soleymani 20
Medical Nanotechnology Department
Injector
Detector
Column
Solvents
Mixer
Pumps
High Performance Liquid Chromatograph
Separation in based upon differential
migration between the stationary and
mobile phases.
Stationary Phase - the phase which
remains fixed in the column, e.g. C18,
Silica
Mobile Phase - carries the sample
through the stationary phase as it
moves through the column.
21. 21
Injector
Detector
Column
Solvents
Mixer
Pumps
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
22. 22
Injector
Detector
Column
Solvents
Mixer
Pumps
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
23. 23
Injector
Detector
Column
Solvents
Mixer
Pumps
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
24. 24
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
25. 25
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
26. 26
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
27. 27
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
28. 28
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
29. 29
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
30. 30
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
31. 31
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
32. 32
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
33. 33
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
34. 34
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
35. 35
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
36. 36
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
37. 37
Injector
Detector
Column
Pumps
Solvents
Mixer
Chromatogram
Start Injection
mAU
time
High Performance Liquid Chromatograph
Mehdi Soleymani
Medical Nanotechnology Department
38. 38
to
Injection
tR
mAU
tR
time
to - elution time of unretained peak
tR- retention time - determines sample identity
Area or height is proportional
to the quantity of analyte.
Mehdi Soleymani
Medical Nanotechnology Department
39. 39
Mobile Phases
Flow Rate
Composition
Injection Volume
Column
Oven Temperature
Wavelength
Time Constant
Mehdi Soleymani
Medical Nanotechnology Department
40. Column Parameters
Column Material
Deactivation
Stationary Phase
Coating Material
Instrument Parameters
Temperature
Flow
Signal
Sample Sensitivity
Detector
Mehdi Soleymani 42
Medical Nanotechnology Department
41. Sample Parameters
Concentration
Matrix
Solvent Effect
Sample Effect
Mehdi Soleymani 43
Medical Nanotechnology Department
42. Normal phase
Reverse phase
Size exclusion
Ion exchange
Mehdi Soleymani 44
Medical Nanotechnology Department
43. Normal Phase.
- Polar stationary phase and non-polar solvent.
• Reverse Phase.
- Non-polar stationary phase and a polar
solvent.
Mehdi Soleymani
Medical Nanotechnology Department
45
44. 44
Types of Compounds Mode Stationary
Phase
Mobile Phase
Neutrals
Weak Acids
Weak Bases
Reversed
Phase
C18, C8, C4
cyano, amino
Water/Organic
Modifiers
Ionics, Bases, Acids Ion
Pair
C-18, C-8 Water/Organic
Ion-Pair Reagent
Compounds not
soluble in water
Normal
Phase
Silica, Amino,
Cyano, Diol
Organics
Ionics Inorganic Ions Ion
Exchange
Anion or Cation
Exchange
Resin
Aqueous/Buffer
Counter Ion
High Molecular Weight
Compounds
Polymers
Size
Exclusion
Polystyrene
Silica
Gel Filtration-
Aqueous
Gel Permeation-
Organic
Mehdi Soleymani
Medical Nanotechnology Department
45. Chemical
Pharmaceuticals
Environmental
Bioscience
proteins
peptides
nucleotides
Consumer Products
Clinical
polystyrenes
dyes
phthalates
tetracyclines
corticosteroids
antidepressants
barbiturates
lipids
antioxidants
sugars
amino acids
vitamins
homocysteine
polyaromatic hydrocarbons
Inorganic ions
herbicides
Mehdi Soleymani 45
Medical Nanotechnology Department
46. I need a quantitative separation of
carbohydrates in some of our
products as soon as possible.
I’ll need a separation technique.
Mehdi Soleymani 46
Medical Nanotechnology Department
I’ll get on it!
47. I have two separation techniques in my lab,
High Performance Liquid Chromatography
and Gas Chromatography. Which one should I use?
Mehdi Soleymani 47
Medical Nanotechnology Department
48. 48
High separation capacity, enabling the batch analysis
of multiple components
Superior quantitative capability and reproducibility
Moderate analytical conditions
Unlike GC, the sample does not need to be vaporized.
Generally high sensitivity
Low sample consumption
Easy preparative separation and purification of
samples
Mehdi Soleymani
Medical Nanotechnology Department