Motion artifact detection in respiratory
signals based on Teager energy operator
and accelerometer signals
Marcel Młyńczak and Gerard Cybulski

Warsaw University of Technology, Faculty of Mechatronics,
Institute of Metrology and Biomedical Engineering
Tampere, June 14, 2017
Introduction
2
Ambulatory measurements

Sleep recordings

Indirect but quantitative method
Sense of using impedance pneumography
Introduction
3
Contact method using electrodes

Need for calibration, dependent on body position

Motion artifacts appearances
Remarks / Drawbacks
Introduction
4
Motion artifacts
Frequency range of an artifact signal 

partly overlaps that of breathing

Amplitude is usually uncorrelated with 

motion intensity and characteristics

Shape may be repeatable, 

but not in the same way in each subject
Introduction
• Spectral or wavelet features
• Signal decomposition
• Kalman filtering
• Adaptive method based on ε-Tube
The problem had not been studied in depth for 

estimation of quantitative respiratory parameters!
5
Related works
Objectives
•Presentation of novel method
• relatively computationally simple
• adapted to IP signal characteristics
to detect motion artifacts using
• Teager energy operator
• additional 3-axis accelerometer signals
•Assessment of its accuracy relative to manual marking
6
Materials and Methods
7
Study group - 24 students (12 females and 12 males)
Female Male
Mean SD Mean SD
Mass [kg] 58.6 5.6 76.2 9.5
Weight [cm] 168.2 6.2 178.8 5.6
BMI 20.7 1.6 23.9 3.3
Age 22.3 5.3 22.9 3.2
8
Study protocol
4 free breaths 

for consecutive 

body positions:
➡ supine

➡ side

➡ prone

➡ side

➡ supine

➡ sitting

➡ standing

➡ sitting

➡ supine
Materials and Methods
Materials and Methods
Pneumonitor 2
9
• ECG signal to estimate heart rate
and tachogram
• Impedance signal relating to 

main breathing activity
• Portable

• Recording on SD card

• Rechargeable battery
• Motion signal from 3-axis
accelerometer to indicate 

subject’s activity and body position
Materials and Methods
Electrode configuration - as suggested by Seppa et al.
IP electrodes
ECG electrodes
Motion sensor
10
Algorithm
11
Raw IP signal
Motion 3-axis signals
Teager-Kaiser
Energy Operator
Aggregation TKE
Moving average smoothing
40 milliseconds window
Moving average smoothing
48 milliseconds window
…
Moving average smoothing
1 second window
-||-
-||-
-||-
Summation of absolute
derivatives of each axis
MRD
Spectrogram estimation
Summation of frequency
content for consecutive
time portions
TKEspecNormalization
Interpolation
Normalization
Avg Std
Weight [kg] 58.6 5.6
Height [cm] 168.2 6.2
BMI 20.7 1.6
Age 22.3 5.3
the arms at the same level [12]. Movements
the belt. A sampling frequency of fs = 250H
B. Motion artifact detection
A schematic of the proposed motion a
method is presented in Figure 1. The pro
based on the continuous Teager–Kaiser en
was originally introduced by Kaiser et al. as
ator to measure instantaneous energy change
sisting of a single time-varying frequency [13
eral applications to detect sudden changes,
sets, were later proposed [15, 16].
The continuous Teager-Kaiser energy ope
x in time t is calculated as follows:
TKE (x) =
✓
dx
dt
◆2
+x
d2x
dt2
It was found that the calculated energy
both amplitude and frequency. Therefore, th
phasizes both instantaneous properties [14].
* * * * * * *
1st step
Manually annotated
motion artifacts
Supine Side Prone Side Supine Sitting Standing Sitting
Algorithm
12
Kaiser
perator
Aggregation TKE
on of absolute
s of each axis
MRD
Spectrogram estimation
Summation of frequency
content for consecutive
time portions
TKEspec
Final reas
Normalization
Interpolation
Envelope d
TKEe
THR
Normalization
Hann window 

with 32 samples
2nd step
Algorithm
13
3rd step
* * * * * * *
Manually annotated
motion artifacts
Algorithm
14
Smoothed envelope estimated 

using the Hilbert transform
4th step
Algorithm
15
Raw IP signal
Motion 3-axis signals
Teager-Kaiser
Energy Operator
Aggregation TKE
Moving average smoothing
40 milliseconds window
Moving average smoothing
48 milliseconds window
…
Moving average smoothing
1 second window
-||-
-||-
-||-
Summation of absolute
derivatives of each axis
MRD
Spectrogram estimation
Summation of frequency
content for consecutive
time portions
TKEspec
Final reasoning
Normalization
Interpolation
Envelope detection
TKEenv
THR
Normalization
Sample signal and analysis
16
Results
17
Classification results for the best setting
Accuracy [%] 81.3
Cohen’s kappa coefficient 0.63
Sensitivity [%] 80.9
Specificity [%] 81.3
Discussion
18
State-of-the-art motion artifact detection methods are poorly suited for 

respiratory data, because they usually assume some regularities in the signals.

Accelerometer signals are certainly not necessary, but could be used to 

adapt the threshold, intended to determine whether some part of a signal 

is more likely to have motion artifacts.

In our opinion, presented method could be improved by:

• changing the thresholding approach into continuous analysis using time series
methods, machine learning heuristics, and/or Kalman filtering fusion method;

• verifying the accuracy improvement by making the whole analysis 

fully independent on the respiratory-related impedance amplitude;

• adding wavelet-based parameters to the analysis, with wavelet shapes closely
resembling those most often observed as motion artifacts.
Summary
We obtained 82% accuracy versus manual marking 

for the optimal combination of threshold level estimation
and assumptions made to utilize motion signals.

Proposed method could be applied in real time during
ambulatory signal acquisitions.
19
Motion artifact detection in respiratory
signals based on Teager energy operator
and accelerometer signals
Tampere, June 14, 2017
Marcel Młyńczak

mlynczak@mchtr.pw.edu.pl

More Related Content

What's hot

Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
RSIS International
 
Controls Based Q Measurement Report
Controls Based Q Measurement ReportControls Based Q Measurement Report
Controls Based Q Measurement Report
Louis Gitelman
 
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
International Journal of Technical Research & Application
 

What's hot (10)

Rolling Element Bearing Condition Monitoring using Filtered Acoustic Emission
Rolling Element Bearing Condition Monitoring using Filtered Acoustic Emission Rolling Element Bearing Condition Monitoring using Filtered Acoustic Emission
Rolling Element Bearing Condition Monitoring using Filtered Acoustic Emission
 
Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
Gearbox Fault Diagnosis using Independent Angular Re-Sampling Technique, Wave...
 
Detecting stable phase structures in eeg signals to classify brain activity a...
Detecting stable phase structures in eeg signals to classify brain activity a...Detecting stable phase structures in eeg signals to classify brain activity a...
Detecting stable phase structures in eeg signals to classify brain activity a...
 
Analysis of Adaptive Algorithms
Analysis of Adaptive AlgorithmsAnalysis of Adaptive Algorithms
Analysis of Adaptive Algorithms
 
A NEW METHOD OF SMALL-SIGNAL CALIBRATION BASED ON KALMAN FILTER
A NEW METHOD OF SMALL-SIGNAL CALIBRATION BASED ON KALMAN FILTERA NEW METHOD OF SMALL-SIGNAL CALIBRATION BASED ON KALMAN FILTER
A NEW METHOD OF SMALL-SIGNAL CALIBRATION BASED ON KALMAN FILTER
 
Controls Based Q Measurement Report
Controls Based Q Measurement ReportControls Based Q Measurement Report
Controls Based Q Measurement Report
 
Ecg programa simulado
Ecg programa simuladoEcg programa simulado
Ecg programa simulado
 
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
SPACE TIME ADAPTIVE PROCESSING FOR CLUTTER SUPPRESSION IN RADAR USING SUBSPAC...
 
F3602045049
F3602045049F3602045049
F3602045049
 
SENSOR SELECTION SCHEME IN TEMPERATURE WIRELESS SENSOR NETWORK
SENSOR SELECTION SCHEME IN TEMPERATURE WIRELESS SENSOR NETWORKSENSOR SELECTION SCHEME IN TEMPERATURE WIRELESS SENSOR NETWORK
SENSOR SELECTION SCHEME IN TEMPERATURE WIRELESS SENSOR NETWORK
 

Similar to Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals

Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
CSCJournals
 
Novel method to find the parameter for noise removal from multi channel ecg w...
Novel method to find the parameter for noise removal from multi channel ecg w...Novel method to find the parameter for noise removal from multi channel ecg w...
Novel method to find the parameter for noise removal from multi channel ecg w...
eSAT Journals
 
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
ijtsrd
 
The vlsi implementation of ECG Signal compression algorithm
The vlsi implementation of ECG Signal compression algorithmThe vlsi implementation of ECG Signal compression algorithm
The vlsi implementation of ECG Signal compression algorithm
VarunDev41
 

Similar to Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals (20)

Classification and Detection of ECG-signals using Artificial Neural Networks
Classification and Detection of ECG-signals using Artificial Neural NetworksClassification and Detection of ECG-signals using Artificial Neural Networks
Classification and Detection of ECG-signals using Artificial Neural Networks
 
Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
Analysis of Human Electrocardiogram for Biometric Recognition Using Analytic ...
 
Classification of ecg signal using artificial neural network
Classification of ecg signal using artificial neural networkClassification of ecg signal using artificial neural network
Classification of ecg signal using artificial neural network
 
Ecg
EcgEcg
Ecg
 
Fo3610221025
Fo3610221025Fo3610221025
Fo3610221025
 
Decomposition of the Cardiac and Respiratory Components from Impedance Pneumo...
Decomposition of the Cardiac and Respiratory Components from Impedance Pneumo...Decomposition of the Cardiac and Respiratory Components from Impedance Pneumo...
Decomposition of the Cardiac and Respiratory Components from Impedance Pneumo...
 
Less computational approach to detect QRS complexes in ECG rhythms
Less computational approach to detect QRS complexes in ECG rhythmsLess computational approach to detect QRS complexes in ECG rhythms
Less computational approach to detect QRS complexes in ECG rhythms
 
Ambulatory Devices Measuring Cardiorespiratory Activity with Motion
Ambulatory Devices Measuring Cardiorespiratory Activity with MotionAmbulatory Devices Measuring Cardiorespiratory Activity with Motion
Ambulatory Devices Measuring Cardiorespiratory Activity with Motion
 
2
22
2
 
CLASSIFICATION OF ECG ARRHYTHMIAS USING /DISCRETE WAVELET TRANSFORM AND NEURA...
CLASSIFICATION OF ECG ARRHYTHMIAS USING /DISCRETE WAVELET TRANSFORM AND NEURA...CLASSIFICATION OF ECG ARRHYTHMIAS USING /DISCRETE WAVELET TRANSFORM AND NEURA...
CLASSIFICATION OF ECG ARRHYTHMIAS USING /DISCRETE WAVELET TRANSFORM AND NEURA...
 
Novel method to find the parameter for noise removal
Novel method to find the parameter for noise removalNovel method to find the parameter for noise removal
Novel method to find the parameter for noise removal
 
Novel method to find the parameter for noise removal from multi channel ecg w...
Novel method to find the parameter for noise removal from multi channel ecg w...Novel method to find the parameter for noise removal from multi channel ecg w...
Novel method to find the parameter for noise removal from multi channel ecg w...
 
Analysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approachAnalysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approach
 
IEEE Bio medical engineering 2016 Title and Abstract
IEEE Bio medical engineering  2016 Title and Abstract IEEE Bio medical engineering  2016 Title and Abstract
IEEE Bio medical engineering 2016 Title and Abstract
 
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
An Automated ECG Signal Diagnosing Methodology using Random Forest Classifica...
 
Electrocardiograph signal recognition using wavelet transform based on optim...
Electrocardiograph signal recognition using wavelet transform  based on optim...Electrocardiograph signal recognition using wavelet transform  based on optim...
Electrocardiograph signal recognition using wavelet transform based on optim...
 
The vlsi implementation of ECG Signal compression algorithm
The vlsi implementation of ECG Signal compression algorithmThe vlsi implementation of ECG Signal compression algorithm
The vlsi implementation of ECG Signal compression algorithm
 
IRJET- Arrhythmia Detection using One Dimensional Convolutional Neural Network
IRJET- Arrhythmia Detection using One Dimensional Convolutional Neural NetworkIRJET- Arrhythmia Detection using One Dimensional Convolutional Neural Network
IRJET- Arrhythmia Detection using One Dimensional Convolutional Neural Network
 
Automatic ECG signal denoising and arrhythmia classification using deep learning
Automatic ECG signal denoising and arrhythmia classification using deep learningAutomatic ECG signal denoising and arrhythmia classification using deep learning
Automatic ECG signal denoising and arrhythmia classification using deep learning
 
New Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet TransformNew Method of R-Wave Detection by Continuous Wavelet Transform
New Method of R-Wave Detection by Continuous Wavelet Transform
 

Recently uploaded

(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
Scintica Instrumentation
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
Silpa
 
The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
seri bangash
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Sérgio Sacani
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
Silpa
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 
Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
MohamedFarag457087
 

Recently uploaded (20)

(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
(May 9, 2024) Enhanced Ultrafast Vector Flow Imaging (VFI) Using Multi-Angle ...
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
 
Use of mutants in understanding seedling development.pptx
Use of mutants in understanding seedling development.pptxUse of mutants in understanding seedling development.pptx
Use of mutants in understanding seedling development.pptx
 
The Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptxThe Mariana Trench remarkable geological features on Earth.pptx
The Mariana Trench remarkable geological features on Earth.pptx
 
module for grade 9 for distance learning
module for grade 9 for distance learningmodule for grade 9 for distance learning
module for grade 9 for distance learning
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Chemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdfChemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdf
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
 
Exploring Criminology and Criminal Behaviour.pdf
Exploring Criminology and Criminal Behaviour.pdfExploring Criminology and Criminal Behaviour.pdf
Exploring Criminology and Criminal Behaviour.pdf
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Stages in the normal growth curve
Stages in the normal growth curveStages in the normal growth curve
Stages in the normal growth curve
 
Velocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.pptVelocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.ppt
 
Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
 
Zoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdfZoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdf
 
300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)
 
GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)
 
Factory Acceptance Test( FAT).pptx .
Factory Acceptance Test( FAT).pptx       .Factory Acceptance Test( FAT).pptx       .
Factory Acceptance Test( FAT).pptx .
 
COMPUTING ANTI-DERIVATIVES (Integration by SUBSTITUTION)
COMPUTING ANTI-DERIVATIVES(Integration by SUBSTITUTION)COMPUTING ANTI-DERIVATIVES(Integration by SUBSTITUTION)
COMPUTING ANTI-DERIVATIVES (Integration by SUBSTITUTION)
 

Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals

  • 1. Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals Marcel Młyńczak and Gerard Cybulski Warsaw University of Technology, Faculty of Mechatronics, Institute of Metrology and Biomedical Engineering Tampere, June 14, 2017
  • 2. Introduction 2 Ambulatory measurements Sleep recordings Indirect but quantitative method Sense of using impedance pneumography
  • 3. Introduction 3 Contact method using electrodes Need for calibration, dependent on body position Motion artifacts appearances Remarks / Drawbacks
  • 4. Introduction 4 Motion artifacts Frequency range of an artifact signal 
 partly overlaps that of breathing Amplitude is usually uncorrelated with 
 motion intensity and characteristics Shape may be repeatable, 
 but not in the same way in each subject
  • 5. Introduction • Spectral or wavelet features • Signal decomposition • Kalman filtering • Adaptive method based on ε-Tube The problem had not been studied in depth for 
 estimation of quantitative respiratory parameters! 5 Related works
  • 6. Objectives •Presentation of novel method • relatively computationally simple • adapted to IP signal characteristics to detect motion artifacts using • Teager energy operator • additional 3-axis accelerometer signals •Assessment of its accuracy relative to manual marking 6
  • 7. Materials and Methods 7 Study group - 24 students (12 females and 12 males) Female Male Mean SD Mean SD Mass [kg] 58.6 5.6 76.2 9.5 Weight [cm] 168.2 6.2 178.8 5.6 BMI 20.7 1.6 23.9 3.3 Age 22.3 5.3 22.9 3.2
  • 8. 8 Study protocol 4 free breaths 
 for consecutive 
 body positions: ➡ supine ➡ side ➡ prone ➡ side ➡ supine ➡ sitting ➡ standing ➡ sitting ➡ supine Materials and Methods
  • 9. Materials and Methods Pneumonitor 2 9 • ECG signal to estimate heart rate and tachogram • Impedance signal relating to 
 main breathing activity • Portable • Recording on SD card • Rechargeable battery • Motion signal from 3-axis accelerometer to indicate 
 subject’s activity and body position
  • 10. Materials and Methods Electrode configuration - as suggested by Seppa et al. IP electrodes ECG electrodes Motion sensor 10
  • 11. Algorithm 11 Raw IP signal Motion 3-axis signals Teager-Kaiser Energy Operator Aggregation TKE Moving average smoothing 40 milliseconds window Moving average smoothing 48 milliseconds window … Moving average smoothing 1 second window -||- -||- -||- Summation of absolute derivatives of each axis MRD Spectrogram estimation Summation of frequency content for consecutive time portions TKEspecNormalization Interpolation Normalization Avg Std Weight [kg] 58.6 5.6 Height [cm] 168.2 6.2 BMI 20.7 1.6 Age 22.3 5.3 the arms at the same level [12]. Movements the belt. A sampling frequency of fs = 250H B. Motion artifact detection A schematic of the proposed motion a method is presented in Figure 1. The pro based on the continuous Teager–Kaiser en was originally introduced by Kaiser et al. as ator to measure instantaneous energy change sisting of a single time-varying frequency [13 eral applications to detect sudden changes, sets, were later proposed [15, 16]. The continuous Teager-Kaiser energy ope x in time t is calculated as follows: TKE (x) = ✓ dx dt ◆2 +x d2x dt2 It was found that the calculated energy both amplitude and frequency. Therefore, th phasizes both instantaneous properties [14]. * * * * * * * 1st step Manually annotated motion artifacts Supine Side Prone Side Supine Sitting Standing Sitting
  • 12. Algorithm 12 Kaiser perator Aggregation TKE on of absolute s of each axis MRD Spectrogram estimation Summation of frequency content for consecutive time portions TKEspec Final reas Normalization Interpolation Envelope d TKEe THR Normalization Hann window 
 with 32 samples 2nd step
  • 13. Algorithm 13 3rd step * * * * * * * Manually annotated motion artifacts
  • 14. Algorithm 14 Smoothed envelope estimated 
 using the Hilbert transform 4th step
  • 15. Algorithm 15 Raw IP signal Motion 3-axis signals Teager-Kaiser Energy Operator Aggregation TKE Moving average smoothing 40 milliseconds window Moving average smoothing 48 milliseconds window … Moving average smoothing 1 second window -||- -||- -||- Summation of absolute derivatives of each axis MRD Spectrogram estimation Summation of frequency content for consecutive time portions TKEspec Final reasoning Normalization Interpolation Envelope detection TKEenv THR Normalization
  • 16. Sample signal and analysis 16
  • 17. Results 17 Classification results for the best setting Accuracy [%] 81.3 Cohen’s kappa coefficient 0.63 Sensitivity [%] 80.9 Specificity [%] 81.3
  • 18. Discussion 18 State-of-the-art motion artifact detection methods are poorly suited for 
 respiratory data, because they usually assume some regularities in the signals. Accelerometer signals are certainly not necessary, but could be used to 
 adapt the threshold, intended to determine whether some part of a signal 
 is more likely to have motion artifacts. In our opinion, presented method could be improved by: • changing the thresholding approach into continuous analysis using time series methods, machine learning heuristics, and/or Kalman filtering fusion method; • verifying the accuracy improvement by making the whole analysis 
 fully independent on the respiratory-related impedance amplitude; • adding wavelet-based parameters to the analysis, with wavelet shapes closely resembling those most often observed as motion artifacts.
  • 19. Summary We obtained 82% accuracy versus manual marking 
 for the optimal combination of threshold level estimation and assumptions made to utilize motion signals. Proposed method could be applied in real time during ambulatory signal acquisitions. 19
  • 20. Motion artifact detection in respiratory signals based on Teager energy operator and accelerometer signals Tampere, June 14, 2017 Marcel Młyńczak mlynczak@mchtr.pw.edu.pl