General European OMCL Network (GEON)
QUALITY MANAGEMENT DOCUMENT
PA/PH/OMCL (21) 60 R2
Recommendations on setting the expiry period for commercial and
in-house-prepared reagents used in the laboratories of the OMCL Network
Full document title
and reference
Recommendations on setting the expiry period for
commercial and in-house-prepared reagents used in the
laboratories of the OMCL Network, PA/PH/OMCL (21) 60 R2
Document type Recommendation Document
Legislative basis Council Directive 2001/83/EC and 2001/82/EC, as amended
Date of first adoption 31 October 2021
Date of original entry
into force
1 January 2022
Date of entry into
force of revised
document
Previous titles/other
references / last valid
version
N/A
Custodian
Organisation
The present document was elaborated by the OMCL Network /
EDQM of the Council of Europe
Concerned Network GEON
N.B. This OMCL Quality Management System document is applicable to members of the European OMCL
Network only. Other laboratories might use the document on a voluntary basis. However, please note that the
EDQM cannot treat any questions related to the application of the documents submitted by laboratories other
than the OMCLs of the Network.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 2/15
Table of Contents
Preface...................................................................................................................................3
1. Purpose...........................................................................................................................3
2. Scope ..............................................................................................................................4
3. Reagents (chemicals).....................................................................................................4
3.1. Definitions ...............................................................................................................4
3.2. Establishing shelf life or an expiry date/period for a reagent (ready-to-use) ...........4
3.2.1. Unopened reagent container ............................................................................4
3.2.2. After opening of the reagent container ............................................................5
3.3. Decision tree for establishing the expiry period after opening of the reagent...........6
4. In-house-prepared reagent solutions ............................................................................6
4.1. Quality check...............................................................................................................6
4.2. Recommended expiry periods ....................................................................................6
4.3. Extension of adopted expiry period............................................................................6
5. Labelling and storage .....................................................................................................7
5.1. Examples of annotations for declared storage conditions..........................................7
6. Conclusions.....................................................................................................................7
7. References......................................................................................................................8
Annex 1..................................................................................................................................9
Annex 2................................................................................................................................10
Annex 3. Risk analysis.........................................................................................................14
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 3/15
Preface
Reagents are an integral part of any laboratory and as such play a vital role in ensuring the validity of analytical
results. Reagents must be suitable for the intended use. Consideration should be given to the accompanying
documentation and appropriate physico-chemical or biological critical quality attributes for the intended
purpose/s.
In the laboratory, two major types of reagents are generally distinguished: commercial (off-the-shelf, ready-
to-use) and those prepared in the laboratory (in-house reagents) using the former. These two types of
reagents are mainly dealt with separately throughout this document.
It should be noted that the principles of FIFO (First In-First Out) or FEFO (First Expired-First Out) should be
applied when using reagents.
The shelf life for an unopened product or expiry period after first opening is usually set taking into account its
stability, i.e. how long the quality attributes and chemical integrity of the reagent can be maintained for the
intended analytical application. This can be a challenging task because most manufacturers give the shelf life
of the unopened product and do not provide information on how long the reagent is stable once the container
has been opened.
The recommendations for the shelf life of reagent solutions prepared and used in laboratories and presented
in this document were established based on scientific data (including validation data obtained from quality
control charts, guidelines, standards, pharmacopoeias, publications, etc.) and on the analytical experience and
knowledge of OMCLs. Preparation times, costs, sustainable waste management policies and environmental
protection concerns were also considered during the preparation of these recommendations.
This document is based on the laboratory experience of the OMCL Network. Its purpose is to help maintain
the quality of the work carried out in the laboratory on a daily basis and ensure that the results obtained are
valid.
The recommendations on the shelf life of reagent solutions provided herein should allow laboratories to assess
whether a given solution can be used safely and reliably (provided that the storage conditions comply with
requirements) as they take into account various factors that contribute to the degradation of some reagents
(temperature, air exposure, moisture, etc.).
For the purpose of this document, the expiry period of a reagent solution should be understood as the period
of time during which the reagent is stable and can be used for a given type of analysis. It should be noted
that the same reagent may be unfit for use for some purposes because it has expired, but still suitable for
other purposes.
Purchase or preparation of reagents and reagent solutions in quantities (volumes) not adapted to current
needs should be avoided, as unnecessary disposal that does not take into account the possibility of prolonged
use potentially constitutes a direct threat to the environment. Hence, if prolonging the use of a given reagent
solution for specific purposes is safe and does not compromise the results, such prolongation should always
be considered and justified.
1. Purpose
The aim of this document is to describe and, where possible, define how to set expiry periods for reagents
(chemicals) used in the OMCL Network (and to guide OMCLs in applying them).
It is also intended to provide guidance on regulating the use, labelling, storage and expiry periods of reagent
solutions prepared in OMCLs (in-house solutions) for qualitative, semi-quantitative and quantitative purposes.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 4/15
2. Scope
Use of commercial reagents (chemicals) and reagent solutions prepared in OMCLs within the OMCL Network.
Commercial reagents or reagent preparations for immediate use are out of the scope of this document as
they are not stored and are used within the same day.
In addition, solutions for which there is a certificate by which the traceability to NIST or any other primary
etalon can be established are also out of the scope of this document.
3. Reagents (chemicals)
3.1. Definitions
Reagents
Reagents are solids, gases, liquids or solutions of pure chemical or biological substances, organic
solvents and water.
Quality criteria
Quality criteria are assumed to be covered by appropriate internal standard operating procedures (SOPs).
In general, special consideration should be given to the selection, purchase and receipt of reagents and
their storage. The emphasis is on the crucial step of purchasing reagents of an expected quality and, upon
receipt, on inspection and verification of compliance with the previously set specifications. These steps
precede the labelling and storage of the reagents.
Shelf life of the reagent is a set period or a date given by the manufacturer, which can usually be found
on the certificate of analysis (CoA) for that particular reagent. During this period or until that date, it is
expected that the reagent will meet the quality specifications found in the CoA if the storage conditions in
the material safety data sheet (MSDS) or in any other relevant document, like CoA or label, are respected.
Reagent without a shelf life or a retest date
Sometimes a shelf life is not defined by the manufacturer because the reagent is considered to have an
extensive shelf life when stored unopened under the proper conditions, or the manufacturer has no
supporting stability data. As the quality of a reagent may be altered due to use (effect of environmental
conditions), setting an expiry date is highly recommended. This can be done, for example, by carrying out
a risk assessment.
Recommended retest date for an unopened reagent
If it is possible to obtain a new certificate from the manufacturer, the shelf life of an unopened reagent
can be extended until a later date.
Expiry period is a period or a specific date after opening of the reagent during which it is expected to
maintain the appropriate physico-chemical or biological critical quality attributes for the intended purpose,
in spite of any physical and chemical changes that may occur. The MSDS should be used as a guide on
how to store the reagent after opening.
Extension of the expiry period for a reagent
The expiry period can be extended following a retest taking into account the critical quality attributes
contributing to the functional characteristics of the reagent, checked by appropriate means or by scientific
judgement.
3.2. Establishing shelf life or an expiry date/period for a reagent (ready-to-use)
3.2.1. Unopened reagent container
Usually a reagent is delivered with a stated shelf life for the unopened container.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 5/15
Where no information is available from the manufacturer, the shelf life for unopened reagent containers
can be taken to be the date of receipt plus up to 5 years. This is only applicable if the unopened reagent
container is stored under specified conditions found in relevant documents.
3.2.2. After opening of the reagent container
It is necessary to set a rational expiry period in the absence of analytical data on the stability of the product
after opening. This could be achieved by “real-time monitoring” of the reagent after opening by gathering
experimental analytical data on critical parameters, both chemical and physical, during laboratory use.
In addition, possible sources of information on the stability of a particular reagent are:
 the CoA or MSDS for the storage conditions and possible instability;
 scientific papers;
 chemical knowledge about the nature of the reagents;
 stability data – additional information from the manufacturer or supplier (e.g. certificates, which
can be found on manufacturers or suppliers website), guidelines, DIN/ISO Standards,
pharmacopoeial texts, etc.
The expiry period after opening should not exceed the shelf life of the unopened reagent container. If it
does, a risk assessment should be done based on the scientific literature and experimental data to show
that use of the reagent after the end of its shelf life does not compromise its safety or the quality of the
measurement results.
Each time a container is opened, the reagent is exposed to various external factors and environmental
conditions that may alter its integrity and affect its quality, which may compromise the validity of the
results.
Reagent properties, storage conditions, handling and the intended use that have to be considered during
the process of establishing the expiry period after opening are:
 Purity of the reagent – this property is very important for establishing the expiry period after
opening. In some cases the higher the purity of the reagent, the shorter the expiry period after
opening. For example, sodium chloride p.a. will have a longer expiry period after opening than
supra-pure sodium chloride, because it is rational to expect that supra-pure solids will deteriorate
more quickly and lose their quality attributes. However, the degree of purity needed for the testing
should be commensurate with the intended use of the reagent: it is important to select the quality
grade that will yield satisfactory results.
 Temperature – should be kept controlled within the range given by the manufacturer/stated in
the MSDS and/or in internal SOPs.
 Humidity – should meet the requirements given by the manufacturer/stated in the MSDS and/or
in internal SOPs.
 Light exposure.
 Contamination during use of the reagent – each time a reagent is opened, accidental
contamination is possible (switching of bottle caps, placing bottle caps on soiled surfaces,
transferring reagents to other containers, use of pipettes for withdrawal, etc.)
It is important to state that, regardless of the expiry period established for the opened reagent, a visual
inspection/examination should be done before use and if any of the physical properties that can be
determined by visual inspection have changed, the reagent should be discarded.
Suggested expiry periods for reagents after opening or for ready-to-use reagent solutions in the absence
of manufacturers' instructions (e.g. certificates) or official regulations (DIN/ISO Standards,
pharmacopoeial texts) or literature data are presented in Annex 1.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 6/15
3.3. Decision tree for establishing the expiry period after opening of the reagent
The decision tree presented in Annex 3 demonstrates the decision-making process on how to approach
setting the expiry period after opening.
4. In-house-prepared reagent solutions
Preparation of reagent solutions is carried out in accordance with applicable guidelines, compendial
methods (e.g. pharmacopoeial texts), test method specifications and procedures (Common Technical
Documents [CTDs], SOPs).
4.1. Quality check
It is necessary to ensure that a reagent solution has the properties required for the intended analysis.
Before using prepared reagent solutions, it is strongly recommended to:
 perform a visual inspection – any reagent should be discarded if there is a change in colour or
opalescence, or if there is any precipitation (unless it is necessary),
 perform a pH check before the use of buffer solutions (if applicable),
 determine the titre of volumetric solutions (if applicable),
 perform a Ph. Eur. sensitivity test for indicators.
Perform any other suitable qualification test specific to the prepared reagents. The environmental aspect
must always be taken into account when deciding to shorten the expiry period and/or dispose of a reagent
solution. A solution that is no longer suitable for a given analytical procedure may still be suitable for
another purpose.
4.2. Recommended expiry periods
Consideration should always be given to expiry periods:
 for reagent solutions described in compendial methods (e.g. 2 months for Primary opalescent
suspension in Ph. Eur. chapter 2.2.1);
 provided in drug manufacturers' documentation;
 for reagent solutions for immediate use given in the test method specification, (e.g. data from
the marketing authorisation dossiers).
If no stability data is available, it is recommended to use the expiry periods provided in Annex 2. The
expiry periods presented in Annexes 1 and 2 were established on the basis of the analytical experience of
OMCLs, method validation, guidelines, scientific publications [1, 4] and chemical knowledge.
OMCLs can also perform an in-house validation study to determine the initial expiry date.
4.3. Extension of adopted expiry period
It may be acceptable to extend the expiry periods provided in this document under the following
conditions:
 risk analysis has been performed (see Annex 3);
and/or
 test criteria are established, supporting the extension;
 an in-house validation study to determine the expiry period has been performed.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 7/15
The extension of the expiry period and the test criteria should be documented according to internal OMCL
procedures.
5. Labelling and storage
General information on labelling and storage requirements for reagents is given in GL Management of
Reagents, PA/PH/OMCL (11) 157 [3]. Whenever possible, reagent labels should state the expiry period for
an opened container.
More specifically, for reagent solutions:
 reagent solutions should be stored under conditions appropriate for their chemical properties;
 reagent solutions should be stored under the prescribed storage conditions: temperature,
protected from light, protected from moisture;
 reagent solutions should be stored in tightly closed containers;
 containers must be closed immediately after use;
 alkaline solutions should be stored in plastic containers;
 flammable liquids should only be stored in small quantities.
5.1. Examples of annotations for declared storage conditions
 protect from light
 protect from moisture
 use tightly sealed containers (e.g. bottles with lip seals)
 laboratory cabinets
 refrigerator (2 – 8 °C)
 freezer (below – 15 °C)
 under vacuum
 liquid nitrogen storage
6. Conclusions
In this document, general common practices – mainly derived from published literature and real-life experience
– are set out to help OMCL Network members define expiry periods for commercial reagents, ready-to-use
solutions and in-house-prepared reagents.
As shown in Annexes 1 and 2, different expiry periods can be set for the same type of reagent or in-house-
prepared solution depending on their characteristics and purpose.
Most importantly, when setting the expiry date, traceability and quality must be ensured in order to avoid
compromising the final results of the analytical process, meaning the worst-case scenario should be assumed
when setting the expiry period of a reagent after opening.
Under the worst-case scenario, it should not be presumed that the opened reagent can be used after the
expiry date of the unopened reagent unless valid scientific and experimental analytical evidence of its stability
has been provided and a valid risk assessment has been performed.
For in-house-prepared reagent solutions, the expiry period is set mainly taking into account the intended use
of the solution, the purity of the starting reagent, scientific data and any validation data that has been properly
documented and approved.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 8/15
Periodic and pre-use quality checks should be conducted regardless of the previously set expiry period both
for opened reagents and for in-house-prepared reagent solutions.
As there are few relevant studies demonstrating stability after opening or stability after preparation of in-
house reagents, caution is required when setting the expiry period as it may compromise the quality of the
final analytical result.
7. References
1. Nogueira JMF, Serôdio P. Determination of the expiration date of chemical solutions. Accred Qual Assur
2003;8:231-4.
2. Barwick V, editor. Eurachem/CITAC Guide: Guide to quality in analytical chemistry: an aid to
accreditation. 3rd edition. 2016. ISBN 978-0-948926-32-7. Available at: https://www.eurachem.org/
3. GL Management of Reagents PA/PH/OMCL 11 (157), EDQM
4. Clontz, L. Quality control systems for the microbiology laboratory. The key to successful inspections. PDA,
Baltimore, Maryland, USA; Davis Horwood International Publishing, Godalming, Surrey, UK, 2000.
5. Bartel, D. Use and storage of chemicals, reagents, reagent solutions, samples, primary and secondary
standards, and laboratory calibration reagents “The PEI approach” Paul-Ehrlich-Institut.
6. DIN/ISO standards (e.g. DIN ISO EN 11732: Water quality: Determination of ammonium nitrogen
method by flow analysis.)
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 9/15
Annex 1.
Suggested expiry periods for reagents after opening or for ready-to-use reagent solutions in the absence of
manufacturers' instructions (e.g. certificates) or official regulations (DIN/ISO standardspharmacopoeial texts) or
literature data.
Reagent / Reagent
solution
Expiry period after
opening
Exceptions Risk analysis
Acid and alkaline solutions
– inorganic, organic
up to 36 months Shorter expiry period of max.
12 months is recommended
for supra-pure and max. 6
months for ultra-pure
solutions after opening
Intake of CO2 or fuming of gases
changing concentration
Ammonia solution up to 6 months Depending on the purity:
supra- and ultra-pure – max.
1 month
Concentration of ammonia
decreases with time
Aqueous salt solutions up to 6 months / Degradation unlikely
Buffer solutions up to 12 months Phosphate buffer – 3
months; ready-to-use buffer
solutions – manufacturer's
CoA and discard if turbid
Degradation possible after
longer storage, especially for
buffers with pH > 6;
Microbiological contamination
Hydrogen peroxide up to 6 months Depending on the purity:
supra- and ultra-pure – max.
1 month
Concentration of hydrogen
peroxide decreases with time
Indicators – organic,
inorganic
up to 36 months
/
Visual examination before use;
avoid oxidants
Ion-pairs – inorganic,
organic
up to 60 months Sodium lauryl sulfonate –
max. 12 months
Possible disintegration,
check with SST
Inorganic reagents –
solids, salts (not
hygroscopic and light-
degradable salts and
oxides)
from 12 to 36 months Shorter expiry period of max.
24 months is recommended
for supra-pure and max. 12
months for ultra-pure solids
after opening
Disintegration unlikely (salts,
oxides);
Contamination and/or cross-
contamination for supra- and
ultra-pure reagents
Inorganic reagents –
solids, salts (hygroscopic
and light-degradable salts
and oxides)
Should not exceed expiry
date of the unopened
containers
Shorter expiry period of max
12 months is recommended
for supra-pure and ultra-pure
solids after opening
Degradation
Liquid media
Culture media
Discard if turbid;
shelf life: up to 6 months
Note special preparations Degradation and sweating
possible, depends greatly on the
chemical composition of the
media
Organic reagents – liquids up to 24 months
/
Degradation possible only after
longer storage
Organic reagents – solids,
acids, bases
up to 36 months Formaldehyde,
Glutaraldehyde,
Trifluoroacetic acid – expiry
period max. 12 months
Degradation and polymerisation
possible
Organic reagents – solids,
salts (hygroscopic and
light-degradable salts and
oxides)
Should not exceed expiry
date of the unopened
containers
Shorter expiry period of max.
12 months is recommended
for supra- and ultra-pure
solids after opening
Disintegration
Organic reagents – -solids,
salts (not hygroscopic and
light-degradable salts and
oxides)
from 12 to 36 months Shorter expiry period of max.
24 months is recommended
for supra-pure and max. 12
months for ultra-pure solids
after opening
Contamination and/or cross-
contamination for supra- and
ultra-pure reagents
Organic solvents from 12 to 36 months Acetone – max. 24 months
and Ethanol – max. 36
months expiry period after
opening; it also depends on
the purity of the reagent: for
sensitive methods like LC-
MS max. 12 months
Degradation or concentration
change due to intake of water
due to humidity
Sera (animal or human
origin)
up to 10 years, stored as
aliquots in air-tight
containers at min. −20°C
Modified sera may have a
shorter shelf life
Sublimation and disintegration or
clotting possible; reduction in
activity possible
REMARKS:
1. Expiry period of the opened container must not exceed the shelf life of the unopened reagent!
2. All proposed periods are based on the experience of OMCLs (exercise performed by OMCLs) and a
document by Detlef Bartel, Paul-Ehrlich-Institut [5].
3. This list is not exhaustive.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 10/15
Annex 2.
Table 1. Suggested expiry periods for in-house-prepared reagent solutions used for physico-chemical tests,
established on the basis of the analytical experience of OMCLs, method validation, guidelines and available
literature [1].
REAGENT SOLUTION WITH AN IMPACT ON A RESULT
REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS
WATER
Water for chromatography (Ph. Eur.
1095503)
(resistivity [25°C] ≥ 18 MΩ/cm)
Water for preparing mobile phases
without buffers
and
Deionised water
(Ph. Eur. 1095508)
(resistivity [25°C] ≥ 18 MΩ/cm)
24 hours after production
High-purity water
rapidly picks up
contaminants
Bacterial
contamination
Visual inspection
Discard if turbid
Use from fresh and
closed containers
SST criteria must be
fulfilled
Water, purified
(Ph. Eur. 04/2018:0008)
(conductivity [25°C] 5.1 µS/cm)
Water for preparing buffers, diluents
(solvents), dissolution media
3-5 days after production
Bacterial
contamination
Water for injections
and
Sterilised water for injections
(Ph. Eur. 04/2017:0169)
(conductivity [25°C] 1.3 µS/cm)
2 months after production
Bacterial
contamination
Visual inspection
Discard if turbid
Use from fresh and
closed containers
MOBILE
PHASES
and
DILUENTS
Pure organic solvent
(aliquot transferred from the original
container)
6 months or original expiry date
(whichever is shorter)
Depletion
Precipitation
Evaporation of organic
phase
Visual inspection
SST criteria must be
fulfilled
Organic solvents + water
(e.g. acetonitrile/water,
methanol/ water)
SST criteria must be
fulfilled
≥ 50% of organic component 6 months
< 50% of organic component 3 months
Organic solvent + buffer or buffers
for HPLC
2 weeks
SST criteria must be
fulfilled
pH check (if applicable)
Organic solvent + acid/base solution
(e.g. phosphoric acid, perchloric
acid, methanesulfonic acid,
trifluoroacetic acid)
3 months
ACID
SOLUTIONS
for pH adjustment 12 months
Degradation
Visual inspection
SST criteria must be
fulfilled
> 0.1 M HCL 7 months [1]
≤ 0.1 M HCL 2 months [1]
> 0.5 M H2SO4 7 months [1]
≤ 0.5 M H2SO4 2 months [1]
Perchloric acid 1 month [1]
ALKALINE
SOLUTIONS
For pH adjustment 6 months
Absorbs CO2
Keep in tightly closed
container
Plastic container only
SST criteria must be
fulfilled
0.2 M - 5 M NaOH 5 months [1]
≤ 0.1 M NaOH 1 month
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 11/15
REAGENT SOLUTION WITH AN IMPACT ON A RESULT
REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS
Ammonia
3 months [1]
SALT
SOLUTIONS
iron(II) and silver salt solutions 6 months
Oxidation
Colour change
Visual inspection
SST criteria must be
fulfilled
2 M (NH4)(CH3COO) 5 months [1]
0.1 M NaCl and 5 ppm NaCl 5 months [1]
Microbial growth
0.1 M Na2SO4 and 10 ppm Na2SO4 5 months [1]
VOLUMETRIC
SOLUTIONS
(*)
COMPLEXOMETRY
0.1 M EDTA 7.5 months [1]
Precipitation
After visual check the
titre must be
determined before use
Plastic container only
0.01 M EDTA(**)
2 months [1]
ARGENTOME
TRY
0.1 M AgNO3 7 months [1]
Photochemical
degradation
After visual check the
titre should be
determined before use
Protect from light
0.01 M AgNO (**)
3 5 months [1]
0.1 M NH4SCN 7 months [1]
REDOXOMETRY
0.1 M I2 7.5 months [1]
Volatility
Air-oxidation
processes
After visual check the
titre must be
determined before use
Keep in tightly closed
container
0.1 M Na2S2O3:
prepare at least 2
weeks before use
0.01 M I (**)
2 5 months [1]
0.1 M Na2S2O3 5 months [1]
0.1 M KMnO4 1 month [1]
After visual check the
titre must be
determined before use
Prepare at least 2
weeks before use
Filter before use
*
For all volumetric solutions, after visual check the titre must be determined before use and the Ph. Eur. criteria for accuracy
must be fulfilled.
**
Diluted solutions should be preferably prepared for immediate and single use.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 12/15
REAGENT SOLUTION WITH AN IMPACT ON A RESULT
REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS
NDICATORS
Ferroin 3 years
Perform the sensitivity
test after preparation
and then periodically
(according to Ph. Eur.)
Visual inspection
Protect from light
Phenolphthalein solution 3 years
Phenol red solution 3 years
Methylene blue 3 years
Methylthymol blue mixture 2 years
Methyl red solution 3 years
BUFFER
SOLUTIONS
Visual inspection
pH or conductivity
calibration/verification buffers
(aliquot transferred from the original
container)
3 months or original expiry date
(whichever is shortest)
Absorbs CO2
depending on pH
Keep in tightly closed
containers
Protect from light
Absorbs CO2
depending on pH Visual inspection
Buffer solutions
(others, e.g. ammonium buffer,
acetate buffer)
2 months
Prone to microbial
growth depending on
pH
Keep in tightly closed
containers
Protect from light
Depletion possible
0.01 M and 0.2 M K2HPO4 1 month and 4 months, respectively [1]
Absorbs CO2
depending on pH
Prone to microbial
growth, depending on
pH
Visual inspection
Keep in tightly closed
containers
Protect from light
0.01 M and 1 M KH2PO4 1 month and 4 months, respectively [1]
0.05 and 0.1 M NaH2PO4 5 months [1]
2 M and 0.03 M (NH4)2HPO4 5 months [1]
10% Na2HPO4 1 day [1]
Very unstable Prepare fresh
SDS buffers or
other surfactants
up to 6 months Precipitation
Visual inspection
[0.2% m/v SDS
solution is particularly
prone to precipitation
depending on the
quality of the starting
reagent]
UREA
Urea solution 1 day
Prepare fresh before
use
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 13/15
REAGENT SOLUTION WITHOUT AN IMPACT ON A RESULT
REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS
Organic solvents and water used as supplementary solutions (e.g. for flushing columns, cleaning HPLC system)
≥ 50% of organic component 1 year
Precipitation
Visual inspection,
discard if turbid
< 50% of organic component 6 months
Table 2. Suggested expiry periods for in-house solutions, buffers and media used for microbiological tests or
cell cultivation. Established on the basis of analytical experience of OMCLs and available literature [4].
FOR MICROBIOLOGICAL AND CELL-BASED METHODS
REAGENT SOLUTION MAX. EXPIRY PERIOD RECOMMENDATION
Solid or liquid growth media stored in tightly
sealed containers (e.g. screw capped flasks)
6 months*
Below 25°C or 2-8°C depending on the
medium
Solid or liquid growth media stored in unsealed
containers (e.g. bottles with cellulose plug and
aluminium foil)
1 month
BUFFER SOLUTION STORED IN TIGHT CONTAINERS
Solution of 0.2 M potassium dihydrogen
phosphate 6 months Below 25°C
Solution of 0.2 M sodium hydroxide
Buffers prepared from buffer solutions** 1 month Below 25°C
Glucose 6-phosphate solution in various
concentrations 1 year ≤ − 15°C
Nicotinamide adenine dinucleotide phosphate
(NADP) in various concentrations
1 year
≤ − 15°C
Agarose 5% 3 months
RESEARCH ON ANTISEPTICS AND DISINFECTANTS
REAGENT SOLUTION MAX. EXPIRY PERIOD RISK FACTORS/RECOMMENDATION
Neutraliser stored in tightly sealed containers 6 months
Below 25°C or 2-8°C depending on the
composition
Diluent stored in tightly sealed containers 6 months Below 25°C
INTERFERING SUBSTANCE
Bovine albumin 3 g/100 mL 1 month 2-8°C
Bovine albumin + sheep erythrocytes 3 g/100 mL
+ 3 mL/100 mL
1 week
2-8°C
SOLUTIONS TO BE PREPARED IN HARD WATER
Solution A (magnesium chloride and calcium
chloride)
1 month 2-8°C
Solution B (sodium bicarbonate) 1 week 2-8°C
* Growth promotion tests must be performed within three months of the time of use.
** Prepared from sterile solution.
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 14/15
Annex 3. Risk analysis
A list of questions supporting the risk analysis of a decision on expiry date extension (can be extended
depending on the intended use) of in-house-prepared reagent solutions (open catalogue):
1. Is the reagent solution used:
a. for quantitative purposes?
b. for qualitative purposes?
c. as an auxiliary reagent?
2. Is the reagent prone to:
a. oxidation?
b. hydrolysis?
c. photo- or other degradation?
d. evaporation?
e. pH change (e.g. as a result of CO2 absorption)?
f. microbial growth?
3. Is the reagent solution stored properly?
4. Is the buffering capacity/ion strength important for the purpose of use?
5. Is the pH of the solution important for the purpose of use?
6. Is the exact concentration of the solution important for the purpose of use?
7. Was the reagent solution prepared in a sterile manner?
8. Is the reagent solution used in a way that preserves sterility?
9. Is sterility important for the intended use?
10. Can the reagent solution attributes be easily verified/controlled before use (e.g. pH check,
osmolality, absorbance)?
11. Will control tests be performed during the use of the reagent solution that will allow its suitability to
be verified?
12. Are the SST criteria fulfilled while the reagent solution is used?
13. Purity of the reagent?
PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network
p. 15/15
Figure 1. Graphical representation of the decision-making process for determining the expiry period of a
reagent after opening.

Weitere ähnliche Inhalte

Ähnlich wie Recommendations on setting the expiry period for commercial and in-house-prepared reagents used in the laboratories of the OMCL Network.docx

Quality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docxQuality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docxmakdul
 
Documentation with respect to release of finished pharmaceutical product
Documentation with respect to release of finished pharmaceutical productDocumentation with respect to release of finished pharmaceutical product
Documentation with respect to release of finished pharmaceutical productMadhuraNewrekar
 
Presentation on-hold-time-study
Presentation on-hold-time-studyPresentation on-hold-time-study
Presentation on-hold-time-studysachin kumar
 
Q1 a(r2) guideline
Q1 a(r2) guidelineQ1 a(r2) guideline
Q1 a(r2) guidelineluckymed
 
Specialty gases: essential for research, production & quality control
Specialty gases: essential for research, production & quality controlSpecialty gases: essential for research, production & quality control
Specialty gases: essential for research, production & quality controlLinde Gas Benelux
 
Anthony crasto scaleup techniques & pilot plant
Anthony crasto scaleup techniques & pilot plantAnthony crasto scaleup techniques & pilot plant
Anthony crasto scaleup techniques & pilot plantAnthony Melvin Crasto Ph.D
 
F 619 03 3b_rjyxoq__
F 619   03   3b_rjyxoq__F 619   03   3b_rjyxoq__
F 619 03 3b_rjyxoq__Luis Guzman
 
OECD Principle Of Good Laboratory Practice (GLP).pptx
OECD Principle Of Good Laboratory Practice (GLP).pptxOECD Principle Of Good Laboratory Practice (GLP).pptx
OECD Principle Of Good Laboratory Practice (GLP).pptxSIRAJUDDIN MOLLA
 
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 

Ähnlich wie Recommendations on setting the expiry period for commercial and in-house-prepared reagents used in the laboratories of the OMCL Network.docx (20)

Q6 b step4
Q6 b step4Q6 b step4
Q6 b step4
 
Glp
GlpGlp
Glp
 
Drug stability
Drug stability Drug stability
Drug stability
 
Quality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docxQuality DocumentationDocumentation OverviewThe Q.docx
Quality DocumentationDocumentation OverviewThe Q.docx
 
Documentation with respect to release of finished pharmaceutical product
Documentation with respect to release of finished pharmaceutical productDocumentation with respect to release of finished pharmaceutical product
Documentation with respect to release of finished pharmaceutical product
 
Presentation on-hold-time-study
Presentation on-hold-time-studyPresentation on-hold-time-study
Presentation on-hold-time-study
 
Glp
GlpGlp
Glp
 
Q1 a(r2) guideline
Q1 a(r2) guidelineQ1 a(r2) guideline
Q1 a(r2) guideline
 
Q1A1. pdf
Q1A1. pdfQ1A1. pdf
Q1A1. pdf
 
Pharmapeda.pdf
Pharmapeda.pdfPharmapeda.pdf
Pharmapeda.pdf
 
Chương 6. Tài liệu EU GMP.pdf
Chương 6. Tài liệu EU GMP.pdfChương 6. Tài liệu EU GMP.pdf
Chương 6. Tài liệu EU GMP.pdf
 
Gmp eu chapter 6
Gmp eu chapter 6Gmp eu chapter 6
Gmp eu chapter 6
 
Specialty gases: essential for research, production & quality control
Specialty gases: essential for research, production & quality controlSpecialty gases: essential for research, production & quality control
Specialty gases: essential for research, production & quality control
 
Anthony crasto scaleup techniques & pilot plant
Anthony crasto scaleup techniques & pilot plantAnthony crasto scaleup techniques & pilot plant
Anthony crasto scaleup techniques & pilot plant
 
F 619 03 3b_rjyxoq__
F 619   03   3b_rjyxoq__F 619   03   3b_rjyxoq__
F 619 03 3b_rjyxoq__
 
S4 KAVANA BB APA-STABILITY STUDIESs.pptx
S4 KAVANA BB APA-STABILITY STUDIESs.pptxS4 KAVANA BB APA-STABILITY STUDIESs.pptx
S4 KAVANA BB APA-STABILITY STUDIESs.pptx
 
OECD Principle Of Good Laboratory Practice (GLP).pptx
OECD Principle Of Good Laboratory Practice (GLP).pptxOECD Principle Of Good Laboratory Practice (GLP).pptx
OECD Principle Of Good Laboratory Practice (GLP).pptx
 
gmp-WHO.pdf
gmp-WHO.pdfgmp-WHO.pdf
gmp-WHO.pdf
 
Yêu cầu cho thực hành tốt sản xuất thuốc (gmp who) – phần 1 (e)
Yêu cầu cho thực hành tốt sản xuất thuốc (gmp who) – phần 1 (e)Yêu cầu cho thực hành tốt sản xuất thuốc (gmp who) – phần 1 (e)
Yêu cầu cho thực hành tốt sản xuất thuốc (gmp who) – phần 1 (e)
 
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...
Phụ lục 5. Hướng dẫn đánh giá độ ổn định vaccine dưới 1 số điều kiện nhiệt độ...
 

Mehr von Công ty Cổ phần Tư vấn Thiết kế GMP EU

Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...
Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...
Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...Công ty Cổ phần Tư vấn Thiết kế GMP EU
 

Mehr von Công ty Cổ phần Tư vấn Thiết kế GMP EU (20)

Quyết định 878/QĐ-QLD 2023 công bố Danh mục thuốc biệt dược gốc Đợt 5
Quyết định 878/QĐ-QLD 2023 công bố Danh mục thuốc biệt dược gốc Đợt 5Quyết định 878/QĐ-QLD 2023 công bố Danh mục thuốc biệt dược gốc Đợt 5
Quyết định 878/QĐ-QLD 2023 công bố Danh mục thuốc biệt dược gốc Đợt 5
 
DANH MỤC DỰ ÁN GMP - ISO.pptx
DANH MỤC DỰ ÁN GMP - ISO.pptxDANH MỤC DỰ ÁN GMP - ISO.pptx
DANH MỤC DỰ ÁN GMP - ISO.pptx
 
ICH Q11 Combine.pdf
ICH Q11 Combine.pdfICH Q11 Combine.pdf
ICH Q11 Combine.pdf
 
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần II.pptx
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần II.pptxThực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần II.pptx
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần II.pptx
 
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần I.pptx
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần I.pptxThực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần I.pptx
Thực hành tốt bảo quản nguyên liệu làm thuốc (GSP) - Phần I.pptx
 
GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS PART I.pdf
GUIDE TO GOOD MANUFACTURING PRACTICE  FOR MEDICINAL PRODUCTS PART I.pdfGUIDE TO GOOD MANUFACTURING PRACTICE  FOR MEDICINAL PRODUCTS PART I.pdf
GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS PART I.pdf
 
Danh sách các cơ sở trong nước đạt tiêu chuẩn nguyên tắc GLP. pdf
Danh sách các cơ sở trong nước đạt tiêu chuẩn nguyên tắc GLP. pdfDanh sách các cơ sở trong nước đạt tiêu chuẩn nguyên tắc GLP. pdf
Danh sách các cơ sở trong nước đạt tiêu chuẩn nguyên tắc GLP. pdf
 
ISPE Good manufacturing practice
ISPE Good manufacturing practiceISPE Good manufacturing practice
ISPE Good manufacturing practice
 
Hồ sơ năng lực GMP EU
Hồ sơ năng lực GMP EUHồ sơ năng lực GMP EU
Hồ sơ năng lực GMP EU
 
Prequalified Active Pharmaceutical Ingredients.pptx
Prequalified Active Pharmaceutical Ingredients.pptxPrequalified Active Pharmaceutical Ingredients.pptx
Prequalified Active Pharmaceutical Ingredients.pptx
 
Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...
Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...
Danh mục 37 thuốc sản xuất trong nước được cấp giấy phép lưu hành tại Việt Na...
 
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...
Danh mục 259 thuốc sản xuất trong nước được cấp giấy đăng ký lưu hành tại Việ...
 
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...
QUYẾT ĐỊNH 438/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC - ĐỢT 2 NĂ...
 
FDA: Practices to Prevent Unsafe Contamunation of Animal Feed from Drug Carvover
FDA: Practices to Prevent Unsafe Contamunation of Animal Feed from Drug CarvoverFDA: Practices to Prevent Unsafe Contamunation of Animal Feed from Drug Carvover
FDA: Practices to Prevent Unsafe Contamunation of Animal Feed from Drug Carvover
 
FDA: Guidance-Labeling Infant Formula-March2023
FDA: Guidance-Labeling Infant Formula-March2023FDA: Guidance-Labeling Infant Formula-March2023
FDA: Guidance-Labeling Infant Formula-March2023
 
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...
QUYẾT ĐỊNH 353/QĐ-QLD: Ban hành danh mục 69 thuốc sản xuất trong nước được cấ...
 
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...
QUYẾT ĐỊNH SỐ 352/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 231 THUỐC SẢN XUẤT TRONG N...
 
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...
QUYẾT ĐỊNH 371/QĐ-QLD VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC BIỆT DƯỢC GỐC ĐỢT 1 - NĂ...
 
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...
QUYẾT ĐỊNH: 370/QĐ-QLD VỀ VIỆC BAN HÀNH DANH MỤC 50 THUỐC NƯỚC NGOÀI ĐƯỢC CẤP...
 
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...
QUYẾT ĐỊNH VỀ VIỆC CÔNG BỐ DANH MỤC THUỐC CÓ CHỨNG MINH TƯƠNG ĐƯƠNG SINH HỌC ...
 

Kürzlich hochgeladen

Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknownarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipurparulsinha
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 

Kürzlich hochgeladen (20)

Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 

Recommendations on setting the expiry period for commercial and in-house-prepared reagents used in the laboratories of the OMCL Network.docx

  • 1. General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (21) 60 R2 Recommendations on setting the expiry period for commercial and in-house-prepared reagents used in the laboratories of the OMCL Network Full document title and reference Recommendations on setting the expiry period for commercial and in-house-prepared reagents used in the laboratories of the OMCL Network, PA/PH/OMCL (21) 60 R2 Document type Recommendation Document Legislative basis Council Directive 2001/83/EC and 2001/82/EC, as amended Date of first adoption 31 October 2021 Date of original entry into force 1 January 2022 Date of entry into force of revised document Previous titles/other references / last valid version N/A
  • 2. Custodian Organisation The present document was elaborated by the OMCL Network / EDQM of the Council of Europe Concerned Network GEON N.B. This OMCL Quality Management System document is applicable to members of the European OMCL Network only. Other laboratories might use the document on a voluntary basis. However, please note that the EDQM cannot treat any questions related to the application of the documents submitted by laboratories other than the OMCLs of the Network.
  • 3. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 2/15 Table of Contents Preface...................................................................................................................................3 1. Purpose...........................................................................................................................3 2. Scope ..............................................................................................................................4 3. Reagents (chemicals).....................................................................................................4 3.1. Definitions ...............................................................................................................4 3.2. Establishing shelf life or an expiry date/period for a reagent (ready-to-use) ...........4 3.2.1. Unopened reagent container ............................................................................4 3.2.2. After opening of the reagent container ............................................................5 3.3. Decision tree for establishing the expiry period after opening of the reagent...........6 4. In-house-prepared reagent solutions ............................................................................6 4.1. Quality check...............................................................................................................6 4.2. Recommended expiry periods ....................................................................................6 4.3. Extension of adopted expiry period............................................................................6 5. Labelling and storage .....................................................................................................7 5.1. Examples of annotations for declared storage conditions..........................................7 6. Conclusions.....................................................................................................................7 7. References......................................................................................................................8 Annex 1..................................................................................................................................9 Annex 2................................................................................................................................10 Annex 3. Risk analysis.........................................................................................................14
  • 4. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 3/15 Preface Reagents are an integral part of any laboratory and as such play a vital role in ensuring the validity of analytical results. Reagents must be suitable for the intended use. Consideration should be given to the accompanying documentation and appropriate physico-chemical or biological critical quality attributes for the intended purpose/s. In the laboratory, two major types of reagents are generally distinguished: commercial (off-the-shelf, ready- to-use) and those prepared in the laboratory (in-house reagents) using the former. These two types of reagents are mainly dealt with separately throughout this document. It should be noted that the principles of FIFO (First In-First Out) or FEFO (First Expired-First Out) should be applied when using reagents. The shelf life for an unopened product or expiry period after first opening is usually set taking into account its stability, i.e. how long the quality attributes and chemical integrity of the reagent can be maintained for the intended analytical application. This can be a challenging task because most manufacturers give the shelf life of the unopened product and do not provide information on how long the reagent is stable once the container has been opened. The recommendations for the shelf life of reagent solutions prepared and used in laboratories and presented in this document were established based on scientific data (including validation data obtained from quality control charts, guidelines, standards, pharmacopoeias, publications, etc.) and on the analytical experience and knowledge of OMCLs. Preparation times, costs, sustainable waste management policies and environmental protection concerns were also considered during the preparation of these recommendations. This document is based on the laboratory experience of the OMCL Network. Its purpose is to help maintain the quality of the work carried out in the laboratory on a daily basis and ensure that the results obtained are valid. The recommendations on the shelf life of reagent solutions provided herein should allow laboratories to assess whether a given solution can be used safely and reliably (provided that the storage conditions comply with requirements) as they take into account various factors that contribute to the degradation of some reagents (temperature, air exposure, moisture, etc.). For the purpose of this document, the expiry period of a reagent solution should be understood as the period of time during which the reagent is stable and can be used for a given type of analysis. It should be noted that the same reagent may be unfit for use for some purposes because it has expired, but still suitable for other purposes. Purchase or preparation of reagents and reagent solutions in quantities (volumes) not adapted to current needs should be avoided, as unnecessary disposal that does not take into account the possibility of prolonged use potentially constitutes a direct threat to the environment. Hence, if prolonging the use of a given reagent solution for specific purposes is safe and does not compromise the results, such prolongation should always be considered and justified. 1. Purpose The aim of this document is to describe and, where possible, define how to set expiry periods for reagents (chemicals) used in the OMCL Network (and to guide OMCLs in applying them). It is also intended to provide guidance on regulating the use, labelling, storage and expiry periods of reagent solutions prepared in OMCLs (in-house solutions) for qualitative, semi-quantitative and quantitative purposes.
  • 5. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 4/15 2. Scope Use of commercial reagents (chemicals) and reagent solutions prepared in OMCLs within the OMCL Network. Commercial reagents or reagent preparations for immediate use are out of the scope of this document as they are not stored and are used within the same day. In addition, solutions for which there is a certificate by which the traceability to NIST or any other primary etalon can be established are also out of the scope of this document. 3. Reagents (chemicals) 3.1. Definitions Reagents Reagents are solids, gases, liquids or solutions of pure chemical or biological substances, organic solvents and water. Quality criteria Quality criteria are assumed to be covered by appropriate internal standard operating procedures (SOPs). In general, special consideration should be given to the selection, purchase and receipt of reagents and their storage. The emphasis is on the crucial step of purchasing reagents of an expected quality and, upon receipt, on inspection and verification of compliance with the previously set specifications. These steps precede the labelling and storage of the reagents. Shelf life of the reagent is a set period or a date given by the manufacturer, which can usually be found on the certificate of analysis (CoA) for that particular reagent. During this period or until that date, it is expected that the reagent will meet the quality specifications found in the CoA if the storage conditions in the material safety data sheet (MSDS) or in any other relevant document, like CoA or label, are respected. Reagent without a shelf life or a retest date Sometimes a shelf life is not defined by the manufacturer because the reagent is considered to have an extensive shelf life when stored unopened under the proper conditions, or the manufacturer has no supporting stability data. As the quality of a reagent may be altered due to use (effect of environmental conditions), setting an expiry date is highly recommended. This can be done, for example, by carrying out a risk assessment. Recommended retest date for an unopened reagent If it is possible to obtain a new certificate from the manufacturer, the shelf life of an unopened reagent can be extended until a later date. Expiry period is a period or a specific date after opening of the reagent during which it is expected to maintain the appropriate physico-chemical or biological critical quality attributes for the intended purpose, in spite of any physical and chemical changes that may occur. The MSDS should be used as a guide on how to store the reagent after opening. Extension of the expiry period for a reagent The expiry period can be extended following a retest taking into account the critical quality attributes contributing to the functional characteristics of the reagent, checked by appropriate means or by scientific judgement. 3.2. Establishing shelf life or an expiry date/period for a reagent (ready-to-use) 3.2.1. Unopened reagent container Usually a reagent is delivered with a stated shelf life for the unopened container.
  • 6. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 5/15 Where no information is available from the manufacturer, the shelf life for unopened reagent containers can be taken to be the date of receipt plus up to 5 years. This is only applicable if the unopened reagent container is stored under specified conditions found in relevant documents. 3.2.2. After opening of the reagent container It is necessary to set a rational expiry period in the absence of analytical data on the stability of the product after opening. This could be achieved by “real-time monitoring” of the reagent after opening by gathering experimental analytical data on critical parameters, both chemical and physical, during laboratory use. In addition, possible sources of information on the stability of a particular reagent are:  the CoA or MSDS for the storage conditions and possible instability;  scientific papers;  chemical knowledge about the nature of the reagents;  stability data – additional information from the manufacturer or supplier (e.g. certificates, which can be found on manufacturers or suppliers website), guidelines, DIN/ISO Standards, pharmacopoeial texts, etc. The expiry period after opening should not exceed the shelf life of the unopened reagent container. If it does, a risk assessment should be done based on the scientific literature and experimental data to show that use of the reagent after the end of its shelf life does not compromise its safety or the quality of the measurement results. Each time a container is opened, the reagent is exposed to various external factors and environmental conditions that may alter its integrity and affect its quality, which may compromise the validity of the results. Reagent properties, storage conditions, handling and the intended use that have to be considered during the process of establishing the expiry period after opening are:  Purity of the reagent – this property is very important for establishing the expiry period after opening. In some cases the higher the purity of the reagent, the shorter the expiry period after opening. For example, sodium chloride p.a. will have a longer expiry period after opening than supra-pure sodium chloride, because it is rational to expect that supra-pure solids will deteriorate more quickly and lose their quality attributes. However, the degree of purity needed for the testing should be commensurate with the intended use of the reagent: it is important to select the quality grade that will yield satisfactory results.  Temperature – should be kept controlled within the range given by the manufacturer/stated in the MSDS and/or in internal SOPs.  Humidity – should meet the requirements given by the manufacturer/stated in the MSDS and/or in internal SOPs.  Light exposure.  Contamination during use of the reagent – each time a reagent is opened, accidental contamination is possible (switching of bottle caps, placing bottle caps on soiled surfaces, transferring reagents to other containers, use of pipettes for withdrawal, etc.) It is important to state that, regardless of the expiry period established for the opened reagent, a visual inspection/examination should be done before use and if any of the physical properties that can be determined by visual inspection have changed, the reagent should be discarded. Suggested expiry periods for reagents after opening or for ready-to-use reagent solutions in the absence of manufacturers' instructions (e.g. certificates) or official regulations (DIN/ISO Standards, pharmacopoeial texts) or literature data are presented in Annex 1.
  • 7. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 6/15 3.3. Decision tree for establishing the expiry period after opening of the reagent The decision tree presented in Annex 3 demonstrates the decision-making process on how to approach setting the expiry period after opening. 4. In-house-prepared reagent solutions Preparation of reagent solutions is carried out in accordance with applicable guidelines, compendial methods (e.g. pharmacopoeial texts), test method specifications and procedures (Common Technical Documents [CTDs], SOPs). 4.1. Quality check It is necessary to ensure that a reagent solution has the properties required for the intended analysis. Before using prepared reagent solutions, it is strongly recommended to:  perform a visual inspection – any reagent should be discarded if there is a change in colour or opalescence, or if there is any precipitation (unless it is necessary),  perform a pH check before the use of buffer solutions (if applicable),  determine the titre of volumetric solutions (if applicable),  perform a Ph. Eur. sensitivity test for indicators. Perform any other suitable qualification test specific to the prepared reagents. The environmental aspect must always be taken into account when deciding to shorten the expiry period and/or dispose of a reagent solution. A solution that is no longer suitable for a given analytical procedure may still be suitable for another purpose. 4.2. Recommended expiry periods Consideration should always be given to expiry periods:  for reagent solutions described in compendial methods (e.g. 2 months for Primary opalescent suspension in Ph. Eur. chapter 2.2.1);  provided in drug manufacturers' documentation;  for reagent solutions for immediate use given in the test method specification, (e.g. data from the marketing authorisation dossiers). If no stability data is available, it is recommended to use the expiry periods provided in Annex 2. The expiry periods presented in Annexes 1 and 2 were established on the basis of the analytical experience of OMCLs, method validation, guidelines, scientific publications [1, 4] and chemical knowledge. OMCLs can also perform an in-house validation study to determine the initial expiry date. 4.3. Extension of adopted expiry period It may be acceptable to extend the expiry periods provided in this document under the following conditions:  risk analysis has been performed (see Annex 3); and/or  test criteria are established, supporting the extension;  an in-house validation study to determine the expiry period has been performed.
  • 8. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 7/15 The extension of the expiry period and the test criteria should be documented according to internal OMCL procedures. 5. Labelling and storage General information on labelling and storage requirements for reagents is given in GL Management of Reagents, PA/PH/OMCL (11) 157 [3]. Whenever possible, reagent labels should state the expiry period for an opened container. More specifically, for reagent solutions:  reagent solutions should be stored under conditions appropriate for their chemical properties;  reagent solutions should be stored under the prescribed storage conditions: temperature, protected from light, protected from moisture;  reagent solutions should be stored in tightly closed containers;  containers must be closed immediately after use;  alkaline solutions should be stored in plastic containers;  flammable liquids should only be stored in small quantities. 5.1. Examples of annotations for declared storage conditions  protect from light  protect from moisture  use tightly sealed containers (e.g. bottles with lip seals)  laboratory cabinets  refrigerator (2 – 8 °C)  freezer (below – 15 °C)  under vacuum  liquid nitrogen storage 6. Conclusions In this document, general common practices – mainly derived from published literature and real-life experience – are set out to help OMCL Network members define expiry periods for commercial reagents, ready-to-use solutions and in-house-prepared reagents. As shown in Annexes 1 and 2, different expiry periods can be set for the same type of reagent or in-house- prepared solution depending on their characteristics and purpose. Most importantly, when setting the expiry date, traceability and quality must be ensured in order to avoid compromising the final results of the analytical process, meaning the worst-case scenario should be assumed when setting the expiry period of a reagent after opening. Under the worst-case scenario, it should not be presumed that the opened reagent can be used after the expiry date of the unopened reagent unless valid scientific and experimental analytical evidence of its stability has been provided and a valid risk assessment has been performed. For in-house-prepared reagent solutions, the expiry period is set mainly taking into account the intended use of the solution, the purity of the starting reagent, scientific data and any validation data that has been properly documented and approved.
  • 9. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 8/15 Periodic and pre-use quality checks should be conducted regardless of the previously set expiry period both for opened reagents and for in-house-prepared reagent solutions. As there are few relevant studies demonstrating stability after opening or stability after preparation of in- house reagents, caution is required when setting the expiry period as it may compromise the quality of the final analytical result. 7. References 1. Nogueira JMF, Serôdio P. Determination of the expiration date of chemical solutions. Accred Qual Assur 2003;8:231-4. 2. Barwick V, editor. Eurachem/CITAC Guide: Guide to quality in analytical chemistry: an aid to accreditation. 3rd edition. 2016. ISBN 978-0-948926-32-7. Available at: https://www.eurachem.org/ 3. GL Management of Reagents PA/PH/OMCL 11 (157), EDQM 4. Clontz, L. Quality control systems for the microbiology laboratory. The key to successful inspections. PDA, Baltimore, Maryland, USA; Davis Horwood International Publishing, Godalming, Surrey, UK, 2000. 5. Bartel, D. Use and storage of chemicals, reagents, reagent solutions, samples, primary and secondary standards, and laboratory calibration reagents “The PEI approach” Paul-Ehrlich-Institut. 6. DIN/ISO standards (e.g. DIN ISO EN 11732: Water quality: Determination of ammonium nitrogen method by flow analysis.)
  • 10. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 9/15 Annex 1. Suggested expiry periods for reagents after opening or for ready-to-use reagent solutions in the absence of manufacturers' instructions (e.g. certificates) or official regulations (DIN/ISO standardspharmacopoeial texts) or literature data. Reagent / Reagent solution Expiry period after opening Exceptions Risk analysis Acid and alkaline solutions – inorganic, organic up to 36 months Shorter expiry period of max. 12 months is recommended for supra-pure and max. 6 months for ultra-pure solutions after opening Intake of CO2 or fuming of gases changing concentration Ammonia solution up to 6 months Depending on the purity: supra- and ultra-pure – max. 1 month Concentration of ammonia decreases with time Aqueous salt solutions up to 6 months / Degradation unlikely Buffer solutions up to 12 months Phosphate buffer – 3 months; ready-to-use buffer solutions – manufacturer's CoA and discard if turbid Degradation possible after longer storage, especially for buffers with pH > 6; Microbiological contamination Hydrogen peroxide up to 6 months Depending on the purity: supra- and ultra-pure – max. 1 month Concentration of hydrogen peroxide decreases with time Indicators – organic, inorganic up to 36 months / Visual examination before use; avoid oxidants Ion-pairs – inorganic, organic up to 60 months Sodium lauryl sulfonate – max. 12 months Possible disintegration, check with SST Inorganic reagents – solids, salts (not hygroscopic and light- degradable salts and oxides) from 12 to 36 months Shorter expiry period of max. 24 months is recommended for supra-pure and max. 12 months for ultra-pure solids after opening Disintegration unlikely (salts, oxides); Contamination and/or cross- contamination for supra- and ultra-pure reagents Inorganic reagents – solids, salts (hygroscopic and light-degradable salts and oxides) Should not exceed expiry date of the unopened containers Shorter expiry period of max 12 months is recommended for supra-pure and ultra-pure solids after opening Degradation Liquid media Culture media Discard if turbid; shelf life: up to 6 months Note special preparations Degradation and sweating possible, depends greatly on the chemical composition of the media Organic reagents – liquids up to 24 months / Degradation possible only after longer storage Organic reagents – solids, acids, bases up to 36 months Formaldehyde, Glutaraldehyde, Trifluoroacetic acid – expiry period max. 12 months Degradation and polymerisation possible Organic reagents – solids, salts (hygroscopic and light-degradable salts and oxides) Should not exceed expiry date of the unopened containers Shorter expiry period of max. 12 months is recommended for supra- and ultra-pure solids after opening Disintegration Organic reagents – -solids, salts (not hygroscopic and light-degradable salts and oxides) from 12 to 36 months Shorter expiry period of max. 24 months is recommended for supra-pure and max. 12 months for ultra-pure solids after opening Contamination and/or cross- contamination for supra- and ultra-pure reagents Organic solvents from 12 to 36 months Acetone – max. 24 months and Ethanol – max. 36 months expiry period after opening; it also depends on the purity of the reagent: for sensitive methods like LC- MS max. 12 months Degradation or concentration change due to intake of water due to humidity Sera (animal or human origin) up to 10 years, stored as aliquots in air-tight containers at min. −20°C Modified sera may have a shorter shelf life Sublimation and disintegration or clotting possible; reduction in activity possible REMARKS: 1. Expiry period of the opened container must not exceed the shelf life of the unopened reagent! 2. All proposed periods are based on the experience of OMCLs (exercise performed by OMCLs) and a document by Detlef Bartel, Paul-Ehrlich-Institut [5]. 3. This list is not exhaustive.
  • 11. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 10/15 Annex 2. Table 1. Suggested expiry periods for in-house-prepared reagent solutions used for physico-chemical tests, established on the basis of the analytical experience of OMCLs, method validation, guidelines and available literature [1]. REAGENT SOLUTION WITH AN IMPACT ON A RESULT REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS WATER Water for chromatography (Ph. Eur. 1095503) (resistivity [25°C] ≥ 18 MΩ/cm) Water for preparing mobile phases without buffers and Deionised water (Ph. Eur. 1095508) (resistivity [25°C] ≥ 18 MΩ/cm) 24 hours after production High-purity water rapidly picks up contaminants Bacterial contamination Visual inspection Discard if turbid Use from fresh and closed containers SST criteria must be fulfilled Water, purified (Ph. Eur. 04/2018:0008) (conductivity [25°C] 5.1 µS/cm) Water for preparing buffers, diluents (solvents), dissolution media 3-5 days after production Bacterial contamination Water for injections and Sterilised water for injections (Ph. Eur. 04/2017:0169) (conductivity [25°C] 1.3 µS/cm) 2 months after production Bacterial contamination Visual inspection Discard if turbid Use from fresh and closed containers MOBILE PHASES and DILUENTS Pure organic solvent (aliquot transferred from the original container) 6 months or original expiry date (whichever is shorter) Depletion Precipitation Evaporation of organic phase Visual inspection SST criteria must be fulfilled Organic solvents + water (e.g. acetonitrile/water, methanol/ water) SST criteria must be fulfilled ≥ 50% of organic component 6 months < 50% of organic component 3 months Organic solvent + buffer or buffers for HPLC 2 weeks SST criteria must be fulfilled pH check (if applicable) Organic solvent + acid/base solution (e.g. phosphoric acid, perchloric acid, methanesulfonic acid, trifluoroacetic acid) 3 months ACID SOLUTIONS for pH adjustment 12 months Degradation Visual inspection SST criteria must be fulfilled > 0.1 M HCL 7 months [1] ≤ 0.1 M HCL 2 months [1] > 0.5 M H2SO4 7 months [1] ≤ 0.5 M H2SO4 2 months [1] Perchloric acid 1 month [1] ALKALINE SOLUTIONS For pH adjustment 6 months Absorbs CO2 Keep in tightly closed container Plastic container only SST criteria must be fulfilled 0.2 M - 5 M NaOH 5 months [1] ≤ 0.1 M NaOH 1 month
  • 12. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 11/15 REAGENT SOLUTION WITH AN IMPACT ON A RESULT REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS Ammonia 3 months [1] SALT SOLUTIONS iron(II) and silver salt solutions 6 months Oxidation Colour change Visual inspection SST criteria must be fulfilled 2 M (NH4)(CH3COO) 5 months [1] 0.1 M NaCl and 5 ppm NaCl 5 months [1] Microbial growth 0.1 M Na2SO4 and 10 ppm Na2SO4 5 months [1] VOLUMETRIC SOLUTIONS (*) COMPLEXOMETRY 0.1 M EDTA 7.5 months [1] Precipitation After visual check the titre must be determined before use Plastic container only 0.01 M EDTA(**) 2 months [1] ARGENTOME TRY 0.1 M AgNO3 7 months [1] Photochemical degradation After visual check the titre should be determined before use Protect from light 0.01 M AgNO (**) 3 5 months [1] 0.1 M NH4SCN 7 months [1] REDOXOMETRY 0.1 M I2 7.5 months [1] Volatility Air-oxidation processes After visual check the titre must be determined before use Keep in tightly closed container 0.1 M Na2S2O3: prepare at least 2 weeks before use 0.01 M I (**) 2 5 months [1] 0.1 M Na2S2O3 5 months [1] 0.1 M KMnO4 1 month [1] After visual check the titre must be determined before use Prepare at least 2 weeks before use Filter before use * For all volumetric solutions, after visual check the titre must be determined before use and the Ph. Eur. criteria for accuracy must be fulfilled. ** Diluted solutions should be preferably prepared for immediate and single use.
  • 13. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 12/15 REAGENT SOLUTION WITH AN IMPACT ON A RESULT REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS NDICATORS Ferroin 3 years Perform the sensitivity test after preparation and then periodically (according to Ph. Eur.) Visual inspection Protect from light Phenolphthalein solution 3 years Phenol red solution 3 years Methylene blue 3 years Methylthymol blue mixture 2 years Methyl red solution 3 years BUFFER SOLUTIONS Visual inspection pH or conductivity calibration/verification buffers (aliquot transferred from the original container) 3 months or original expiry date (whichever is shortest) Absorbs CO2 depending on pH Keep in tightly closed containers Protect from light Absorbs CO2 depending on pH Visual inspection Buffer solutions (others, e.g. ammonium buffer, acetate buffer) 2 months Prone to microbial growth depending on pH Keep in tightly closed containers Protect from light Depletion possible 0.01 M and 0.2 M K2HPO4 1 month and 4 months, respectively [1] Absorbs CO2 depending on pH Prone to microbial growth, depending on pH Visual inspection Keep in tightly closed containers Protect from light 0.01 M and 1 M KH2PO4 1 month and 4 months, respectively [1] 0.05 and 0.1 M NaH2PO4 5 months [1] 2 M and 0.03 M (NH4)2HPO4 5 months [1] 10% Na2HPO4 1 day [1] Very unstable Prepare fresh SDS buffers or other surfactants up to 6 months Precipitation Visual inspection [0.2% m/v SDS solution is particularly prone to precipitation depending on the quality of the starting reagent] UREA Urea solution 1 day Prepare fresh before use
  • 14. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 13/15 REAGENT SOLUTION WITHOUT AN IMPACT ON A RESULT REAGENT SOLUTION MAXIMUM EXPIRY PERIOD RISK FACTORS RECOMMENDATIONS Organic solvents and water used as supplementary solutions (e.g. for flushing columns, cleaning HPLC system) ≥ 50% of organic component 1 year Precipitation Visual inspection, discard if turbid < 50% of organic component 6 months Table 2. Suggested expiry periods for in-house solutions, buffers and media used for microbiological tests or cell cultivation. Established on the basis of analytical experience of OMCLs and available literature [4]. FOR MICROBIOLOGICAL AND CELL-BASED METHODS REAGENT SOLUTION MAX. EXPIRY PERIOD RECOMMENDATION Solid or liquid growth media stored in tightly sealed containers (e.g. screw capped flasks) 6 months* Below 25°C or 2-8°C depending on the medium Solid or liquid growth media stored in unsealed containers (e.g. bottles with cellulose plug and aluminium foil) 1 month BUFFER SOLUTION STORED IN TIGHT CONTAINERS Solution of 0.2 M potassium dihydrogen phosphate 6 months Below 25°C Solution of 0.2 M sodium hydroxide Buffers prepared from buffer solutions** 1 month Below 25°C Glucose 6-phosphate solution in various concentrations 1 year ≤ − 15°C Nicotinamide adenine dinucleotide phosphate (NADP) in various concentrations 1 year ≤ − 15°C Agarose 5% 3 months RESEARCH ON ANTISEPTICS AND DISINFECTANTS REAGENT SOLUTION MAX. EXPIRY PERIOD RISK FACTORS/RECOMMENDATION Neutraliser stored in tightly sealed containers 6 months Below 25°C or 2-8°C depending on the composition Diluent stored in tightly sealed containers 6 months Below 25°C INTERFERING SUBSTANCE Bovine albumin 3 g/100 mL 1 month 2-8°C Bovine albumin + sheep erythrocytes 3 g/100 mL + 3 mL/100 mL 1 week 2-8°C SOLUTIONS TO BE PREPARED IN HARD WATER Solution A (magnesium chloride and calcium chloride) 1 month 2-8°C Solution B (sodium bicarbonate) 1 week 2-8°C * Growth promotion tests must be performed within three months of the time of use. ** Prepared from sterile solution.
  • 15. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 14/15 Annex 3. Risk analysis A list of questions supporting the risk analysis of a decision on expiry date extension (can be extended depending on the intended use) of in-house-prepared reagent solutions (open catalogue): 1. Is the reagent solution used: a. for quantitative purposes? b. for qualitative purposes? c. as an auxiliary reagent? 2. Is the reagent prone to: a. oxidation? b. hydrolysis? c. photo- or other degradation? d. evaporation? e. pH change (e.g. as a result of CO2 absorption)? f. microbial growth? 3. Is the reagent solution stored properly? 4. Is the buffering capacity/ion strength important for the purpose of use? 5. Is the pH of the solution important for the purpose of use? 6. Is the exact concentration of the solution important for the purpose of use? 7. Was the reagent solution prepared in a sterile manner? 8. Is the reagent solution used in a way that preserves sterility? 9. Is sterility important for the intended use? 10. Can the reagent solution attributes be easily verified/controlled before use (e.g. pH check, osmolality, absorbance)? 11. Will control tests be performed during the use of the reagent solution that will allow its suitability to be verified? 12. Are the SST criteria fulfilled while the reagent solution is used? 13. Purity of the reagent?
  • 16. PA/PH/OMCL (21) 60 R2 - Recommendations on setting the expiry period for reagents used in the laboratories of the OMCL Network p. 15/15 Figure 1. Graphical representation of the decision-making process for determining the expiry period of a reagent after opening.