ICD System
Troubleshooting
Word meaning
• ICD
Word meaning
• ICD
– Intracardiac
Word meaning
• ICD
– Intracardiac cardioverter
Word meaning
• ICD
– Intracardiac cardioverter defibrillator
• ICD brief history
– 1947 First human internal defibrillation
– 1956 First human external defibrillation
– 1966 First ICD conception
– 1969 First external canine prototype tested
– 1970 First implantable prototype (895 g)
– 1975 First implantable Defib in canines (250 g)
– 1980 First human implant @ Johns Hopkins – Mirowski et al
– 1985 ICD market released (350 units) – (FDA approval)
ICD: Conception 1966
Dr Mirowski
ICD: Conception 1966
Over-the-counter electronic components
ICD: First Human Implant 1980
Dr Winkle, Dr Mirowski, the first ICD patient, Dr Heilman
ICD: First Human Implant 1980
A transvenous lead in SVC and a ventricular patch lead
Evolution of ICD Therapy
Word meaning
• Troubleshooting
Word meaning
• Troubleshooting
Word meaning
• Troubleshooting
ICD FUNCTIONNG
Sensing Detection Therapy
ICD treats ventricular tachyarrhythmias
SENSING IN ICD
Sensing in ICD: A major challenge…
Sensing - process by which an ICD determines the timing of each
atrial or ventricular depolarization from electrogram signals.
A wide variation in the size of signals
• stable and large normal ventricular signals
• low amplitude VF signals
ICD sensing and EGMs
Sensing auto-adjust
Beat-to-beat auto-adjusting sensitivity
Sensing auto-adjust
Beat-to-beat auto-adjusting sensitivity
Sensing auto-adjust
Sinus rhythm – To avoid oversensing T waves
VF – To avoid undersensing
Sensing auto-adjust
Sensing auto-adjust
Detection
Algorithm by which ICD processes sensed signals to classify the
rhythm and determine if therapy should be delivered.
Up to 3 programmable rate detection zones to permit zone-
specific therapies for slower VT, faster VT, and ventricular
fibrillation (VF).
The minimum duration of tachycardia required for detection is
programmable, either in seconds or in number of ventricular
intervals required for detection.
The specific method used to count ventricular intervals
influences the sensitivity and specificity of VT detection
SVT-VT discriminators
Programmable subset of the VT/VF detection algorithm that
withholds ventricular therapy for SVT to improve specificity.
They usually differ from SVT detection algorithms used to mode-
switch during bradycardia pacing or to deliver atrial therapy
for atrial fibrillation or atrial flutter.
Confirmation or Reconfirmation
Brief process by which ICDs determine whether to deliver or
abort a shock after the high-voltage capacitor is charged.
Redetection
Process by which ICDs determine whether VT or VF detection
criteria remain satisfied after therapy is delivered.
Termination
Termination of a VT or VF episode occurs when the ICD
reclassifies the rhythm as sinus after VT or VF has been
detected.
DETECTION
Detection
Ventricular Fibrillation (VF)
• Rate cut-off (VF zone)
• Number of intervals for detection
(usually consecutive intervals not required; high sensitivity)
Detection
Ventricular Tachycardia (VT)
• Rate cut-off [VT zone(s)]
• Number of intervals for detection
• Detection Enhancement/SVT-VT Discriminators
Detection
Electrogram Morphology
• Current electrogram signal compared with stored normal
template
• Match percentage = 1 - (area of difference)
SVT-VT Discriminators
Stability
• To reduce inappropriate detection of atrial arrhythmias eg
atrial fibrillation (rhythm not “stable” i.e. variable cycle length)
• Each interval compared to previous intervals
• Count as VT if the difference  the programmed stability
interval i.e. VT is “stable”
VT detection interval: 500ms Stability 30ms
SVT-VT Discriminators
Onset
• To reduce inappropriate detection of sinus tachycardia
• Average of current 4 intervals is compared to average of
previous 4 intervals
Current average programmed
Previous average onset percentage
i.e. onset is “sudden” in VT

SVT-VT Discriminators
Onset percentage: 81%
SVT-VT Discriminators
SVT-VT Discriminators
SVT-VT Discriminators
Lead noise discriminator
 Caused by pace-sense conductor fracture, loose
set screw, lead insulation breach.
 Therapy avoided by comparing near-field channel
(RV tip to RV ring) to far-field channel (RV tip to
Can/RV coil to Can)
SVT-VT Discriminators
SVT-VT Discriminators
* - AT may have sudden onset, weaker criteria
Boston scientific: RHYTHM ID
St Jude: Rate Branch
Biotronik: SMART
Medtronik: PR Logic
ICD TROUBLESHOOTING
ICD troubleshooting
• Troubleshooting problems identified at routine follow-up
ICD troubleshooting
• Troubleshooting problems identified at routine follow-up
• Troubleshooting an ICD shock
ICD troubleshooting
• Troubleshooting problems identified at routine follow-up
• Troubleshooting an ICD shock
• Remote monitoring
TROUBLESHOOTING PROBLEMS
IDENTIFIED AT ROUTINE FOLLOW-UP
Battery depletion
Premature battery depletion
• Latest iterations of ICDs claim battery lifespan of more than 10 yrs
with 2 shocks per year and upto 40% pacing support
Battery depletion
Premature battery depletion
• Latest iterations of ICDs claim battery lifespan of more than 10 yrs
with 2 shocks per year and upto 40% pacing support
• Causes related to pacing
– Unnecessary ventricular pacing
– High pacing outputs
– Lead insulation failure
Battery depletion
Premature battery depletion
• Latest iterations of ICDs claim battery lifespan of more than 10 yrs
with 2 shocks per year and upto 40% pacing support
• Causes related to pacing
– Unnecessary ventricular pacing
– High pacing outputs
– Lead insulation failure
• Most common cause of asymptomatic battery depletion
– Repeated capacitor charging due to aborted shocks due to
repetitive nonsustained VT or oversensing due to lead-
connector problems
• Repeated shocks due to VT storm
Lead problems
Evaluation in all patients usually
• Electrograms from all electrodes on the lead
– Sensed amplitude (P/R)
– Any oversensing
Lead problems
Evaluation in all patients usually
• Electrograms from all electrodes on the lead
• Pacing threshold, pacing impedance
– Impedance varies with manufacturer/lead type
– 200-2000 ohms
Lead problems
Evaluation in all patients usually
• Electrograms from all electrodes on the lead
• Pacing threshold, pacing impedance
• Painless high voltage electrode impedance
– 25-75 ohms
– Review of periodic assessment
Lead problems
Evaluation in all patients usually
• Electrograms from all electrodes on the lead
• Pacing threshold, pacing impedance
• Painless high voltage electrode impedance
Selective patients
• Radiography
• Stored episode electrograms, data logs, patient alerts
• Real time telemetry – provocative maneuvers
Undersensing VT/VF
Sustained Ventricular Arrhythmias Without ICD Therapy:
• Programming error
– VT rate limit is set above that of the clinical VT.
– Discrimainator inhibits therapy thinking it is SVT
• System component failure
– Lead failure - conducture fracture, insulation failure, decrease in
sensing characteristics of the lead
– Battery or capacitor failure – prolonged charge time
TROUBLESHOOTING AN ICD SHOCK
Evaluate ICD shock
Evaluate ICD shock
Inappropriate Shock
Tachycardia
Absent
(Oversensing)
Intracardiac
T/P wave
Double count
Extracardiac
EMI
Myopotentials
Present
(Inappropriate SVT
detection)
Extracardiac oversensing
Extracardiac oversensing
Intracardiac Oversensing
Lead failure
Tachycardia present
• VF? – Analyse EGM to verify
• VT? – Analyse EGM to differentiate SVT/VT
EGM to differentiate SVT/VT
• Single chamber ICD
– Unstable rhythm is more likely to represent AF,
– Gradual acceleration is more compatible with sinus tachycardia
– Sudden onset regular tachycardias – SVT (AVNRT, AT, Afl) or VT
• Compare morphology of far-field channel from sinus rhythm with the one
in arrhythmia
• Morphology of beats that immediately follow shocks should not be relied
on for comparison because of possible EGM distortion
• Beats that immediately follow ATP can be analysed
EGM to differentiate SVT/VT
• ATP response
• In V=A Tachycardia, During ATP
– If AV dissociation – AT > AVNRT >> VT
– If retrograde AV blocks – AVNRT > VT
– Arrhythmia termination by ATP – supports VT >> AVRT, AVNRT
– If atrial acceleration to ATP (entrainment), then after ATP stops
• VAAV – AT
• VVA – VT
• VAV – Not helpful
EGM to differentiate SVT/VT
• Dual chamber ICD
– Atrial sensing
• V>A vs. V<A
• PVC vs. PAC initiated
• If V=A, Driving chamber
– A-A followed by V-V vs. V-V followed by A-A
Appropriate Shock
• Single shock
• Multiple shocks
Appropriate Shock
• Single shock
• Multiple shocks
– Ineffective Initial shocks – High DFT
– Incessant VT
– VT Storm
– Appropriate shocks for AF – If AF defibrillation ‘ON’
Multiple shocks / VT storm
Multiple shocks / VT storm
VT storm
• Treat ischemia or other precipitating factors such as HF
• Judicious use of a magnet or inactivation of ICD therapy
• Pharmacological therapy for VT (e.g. β-blockers, amiodarone);
possible catheter ablation
• IABP or hemodynamic support for hypotension
• Replace electrolytes (K+, Mg2+)
• Left stellate ganglionic blockade in selected patients
• Identify specific diseases that may need specific therapy (e.g.
recurrent VT in Brugada syndrome - isoproterenol and
quinidine)
SYNCOPE WITH NO APPARENT
SHOCK
PERIOPERATIVE MANAGEMENT
Perioperative management
PROGRAMMING ICD
Is the ICD programming appropriate to the
patient needs?
Traditional/Older Concepts
 Quick detection: the longer the arrhythmia, the more likely a
patient to have symptoms e.g. syncope
 Lower cut-off rate: if a higher cut-off rate is used, a
hemodynamically significant VT will be missed, leading to
syncope or VF.
 But primary prevention indications were uncommon in the
past.
 Atrial arrhythmias (AF, atrial flutter, SVT)
 Oversensing due to lead fracture noise or overcounting (T-wave
oversensing, EMI, myopotentials)
Causes for inappropriate therapy (Shock or ATP)
Too aggressive treatment of VT than absolutely required
 Shock or ATP for premature detection of non-sustained VT;
premature ATP may accelerate NSVT
 Shock for sustained pace-terminable VT
Causes for unnecessary therapy (Shock or ATP)
 A randomized, single-blind, multicenter clinical study
 1500 patients with either ischaemic or nonischaemic disease with an
indication for a primary-prevention dual-chamber ICD or CRTD.
Patients with atrial fibrillation or device replacement excluded.
 Patients were assigned to one of three ICD programming groups with
the primary objective of finding their rate of a first occurrence of
inappropriate ATP or shocks.
Reduction in Inappropriate Therapy and
mortality through ICD Programming
(MADIT-RIT trial)
Moss A., etal. NEJM 2012; 367: 2275-83
 The ICD was not a new type of ICD.
 The programming choices were not new.
 Aim - To ignore the slower tachyarrhythmias in High-rate
and those of shorter duration in Duration-delay.
MADIT-RIT: Three Treatment Arms
Arm A
(Conventional)
Arm B
(High-rate)
Arm C
(Duration-delay)
Zone 1:
170 bpm, 2.5s delay
Onset/Stability Detection
Enhancements ON
ATP + Shock
Zone 1:
170 bpm
Monitor only
Zone 1:
170 bpm, 60s delay
Rhythm ID Detection
Enhancements ON
ATP + Shock
Zone 2:
200 bpm, 1s delay
Quick Convert ATP
Shock
Zone 2:
200 bpm, 2.5s delay
Quick Convert ATP
Shock
Zone 2:
200 bpm, 12s delay
Rhythm ID Detection
Enhancements ON
ATP + Shock
Zone 3:
250 bpm, 2.5s delay
Quick Convert ATP +
Shock
 During an average follow-up of 1.4 years, high-rate
therapy and delayed ICD therapy, as compared with
conventional device programming, associated with
- reduction in a first occurrence of inappropriate therapy
- reduction in all-cause mortality
- no significant difference in adverse events
e.g. syncope
 There was significant reduction by more than 70% of
inappropriate therapy in both high-rate and delayed
therapy groups
 Mortality was reduced by 55% in high-rate group
(p=0.01) and by 44% in the delayed-therapy group
(p=0.06).
MADIT-RIT Results
Lesson: Treat sustained tachyarrhythmias only. It does not favor
quick detection and therapy.
MADIT-RIT
Note that in MADIT-RIT:
•Primary prevention indication only
•Dual-chamber ICD and CRT-D only
•Excluded AF patients.
 A randomized, single-blind study to determine whether using 30
of 40 intervals to detect VT (long detection) reduces ATP and
Shock compared with 18 of 24 intervals (standard detection).
 Patients with
 Primary or secondary prevention,
 Single- or dual-chamber ICD or CRT-D
 Sinus or AF.
Effects of Long-detection Interval vs Standard-
detection Interval for ICDs on Antitachycardia
Pacing and Shock Delivery (ADVANCE III trial)
Gasparini M, et al. JAMA 2013; 309:1903-11
 After a median follow-up of 1 year, the long-detection group had
significantly less ICD therapies (ATP and shocks) and inappropriate
shocks.
There was significant reduction in all-cause hospitalizations.
No difference in arrhythmia syncope and mortality.
Lession: Broaden the long detection applicability to include
secondary prevention, single-chamber ICD and AF patients.
ADVANCE III trial Results
 A non-randomized trial with primary endpoint of the rate of
inappropriate shocks at one year post implant.
 A special detection algorithm includes wavelet morphology
analysis, discriminating T wave, assessment of lead integrity or
noise and improved recognition of nonsustained episodes.
 Patients with
 Single chamber ICD,
 Primary or secondary prevention,
 Sinus or AF.
Inappropriate shock rates in patients with
single chamber ICDs using a novel suite of
detection algorithms (PainFree SST study)
Meijer A, et al. Europace June 2013
 757 patients with single-chamber ICD
 97.6% of patients free of inappropriate shocks during the first
year post implant.
Lession: reducing inappropriate shock rate below 3% at 1 year is
possible
PainFree SST study Results
 A randomized study of primary prevention ICD (single or dual
chamber or CRTD) in 1670 patients.
 A combination of programmed parameters (higher detection
rates, longer detection intervals, empiric ATP, SVT discriminators)
vs conventional parameters.
 Result: Reduced ICD therapies without increasing arrhythmic
syncope and reduced all-cause mortality.
Programming ICDs in patients with primary
prevention indication to prolong time to first
shock (PROVIDE study)
Saeed M, et al. J Cardiovasc Electrophysiol 2014; 25(1): 52-59
 4896 patients from MADIT-RIT, ADVANCE 3, PROVIDE, RELEVANT.
 Reduced mortality by 23%
 Reduced inappropriate shocks and both appropriate and
inappropriate ATP significantly.
 64% risk reduction in primary endpoint (composite of death and
appropriate shocks) and 70% reduction in inappropriate shocks.
The impact of prolonged arrhythmia detection
times on outcomes: a meta-analysis
Scott PA, et al. Heart Rhythm 2014; D01: 10.1016. Abstract.
Buber J, et al. Europace 2014; 16(2): 227-234
1. ICD therapy has morbidity. The benefits of an ICD are greatly affected
by its programming.
2. Avoid inappropriate therapy and unnecessary therapy.
(less pain, less hospitalization, better QOL, improved survival)
3. To be certain that there is a sustained tachyarrhythmia before
treating the rhythm
 Prolonged duration of arrhythmia detection
 Faster rate of arrhythmia detection
 Use of ATP
 Algorithms for discrimination of SVT
Aims of ICD programming
CONCLUSION
• ICD is a medical breakthrough but has a complex functioning
• ICD is a medical breakthrough but has a complex functioning
• Adequate therapy prolongs survival but unnecessary therapy
increases morbidity and mortality
• ICD is a medical breakthrough but has a complex functioning
• Adequate therapy prolongs survival but unnecessary therapy
increases morbidity and mortality
• Appropriate function requires adequate mechanical
components (Lead), electrical function (battery and
capacitor), software functioning (detection algorithms), and
substrate (appropriate DFT and safety margin)
• ICD is a medical breakthrough but has a complex functioning
• Adequate therapy prolongs survival but unnecessary therapy
increases morbidity and mortality
• Appropriate function requires adequate mechanical
components (Lead), electrical function (battery and
capacitor), software functioning (detection algorithms), and
substrate (appropriate DFT and safety margin)
• Knowledge and know-how of all 4 components of ICD therapy
is essential for the caregiver
An ICD Prayer
An ICD Prayer
O Creator,
Grant me the serenity to accept the things I should
not(cannot) change,
the voltage (courage) to change the things I can,
and the algorithm (wisdom) to know the difference.
Amen.

More Related Content

What's hot

Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
drskd6
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrt
Rahul Chalwade
 
Shunt Detection And Quantification
Shunt Detection And QuantificationShunt Detection And Quantification
Shunt Detection And Quantification
Dang Thanh Tuan
 

What's hot (20)

CRT Case-Based Troubleshooting
CRT Case-Based TroubleshootingCRT Case-Based Troubleshooting
CRT Case-Based Troubleshooting
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
 
Conduction system pacing as resynchronization
Conduction system pacing as resynchronizationConduction system pacing as resynchronization
Conduction system pacing as resynchronization
 
CATHETER ABLATION IN VT
CATHETER ABLATION IN VTCATHETER ABLATION IN VT
CATHETER ABLATION IN VT
 
Cardiac resynchronization therapy
Cardiac resynchronization therapyCardiac resynchronization therapy
Cardiac resynchronization therapy
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
 
ECG interpretation postCRT
ECG interpretation postCRT ECG interpretation postCRT
ECG interpretation postCRT
 
ELectrophysiology basics part4
ELectrophysiology basics part4ELectrophysiology basics part4
ELectrophysiology basics part4
 
PTMC/PBMC
PTMC/PBMCPTMC/PBMC
PTMC/PBMC
 
Pre-Procedural Preparation and CRT Implantation Tips and Tricks
Pre-Procedural Preparation and CRT Implantation Tips and TricksPre-Procedural Preparation and CRT Implantation Tips and Tricks
Pre-Procedural Preparation and CRT Implantation Tips and Tricks
 
Percutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve InterventionsPercutaneous Pulmonary Valve Interventions
Percutaneous Pulmonary Valve Interventions
 
Trans septal puncture
Trans septal punctureTrans septal puncture
Trans septal puncture
 
CATH MEET PDA
CATH MEET PDACATH MEET PDA
CATH MEET PDA
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrt
 
Coronary anatomy and angiographic views
Coronary anatomy and angiographic viewsCoronary anatomy and angiographic views
Coronary anatomy and angiographic views
 
Electrophysiology study basics
Electrophysiology study basicsElectrophysiology study basics
Electrophysiology study basics
 
Mitra clip
Mitra clipMitra clip
Mitra clip
 
Pacemaker basics
Pacemaker basicsPacemaker basics
Pacemaker basics
 
Shunt Detection And Quantification
Shunt Detection And QuantificationShunt Detection And Quantification
Shunt Detection And Quantification
 
POST CRT OPTIMISATION
POST CRT OPTIMISATIONPOST CRT OPTIMISATION
POST CRT OPTIMISATION
 

Viewers also liked

Viewers also liked (20)

Assessment of mitral valve for PTMC
Assessment of mitral valve for PTMCAssessment of mitral valve for PTMC
Assessment of mitral valve for PTMC
 
Coronary revascularization in diabetes mellitus and multivessel cad
Coronary revascularization in diabetes mellitus and multivessel cadCoronary revascularization in diabetes mellitus and multivessel cad
Coronary revascularization in diabetes mellitus and multivessel cad
 
Pharmacological stress echocardiography
Pharmacological stress echocardiographyPharmacological stress echocardiography
Pharmacological stress echocardiography
 
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
 
Longitudinal stent deformation in PCI
Longitudinal stent deformation in PCILongitudinal stent deformation in PCI
Longitudinal stent deformation in PCI
 
Use of Vascular plugs in cardiovascular medicine
Use of Vascular plugs in cardiovascular medicineUse of Vascular plugs in cardiovascular medicine
Use of Vascular plugs in cardiovascular medicine
 
Electrophysiology study protocol
Electrophysiology study protocolElectrophysiology study protocol
Electrophysiology study protocol
 
Electrophysiology AVRT
Electrophysiology AVRTElectrophysiology AVRT
Electrophysiology AVRT
 
Newer Oral Anticoagulants or warfarin in DVT/PE
Newer Oral Anticoagulants or warfarin in DVT/PENewer Oral Anticoagulants or warfarin in DVT/PE
Newer Oral Anticoagulants or warfarin in DVT/PE
 
Clinical approach to multi valvular heart disease
Clinical approach to multi valvular heart diseaseClinical approach to multi valvular heart disease
Clinical approach to multi valvular heart disease
 
Approach to TOF physiology
Approach to TOF physiologyApproach to TOF physiology
Approach to TOF physiology
 
Electrophysiology AVNRT
Electrophysiology AVNRTElectrophysiology AVNRT
Electrophysiology AVNRT
 
DRUG ELUTING BALLOONS (DCB/DEB)
DRUG ELUTING BALLOONS (DCB/DEB)DRUG ELUTING BALLOONS (DCB/DEB)
DRUG ELUTING BALLOONS (DCB/DEB)
 
Beta blockers in Acute MI
Beta blockers in Acute MIBeta blockers in Acute MI
Beta blockers in Acute MI
 
Choice of guiding catheters in PCI
Choice of guiding catheters in PCIChoice of guiding catheters in PCI
Choice of guiding catheters in PCI
 
Marfan syndrome
Marfan syndromeMarfan syndrome
Marfan syndrome
 
Cardiac tamponade
Cardiac tamponadeCardiac tamponade
Cardiac tamponade
 
The Cycle of Cardiac Care and the Uninformed Patient
The Cycle of Cardiac Care and the Uninformed PatientThe Cycle of Cardiac Care and the Uninformed Patient
The Cycle of Cardiac Care and the Uninformed Patient
 
Cardiology ICD-10 records with Dual Coding-ICD-10 Training
Cardiology ICD-10 records with Dual Coding-ICD-10 TrainingCardiology ICD-10 records with Dual Coding-ICD-10 Training
Cardiology ICD-10 records with Dual Coding-ICD-10 Training
 
History of ICDs
History of ICDsHistory of ICDs
History of ICDs
 

Similar to ICD troubleshooting

How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptx
Kush Bhagat
 
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
Taiwan Heart Rhythm Society
 

Similar to ICD troubleshooting (20)

How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptx
 
Tachycardia discriminating algorithms and trouble shooting of ICDs
Tachycardia discriminating algorithms and trouble shooting of ICDsTachycardia discriminating algorithms and trouble shooting of ICDs
Tachycardia discriminating algorithms and trouble shooting of ICDs
 
ICD System High Voltage Component Failure (Presented to FDA)
ICD System High Voltage Component Failure (Presented to FDA)ICD System High Voltage Component Failure (Presented to FDA)
ICD System High Voltage Component Failure (Presented to FDA)
 
LEADLESS PACEMAKER AND SUBCUTANEOUS ICD
LEADLESS PACEMAKER AND SUBCUTANEOUS ICDLEADLESS PACEMAKER AND SUBCUTANEOUS ICD
LEADLESS PACEMAKER AND SUBCUTANEOUS ICD
 
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
 
Icd basic 042411(dr chang)
Icd basic 042411(dr chang)Icd basic 042411(dr chang)
Icd basic 042411(dr chang)
 
Basics of ICD - Dr Hafeesh Fazulu - Pushpagiri
Basics of ICD - Dr Hafeesh Fazulu - PushpagiriBasics of ICD - Dr Hafeesh Fazulu - Pushpagiri
Basics of ICD - Dr Hafeesh Fazulu - Pushpagiri
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Ventricular tachycardia
Ventricular tachycardiaVentricular tachycardia
Ventricular tachycardia
 
Emergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesEmergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issues
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Perioperative anaesthetic management of a patient with an implanted cardiac d...
Perioperative anaesthetic management of a patient with an implanted cardiac d...Perioperative anaesthetic management of a patient with an implanted cardiac d...
Perioperative anaesthetic management of a patient with an implanted cardiac d...
 
Perioperative anaesthetic management of a patient with an implanted cardiac d...
Perioperative anaesthetic management of a patient with an implanted cardiac d...Perioperative anaesthetic management of a patient with an implanted cardiac d...
Perioperative anaesthetic management of a patient with an implanted cardiac d...
 
Pacer ppt
Pacer pptPacer ppt
Pacer ppt
 
Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabg
 
pace1.ppt
pace1.pptpace1.ppt
pace1.ppt
 
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
 
Approach to cardiac arrhythmias
Approach to cardiac arrhythmiasApproach to cardiac arrhythmias
Approach to cardiac arrhythmias
 

More from Satyam Rajvanshi (7)

How to avoid seeing a cardiologist
How to avoid seeing a cardiologistHow to avoid seeing a cardiologist
How to avoid seeing a cardiologist
 
Management of Carotid Artery Stenosis - Evidence and guidelines
Management of Carotid Artery Stenosis - Evidence and guidelinesManagement of Carotid Artery Stenosis - Evidence and guidelines
Management of Carotid Artery Stenosis - Evidence and guidelines
 
STEMI Late Presentation - Management and practical approach
STEMI Late Presentation - Management and practical approachSTEMI Late Presentation - Management and practical approach
STEMI Late Presentation - Management and practical approach
 
Coronary Intramural Hematoma
Coronary Intramural HematomaCoronary Intramural Hematoma
Coronary Intramural Hematoma
 
Endovascular management of Aortic Dissection
Endovascular management of Aortic DissectionEndovascular management of Aortic Dissection
Endovascular management of Aortic Dissection
 
Are all sartans equal
Are all sartans equalAre all sartans equal
Are all sartans equal
 
Digoxin and its Toxicity
Digoxin and its ToxicityDigoxin and its Toxicity
Digoxin and its Toxicity
 

Recently uploaded

Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 

Recently uploaded (20)

Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadO898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
O898O367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
 

ICD troubleshooting

  • 4. Word meaning • ICD – Intracardiac cardioverter
  • 5. Word meaning • ICD – Intracardiac cardioverter defibrillator
  • 6. • ICD brief history – 1947 First human internal defibrillation – 1956 First human external defibrillation – 1966 First ICD conception – 1969 First external canine prototype tested – 1970 First implantable prototype (895 g) – 1975 First implantable Defib in canines (250 g) – 1980 First human implant @ Johns Hopkins – Mirowski et al – 1985 ICD market released (350 units) – (FDA approval)
  • 9. ICD: First Human Implant 1980 Dr Winkle, Dr Mirowski, the first ICD patient, Dr Heilman
  • 10. ICD: First Human Implant 1980 A transvenous lead in SVC and a ventricular patch lead
  • 11. Evolution of ICD Therapy
  • 16. Sensing Detection Therapy ICD treats ventricular tachyarrhythmias
  • 18. Sensing in ICD: A major challenge… Sensing - process by which an ICD determines the timing of each atrial or ventricular depolarization from electrogram signals. A wide variation in the size of signals • stable and large normal ventricular signals • low amplitude VF signals
  • 22. Sensing auto-adjust Sinus rhythm – To avoid oversensing T waves VF – To avoid undersensing
  • 25. Detection Algorithm by which ICD processes sensed signals to classify the rhythm and determine if therapy should be delivered. Up to 3 programmable rate detection zones to permit zone- specific therapies for slower VT, faster VT, and ventricular fibrillation (VF). The minimum duration of tachycardia required for detection is programmable, either in seconds or in number of ventricular intervals required for detection. The specific method used to count ventricular intervals influences the sensitivity and specificity of VT detection
  • 26. SVT-VT discriminators Programmable subset of the VT/VF detection algorithm that withholds ventricular therapy for SVT to improve specificity. They usually differ from SVT detection algorithms used to mode- switch during bradycardia pacing or to deliver atrial therapy for atrial fibrillation or atrial flutter.
  • 27. Confirmation or Reconfirmation Brief process by which ICDs determine whether to deliver or abort a shock after the high-voltage capacitor is charged.
  • 28. Redetection Process by which ICDs determine whether VT or VF detection criteria remain satisfied after therapy is delivered.
  • 29. Termination Termination of a VT or VF episode occurs when the ICD reclassifies the rhythm as sinus after VT or VF has been detected.
  • 31. Detection Ventricular Fibrillation (VF) • Rate cut-off (VF zone) • Number of intervals for detection (usually consecutive intervals not required; high sensitivity)
  • 32. Detection Ventricular Tachycardia (VT) • Rate cut-off [VT zone(s)] • Number of intervals for detection • Detection Enhancement/SVT-VT Discriminators
  • 34.
  • 35. Electrogram Morphology • Current electrogram signal compared with stored normal template • Match percentage = 1 - (area of difference) SVT-VT Discriminators
  • 36. Stability • To reduce inappropriate detection of atrial arrhythmias eg atrial fibrillation (rhythm not “stable” i.e. variable cycle length) • Each interval compared to previous intervals • Count as VT if the difference  the programmed stability interval i.e. VT is “stable” VT detection interval: 500ms Stability 30ms SVT-VT Discriminators
  • 37. Onset • To reduce inappropriate detection of sinus tachycardia • Average of current 4 intervals is compared to average of previous 4 intervals Current average programmed Previous average onset percentage i.e. onset is “sudden” in VT  SVT-VT Discriminators
  • 42. Lead noise discriminator  Caused by pace-sense conductor fracture, loose set screw, lead insulation breach.  Therapy avoided by comparing near-field channel (RV tip to RV ring) to far-field channel (RV tip to Can/RV coil to Can) SVT-VT Discriminators
  • 43. SVT-VT Discriminators * - AT may have sudden onset, weaker criteria
  • 45. St Jude: Rate Branch
  • 49. ICD troubleshooting • Troubleshooting problems identified at routine follow-up
  • 50. ICD troubleshooting • Troubleshooting problems identified at routine follow-up • Troubleshooting an ICD shock
  • 51. ICD troubleshooting • Troubleshooting problems identified at routine follow-up • Troubleshooting an ICD shock • Remote monitoring
  • 53. Battery depletion Premature battery depletion • Latest iterations of ICDs claim battery lifespan of more than 10 yrs with 2 shocks per year and upto 40% pacing support
  • 54. Battery depletion Premature battery depletion • Latest iterations of ICDs claim battery lifespan of more than 10 yrs with 2 shocks per year and upto 40% pacing support • Causes related to pacing – Unnecessary ventricular pacing – High pacing outputs – Lead insulation failure
  • 55. Battery depletion Premature battery depletion • Latest iterations of ICDs claim battery lifespan of more than 10 yrs with 2 shocks per year and upto 40% pacing support • Causes related to pacing – Unnecessary ventricular pacing – High pacing outputs – Lead insulation failure • Most common cause of asymptomatic battery depletion – Repeated capacitor charging due to aborted shocks due to repetitive nonsustained VT or oversensing due to lead- connector problems • Repeated shocks due to VT storm
  • 56. Lead problems Evaluation in all patients usually • Electrograms from all electrodes on the lead – Sensed amplitude (P/R) – Any oversensing
  • 57. Lead problems Evaluation in all patients usually • Electrograms from all electrodes on the lead • Pacing threshold, pacing impedance – Impedance varies with manufacturer/lead type – 200-2000 ohms
  • 58. Lead problems Evaluation in all patients usually • Electrograms from all electrodes on the lead • Pacing threshold, pacing impedance • Painless high voltage electrode impedance – 25-75 ohms – Review of periodic assessment
  • 59. Lead problems Evaluation in all patients usually • Electrograms from all electrodes on the lead • Pacing threshold, pacing impedance • Painless high voltage electrode impedance Selective patients • Radiography • Stored episode electrograms, data logs, patient alerts • Real time telemetry – provocative maneuvers
  • 60. Undersensing VT/VF Sustained Ventricular Arrhythmias Without ICD Therapy: • Programming error – VT rate limit is set above that of the clinical VT. – Discrimainator inhibits therapy thinking it is SVT • System component failure – Lead failure - conducture fracture, insulation failure, decrease in sensing characteristics of the lead – Battery or capacitor failure – prolonged charge time
  • 64.
  • 65. Inappropriate Shock Tachycardia Absent (Oversensing) Intracardiac T/P wave Double count Extracardiac EMI Myopotentials Present (Inappropriate SVT detection)
  • 66.
  • 68.
  • 72.
  • 73. Tachycardia present • VF? – Analyse EGM to verify • VT? – Analyse EGM to differentiate SVT/VT
  • 74. EGM to differentiate SVT/VT • Single chamber ICD – Unstable rhythm is more likely to represent AF, – Gradual acceleration is more compatible with sinus tachycardia – Sudden onset regular tachycardias – SVT (AVNRT, AT, Afl) or VT • Compare morphology of far-field channel from sinus rhythm with the one in arrhythmia • Morphology of beats that immediately follow shocks should not be relied on for comparison because of possible EGM distortion • Beats that immediately follow ATP can be analysed
  • 75. EGM to differentiate SVT/VT • ATP response • In V=A Tachycardia, During ATP – If AV dissociation – AT > AVNRT >> VT – If retrograde AV blocks – AVNRT > VT – Arrhythmia termination by ATP – supports VT >> AVRT, AVNRT – If atrial acceleration to ATP (entrainment), then after ATP stops • VAAV – AT • VVA – VT • VAV – Not helpful
  • 76. EGM to differentiate SVT/VT • Dual chamber ICD – Atrial sensing • V>A vs. V<A • PVC vs. PAC initiated • If V=A, Driving chamber – A-A followed by V-V vs. V-V followed by A-A
  • 77.
  • 78. Appropriate Shock • Single shock • Multiple shocks
  • 79. Appropriate Shock • Single shock • Multiple shocks – Ineffective Initial shocks – High DFT – Incessant VT – VT Storm – Appropriate shocks for AF – If AF defibrillation ‘ON’
  • 80. Multiple shocks / VT storm
  • 81. Multiple shocks / VT storm
  • 82. VT storm • Treat ischemia or other precipitating factors such as HF • Judicious use of a magnet or inactivation of ICD therapy • Pharmacological therapy for VT (e.g. β-blockers, amiodarone); possible catheter ablation • IABP or hemodynamic support for hypotension • Replace electrolytes (K+, Mg2+) • Left stellate ganglionic blockade in selected patients • Identify specific diseases that may need specific therapy (e.g. recurrent VT in Brugada syndrome - isoproterenol and quinidine)
  • 83. SYNCOPE WITH NO APPARENT SHOCK
  • 84.
  • 88. Is the ICD programming appropriate to the patient needs?
  • 89. Traditional/Older Concepts  Quick detection: the longer the arrhythmia, the more likely a patient to have symptoms e.g. syncope  Lower cut-off rate: if a higher cut-off rate is used, a hemodynamically significant VT will be missed, leading to syncope or VF.  But primary prevention indications were uncommon in the past.
  • 90.  Atrial arrhythmias (AF, atrial flutter, SVT)  Oversensing due to lead fracture noise or overcounting (T-wave oversensing, EMI, myopotentials) Causes for inappropriate therapy (Shock or ATP) Too aggressive treatment of VT than absolutely required  Shock or ATP for premature detection of non-sustained VT; premature ATP may accelerate NSVT  Shock for sustained pace-terminable VT Causes for unnecessary therapy (Shock or ATP)
  • 91.  A randomized, single-blind, multicenter clinical study  1500 patients with either ischaemic or nonischaemic disease with an indication for a primary-prevention dual-chamber ICD or CRTD. Patients with atrial fibrillation or device replacement excluded.  Patients were assigned to one of three ICD programming groups with the primary objective of finding their rate of a first occurrence of inappropriate ATP or shocks. Reduction in Inappropriate Therapy and mortality through ICD Programming (MADIT-RIT trial) Moss A., etal. NEJM 2012; 367: 2275-83
  • 92.  The ICD was not a new type of ICD.  The programming choices were not new.  Aim - To ignore the slower tachyarrhythmias in High-rate and those of shorter duration in Duration-delay.
  • 93. MADIT-RIT: Three Treatment Arms Arm A (Conventional) Arm B (High-rate) Arm C (Duration-delay) Zone 1: 170 bpm, 2.5s delay Onset/Stability Detection Enhancements ON ATP + Shock Zone 1: 170 bpm Monitor only Zone 1: 170 bpm, 60s delay Rhythm ID Detection Enhancements ON ATP + Shock Zone 2: 200 bpm, 1s delay Quick Convert ATP Shock Zone 2: 200 bpm, 2.5s delay Quick Convert ATP Shock Zone 2: 200 bpm, 12s delay Rhythm ID Detection Enhancements ON ATP + Shock Zone 3: 250 bpm, 2.5s delay Quick Convert ATP + Shock
  • 94.  During an average follow-up of 1.4 years, high-rate therapy and delayed ICD therapy, as compared with conventional device programming, associated with - reduction in a first occurrence of inappropriate therapy - reduction in all-cause mortality - no significant difference in adverse events e.g. syncope  There was significant reduction by more than 70% of inappropriate therapy in both high-rate and delayed therapy groups  Mortality was reduced by 55% in high-rate group (p=0.01) and by 44% in the delayed-therapy group (p=0.06). MADIT-RIT Results
  • 95. Lesson: Treat sustained tachyarrhythmias only. It does not favor quick detection and therapy. MADIT-RIT Note that in MADIT-RIT: •Primary prevention indication only •Dual-chamber ICD and CRT-D only •Excluded AF patients.
  • 96.  A randomized, single-blind study to determine whether using 30 of 40 intervals to detect VT (long detection) reduces ATP and Shock compared with 18 of 24 intervals (standard detection).  Patients with  Primary or secondary prevention,  Single- or dual-chamber ICD or CRT-D  Sinus or AF. Effects of Long-detection Interval vs Standard- detection Interval for ICDs on Antitachycardia Pacing and Shock Delivery (ADVANCE III trial) Gasparini M, et al. JAMA 2013; 309:1903-11
  • 97.  After a median follow-up of 1 year, the long-detection group had significantly less ICD therapies (ATP and shocks) and inappropriate shocks. There was significant reduction in all-cause hospitalizations. No difference in arrhythmia syncope and mortality. Lession: Broaden the long detection applicability to include secondary prevention, single-chamber ICD and AF patients. ADVANCE III trial Results
  • 98.  A non-randomized trial with primary endpoint of the rate of inappropriate shocks at one year post implant.  A special detection algorithm includes wavelet morphology analysis, discriminating T wave, assessment of lead integrity or noise and improved recognition of nonsustained episodes.  Patients with  Single chamber ICD,  Primary or secondary prevention,  Sinus or AF. Inappropriate shock rates in patients with single chamber ICDs using a novel suite of detection algorithms (PainFree SST study) Meijer A, et al. Europace June 2013
  • 99.  757 patients with single-chamber ICD  97.6% of patients free of inappropriate shocks during the first year post implant. Lession: reducing inappropriate shock rate below 3% at 1 year is possible PainFree SST study Results
  • 100.  A randomized study of primary prevention ICD (single or dual chamber or CRTD) in 1670 patients.  A combination of programmed parameters (higher detection rates, longer detection intervals, empiric ATP, SVT discriminators) vs conventional parameters.  Result: Reduced ICD therapies without increasing arrhythmic syncope and reduced all-cause mortality. Programming ICDs in patients with primary prevention indication to prolong time to first shock (PROVIDE study) Saeed M, et al. J Cardiovasc Electrophysiol 2014; 25(1): 52-59
  • 101.  4896 patients from MADIT-RIT, ADVANCE 3, PROVIDE, RELEVANT.  Reduced mortality by 23%  Reduced inappropriate shocks and both appropriate and inappropriate ATP significantly.  64% risk reduction in primary endpoint (composite of death and appropriate shocks) and 70% reduction in inappropriate shocks. The impact of prolonged arrhythmia detection times on outcomes: a meta-analysis Scott PA, et al. Heart Rhythm 2014; D01: 10.1016. Abstract. Buber J, et al. Europace 2014; 16(2): 227-234
  • 102. 1. ICD therapy has morbidity. The benefits of an ICD are greatly affected by its programming. 2. Avoid inappropriate therapy and unnecessary therapy. (less pain, less hospitalization, better QOL, improved survival) 3. To be certain that there is a sustained tachyarrhythmia before treating the rhythm  Prolonged duration of arrhythmia detection  Faster rate of arrhythmia detection  Use of ATP  Algorithms for discrimination of SVT Aims of ICD programming
  • 103.
  • 104.
  • 105.
  • 106.
  • 108. • ICD is a medical breakthrough but has a complex functioning
  • 109. • ICD is a medical breakthrough but has a complex functioning • Adequate therapy prolongs survival but unnecessary therapy increases morbidity and mortality
  • 110. • ICD is a medical breakthrough but has a complex functioning • Adequate therapy prolongs survival but unnecessary therapy increases morbidity and mortality • Appropriate function requires adequate mechanical components (Lead), electrical function (battery and capacitor), software functioning (detection algorithms), and substrate (appropriate DFT and safety margin)
  • 111. • ICD is a medical breakthrough but has a complex functioning • Adequate therapy prolongs survival but unnecessary therapy increases morbidity and mortality • Appropriate function requires adequate mechanical components (Lead), electrical function (battery and capacitor), software functioning (detection algorithms), and substrate (appropriate DFT and safety margin) • Knowledge and know-how of all 4 components of ICD therapy is essential for the caregiver
  • 113. An ICD Prayer O Creator, Grant me the serenity to accept the things I should not(cannot) change, the voltage (courage) to change the things I can, and the algorithm (wisdom) to know the difference. Amen.