EU MDR Preparation
OMTEC
Mike Wolf CBA, CQA, RAC
Chicago, IL 13 June 2018
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 2
What is the new EU MDR?
 Most significant regulatory change in Europe in over 20 years!
 Compliance to the EU MDR required to gain and retain CE Mark
 80+ delegating or implementing acts to come
 Notified Bodies to be re-designated
 Focus on Clinical Evidence and Post Market Clinical Follow-Up and Surveillance
 Scrutiny and review of high risk devices
 Improved transparency between NBs, CAs, Mfgs, and Users (UDI and EUDAMED)
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 3
The new European Union Medical Device Regulation (EU MDR)
AIMD
17 Articles
35 Pages
9 Annexes
MDD
20 Articles
60 Pages
12 Annexes
Directives Regulation
MDR
123 Articles
178 Pages
17 Annexes
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 4
Key Players - MDR
EU National
Competent Authorities
Notified Bodies
Reviews Testing Auditing
Clinical
Evaluations
Records
Data
QMS
Certificate
CE
Mark
Medical Device
Coordination Group
(MDCG)
Expert Panels
Suppliers
Importer
Distributors
Medical
Devices
with UDI Authorized
Representative
Manufacturer
Public
EUDAMED
= Economic Operator
PMS/PMCF
Reporting
= New or Changed Element
MDR
Person Responsible
for Regulatory
Compliance
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 5
Timelines – how long do we have?
Transition Timeline
2017 [May 25]
EU MDR
Entered into Force
2017 [Nov 25]
Notified Bodies
apply for MDR
Designation
2020 [May 25]
Notified Bodies stop issuing
MDD Certs
2017 2018 2019 2020 2024
2020 [May 26]
New or Renewing Devices
must have MDR certs
MDD Certs Issued Existing MDD Certs Valid
May 26, 2020
Date of Application
May 26, 2024
Any Remaining MDD
Certs Voided
May 26, 2025
No further placing
MDD-certified
products on the market
MDR Certs Issued
6
Timeline Convergence – Competing Transitions
ISO 13485:2016
Both ISO 13485:2003 or
2016 Certifications
New ISO 13485:2003
Certifications Discouraged
No 13485:2003 Certifications or
Re-Certifications Allowed
01 Mar 2016 01 Mar 2017 01 Mar 2018 01 Mar 2019
MDSAP
01 Jan 2016 01 Jan 2017 01 Jan 2018 01 Jan 2019
MDSAP Audit Certified
Mandatory by Health Canada
EU MDR
26 May 2020
Present Day
New or Renewing Devices
must have MDR certificates
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc.
Important for orthopedic device manufacturers
Revised device classification rules
Extended process of conformance assessment by NB and MDCG
Increased scrutiny of technical documentation - Class III and Implantable
New labeling requirements
Restriction of hazardous substances
Risks associated with interaction between software and the IT environment within
which it operates
Conformance with harmonized standards
 And Common Specifications yet to be published – 26 May 2020
Clinical evaluation and adequate clinical data
Particular Requirements for Active Implantable Medical Devices
 Devices and their component parts shall be identifiable
 Shall bear a code by which the device and manufacturer can be unequivocally identified
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 7
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 8
Significant Change – Legacy Products
CHANGE
 There is no provision for
grandfathering certification
(CE marks) obtained under the
previous Directives
 All products have to be CE
marked under the new
Regulation 2017/745, in order
to be placed on the market or
put into service (MDR Article
5), after the transition period
Market Access of Legacy Products
IMPLICATION FOR MANUFACTURER
A comprehensive plan needs to be put in place to
ensure that all products that will be maintained on the
EU market are CE marked in accordance with the full
requirements of the new MDR. This should include
products currently under development. This review
may provide an opportunity for rationalization of the
product portfolio and elimination of any marginal
products.
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 9
Significant Change - Reclassification
CHANGE
 Certain products have received
special consideration in the MDR
and are subject to reclassification.
(MDR Article 4)
 These new provisions variously
apply to cosmetic implants,
standalone software, products
without an intended medical
purpose (MDR Annex XVI), certain
spine products and reusable Class 1
devices
Reclassification of Devices
IMPLICATION FOR MANUFACTURER
Manufacturers of medical devices should carefully
examine the MDR Classification rules in Annex VIII to
determine whether new conformity assessment
routes are now applicable to their product portfolio. If
so, they should engage their Notified Body (where
necessary) and take steps to evaluate the necessary
timescales involved in implementing this change. In
some cases, where reclassification has occurred, this
may be the first time that such relationships and
agreements with Notified Bodies have been required.
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 10
Significant Change - Reclassification
CHANGE
 Changes in the classification rules of
medical devices might lead to
reclassification of certain medical devices
 Rules organized by
‒ Non-Invasive Devices (Rules 1 – 4)
‒ Invasive Devices (Rules 5 – 8)
‒ Active Devices (Rules 9 – 13)
‒ Special Rules (Rules 14 – 22)
 There is a process to handle classifications
disputes
‒ Steps depend on if Manufacturer and notified
body are in the same EU Member State
‒ May involve consultation with competent
authority of the member state
Check new classification rules
Classification Change Examples
• In vitro contact with cells/embryos going back into
the body to be Class III (or IIa)
• Apheresis devices to be Class III
• IVF and ART non-invasive devices can be IIb
• Spinal implants to be Class III
• Total and partial joint replacements to be Class III
• Devices recording diagnostic images to be IIa
• Nanomaterial devices to be Class III
• AIMD accessories to be Class III
• Devices which are intended to be introduced into
the human body via a body orifice, or applied on
skin and that are absorbed by or locally dispersed in
the human body are to be Class III
• Reusable surgical Instruments are no longer class I
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 11
Significant Change – General Safety & Performance Requirements
CHANGE
The spirit of the “Essential
Requirements” has not changed,
however there are new provisions
and more details required (Annex I),
for example:
 Hazardous substances
 Devices with non-viable human cells
or tissues, and other “non-viable
biological substances”
 IT security and prevention of
unauthorized access
 Risk Management
 Labeling
 Alignment with Global Harmonization
Task Force (GHTF)
IMPLICATION FOR MANUFACTURER
Manufacturers need to review and update
technical documentation in preparation for a
complete conformity assessment against
requirements of GSPR. Evidence of
conformance with each GSPR should be
critically assessed. Emphasis on improved risk
management and risk controls will also need to
be incorporated in practices and processes.
Some new requirements – others just more detailed
Significant Change – Technical Documentation
CHANGE
The MDR is significantly more prescriptive about the required content of technical documentation
(Technical File/Design Dossier):
 Essential Requirements (ERs) are replaced by “General Safety Requirements (Annex I) and
the number of requirements has been expanded
 A Presumption of conformity still applies for devices that are in conformity with relevant
harmonised standards (Article 8), but the Commission may define Common Specifications
where no harmonized standards exist or where they are considered insufficient
Additionally, there are more detailed requirements for the Quality Management Systems (QMS)
(Annex IX)
 EN ISO 13485:2016 was re-written and issued in 2016 with the new MDR very much in mind
 It should be noted that while EN ISO 13485:2016 is not an absolute requirement there will be a
general expectation that this standard will be used
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 12
Technical Documentation
Significant Change – Technical Documentation
 Technical Documents shall include:
 Device description and specification
 Reference to previous and similar generations of device
 Labels & IFUs
 Design and Manufacturing information
 Demonstration of conformance to General Safety and Performance Requirements
 Benefit-Risk analysis and Risk Management
 Product verification and validation
 Additional Requirements for Technical Documents on Post Market Surveillance
 Post Market Surveillance Plan, per Article 84
 Periodic Safety Update Report, per Article 86
 Post Market Surveillance Report, per Article 85
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 13
Detailed Requirements for Technical Documents
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 14
Significant Change – Vigilance and PMS
CHANGE
 Under the new Regulation 2017/745,
device manufacturers will be required
to collect post-market clinical data as
part of their on-going assessment of
potential safety risks
 Post Market Clinical Follow up (PMCF) is
a continuous process with the objective
of constantly updating the clinical
evaluation (Annex XIV, Part B)
 Additionally, reporting timeframes are
tightened from 30 days to 15 days for
reporting serious incidents (Article 87)
Vigilance and Post Market Surveillance
IMPLICATION FOR MANUFACTURER
Manufacturers need to review their procedures
for PMS and ensure that the responsibility for
the provision of this additional data and
associated support is clearly established. This is
likely to require additional resources in
functions that support products on the market
such as Regulatory and Medical Affairs.
 There will be new electronic vigilance
reporting (Article 92) and Periodic Safety
Update Reports (PSUR) for all devices (Article
86) subject to differing frequency and
submission requirements
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 15
Significant Change – Vigilance and PMS
Summary of safety and clinical performance
 Draft is part of submission to notified
body for conformity assessment
 Is validated by the notified body
 After validation, is uploaded to EUDAMED
Two new deliverables required by the EU MDR
Periodic safety update report (PSUR)
 Throughout the lifetime of the device
concerned, the PSUR shall set out
‒ Conclusions of the benefit-risk
determination
‒ Main findings of the PMCF
‒ Volume of sales of the device and
estimated evaluation of the size and other
characteristics of the population using
device, and where practicable, the usage
frequency
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 16
Significant Change – Clinical Evaluations
CHANGE
 The MDR is more specific about
the need for clinical evidence
and clinical evaluation,
proportionate with the risk
associated with a given device
(Annex XIV, Part A)
 Reliance on the scientific
literature to demonstrate
equivalence will be more tightly
regulated, and clinical
evaluations will be more closely
aligned with clinical trials
associated with medicinal
products
Clinical Evidence and Clinical Evaluations
IMPLICATION FOR MANUFACTURER
This may require manufacturers to obtain additional
clinical data from clinical studies. There will be
additional scrutiny of Clinical Evaluation Reports
(CERs) by Notified Bodies as outlined by new
guidance (MEDDEV 2.7.1 rev. 4). CERs that previously
were of a suitable level and standard and accepted
by Notified Bodies, may no longer be accepted.
Manufacturers should plan to review all of their CERs
if not reviewed within the last 1-2 years and ensure
that CERs include Post Market Surveillance data;
especially where devices are new to market and
where clinical data was limited at the time of first CE
Marking.
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 17
Significant Change – Labeling
CHANGE
Requirements for product labeling are
more prescriptive
 For example, information supplied
by the manufacturer shall be
made available and kept up to
date on the manufacturer’s
website (Chapter III, 23.1)
IMPLICATION FOR MANUFACTURER
Manufacturers should carefully review the adequacy of
their product labelling and precautionary statements and
consider how this will be reflected on their websites.
There are requirements for
 Specific details for labels and for sterile packages (Annex I).
 Inclusion of information on residual risks for vulnerable patient
groups (e.g. children, pregnant or nursing women) and, if
applicable, on appropriate precautionary measures in the
instructions for use.
 Hazardous Substances
All EU labeling and IFU’s likely to be impacted
18
Significant Change - EUDAMED
What is EUDAMED?
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc.
• SU
 EUDAMED is a central European databank that will be accessible to notified bodies,
economic operators, sponsors, and to the public to the extent specified
 It will include protections for personal data
 What’s driving EUDAMED?
 Article 14a (MDD 93/42/EEC)
 Clinical investigations
 Vigilance and incident reporting
 Status of conformity on issued certificates
 Registration of manufacturers and EUAR
Technical Docs to MDR Certification Timeline
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 19
Many factors impact the timeline for transition of existing product submissions
MDCG Consultation for Class III &
Implantable Devices Only
Technical Documentation Updates for MDR likely to include:
• Design Requirements traced to new Safety and Performance Requirements
• Risk Management File/Report
• Clinical Evaluation Plan/Report
• Gaps in Clinical Evidence
• Any equivalence to competitive device data replaced with company’s own data
• Post Market Surveillance Reporting
• PMS/PMCF Plan/Report, PSUR/SSCP Reports (when needed + translations)
• Labeling +UDI + Authorized Rep/Branding Changes + Translations
• Declaration of Conformity + UDI inspections in place at DCs
Technical
Documentation Update
(? weeks)
Notified Body
Notification & Planning
(? weeks)
First evaluation by
Notified Body
(40 days)
Response
(15 days)
Second
evaluation
(20 days)
MDCG
(21 days)
Scrutiny
(60 days)
Certification
decision
(15 days)
Submission required for:
• First time products
• Change to products
What factors will slow down the process?
Delays in Notified Bodies receiving their designation to MDR
Notified Body capacity constraints
Additional scrutiny of Notified Body performance
Additional scrutiny of technical documentation by Notified Bodies
Extended process of conformity assessment
 Especially for Class III and implantable devices
Insufficient evidence of compliance with GSPR – remediation and delay
Inadequate clinical data -- remediation and delay
Natural tendency of manufacturers to wait until last moment
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 20
Increased scrutiny, NB capacity, insufficient evidence from manufacturers
Significant Change – Notified Body Scrutiny
Articles 35-50 detail the significant changes to Notified Bodies, their designation and
continuous assessment
The process of designation of a Notified Body is complex and may take a long time
 All Notified Bodies are going through this process over ~18 months
 Notified Bodies could apply for designation beginning in November, 2017
Critical review of Notified Bodies assessments of technical documentation and
clinical evaluation documentation
 Translates into pain for industry in CER and technical documentation review
activities
Existing Notified Body structure will undergo a major change, expecting many
Notified Bodies impacted resulting in a potential significant reduction in available
Notified Bodies
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 21
22
Notified Body & Talent Capacity
 Now only 59 Notified Bodies – reduced from 83 prior to MDR publication
 Reducing number with Increased workload
 Unannounced inspections
o Manufacturers
o Critical suppliers
 Greater scrutiny in general (e.g. Clinical Evaluations)
 NBOG codes granted to Notified Bodies with MDR designation uncertain
• Will likely affect the scope of some Notified Bodies
 Notified Bodies competing with Manufacturers and Consultants for qualified people
 EN ISO 13485:2016 and MDSAP are competing for NB time and resources
 Some NBs cancelling contracts
 Some NBs not taking on new work
Status and availability of Notified Bodies is uncertain
©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc.
Significant Change – The Role of EUAR
CHANGE
 There are also more prescriptive requirements
placed on EU Authorized Representatives (EUAR)
(Article 11)
 The MDR states that “the authorised
representative should be jointly and severally
liable with the importer and the manufacturer“
 The EUAR will also be required to be registered
and must also have a PRRC
 The MDR extends the scope of the regulation
beyond the device manufacturer to other
Economic Operators in the supply chain
 Distributors and Importers are now specifically
regulated (Article 25) and have specific regulatory
obligations too
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 23
EU Authorized Representative
IMPLICATION FOR MANUFACTURER
Manufacturers should consider their
entire Supply Chain to ensure that these
provisions are adequately addressed
and agreed with their business partners.
In addition, they should identify a
person who is suitably qualified to
assume the PRRC role. Agreements with
EUARs, if used, should be revisited and
updated together ensuring that EUARs
have defined plans to be registered and
also have appropriate levels of
insurance to cover their new liability.
What about Brexit?
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 24
Still so many unknowns – difficult to forecast outcomes
UK policy makers and industry have indicated support for convergence with EU
device regulatory framework i.e. MDR
While the UK remains part of the EU (during transition period) the MDR will be
implemented
Post Brexit, UK regulatory framework will be required
ABHI (UK Trade Association) is working closely with MHRA and policy makers to
ensure harmonization is achieved with EU and rest of global markets
Still unclear regarding roles and responsibilities and impact to UK stakeholders –
some UK Notified Bodies are making plans to relocate Headquarters to EU
High % of total CE markings are issued by Notified Bodies based in UK
 Even higher % of CE markings for Class III devices
Majority of European Authorized Representatives are based in UK
After Brexit, they will no longer be in Europe
25
What should we do to implement MDR?
Some advice…..
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc.
Rule 8 Engage Senior Management internally
• Overall plan for timeline and resources
 Review all of your devices against revised classification rules
• Existing products and those under development
• Identify path to CE Mark for any devices with new classification
 Open a dialogue with your Notified Body
• Understand their intentions relative to designation as NB for MDR
• Confirm your interpretation of device classification
• Discuss scheduling considerations
• Agree on path forward and tentative schedule
 Discuss with your current European Authorized Representative
• Confirm their plans in light of new requirements
• Identify potential alternates
 Decide timeline for ISO 13485:2016 readiness
 Identify Person Responsible for Regulatory Compliance (PRRC)
26
What should we do to implement MDR?
Some advice…..
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc.
Rule 8 Revise Technical File for each product family
• Annex I – General Safety and Performance Requirements
• Create new checklist to replace “Essential Requirements Checklist”
 Revise and upgrade each Clinical Evaluation Report (CER)
• Complete and comprehensive – Start early!
• Critically examine the clinical data you have, and its adequacy
• Get help from internal and/or external SMEs
 Enhance Post Market Surveillance processes
• PMCF and PSUR
 Review labelling against expanded requirements
• Annex I Chapter III
• UDI
 Determine status of devices with regard to Restricted Substances
• If above threshold, decide alternate materials or strong risk/benefit rationale
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 27
Review your product portfolio for possible rationalization
Additional Recommended Activities
Rule 8 Perform a portfolio review to determine what products will remain in the European Market or
require CE Mark for country market use
 Develop exit plan for existing CE Marked devices where re-certification will not be pursued –
complete product EOL prior to certification expiration date
 Review all remaining devices against revised classification rules
• Review existing products and those under development, current MDD classification vs. new
MDR Classification
• Identify path to CE Mark for any devices with new classification
Finding business advantage in MDR
Compliance pioneers can be perceived as having superior systems
 “They’ve cracked it.”
Gain market share when laggards get trapped in bottleneck and fail
Product portfolio rationalization simplifies product support, reduces cost
 Accelerate end-of-life for marginal products
Strengthening technical documentation improves knowledge and confidence
 GSPR evidence
 Clinical evidence
Process of achieving MDR compliance will build new knowledge among key staff
 Internal study and “figuring it out”
 Transfer of knowledge from external resources
See Maetrics whitepaper “Market Opportunity Value of MDR”
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 28
Go early, exploit distance from slow adopters, maximize & leverage new knowledge
In Summary
Conformance with the MDR is a significant business challenge that must be
overcome in order to enjoy continued access to the EU market
To meet this challenge, manufacturers must be proactive and begin preparing now
Engage immediately with Notified Body regarding intentions and timeline
Manufacturers must closely follow the emerging regulatory landscape, keeping alert
for the detailed Implementing and Delegating Acts that are required in order to
make the MDR operational
The availability of resources, both internal and external, are likely to prove crucial to
effective implementation
Success will likely require strengthening of technical evidence documents
 GSPR and CER
Executive support and proactive program management of MDR implementation is a
prerequisite for a manufacturer’s success.
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 29
Questions & Close
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 30
BACKUP SLIDES
31
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc.
©2018 Maetrics. All Rights Reserved. ® Trademark of
Theoris Group, Inc. 3
How Can Maetrics Help You?
Our firm delivers compliance strategies that enable our clients to achieve new efficiencies and adjust to
compliance standards successfully across the life sciences industries’ product life cycle:
REGULATORY COMPLIANCE
 Global regulatory strategies
(FDA/EU/Other)
 Compliance audits & gap
analyses
 Mock inspections
 FDA warning letter responses &
remediation
 Facilities operations review
 CE Mark
REGULATORY AFFAIRS
 US Class III medical device
submissions; EU Class III MDD
submissions (change amendments,
design dossiers)
 Independent regulatory guidance
to product development teams
 Technical file remediation
 EU MDR / MDSAP / ISO 13485:2016
Compliance
QUALITY MANAGEMENT SYSTEMS
 Quality systems auditing
 Pre-certification auditing
 ISO certification readiness
 Batch records documentation
 Master record & technical file
remediation
 CAPA/NCR services
 Auditing, internal & external
 Complaint handling & recalls
PERFORMANCE IMPROVEMENT
 Kaizen event planning
 Mfg process improvement
 Operations/asset utilization (Mfg,
Inventory, DC & Transportation)
 Value Stream mapping & process
standardization & simplification
PRODUCT LIFE CYCLE MANAGEMENT
 Product risk management
framework & risk assessment
 Risk controls to drive results in
acceptance, mitigation, and
avoidance
 Design verification & activities
M&A INTEGRATION & DEAL SUPPORT
 Due diligence
 Regulatory compliance
 Quality Remediation & Operational
Integration
 Product Transfers
 Facility Consolidation
DATA ANALYTICS & REPORTING
 Performance analytics (Vision
Tool)
 Risk mitigation & compliance
awareness
Clinical &
Development
Manufacturing Distribution
Sales &
Marketing
CLINICAL EVALUATION REPORTS (CERS)
 CERs strategies & guidance against
Annex X MEDDEV 2.7/1 Rev. 4
 Data/literature protocol & search
 Full product CERs or updates to CERs
post review (medical writing)
 Internal CER process improvement
P&PC / VALIDATION
 Validation master plan
 Process & product validation
 Computer systems validation
 Equipment validation
 Special processes & procedures
 Test method validation
 Packaging
 Sterilization
TRAINING
 Portfolio of specialized courses in
regulatory, quality, risk
management, sterilization and
contamination control
MEDICAL DEVICES and DIAGNOSTICS
ORGANIZATIONAL CHANGE MGMT
 Due diligence
 Regulatory compliance
 Quality & Operational Integration
INFORMATION TECHNOLOGY
 Strategy & policy
 510(k) device software
development and testing
 Data migration validation
Note: This represents the highest volume of service offerings, but does not represent our entire service capability
EU MDR Preparation: Seize the Market Opportunity and Avoid the Bottleneck

More Related Content

What's hot

Medical Device Regulation (MDR) overview for Technion, May 25, 2021
Medical Device Regulation (MDR) overview for Technion, May 25, 2021Medical Device Regulation (MDR) overview for Technion, May 25, 2021
Medical Device Regulation (MDR) overview for Technion, May 25, 2021Levi Shapiro
 
Medical Devices Regulation (MDR) 2017/745 - Part I Purpose, Scope, Definitions
Medical Devices Regulation (MDR)  2017/745 - Part I Purpose, Scope, DefinitionsMedical Devices Regulation (MDR)  2017/745 - Part I Purpose, Scope, Definitions
Medical Devices Regulation (MDR) 2017/745 - Part I Purpose, Scope, DefinitionsArete-Zoe, LLC
 
The European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final textThe European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final textpi
 
20210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr202120210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr2021Annet Visscher
 
Medical Devices Regulation (MDR) 2017/745 - Annex II
Medical Devices Regulation (MDR)  2017/745 - Annex IIMedical Devices Regulation (MDR)  2017/745 - Annex II
Medical Devices Regulation (MDR) 2017/745 - Annex IIArete-Zoe, LLC
 
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...Greenlight Guru
 
8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program
8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program
8 Steps You Should Be Taking to Implement an EU-MDR Compliance ProgramGreenlight Guru
 
How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)Greenlight Guru
 
Medical Devices Regulation (MDR) 2017/745 - Annex XI, Conformity assessment,...
Medical Devices Regulation (MDR)  2017/745 - Annex XI, Conformity assessment,...Medical Devices Regulation (MDR)  2017/745 - Annex XI, Conformity assessment,...
Medical Devices Regulation (MDR) 2017/745 - Annex XI, Conformity assessment,...Arete-Zoe, LLC
 
Medical Devices Regulation (MDR) 2017/745 - Annex I
Medical Devices Regulation (MDR)  2017/745 - Annex I Medical Devices Regulation (MDR)  2017/745 - Annex I
Medical Devices Regulation (MDR) 2017/745 - Annex I Arete-Zoe, LLC
 
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...Levi Shapiro
 
Regulatory requirement and approval procedure for medical devices
Regulatory requirement and approval procedure for medical devicesRegulatory requirement and approval procedure for medical devices
Regulatory requirement and approval procedure for medical devicesAtul Bhombe
 
Medical Devices Regulation (MDR) 2017/745 - Clinical Evaluation & Post-Marke...
Medical Devices Regulation (MDR)  2017/745 - Clinical Evaluation & Post-Marke...Medical Devices Regulation (MDR)  2017/745 - Clinical Evaluation & Post-Marke...
Medical Devices Regulation (MDR) 2017/745 - Clinical Evaluation & Post-Marke...Arete-Zoe, LLC
 
Medical Devices Regulation (MDR) 2017/745 - Identification, traceability
Medical Devices Regulation (MDR)  2017/745 - Identification, traceability Medical Devices Regulation (MDR)  2017/745 - Identification, traceability
Medical Devices Regulation (MDR) 2017/745 - Identification, traceability Arete-Zoe, LLC
 
Medical device classification following MDR 2017/745
Medical device classification following MDR 2017/745Medical device classification following MDR 2017/745
Medical device classification following MDR 2017/745Monir EL AZZOUZI
 
The New EU MDR and What You Need to Know
The New EU MDR and What You Need to KnowThe New EU MDR and What You Need to Know
The New EU MDR and What You Need to KnowEMMAIntl
 

What's hot (20)

Medical Device Regulation (MDR) overview for Technion, May 25, 2021
Medical Device Regulation (MDR) overview for Technion, May 25, 2021Medical Device Regulation (MDR) overview for Technion, May 25, 2021
Medical Device Regulation (MDR) overview for Technion, May 25, 2021
 
Medical Devices Regulation (MDR) 2017/745 - Part I Purpose, Scope, Definitions
Medical Devices Regulation (MDR)  2017/745 - Part I Purpose, Scope, DefinitionsMedical Devices Regulation (MDR)  2017/745 - Part I Purpose, Scope, Definitions
Medical Devices Regulation (MDR) 2017/745 - Part I Purpose, Scope, Definitions
 
The European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final textThe European Medical Device Regulations - analysis of the final text
The European Medical Device Regulations - analysis of the final text
 
EU MDR
EU MDR EU MDR
EU MDR
 
Ghtf study group 4
Ghtf study group 4Ghtf study group 4
Ghtf study group 4
 
20210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr202120210413 nvfg acs iso14155 13_apr2021
20210413 nvfg acs iso14155 13_apr2021
 
Medical Devices Regulation (MDR) 2017/745 - Annex II
Medical Devices Regulation (MDR)  2017/745 - Annex IIMedical Devices Regulation (MDR)  2017/745 - Annex II
Medical Devices Regulation (MDR) 2017/745 - Annex II
 
Medical devices CHINA
Medical devices CHINA Medical devices CHINA
Medical devices CHINA
 
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...
The Top 10 Most Significant Changes Introduced by the New EU MDR (and how to ...
 
8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program
8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program
8 Steps You Should Be Taking to Implement an EU-MDR Compliance Program
 
How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)
 
Medical Devices Regulation (MDR) 2017/745 - Annex XI, Conformity assessment,...
Medical Devices Regulation (MDR)  2017/745 - Annex XI, Conformity assessment,...Medical Devices Regulation (MDR)  2017/745 - Annex XI, Conformity assessment,...
Medical Devices Regulation (MDR) 2017/745 - Annex XI, Conformity assessment,...
 
Medical Devices Regulation (MDR) 2017/745 - Annex I
Medical Devices Regulation (MDR)  2017/745 - Annex I Medical Devices Regulation (MDR)  2017/745 - Annex I
Medical Devices Regulation (MDR) 2017/745 - Annex I
 
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...
mHealth Israel_The New Regulatory Challenges in Europe The Clinical Evaluatio...
 
Regulatory requirement and approval procedure for medical devices
Regulatory requirement and approval procedure for medical devicesRegulatory requirement and approval procedure for medical devices
Regulatory requirement and approval procedure for medical devices
 
Medical Devices Regulation (MDR) 2017/745 - Clinical Evaluation & Post-Marke...
Medical Devices Regulation (MDR)  2017/745 - Clinical Evaluation & Post-Marke...Medical Devices Regulation (MDR)  2017/745 - Clinical Evaluation & Post-Marke...
Medical Devices Regulation (MDR) 2017/745 - Clinical Evaluation & Post-Marke...
 
Medical Devices Regulation (MDR) 2017/745 - Identification, traceability
Medical Devices Regulation (MDR)  2017/745 - Identification, traceability Medical Devices Regulation (MDR)  2017/745 - Identification, traceability
Medical Devices Regulation (MDR) 2017/745 - Identification, traceability
 
Medical device classification following MDR 2017/745
Medical device classification following MDR 2017/745Medical device classification following MDR 2017/745
Medical device classification following MDR 2017/745
 
The New EU MDR and What You Need to Know
The New EU MDR and What You Need to KnowThe New EU MDR and What You Need to Know
The New EU MDR and What You Need to Know
 
Medical devices
Medical devicesMedical devices
Medical devices
 

Similar to EU MDR Preparation: Seize the Market Opportunity and Avoid the Bottleneck

New medical device regulation: implications for medical device manufacturers
New medical device regulation: implications for medical device manufacturersNew medical device regulation: implications for medical device manufacturers
New medical device regulation: implications for medical device manufacturersPEPGRA Healthcare
 
PECB Webinar: Proposed changes for medical device quality management systems ...
PECB Webinar: Proposed changes for medical device quality management systems ...PECB Webinar: Proposed changes for medical device quality management systems ...
PECB Webinar: Proposed changes for medical device quality management systems ...PECB
 
New Medical Device Regulation Implications For Medical Device Manufacturers
New Medical Device Regulation  Implications For Medical Device ManufacturersNew Medical Device Regulation  Implications For Medical Device Manufacturers
New Medical Device Regulation Implications For Medical Device ManufacturersPEPGRA Healthcare
 
Q1 Medical Devices Regulation - practical consequences for manufacturers
Q1 Medical Devices Regulation - practical consequences for manufacturersQ1 Medical Devices Regulation - practical consequences for manufacturers
Q1 Medical Devices Regulation - practical consequences for manufacturersErik Vollebregt
 
Changing medical device regulations in Europe and the U.S.
Changing medical device regulations in Europe and the U.S.Changing medical device regulations in Europe and the U.S.
Changing medical device regulations in Europe and the U.S.Maetrics
 
Presentation: Custom-made medical devices: Information for the dental industry
Presentation: Custom-made medical devices: Information for the dental industryPresentation: Custom-made medical devices: Information for the dental industry
Presentation: Custom-made medical devices: Information for the dental industryTGA Australia
 
Prioritizing Documentation for MDR Transition Planning
Prioritizing Documentation for MDR Transition PlanningPrioritizing Documentation for MDR Transition Planning
Prioritizing Documentation for MDR Transition PlanningGreenlight Guru
 
Mdr transmission and business strategy of device
Mdr transmission and business strategy of deviceMdr transmission and business strategy of device
Mdr transmission and business strategy of deviceKishoreSteveAustin
 
IVDR Readiness Checklist
IVDR Readiness ChecklistIVDR Readiness Checklist
IVDR Readiness ChecklistGreenlight Guru
 
Key current and future regulatory challenges in the Medical Device and/or IVD...
Key current and future regulatory challenges in the Medical Device and/or IVD...Key current and future regulatory challenges in the Medical Device and/or IVD...
Key current and future regulatory challenges in the Medical Device and/or IVD...Diane Hatziioanou
 
Lancashire SME workshop - regulation and medical device workshop
Lancashire SME workshop - regulation and medical device workshopLancashire SME workshop - regulation and medical device workshop
Lancashire SME workshop - regulation and medical device workshopInnovation Agency
 
Free webinar on Unique Device Identification
Free webinar on Unique Device IdentificationFree webinar on Unique Device Identification
Free webinar on Unique Device Identificationnykathlen
 
Medical devices for biosimilars
Medical devices for biosimilarsMedical devices for biosimilars
Medical devices for biosimilarsDirk Kreder
 
Presentation: Conformity assessment evidence
Presentation: Conformity assessment evidencePresentation: Conformity assessment evidence
Presentation: Conformity assessment evidenceTGA Australia
 
mHealth Israel_MDR Bootcamp_Ulf Grundmann
mHealth Israel_MDR Bootcamp_Ulf GrundmannmHealth Israel_MDR Bootcamp_Ulf Grundmann
mHealth Israel_MDR Bootcamp_Ulf GrundmannLevi Shapiro
 
industrial cleaners market
industrial cleaners marketindustrial cleaners market
industrial cleaners marketsnehalkasat
 
industrial cleaners market
industrial cleaners marketindustrial cleaners market
industrial cleaners marketsnehalkasat
 
EU Medical Device Regulatory Framework_Dec, 2022
EU Medical Device Regulatory Framework_Dec, 2022EU Medical Device Regulatory Framework_Dec, 2022
EU Medical Device Regulatory Framework_Dec, 2022Levi Shapiro
 
UDI: What & How within the overall Medical Device value chain
UDI: What & How within the overall Medical Device value chainUDI: What & How within the overall Medical Device value chain
UDI: What & How within the overall Medical Device value chainnancykathlen
 

Similar to EU MDR Preparation: Seize the Market Opportunity and Avoid the Bottleneck (20)

New medical device regulation: implications for medical device manufacturers
New medical device regulation: implications for medical device manufacturersNew medical device regulation: implications for medical device manufacturers
New medical device regulation: implications for medical device manufacturers
 
PECB Webinar: Proposed changes for medical device quality management systems ...
PECB Webinar: Proposed changes for medical device quality management systems ...PECB Webinar: Proposed changes for medical device quality management systems ...
PECB Webinar: Proposed changes for medical device quality management systems ...
 
New Medical Device Regulation Implications For Medical Device Manufacturers
New Medical Device Regulation  Implications For Medical Device ManufacturersNew Medical Device Regulation  Implications For Medical Device Manufacturers
New Medical Device Regulation Implications For Medical Device Manufacturers
 
Q1 Medical Devices Regulation - practical consequences for manufacturers
Q1 Medical Devices Regulation - practical consequences for manufacturersQ1 Medical Devices Regulation - practical consequences for manufacturers
Q1 Medical Devices Regulation - practical consequences for manufacturers
 
Changing medical device regulations in Europe and the U.S.
Changing medical device regulations in Europe and the U.S.Changing medical device regulations in Europe and the U.S.
Changing medical device regulations in Europe and the U.S.
 
Presentation: Custom-made medical devices: Information for the dental industry
Presentation: Custom-made medical devices: Information for the dental industryPresentation: Custom-made medical devices: Information for the dental industry
Presentation: Custom-made medical devices: Information for the dental industry
 
Prioritizing Documentation for MDR Transition Planning
Prioritizing Documentation for MDR Transition PlanningPrioritizing Documentation for MDR Transition Planning
Prioritizing Documentation for MDR Transition Planning
 
Mdr transmission and business strategy of device
Mdr transmission and business strategy of deviceMdr transmission and business strategy of device
Mdr transmission and business strategy of device
 
IVDR Readiness Checklist
IVDR Readiness ChecklistIVDR Readiness Checklist
IVDR Readiness Checklist
 
Key current and future regulatory challenges in the Medical Device and/or IVD...
Key current and future regulatory challenges in the Medical Device and/or IVD...Key current and future regulatory challenges in the Medical Device and/or IVD...
Key current and future regulatory challenges in the Medical Device and/or IVD...
 
Lancashire SME workshop - regulation and medical device workshop
Lancashire SME workshop - regulation and medical device workshopLancashire SME workshop - regulation and medical device workshop
Lancashire SME workshop - regulation and medical device workshop
 
Free webinar on Unique Device Identification
Free webinar on Unique Device IdentificationFree webinar on Unique Device Identification
Free webinar on Unique Device Identification
 
Medical devices for biosimilars
Medical devices for biosimilarsMedical devices for biosimilars
Medical devices for biosimilars
 
Presentation: Conformity assessment evidence
Presentation: Conformity assessment evidencePresentation: Conformity assessment evidence
Presentation: Conformity assessment evidence
 
160428_WP_Medical_UDI_EN.PDF
160428_WP_Medical_UDI_EN.PDF160428_WP_Medical_UDI_EN.PDF
160428_WP_Medical_UDI_EN.PDF
 
mHealth Israel_MDR Bootcamp_Ulf Grundmann
mHealth Israel_MDR Bootcamp_Ulf GrundmannmHealth Israel_MDR Bootcamp_Ulf Grundmann
mHealth Israel_MDR Bootcamp_Ulf Grundmann
 
industrial cleaners market
industrial cleaners marketindustrial cleaners market
industrial cleaners market
 
industrial cleaners market
industrial cleaners marketindustrial cleaners market
industrial cleaners market
 
EU Medical Device Regulatory Framework_Dec, 2022
EU Medical Device Regulatory Framework_Dec, 2022EU Medical Device Regulatory Framework_Dec, 2022
EU Medical Device Regulatory Framework_Dec, 2022
 
UDI: What & How within the overall Medical Device value chain
UDI: What & How within the overall Medical Device value chainUDI: What & How within the overall Medical Device value chain
UDI: What & How within the overall Medical Device value chain
 

More from April Bright

The Future of Digital Health and Wearables in Orthopedicsrables
The Future of Digital Health and Wearables in OrthopedicsrablesThe Future of Digital Health and Wearables in Orthopedicsrables
The Future of Digital Health and Wearables in OrthopedicsrablesApril Bright
 
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...April Bright
 
Innovation in Orthopedics: Surgeon Perspectives
Innovation in Orthopedics:  Surgeon PerspectivesInnovation in Orthopedics:  Surgeon Perspectives
Innovation in Orthopedics: Surgeon PerspectivesApril Bright
 
Antimicrobial Coatings: The Research and Regulatory Perspective
Antimicrobial Coatings: The Research and Regulatory PerspectiveAntimicrobial Coatings: The Research and Regulatory Perspective
Antimicrobial Coatings: The Research and Regulatory PerspectiveApril Bright
 
Leverage These Effective Communication Skills to Get Your Message Across
Leverage These Effective Communication Skills to Get Your Message AcrossLeverage These Effective Communication Skills to Get Your Message Across
Leverage These Effective Communication Skills to Get Your Message AcrossApril Bright
 
Operations: Top Reasons for Long Lead Times and What to Do About Them
Operations: Top Reasons for Long Lead Times and What to Do About ThemOperations: Top Reasons for Long Lead Times and What to Do About Them
Operations: Top Reasons for Long Lead Times and What to Do About ThemApril Bright
 
Computational Modeling & Simulation in Orthopedics: Tools to Comply in an Ev...
Computational Modeling & Simulation in Orthopedics:  Tools to Comply in an Ev...Computational Modeling & Simulation in Orthopedics:  Tools to Comply in an Ev...
Computational Modeling & Simulation in Orthopedics: Tools to Comply in an Ev...April Bright
 
Joint Replacement: The Current and Future Impact of Coatings
Joint Replacement: The Current and Future Impact of CoatingsJoint Replacement: The Current and Future Impact of Coatings
Joint Replacement: The Current and Future Impact of CoatingsApril Bright
 
Engineers: Practical Application of Project Management Principles
Engineers: Practical Application of Project Management PrinciplesEngineers: Practical Application of Project Management Principles
Engineers: Practical Application of Project Management PrinciplesApril Bright
 
Regulatory and Quality Affairs: Answers to FDA and ISO Gray Areas
Regulatory and Quality Affairs: Answers to FDA and ISO Gray AreasRegulatory and Quality Affairs: Answers to FDA and ISO Gray Areas
Regulatory and Quality Affairs: Answers to FDA and ISO Gray AreasApril Bright
 
The Future of Orthobiologics in Trauma Procedures
The Future of Orthobiologics in Trauma ProceduresThe Future of Orthobiologics in Trauma Procedures
The Future of Orthobiologics in Trauma ProceduresApril Bright
 
Spine Implants: Porous Coatings vs. Porous Materials vs. Additive Manufacturing
Spine Implants:  Porous Coatings vs. Porous Materials vs. Additive ManufacturingSpine Implants:  Porous Coatings vs. Porous Materials vs. Additive Manufacturing
Spine Implants: Porous Coatings vs. Porous Materials vs. Additive ManufacturingApril Bright
 
How to Influence People: The Value of Employee Engagement
How to Influence People: The Value of Employee EngagementHow to Influence People: The Value of Employee Engagement
How to Influence People: The Value of Employee EngagementApril Bright
 
Real-World Evidence: The Future of Data Generation and Usage
Real-World Evidence: The Future of Data Generation and UsageReal-World Evidence: The Future of Data Generation and Usage
Real-World Evidence: The Future of Data Generation and UsageApril Bright
 
Orthopedic Coatings: Predictions for 2025
Orthopedic Coatings: Predictions for 2025Orthopedic Coatings: Predictions for 2025
Orthopedic Coatings: Predictions for 2025April Bright
 
Engineers: Apply Automation to Increase Quality, Speed to Market
Engineers: Apply Automation to Increase Quality, Speed to MarketEngineers: Apply Automation to Increase Quality, Speed to Market
Engineers: Apply Automation to Increase Quality, Speed to MarketApril Bright
 
OSMA: Orthopedic Industry's Top Regulatory Challenges and Opportunities
OSMA: Orthopedic Industry's Top Regulatory Challenges and OpportunitiesOSMA: Orthopedic Industry's Top Regulatory Challenges and Opportunities
OSMA: Orthopedic Industry's Top Regulatory Challenges and OpportunitiesApril Bright
 
Unique Device Identification: Manufacturer, Hospital and Global Implications
Unique Device Identification: Manufacturer, Hospital and Global ImplicationsUnique Device Identification: Manufacturer, Hospital and Global Implications
Unique Device Identification: Manufacturer, Hospital and Global ImplicationsApril Bright
 
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018Additive Manufacturing - Mechanical Test Methods - OMTEC 2018
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018April Bright
 
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...April Bright
 

More from April Bright (20)

The Future of Digital Health and Wearables in Orthopedicsrables
The Future of Digital Health and Wearables in OrthopedicsrablesThe Future of Digital Health and Wearables in Orthopedicsrables
The Future of Digital Health and Wearables in Orthopedicsrables
 
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...
The Future of Personalized Implants in Joint Replacement: Additive, Robotics,...
 
Innovation in Orthopedics: Surgeon Perspectives
Innovation in Orthopedics:  Surgeon PerspectivesInnovation in Orthopedics:  Surgeon Perspectives
Innovation in Orthopedics: Surgeon Perspectives
 
Antimicrobial Coatings: The Research and Regulatory Perspective
Antimicrobial Coatings: The Research and Regulatory PerspectiveAntimicrobial Coatings: The Research and Regulatory Perspective
Antimicrobial Coatings: The Research and Regulatory Perspective
 
Leverage These Effective Communication Skills to Get Your Message Across
Leverage These Effective Communication Skills to Get Your Message AcrossLeverage These Effective Communication Skills to Get Your Message Across
Leverage These Effective Communication Skills to Get Your Message Across
 
Operations: Top Reasons for Long Lead Times and What to Do About Them
Operations: Top Reasons for Long Lead Times and What to Do About ThemOperations: Top Reasons for Long Lead Times and What to Do About Them
Operations: Top Reasons for Long Lead Times and What to Do About Them
 
Computational Modeling & Simulation in Orthopedics: Tools to Comply in an Ev...
Computational Modeling & Simulation in Orthopedics:  Tools to Comply in an Ev...Computational Modeling & Simulation in Orthopedics:  Tools to Comply in an Ev...
Computational Modeling & Simulation in Orthopedics: Tools to Comply in an Ev...
 
Joint Replacement: The Current and Future Impact of Coatings
Joint Replacement: The Current and Future Impact of CoatingsJoint Replacement: The Current and Future Impact of Coatings
Joint Replacement: The Current and Future Impact of Coatings
 
Engineers: Practical Application of Project Management Principles
Engineers: Practical Application of Project Management PrinciplesEngineers: Practical Application of Project Management Principles
Engineers: Practical Application of Project Management Principles
 
Regulatory and Quality Affairs: Answers to FDA and ISO Gray Areas
Regulatory and Quality Affairs: Answers to FDA and ISO Gray AreasRegulatory and Quality Affairs: Answers to FDA and ISO Gray Areas
Regulatory and Quality Affairs: Answers to FDA and ISO Gray Areas
 
The Future of Orthobiologics in Trauma Procedures
The Future of Orthobiologics in Trauma ProceduresThe Future of Orthobiologics in Trauma Procedures
The Future of Orthobiologics in Trauma Procedures
 
Spine Implants: Porous Coatings vs. Porous Materials vs. Additive Manufacturing
Spine Implants:  Porous Coatings vs. Porous Materials vs. Additive ManufacturingSpine Implants:  Porous Coatings vs. Porous Materials vs. Additive Manufacturing
Spine Implants: Porous Coatings vs. Porous Materials vs. Additive Manufacturing
 
How to Influence People: The Value of Employee Engagement
How to Influence People: The Value of Employee EngagementHow to Influence People: The Value of Employee Engagement
How to Influence People: The Value of Employee Engagement
 
Real-World Evidence: The Future of Data Generation and Usage
Real-World Evidence: The Future of Data Generation and UsageReal-World Evidence: The Future of Data Generation and Usage
Real-World Evidence: The Future of Data Generation and Usage
 
Orthopedic Coatings: Predictions for 2025
Orthopedic Coatings: Predictions for 2025Orthopedic Coatings: Predictions for 2025
Orthopedic Coatings: Predictions for 2025
 
Engineers: Apply Automation to Increase Quality, Speed to Market
Engineers: Apply Automation to Increase Quality, Speed to MarketEngineers: Apply Automation to Increase Quality, Speed to Market
Engineers: Apply Automation to Increase Quality, Speed to Market
 
OSMA: Orthopedic Industry's Top Regulatory Challenges and Opportunities
OSMA: Orthopedic Industry's Top Regulatory Challenges and OpportunitiesOSMA: Orthopedic Industry's Top Regulatory Challenges and Opportunities
OSMA: Orthopedic Industry's Top Regulatory Challenges and Opportunities
 
Unique Device Identification: Manufacturer, Hospital and Global Implications
Unique Device Identification: Manufacturer, Hospital and Global ImplicationsUnique Device Identification: Manufacturer, Hospital and Global Implications
Unique Device Identification: Manufacturer, Hospital and Global Implications
 
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018Additive Manufacturing - Mechanical Test Methods - OMTEC 2018
Additive Manufacturing - Mechanical Test Methods - OMTEC 2018
 
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...
Analyze and Optimize Your Supply Chain Operations for Higher Performance - OM...
 

Recently uploaded

Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...narwatsonia7
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 

Recently uploaded (20)

Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 

EU MDR Preparation: Seize the Market Opportunity and Avoid the Bottleneck

  • 1.
  • 2. EU MDR Preparation OMTEC Mike Wolf CBA, CQA, RAC Chicago, IL 13 June 2018 ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 2
  • 3. What is the new EU MDR?  Most significant regulatory change in Europe in over 20 years!  Compliance to the EU MDR required to gain and retain CE Mark  80+ delegating or implementing acts to come  Notified Bodies to be re-designated  Focus on Clinical Evidence and Post Market Clinical Follow-Up and Surveillance  Scrutiny and review of high risk devices  Improved transparency between NBs, CAs, Mfgs, and Users (UDI and EUDAMED) ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 3 The new European Union Medical Device Regulation (EU MDR) AIMD 17 Articles 35 Pages 9 Annexes MDD 20 Articles 60 Pages 12 Annexes Directives Regulation MDR 123 Articles 178 Pages 17 Annexes
  • 4. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 4 Key Players - MDR EU National Competent Authorities Notified Bodies Reviews Testing Auditing Clinical Evaluations Records Data QMS Certificate CE Mark Medical Device Coordination Group (MDCG) Expert Panels Suppliers Importer Distributors Medical Devices with UDI Authorized Representative Manufacturer Public EUDAMED = Economic Operator PMS/PMCF Reporting = New or Changed Element MDR Person Responsible for Regulatory Compliance
  • 5. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 5 Timelines – how long do we have? Transition Timeline 2017 [May 25] EU MDR Entered into Force 2017 [Nov 25] Notified Bodies apply for MDR Designation 2020 [May 25] Notified Bodies stop issuing MDD Certs 2017 2018 2019 2020 2024 2020 [May 26] New or Renewing Devices must have MDR certs MDD Certs Issued Existing MDD Certs Valid May 26, 2020 Date of Application May 26, 2024 Any Remaining MDD Certs Voided May 26, 2025 No further placing MDD-certified products on the market MDR Certs Issued
  • 6. 6 Timeline Convergence – Competing Transitions ISO 13485:2016 Both ISO 13485:2003 or 2016 Certifications New ISO 13485:2003 Certifications Discouraged No 13485:2003 Certifications or Re-Certifications Allowed 01 Mar 2016 01 Mar 2017 01 Mar 2018 01 Mar 2019 MDSAP 01 Jan 2016 01 Jan 2017 01 Jan 2018 01 Jan 2019 MDSAP Audit Certified Mandatory by Health Canada EU MDR 26 May 2020 Present Day New or Renewing Devices must have MDR certificates ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc.
  • 7. Important for orthopedic device manufacturers Revised device classification rules Extended process of conformance assessment by NB and MDCG Increased scrutiny of technical documentation - Class III and Implantable New labeling requirements Restriction of hazardous substances Risks associated with interaction between software and the IT environment within which it operates Conformance with harmonized standards  And Common Specifications yet to be published – 26 May 2020 Clinical evaluation and adequate clinical data Particular Requirements for Active Implantable Medical Devices  Devices and their component parts shall be identifiable  Shall bear a code by which the device and manufacturer can be unequivocally identified ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 7
  • 8. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 8 Significant Change – Legacy Products CHANGE  There is no provision for grandfathering certification (CE marks) obtained under the previous Directives  All products have to be CE marked under the new Regulation 2017/745, in order to be placed on the market or put into service (MDR Article 5), after the transition period Market Access of Legacy Products IMPLICATION FOR MANUFACTURER A comprehensive plan needs to be put in place to ensure that all products that will be maintained on the EU market are CE marked in accordance with the full requirements of the new MDR. This should include products currently under development. This review may provide an opportunity for rationalization of the product portfolio and elimination of any marginal products.
  • 9. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 9 Significant Change - Reclassification CHANGE  Certain products have received special consideration in the MDR and are subject to reclassification. (MDR Article 4)  These new provisions variously apply to cosmetic implants, standalone software, products without an intended medical purpose (MDR Annex XVI), certain spine products and reusable Class 1 devices Reclassification of Devices IMPLICATION FOR MANUFACTURER Manufacturers of medical devices should carefully examine the MDR Classification rules in Annex VIII to determine whether new conformity assessment routes are now applicable to their product portfolio. If so, they should engage their Notified Body (where necessary) and take steps to evaluate the necessary timescales involved in implementing this change. In some cases, where reclassification has occurred, this may be the first time that such relationships and agreements with Notified Bodies have been required.
  • 10. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 10 Significant Change - Reclassification CHANGE  Changes in the classification rules of medical devices might lead to reclassification of certain medical devices  Rules organized by ‒ Non-Invasive Devices (Rules 1 – 4) ‒ Invasive Devices (Rules 5 – 8) ‒ Active Devices (Rules 9 – 13) ‒ Special Rules (Rules 14 – 22)  There is a process to handle classifications disputes ‒ Steps depend on if Manufacturer and notified body are in the same EU Member State ‒ May involve consultation with competent authority of the member state Check new classification rules Classification Change Examples • In vitro contact with cells/embryos going back into the body to be Class III (or IIa) • Apheresis devices to be Class III • IVF and ART non-invasive devices can be IIb • Spinal implants to be Class III • Total and partial joint replacements to be Class III • Devices recording diagnostic images to be IIa • Nanomaterial devices to be Class III • AIMD accessories to be Class III • Devices which are intended to be introduced into the human body via a body orifice, or applied on skin and that are absorbed by or locally dispersed in the human body are to be Class III • Reusable surgical Instruments are no longer class I
  • 11. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 11 Significant Change – General Safety & Performance Requirements CHANGE The spirit of the “Essential Requirements” has not changed, however there are new provisions and more details required (Annex I), for example:  Hazardous substances  Devices with non-viable human cells or tissues, and other “non-viable biological substances”  IT security and prevention of unauthorized access  Risk Management  Labeling  Alignment with Global Harmonization Task Force (GHTF) IMPLICATION FOR MANUFACTURER Manufacturers need to review and update technical documentation in preparation for a complete conformity assessment against requirements of GSPR. Evidence of conformance with each GSPR should be critically assessed. Emphasis on improved risk management and risk controls will also need to be incorporated in practices and processes. Some new requirements – others just more detailed
  • 12. Significant Change – Technical Documentation CHANGE The MDR is significantly more prescriptive about the required content of technical documentation (Technical File/Design Dossier):  Essential Requirements (ERs) are replaced by “General Safety Requirements (Annex I) and the number of requirements has been expanded  A Presumption of conformity still applies for devices that are in conformity with relevant harmonised standards (Article 8), but the Commission may define Common Specifications where no harmonized standards exist or where they are considered insufficient Additionally, there are more detailed requirements for the Quality Management Systems (QMS) (Annex IX)  EN ISO 13485:2016 was re-written and issued in 2016 with the new MDR very much in mind  It should be noted that while EN ISO 13485:2016 is not an absolute requirement there will be a general expectation that this standard will be used ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 12 Technical Documentation
  • 13. Significant Change – Technical Documentation  Technical Documents shall include:  Device description and specification  Reference to previous and similar generations of device  Labels & IFUs  Design and Manufacturing information  Demonstration of conformance to General Safety and Performance Requirements  Benefit-Risk analysis and Risk Management  Product verification and validation  Additional Requirements for Technical Documents on Post Market Surveillance  Post Market Surveillance Plan, per Article 84  Periodic Safety Update Report, per Article 86  Post Market Surveillance Report, per Article 85 ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 13 Detailed Requirements for Technical Documents
  • 14. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 14 Significant Change – Vigilance and PMS CHANGE  Under the new Regulation 2017/745, device manufacturers will be required to collect post-market clinical data as part of their on-going assessment of potential safety risks  Post Market Clinical Follow up (PMCF) is a continuous process with the objective of constantly updating the clinical evaluation (Annex XIV, Part B)  Additionally, reporting timeframes are tightened from 30 days to 15 days for reporting serious incidents (Article 87) Vigilance and Post Market Surveillance IMPLICATION FOR MANUFACTURER Manufacturers need to review their procedures for PMS and ensure that the responsibility for the provision of this additional data and associated support is clearly established. This is likely to require additional resources in functions that support products on the market such as Regulatory and Medical Affairs.  There will be new electronic vigilance reporting (Article 92) and Periodic Safety Update Reports (PSUR) for all devices (Article 86) subject to differing frequency and submission requirements
  • 15. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 15 Significant Change – Vigilance and PMS Summary of safety and clinical performance  Draft is part of submission to notified body for conformity assessment  Is validated by the notified body  After validation, is uploaded to EUDAMED Two new deliverables required by the EU MDR Periodic safety update report (PSUR)  Throughout the lifetime of the device concerned, the PSUR shall set out ‒ Conclusions of the benefit-risk determination ‒ Main findings of the PMCF ‒ Volume of sales of the device and estimated evaluation of the size and other characteristics of the population using device, and where practicable, the usage frequency
  • 16. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 16 Significant Change – Clinical Evaluations CHANGE  The MDR is more specific about the need for clinical evidence and clinical evaluation, proportionate with the risk associated with a given device (Annex XIV, Part A)  Reliance on the scientific literature to demonstrate equivalence will be more tightly regulated, and clinical evaluations will be more closely aligned with clinical trials associated with medicinal products Clinical Evidence and Clinical Evaluations IMPLICATION FOR MANUFACTURER This may require manufacturers to obtain additional clinical data from clinical studies. There will be additional scrutiny of Clinical Evaluation Reports (CERs) by Notified Bodies as outlined by new guidance (MEDDEV 2.7.1 rev. 4). CERs that previously were of a suitable level and standard and accepted by Notified Bodies, may no longer be accepted. Manufacturers should plan to review all of their CERs if not reviewed within the last 1-2 years and ensure that CERs include Post Market Surveillance data; especially where devices are new to market and where clinical data was limited at the time of first CE Marking.
  • 17. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 17 Significant Change – Labeling CHANGE Requirements for product labeling are more prescriptive  For example, information supplied by the manufacturer shall be made available and kept up to date on the manufacturer’s website (Chapter III, 23.1) IMPLICATION FOR MANUFACTURER Manufacturers should carefully review the adequacy of their product labelling and precautionary statements and consider how this will be reflected on their websites. There are requirements for  Specific details for labels and for sterile packages (Annex I).  Inclusion of information on residual risks for vulnerable patient groups (e.g. children, pregnant or nursing women) and, if applicable, on appropriate precautionary measures in the instructions for use.  Hazardous Substances All EU labeling and IFU’s likely to be impacted
  • 18. 18 Significant Change - EUDAMED What is EUDAMED? ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. • SU  EUDAMED is a central European databank that will be accessible to notified bodies, economic operators, sponsors, and to the public to the extent specified  It will include protections for personal data  What’s driving EUDAMED?  Article 14a (MDD 93/42/EEC)  Clinical investigations  Vigilance and incident reporting  Status of conformity on issued certificates  Registration of manufacturers and EUAR
  • 19. Technical Docs to MDR Certification Timeline ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 19 Many factors impact the timeline for transition of existing product submissions MDCG Consultation for Class III & Implantable Devices Only Technical Documentation Updates for MDR likely to include: • Design Requirements traced to new Safety and Performance Requirements • Risk Management File/Report • Clinical Evaluation Plan/Report • Gaps in Clinical Evidence • Any equivalence to competitive device data replaced with company’s own data • Post Market Surveillance Reporting • PMS/PMCF Plan/Report, PSUR/SSCP Reports (when needed + translations) • Labeling +UDI + Authorized Rep/Branding Changes + Translations • Declaration of Conformity + UDI inspections in place at DCs Technical Documentation Update (? weeks) Notified Body Notification & Planning (? weeks) First evaluation by Notified Body (40 days) Response (15 days) Second evaluation (20 days) MDCG (21 days) Scrutiny (60 days) Certification decision (15 days) Submission required for: • First time products • Change to products
  • 20. What factors will slow down the process? Delays in Notified Bodies receiving their designation to MDR Notified Body capacity constraints Additional scrutiny of Notified Body performance Additional scrutiny of technical documentation by Notified Bodies Extended process of conformity assessment  Especially for Class III and implantable devices Insufficient evidence of compliance with GSPR – remediation and delay Inadequate clinical data -- remediation and delay Natural tendency of manufacturers to wait until last moment ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 20 Increased scrutiny, NB capacity, insufficient evidence from manufacturers
  • 21. Significant Change – Notified Body Scrutiny Articles 35-50 detail the significant changes to Notified Bodies, their designation and continuous assessment The process of designation of a Notified Body is complex and may take a long time  All Notified Bodies are going through this process over ~18 months  Notified Bodies could apply for designation beginning in November, 2017 Critical review of Notified Bodies assessments of technical documentation and clinical evaluation documentation  Translates into pain for industry in CER and technical documentation review activities Existing Notified Body structure will undergo a major change, expecting many Notified Bodies impacted resulting in a potential significant reduction in available Notified Bodies ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 21
  • 22. 22 Notified Body & Talent Capacity  Now only 59 Notified Bodies – reduced from 83 prior to MDR publication  Reducing number with Increased workload  Unannounced inspections o Manufacturers o Critical suppliers  Greater scrutiny in general (e.g. Clinical Evaluations)  NBOG codes granted to Notified Bodies with MDR designation uncertain • Will likely affect the scope of some Notified Bodies  Notified Bodies competing with Manufacturers and Consultants for qualified people  EN ISO 13485:2016 and MDSAP are competing for NB time and resources  Some NBs cancelling contracts  Some NBs not taking on new work Status and availability of Notified Bodies is uncertain ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc.
  • 23. Significant Change – The Role of EUAR CHANGE  There are also more prescriptive requirements placed on EU Authorized Representatives (EUAR) (Article 11)  The MDR states that “the authorised representative should be jointly and severally liable with the importer and the manufacturer“  The EUAR will also be required to be registered and must also have a PRRC  The MDR extends the scope of the regulation beyond the device manufacturer to other Economic Operators in the supply chain  Distributors and Importers are now specifically regulated (Article 25) and have specific regulatory obligations too ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 23 EU Authorized Representative IMPLICATION FOR MANUFACTURER Manufacturers should consider their entire Supply Chain to ensure that these provisions are adequately addressed and agreed with their business partners. In addition, they should identify a person who is suitably qualified to assume the PRRC role. Agreements with EUARs, if used, should be revisited and updated together ensuring that EUARs have defined plans to be registered and also have appropriate levels of insurance to cover their new liability.
  • 24. What about Brexit? ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 24 Still so many unknowns – difficult to forecast outcomes UK policy makers and industry have indicated support for convergence with EU device regulatory framework i.e. MDR While the UK remains part of the EU (during transition period) the MDR will be implemented Post Brexit, UK regulatory framework will be required ABHI (UK Trade Association) is working closely with MHRA and policy makers to ensure harmonization is achieved with EU and rest of global markets Still unclear regarding roles and responsibilities and impact to UK stakeholders – some UK Notified Bodies are making plans to relocate Headquarters to EU High % of total CE markings are issued by Notified Bodies based in UK  Even higher % of CE markings for Class III devices Majority of European Authorized Representatives are based in UK After Brexit, they will no longer be in Europe
  • 25. 25 What should we do to implement MDR? Some advice….. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. Rule 8 Engage Senior Management internally • Overall plan for timeline and resources  Review all of your devices against revised classification rules • Existing products and those under development • Identify path to CE Mark for any devices with new classification  Open a dialogue with your Notified Body • Understand their intentions relative to designation as NB for MDR • Confirm your interpretation of device classification • Discuss scheduling considerations • Agree on path forward and tentative schedule  Discuss with your current European Authorized Representative • Confirm their plans in light of new requirements • Identify potential alternates  Decide timeline for ISO 13485:2016 readiness  Identify Person Responsible for Regulatory Compliance (PRRC)
  • 26. 26 What should we do to implement MDR? Some advice….. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. Rule 8 Revise Technical File for each product family • Annex I – General Safety and Performance Requirements • Create new checklist to replace “Essential Requirements Checklist”  Revise and upgrade each Clinical Evaluation Report (CER) • Complete and comprehensive – Start early! • Critically examine the clinical data you have, and its adequacy • Get help from internal and/or external SMEs  Enhance Post Market Surveillance processes • PMCF and PSUR  Review labelling against expanded requirements • Annex I Chapter III • UDI  Determine status of devices with regard to Restricted Substances • If above threshold, decide alternate materials or strong risk/benefit rationale
  • 27. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 27 Review your product portfolio for possible rationalization Additional Recommended Activities Rule 8 Perform a portfolio review to determine what products will remain in the European Market or require CE Mark for country market use  Develop exit plan for existing CE Marked devices where re-certification will not be pursued – complete product EOL prior to certification expiration date  Review all remaining devices against revised classification rules • Review existing products and those under development, current MDD classification vs. new MDR Classification • Identify path to CE Mark for any devices with new classification
  • 28. Finding business advantage in MDR Compliance pioneers can be perceived as having superior systems  “They’ve cracked it.” Gain market share when laggards get trapped in bottleneck and fail Product portfolio rationalization simplifies product support, reduces cost  Accelerate end-of-life for marginal products Strengthening technical documentation improves knowledge and confidence  GSPR evidence  Clinical evidence Process of achieving MDR compliance will build new knowledge among key staff  Internal study and “figuring it out”  Transfer of knowledge from external resources See Maetrics whitepaper “Market Opportunity Value of MDR” ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 28 Go early, exploit distance from slow adopters, maximize & leverage new knowledge
  • 29. In Summary Conformance with the MDR is a significant business challenge that must be overcome in order to enjoy continued access to the EU market To meet this challenge, manufacturers must be proactive and begin preparing now Engage immediately with Notified Body regarding intentions and timeline Manufacturers must closely follow the emerging regulatory landscape, keeping alert for the detailed Implementing and Delegating Acts that are required in order to make the MDR operational The availability of resources, both internal and external, are likely to prove crucial to effective implementation Success will likely require strengthening of technical evidence documents  GSPR and CER Executive support and proactive program management of MDR implementation is a prerequisite for a manufacturer’s success. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 29
  • 30. Questions & Close ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 30
  • 31. BACKUP SLIDES 31 ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc.
  • 32. ©2018 Maetrics. All Rights Reserved. ® Trademark of Theoris Group, Inc. 3 How Can Maetrics Help You? Our firm delivers compliance strategies that enable our clients to achieve new efficiencies and adjust to compliance standards successfully across the life sciences industries’ product life cycle: REGULATORY COMPLIANCE  Global regulatory strategies (FDA/EU/Other)  Compliance audits & gap analyses  Mock inspections  FDA warning letter responses & remediation  Facilities operations review  CE Mark REGULATORY AFFAIRS  US Class III medical device submissions; EU Class III MDD submissions (change amendments, design dossiers)  Independent regulatory guidance to product development teams  Technical file remediation  EU MDR / MDSAP / ISO 13485:2016 Compliance QUALITY MANAGEMENT SYSTEMS  Quality systems auditing  Pre-certification auditing  ISO certification readiness  Batch records documentation  Master record & technical file remediation  CAPA/NCR services  Auditing, internal & external  Complaint handling & recalls PERFORMANCE IMPROVEMENT  Kaizen event planning  Mfg process improvement  Operations/asset utilization (Mfg, Inventory, DC & Transportation)  Value Stream mapping & process standardization & simplification PRODUCT LIFE CYCLE MANAGEMENT  Product risk management framework & risk assessment  Risk controls to drive results in acceptance, mitigation, and avoidance  Design verification & activities M&A INTEGRATION & DEAL SUPPORT  Due diligence  Regulatory compliance  Quality Remediation & Operational Integration  Product Transfers  Facility Consolidation DATA ANALYTICS & REPORTING  Performance analytics (Vision Tool)  Risk mitigation & compliance awareness Clinical & Development Manufacturing Distribution Sales & Marketing CLINICAL EVALUATION REPORTS (CERS)  CERs strategies & guidance against Annex X MEDDEV 2.7/1 Rev. 4  Data/literature protocol & search  Full product CERs or updates to CERs post review (medical writing)  Internal CER process improvement P&PC / VALIDATION  Validation master plan  Process & product validation  Computer systems validation  Equipment validation  Special processes & procedures  Test method validation  Packaging  Sterilization TRAINING  Portfolio of specialized courses in regulatory, quality, risk management, sterilization and contamination control MEDICAL DEVICES and DIAGNOSTICS ORGANIZATIONAL CHANGE MGMT  Due diligence  Regulatory compliance  Quality & Operational Integration INFORMATION TECHNOLOGY  Strategy & policy  510(k) device software development and testing  Data migration validation Note: This represents the highest volume of service offerings, but does not represent our entire service capability