NON-INVASIVE AND INVASIVE BP
MONITORING
By Dr. S.A.Fatima
Dr. Arsalan A. Zaki
NON-INVASIVE BP MONITORING
WHAT DO WE NEED TO KNOW?
• Techniques
• Principles
• Indications
• Contraindications
• Limitations
• Complications
• cautions.
NON INVASIVE BP MONITORING TECHNIQUES
1. Manual intermittent techniques.
• Palpation
• Auscultation
• Doppler probe
2. Automated intermittent techniques.
• Oscillometry
3. Automated continuous techniques.
• Tonometry
• photoplathysmography
PRINCIPLES OF DIFFERENT NON-INVASIVE
TECHNIQUES
• Palpation: systolic BP is measured by occluding the artery with
cuff
• Auscultation: turbulent flow of blood with partially occluded
artery produces Korotkoff sounds.
• Doppler: movement of RBCs produces a Doppler frequency
shift.
• Oscillometry: oscillations in the pressure amplitude are
measured in an air filled cuff that slowly deflates from the
pressure well in excess of pressure to occlude the artery.
• Tonometery: sense the pressure required to partially flatten the
superficial artery.
• Phoyoplathysmography: quantification of infrared light
INDICATIONS OF NIBP MONITORING
• Any procedure requiring any sort of anesthesia.
• When invasive monitoring is contraindicated.
• Intermittent measurement interval should be 3 to 5 minutes.
CONTRAINDICATIONS:
• Cuff based BP monitoring is contraindicated in
Limb with axillary clearance
Limb with AV fistula for dialysis purpose.
Limb with PICC line or peripheral IV line.
Patients with known focal vascular anomalies.
Burnt extremity.
LIMITATIONS
• Palpation:
• only detects systolic BP.
• Not useful for continued monitoring.
• Not reliable for conditions of low flow.
• Auscultation:
• MAP cannot be measured directly and has to be calculated by formula
MAP=SBP+ 2(DBP)/3
• Not useful for continued monitoring.
• Interpersonal variability of hearing.
• Not reliable for conditions of low flow.
LIMITATIONS
• Ocillometery:
• cuff size is of prime importance in accuracy of the BP.(bladder length 80% of arm ; bladder width 40%
of arm circumference)
• Interpretation varies according to site.
• Underestimate MAP during hypertension and overestimate MAP during hypotension so not reliable
in unstable patients.
• Underestimates SBP and overestimates DBP.
• Dysrhythmias lead to inaccuracy of BP.
Tonomertery
• Need frequent calibration.
• Movement artifacts common.
• Limited applicability in critically ill patients.
COMPLICATIONS:
• Pain
• Patechiae and ecchymoses
• Limb edema
• Venous stasis and thrombophlebitis.
• Peripheral neuropathy (radial nerve common).
• Compartment syndrome.
• CAUSIONS:
• Patients with known peripheral neuropathy.
• Arterial or venous insufficiency.
• Severe coagulopathies.
• Recet use of thrombolytic therapy.
INVASIVE BP MONITORING
WHAT DO WE NEED TO KNOW?
• Basic principle
• Indications
• Percutaneous radial artery cannulation
• Other arteries which can be uses
• Complications
• Components
• Leveling and zeroing
• Normal IBP waveforms
• Abnormal IBP waveforms
• Waveform analysis for prediction of volume responsiveness
PRINCIPLE
• Intra-arterial blood pressure monitoring uses fluid filled
tubing to transmit the force of the pressure pulse wave to a pressure
transducer that converts the displacement of a silicon crystal into voltage
changes.
• These electrical signals are amplified, filtered, and displayed as the
arterial pressure trace.
INDICATIONS:
• Continuos, real time blood pressure monitoring.
• Anticipated pharmacological and/or mechanical cardiovascular manipulation.
• Repeated blood sampling.
• Anticipated respiratory compromise.
• Faliure of indirect BP measurement.
• Supplementary diagnostic information from the arterial waveform.
PERCUTANEOUS RADIAL ARTERY
CANNULATION
Radial artery is the most common artery used for IBP because
• Easy to cannulate due to being superficial
• Technically less incidence of complications.
Modified ALLEN’S Test
OTHER SITES
COMPLICATIONS
• Distal ischemia
• Pseudoaneurysm
• Arteriovenous fistula
• Hemorrhage
• Arterial embolization
• Infection
• Peripheral neuropathy
• Misinterpretaion of data
• Misuse of equipment
COMPONENTS OF IBP MONITORING
EQUIPMENT
• Intra-arterial catheter
• Stopcocks
• In-line blood sampling ports
• Pressure transducer
• Continuoius flush device
• Electronic cable
PROPERTIES OF IBP DEVICE
• NATURAL FREQUENCY of a system determines how rapidly the system oscillates after a perturbation.
• DAMPING COEFFICIENT reflects how rapidly it returns to its prior resting state.
NATURAL FREQUENCY
• It is imporatant that an IBP system has a very high natural frequency i.e 6 to 10
times the fundamental frequency of the arterial waveform (pulse rate).
• e.G for a system to remaine accurate at a heart rate of 120bpm (2cycles/min or
2Hz) ,it would require a natural frequency 6 to 8 times higher than pulse rate i.e
12 TO 20 Hz.
• If Natural frequency is too low, system will resonate and waveform will be
exaggerated.
DAMPING COEFFICIANT
Anything which reduces the energy in an oscillating system will decrease the amplitude of the oscillations. This is
damping.
FACTORES:
Friction
Three way taps
Clots and bubbles
Kinking
Narrow long and compliant tubings
vasospasm
LEVELING AND ZEROING
ZEROING
for an IBP system to optimally work, atmospheric pressure should be discounted from the measurement of
arterial pressure.
Done by exposing the transducer to atm pressure and calibrating the pressure reading to zero.
LEVELLING
Transducer should be placed at the level of patient’s heart in midaxillary line at 4thintercostel space
A transducer too low will over-read and a transducer too high will under-read.
NORMAL ARTERIAL WAVEFORMS
The systolic waveform immediately follows the EKG and consists of steep pressure upstroke, peak
and ensuing decline.
The downslope of arterial pressure waveform is interrupted by the dicrotic notch, continues its
decline during diastole after the EKG T wave and reaches its nadir at the end diastole.
Dicrotic notch of central aortic pressure is sharply defined and represents the closure of aortic valve.
Dicrotic notch in peripheral arterial pressure waveforms is slightly delayed and mildly blunted
depending upon the arterial wall properties.
Systolic upstroke starts 120 to 180 ms after the R wave.
NORMAL IBP WAVEFORM
ABNORMAL IBP WAVEFORMS
ABNORMAL IBP WAVEFORMS
ABNORMAL IBP WAVEFORMS
ARTERIAL PRESSURE WAVEFORM ANALYSIS FOR
PREDICTION OF VOLUME RESPONSIVENESS
Pulse pressure variation (PPV) is the difference between maximum pulse
pressure and minimum pulse pressure during a single respiratory cycle
divided by the average of these two values.
THANK YOU

More Related Content

What's hot

Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptImran Sheikh
 
High pressure system- Anaesthesia Machine
High pressure system- Anaesthesia MachineHigh pressure system- Anaesthesia Machine
High pressure system- Anaesthesia MachineDr.Daber Pareed
 
anaesthesia for lap cholecystectomy
anaesthesia for lap cholecystectomyanaesthesia for lap cholecystectomy
anaesthesia for lap cholecystectomyabhijit wagh
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machineSwadheen Rout
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologySaneesh P J
 
CIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptxCIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptxDrTejaswini7
 
Humidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careHumidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careTuhin Mistry
 
Monitoring in anaesthesia ro
Monitoring in anaesthesia roMonitoring in anaesthesia ro
Monitoring in anaesthesia rofarranajwa
 
Breathing System
Breathing SystemBreathing System
Breathing SystemKhalid
 
caudal anesthesia.pdf
caudal anesthesia.pdfcaudal anesthesia.pdf
caudal anesthesia.pdfKhodifadVijay
 
Monitored anaesthesia care
Monitored anaesthesia careMonitored anaesthesia care
Monitored anaesthesia careAnaestHSNZ
 

What's hot (20)

Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes ppt
 
Machine boyles
Machine boylesMachine boyles
Machine boyles
 
High pressure system- Anaesthesia Machine
High pressure system- Anaesthesia MachineHigh pressure system- Anaesthesia Machine
High pressure system- Anaesthesia Machine
 
anaesthesia for lap cholecystectomy
anaesthesia for lap cholecystectomyanaesthesia for lap cholecystectomy
anaesthesia for lap cholecystectomy
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in Anesthesiology
 
Capnography new
Capnography new Capnography new
Capnography new
 
CIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptxCIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptx
 
Capnography
CapnographyCapnography
Capnography
 
Difficult airway management
Difficult airway managementDifficult airway management
Difficult airway management
 
Humidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careHumidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical care
 
Monitoring in anaesthesia ro
Monitoring in anaesthesia roMonitoring in anaesthesia ro
Monitoring in anaesthesia ro
 
Perioperative hypoxia
Perioperative hypoxiaPerioperative hypoxia
Perioperative hypoxia
 
Breathing System
Breathing SystemBreathing System
Breathing System
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
Spinal & epidural needle
Spinal & epidural needleSpinal & epidural needle
Spinal & epidural needle
 
caudal anesthesia.pdf
caudal anesthesia.pdfcaudal anesthesia.pdf
caudal anesthesia.pdf
 
CAPNOGRAPHY
CAPNOGRAPHYCAPNOGRAPHY
CAPNOGRAPHY
 
Et co2
Et co2Et co2
Et co2
 
Monitored anaesthesia care
Monitored anaesthesia careMonitored anaesthesia care
Monitored anaesthesia care
 

Similar to Non invasive and_invasive_bp_monitoring__copy

017 intraoperative monitoring
017 intraoperative monitoring017 intraoperative monitoring
017 intraoperative monitoringbothyshiri
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxamitkalirawana07
 
Monitoring_New (1).pptx
Monitoring_New (1).pptxMonitoring_New (1).pptx
Monitoring_New (1).pptxNehaMasarkar1
 
Cardiac monitoring ppt
Cardiac monitoring pptCardiac monitoring ppt
Cardiac monitoring pptManu Gupta
 
Haemodynamic monitoring-Minati
Haemodynamic monitoring-MinatiHaemodynamic monitoring-Minati
Haemodynamic monitoring-MinatiMinati Choudhury
 
MONITORING IN ANAESTHESIA.pptx
MONITORING IN ANAESTHESIA.pptxMONITORING IN ANAESTHESIA.pptx
MONITORING IN ANAESTHESIA.pptxAmin Badamosi
 
Cardio Vascular Monitoring, Anesthesia
Cardio Vascular Monitoring, AnesthesiaCardio Vascular Monitoring, Anesthesia
Cardio Vascular Monitoring, AnesthesiaRashmit Shrestha
 
4. monitoring and interpreting medical investigations in icu
4. monitoring and interpreting medical investigations in icu4. monitoring and interpreting medical investigations in icu
4. monitoring and interpreting medical investigations in icuHibaAnis2
 
1 Monitoring of Central Venous Pressure & Its Techniques
1 Monitoring of Central Venous Pressure & Its Techniques1 Monitoring of Central Venous Pressure & Its Techniques
1 Monitoring of Central Venous Pressure & Its Techniquessrinivas8990
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicineanjika
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptxananya nanda
 
intracranial pressure monitoring
intracranial pressure monitoring intracranial pressure monitoring
intracranial pressure monitoring SHAMEEJ MUHAMED KV
 
physiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxphysiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxkiogakimathi
 
Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Muhammad Asim Rana
 

Similar to Non invasive and_invasive_bp_monitoring__copy (20)

3. CVS monitoring.pptx
3. CVS monitoring.pptx3. CVS monitoring.pptx
3. CVS monitoring.pptx
 
017 intraoperative monitoring
017 intraoperative monitoring017 intraoperative monitoring
017 intraoperative monitoring
 
Basic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptxBasic and advanced Cardiovascular monitoring.pptx
Basic and advanced Cardiovascular monitoring.pptx
 
Monitoring_New (1).pptx
Monitoring_New (1).pptxMonitoring_New (1).pptx
Monitoring_New (1).pptx
 
Cardiac monitoring ppt
Cardiac monitoring pptCardiac monitoring ppt
Cardiac monitoring ppt
 
Haemodynamic monitoring-Minati
Haemodynamic monitoring-MinatiHaemodynamic monitoring-Minati
Haemodynamic monitoring-Minati
 
CVP.pptx
CVP.pptxCVP.pptx
CVP.pptx
 
MONITORING IN ANAESTHESIA.pptx
MONITORING IN ANAESTHESIA.pptxMONITORING IN ANAESTHESIA.pptx
MONITORING IN ANAESTHESIA.pptx
 
Nibp and oxygen analyser
Nibp and oxygen analyserNibp and oxygen analyser
Nibp and oxygen analyser
 
Hemodynamic monitorig
Hemodynamic monitorigHemodynamic monitorig
Hemodynamic monitorig
 
Cardio Vascular Monitoring, Anesthesia
Cardio Vascular Monitoring, AnesthesiaCardio Vascular Monitoring, Anesthesia
Cardio Vascular Monitoring, Anesthesia
 
4. monitoring and interpreting medical investigations in icu
4. monitoring and interpreting medical investigations in icu4. monitoring and interpreting medical investigations in icu
4. monitoring and interpreting medical investigations in icu
 
BASIC ANAESTHETIC MONITORING
BASIC ANAESTHETIC MONITORING BASIC ANAESTHETIC MONITORING
BASIC ANAESTHETIC MONITORING
 
Invasive procedures
Invasive proceduresInvasive procedures
Invasive procedures
 
1 Monitoring of Central Venous Pressure & Its Techniques
1 Monitoring of Central Venous Pressure & Its Techniques1 Monitoring of Central Venous Pressure & Its Techniques
1 Monitoring of Central Venous Pressure & Its Techniques
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicine
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptx
 
intracranial pressure monitoring
intracranial pressure monitoring intracranial pressure monitoring
intracranial pressure monitoring
 
physiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxphysiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptx
 
Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses
 

Recently uploaded

Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetsurat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...mahaiklolahd
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali Punjab
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali PunjabGorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali Punjab
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali PunjabSheetaleventcompany
 
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreDeny Daniel
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Sheetaleventcompany
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...Ahmedabad Call Girls
 
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetpalanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsEscorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsAhmedabad Call Girls
 
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...mahaiklolahd
 

Recently uploaded (20)

Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetsurat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
surat Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali Punjab
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali PunjabGorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali Punjab
Gorgeous Call Girls Mohali {7435815124} ❤️VVIP ANGEL Call Girls in Mohali Punjab
 
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mathura Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetpalanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kochi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance PaymentsEscorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
Escorts Service Ahmedabad🌹6367187148 🌹 No Need For Advance Payments
 
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
 

Non invasive and_invasive_bp_monitoring__copy

  • 1.
  • 2. NON-INVASIVE AND INVASIVE BP MONITORING By Dr. S.A.Fatima Dr. Arsalan A. Zaki
  • 4. WHAT DO WE NEED TO KNOW? • Techniques • Principles • Indications • Contraindications • Limitations • Complications • cautions.
  • 5. NON INVASIVE BP MONITORING TECHNIQUES 1. Manual intermittent techniques. • Palpation • Auscultation • Doppler probe 2. Automated intermittent techniques. • Oscillometry 3. Automated continuous techniques. • Tonometry • photoplathysmography
  • 6. PRINCIPLES OF DIFFERENT NON-INVASIVE TECHNIQUES • Palpation: systolic BP is measured by occluding the artery with cuff • Auscultation: turbulent flow of blood with partially occluded artery produces Korotkoff sounds. • Doppler: movement of RBCs produces a Doppler frequency shift. • Oscillometry: oscillations in the pressure amplitude are measured in an air filled cuff that slowly deflates from the pressure well in excess of pressure to occlude the artery. • Tonometery: sense the pressure required to partially flatten the superficial artery. • Phoyoplathysmography: quantification of infrared light
  • 7.
  • 8. INDICATIONS OF NIBP MONITORING • Any procedure requiring any sort of anesthesia. • When invasive monitoring is contraindicated. • Intermittent measurement interval should be 3 to 5 minutes.
  • 9. CONTRAINDICATIONS: • Cuff based BP monitoring is contraindicated in Limb with axillary clearance Limb with AV fistula for dialysis purpose. Limb with PICC line or peripheral IV line. Patients with known focal vascular anomalies. Burnt extremity.
  • 10. LIMITATIONS • Palpation: • only detects systolic BP. • Not useful for continued monitoring. • Not reliable for conditions of low flow. • Auscultation: • MAP cannot be measured directly and has to be calculated by formula MAP=SBP+ 2(DBP)/3 • Not useful for continued monitoring. • Interpersonal variability of hearing. • Not reliable for conditions of low flow.
  • 11. LIMITATIONS • Ocillometery: • cuff size is of prime importance in accuracy of the BP.(bladder length 80% of arm ; bladder width 40% of arm circumference) • Interpretation varies according to site. • Underestimate MAP during hypertension and overestimate MAP during hypotension so not reliable in unstable patients. • Underestimates SBP and overestimates DBP. • Dysrhythmias lead to inaccuracy of BP. Tonomertery • Need frequent calibration. • Movement artifacts common. • Limited applicability in critically ill patients.
  • 12. COMPLICATIONS: • Pain • Patechiae and ecchymoses • Limb edema • Venous stasis and thrombophlebitis. • Peripheral neuropathy (radial nerve common). • Compartment syndrome. • CAUSIONS: • Patients with known peripheral neuropathy. • Arterial or venous insufficiency. • Severe coagulopathies. • Recet use of thrombolytic therapy.
  • 14. WHAT DO WE NEED TO KNOW? • Basic principle • Indications • Percutaneous radial artery cannulation • Other arteries which can be uses • Complications • Components • Leveling and zeroing • Normal IBP waveforms • Abnormal IBP waveforms • Waveform analysis for prediction of volume responsiveness
  • 15. PRINCIPLE • Intra-arterial blood pressure monitoring uses fluid filled tubing to transmit the force of the pressure pulse wave to a pressure transducer that converts the displacement of a silicon crystal into voltage changes. • These electrical signals are amplified, filtered, and displayed as the arterial pressure trace.
  • 16. INDICATIONS: • Continuos, real time blood pressure monitoring. • Anticipated pharmacological and/or mechanical cardiovascular manipulation. • Repeated blood sampling. • Anticipated respiratory compromise. • Faliure of indirect BP measurement. • Supplementary diagnostic information from the arterial waveform.
  • 17. PERCUTANEOUS RADIAL ARTERY CANNULATION Radial artery is the most common artery used for IBP because • Easy to cannulate due to being superficial • Technically less incidence of complications. Modified ALLEN’S Test
  • 18.
  • 20. COMPLICATIONS • Distal ischemia • Pseudoaneurysm • Arteriovenous fistula • Hemorrhage • Arterial embolization • Infection • Peripheral neuropathy • Misinterpretaion of data • Misuse of equipment
  • 21. COMPONENTS OF IBP MONITORING EQUIPMENT • Intra-arterial catheter • Stopcocks • In-line blood sampling ports • Pressure transducer • Continuoius flush device • Electronic cable
  • 22. PROPERTIES OF IBP DEVICE • NATURAL FREQUENCY of a system determines how rapidly the system oscillates after a perturbation. • DAMPING COEFFICIENT reflects how rapidly it returns to its prior resting state.
  • 23. NATURAL FREQUENCY • It is imporatant that an IBP system has a very high natural frequency i.e 6 to 10 times the fundamental frequency of the arterial waveform (pulse rate). • e.G for a system to remaine accurate at a heart rate of 120bpm (2cycles/min or 2Hz) ,it would require a natural frequency 6 to 8 times higher than pulse rate i.e 12 TO 20 Hz. • If Natural frequency is too low, system will resonate and waveform will be exaggerated.
  • 24. DAMPING COEFFICIANT Anything which reduces the energy in an oscillating system will decrease the amplitude of the oscillations. This is damping. FACTORES: Friction Three way taps Clots and bubbles Kinking Narrow long and compliant tubings vasospasm
  • 25. LEVELING AND ZEROING ZEROING for an IBP system to optimally work, atmospheric pressure should be discounted from the measurement of arterial pressure. Done by exposing the transducer to atm pressure and calibrating the pressure reading to zero. LEVELLING Transducer should be placed at the level of patient’s heart in midaxillary line at 4thintercostel space A transducer too low will over-read and a transducer too high will under-read.
  • 26. NORMAL ARTERIAL WAVEFORMS The systolic waveform immediately follows the EKG and consists of steep pressure upstroke, peak and ensuing decline. The downslope of arterial pressure waveform is interrupted by the dicrotic notch, continues its decline during diastole after the EKG T wave and reaches its nadir at the end diastole. Dicrotic notch of central aortic pressure is sharply defined and represents the closure of aortic valve. Dicrotic notch in peripheral arterial pressure waveforms is slightly delayed and mildly blunted depending upon the arterial wall properties. Systolic upstroke starts 120 to 180 ms after the R wave.
  • 31. ARTERIAL PRESSURE WAVEFORM ANALYSIS FOR PREDICTION OF VOLUME RESPONSIVENESS Pulse pressure variation (PPV) is the difference between maximum pulse pressure and minimum pulse pressure during a single respiratory cycle divided by the average of these two values.