Prospective, randomized, single blinded trial of a novel
                                                                                                  sternal fixation device
                                                                                                                                                      Paul Ghareeb1, Kee Lee2, W. Thomas McClellan1
                                                                                         1.              Department of Surgery, Division of Plastic Surgery, West Virginia University School of Medicine, Morgantown WV
                                                                                                      2.    Department of Surgery, Division of Cardiothoracic Surgery, Monongalia General Hospital, Morgantown WV



                                      Abstract                                                                                            Background (Cont)                                                                                                                                                         Results (Cont)                                                                                     Results (Cont)

      Unstable steel wire cerclage following open heart surgery may result in                                                                                                                                                                                                                         Average Patient-Reported Pain                                                                            Surgery to Discharge Cost
increased pain, risk of sternal cut-through, and dehiscence. These complications                                                                                                                                                                             10
                                                                                                                                                                                                                                                                                                                                                                                                       90000


lead to longer hospital stays, increased cost, higher morbidity, and patient                                                                                                                                                                                     9                                                                                                                                                                      $82,896

dissatisfaction. The Figure 8 Sternal Device is a novel construct which is simple,                                                                                                                                                                               8                                                                                                                                     80000
intuitive, and inexpensive. Prior bench-top testing has demonstrated superior




                                                                                                                                                                                                                       Pain (Visual Analog 10 Point Scale)




                                                                                                                                                                                                                                                                                                                                                                                             Dollars
                                                                                                                                                                                                                                                                 7
strength and stiffness compared to standard steel wire. We present the initial                                                                                                                                                                                                                                                                                                                                      $73,187

                                                                                                                                                                                                                                                                 6
clinical data in our prospective, randomized, single blinded study utilizing this                                                                                                                                                                                                                                                                                                                      70000

new FDA approved system.                                                                                                                                                                                                                                         5
                                                                                                                                                                                                                                                                                            4.5
                                                                                                                                                                                                                                                                                                                                                    Figure 8 Device
                                                                                                                                                                                                                                                                                                                                                    Steel Wire
      30 patients undergoing elective complete sternotomies at a single                                                                                                                                                                                          4


institution were randomly assigned to receive either the Figure 8 Sternal Device                                                                                                                                                                                 3
                                                                                                                                                                                                                                                                                            2.8
                                                                                                                                                                                                                                                                                                                                                                                                       60000
                                                                                                                                                                                                                                                                                                                                                                                                                 Figure 8 Device       Steel Wire

or standard steel wire closure. All surgeries were performed by a single board                                                                                                                                                                                   2

certified cardiothoracic surgeon. Data collected included: pump time, pump off                                                                                                                                                                                   1
                                                                                                                                                                                                                                                                                                                          1.5
                                                                                                                                                                                                                                                                                                                          1               1                                Cost was reduced by 12% on average in the Figure 8
to end time, design of closure, length of hospital stay, cost from time of surgery            The Figure 8  Device  shown  above  in  the  Transverse  (left)  and  “X”  (right)  
to discharge, and pain on a visual analog pain scale on the day of discharge, day                                             patterns.
                                                                                                                                                                                                                                                                 0
                                                                                                                                                                                                                                                                                   Day of Discharge                   30 Days
                                                                                                                                                                                                                                                                                                                                          0.22
                                                                                                                                                                                                                                                                                                                                     60 Days
                                                                                                                                                                                                                                                                                                                                                                                             Device cohort.
30, and day 60.
      Patients receiving the Fig. 8 Device had a significant reduction in length of           The authors hypothesized that the superior mechanical aspects                                                                                                                                                                                                                                                            Conclusions
stay compared to patients receiving steel wire (5.67 vs. 7.58 days respectively,         of the Figure 8 Sternal Device would translate into significantly                                                           Patients receiving the Figure 8 Device reported
p<0.03). Additionally those with the Figure 8 Device reported significantly              improved clinical outcomes by reducing post-operative pain, leading                                                    significantly reduced pain at day of discharge compared                                                                                                   The preliminary results presented here suggest that the Figure 8
decreased pain at day of discharge (2.8 vs. 4.5 points on pain scale, p<0.02), with
                                                                                         to decreased length of stay and total cost.                                                                                                                                                                                                                                  Sternal Device provides superior in-vivo stability, resulting in
equal pain scores at 30 and 60 days. Sternal closure time was increased by 25                                                                                                                                   to the steel wire cohort (2.8 vs. 4.5, p<0.02). Pain scores
percent in patients receiving the device vs. steel wires (70.82 vs. 53.17 minutes,                                                                                                                                                                                                                                                                                    improved clinical outcomes and decreased healthcare costs.
p=0.0002). Cost was 12 percent less, on average, for the device cohort.
                                                                                                                                                                                                                        were statistically equal at 30 and 60 days.
      Prior bench-top testing of the Figure 8 Device demonstrated superior                                                                                                                                                                                                                                                                                                 Patients in the device cohort experienced significantly reduced
                                                                                                                                                  Methods                                                                                                                                                                                                             post-operative pain, leading to reduced length of stay. This reduction
strength and stiffness to standard steel wire. Early clinical results suggest that the                                                                                                                                                                                                     Average Length of Post-Operative Stay
Figure 8 Sternal Device provides improved stability, which resulted in
                                                                                                                                                                                                                                                                              9
                                                                                                                                                                                                                                                                                                                                                                      in total number of hospital days resulted in decreased cost and
diminished post-operative pain at discharge, decreased length of stay, and               30 patients scheduled to undergo elective complete sternotomy at a                                                                                                                   8                                                                                       increased patient satisfaction.
                                                                                                                                                                                                                                                                                                                                         7.58

reduced cost. Further clinical research is warranted to expand upon these initial        single institution were randomly divided into two cohorts: one
                                                                                                                                                                                                                                                                              7

trends and validate long term outcomes.                                                  group received the Figure 8 Device, while the other received standard                                                                                                                                                                                                            Sternal closure time was slightly increased in the Figure 8 group,
                                                                                         steel wire cerclage. The patients were blinded to the type of closure                                                                                                                6                        5.66
                                                                                                                                                                                                                                                                                                                                                                      which may be explained by the fact that application of the device is
                                                                                         received. The same board-certified cardiothoracic surgeon performed                                                                                                                  5
                                                                                                                                                                                                                                                                                                                                                                      new to the surgeon. As surgeons become more proficient with the




                                                                                                                                                                                                                                                              Days
                                   Background                                            each procedure, and the configuration of the closure method was                                                                                                                      4                                                                                       device, one could expect closure times to improve.
                                                                                         determined  by  surgeon’s  clinical  judgment.    X-rays at day of                                                                                                                   3
                                                                                                                                                                                                                                                                                                                                                                           Prior biomechanical testing demonstrated superior strength and
     The consequences of sternal separation, cut-through, or                             discharge and 30 days confirmed alignment and proper placement                                                                                                                       2
                                                                                                                                                                                                                                                                                                                                                                      stiffness when compared to standard steel wire. Our results suggest
dehiscence can be profound, with a mortality rate of 10-40%1. The                        of sternal constructs.
                                                                                                                                                                                                                                                                              1                                                                                       that the enhanced mechanical characteristics of the Figure 8 Device
risk of complications increase in complex patients and in the elderly2-                  Pain was measured on a 10 point visual analog pain scale at day of                                                                                                                   0
                                                                                                                                                                                                                                                                                                                                                                      translate into improved clinical outcomes. By improving upon the
3. Over 760,000 procedures requiring sternotomy are performed
                                                                                         discharge, day 30, and day 60. Length of post-operative stay was
                                                                                                                                                                                                                                                                                                  Figure 8 Device                      Steel Wire
                                                                                                                                                                                                                                                                                                                                                                      current standard of care, the Figure 8 Sternal Fixation Device has the
every year, making complications a serious healthcare issue4-5.                          recorded, and total cost from surgery to discharge was calculated.                                                                                                                                                                                                           potential to become the primary method of sternal closure.
                                                                                         Length of closure was estimated by using the cardiopulmonary                                                           Patients receiving the Figure 8 Device had significantly
     The median hospital costs for patients developing sternal wound                                                                                                                                                                                                                                                                                                      Further clinical research is necessary to expand upon our initial
complications following CABG can be up to 2.8 times higher than                          bypass pump off to surgery end time. Complications were                                                                reduced post-operative length of stay compared to those
                                                                                                                                                                                                                                                                                                                                                                      data, and long term outcomes must be validated.
for uncomplicated patients (8). The Department of Health and                             documented.                                                                                                                receiving steel wire (5.67 vs. 7.58 days, p<0.03) .
Human Resources has identified these complications as hospital-                          Statistical analyses were performed using SAS software (Version 9.2;
acquired events (HAC) for which hospitals should not receive                             Cary, NC); a student t-test was utilized to compare pain scores and
                                                                                                                                                                                                                                                                                                Average Bypass Pump off to
                                                                                                                                                                                                                                                                                                     Surgery End Time
additional payment if the condition was not present upon admission.                      length of stay between the two groups.                                                                                                                                               80
                                                                                                                                                                                                                                                                                                      70.82
                                                                                                                                                                                                                                                                                                                                                                                                                          References
     The most important factor in the prevention of sternal events is                                                                                                                                                                                                         70


                                                                                                                                                                                                                                                                              60                                                                                      1.   Shih CC, Shih CM, Su YY, Lin SJ. Potential risk of sternal wires. Eur J Cardiothorac Surg.
a stable sternal approximation following sternotomy 6-7.There have                                                                                                                                                                                                            50
                                                                                                                                                                                                                                                                                                                                      53.16
                                                                                                                                                                                                                                                                                                                                                                           2004 May;25(5):812-8.
                                                                                                                                                                                                                                                             Time (Minutes)



been many attempts to improve upon the standard method of steel                                                                                   Results                                                                                                                     40
                                                                                                                                                                                                                                                                                                                                                                      2.   Ståhle E, Tammelin A, Bergström R, Hambreus A, Nyström SO, Hansson HE. Sternal
wire cerclage, but these systems have failed to gain widespread                                                                                                                                                                                                               30                                                                                           wound complications--incidence, microbiology and risk factors. Eur J Cardiothorac Surg. 1997
adoption.                                                                                                  Average Patient Age                                         Average Patient Body Mass                                                                              20                                                                                           Jun;11(6):1146-53.
                                                                                                      70
                                                                                                                                  62.75
                                                                                                                                                                              Index (BMI)
                                                                                                                62.5                                                                                                                                                          10
                                                                                                                                                                                                                                                                                                                                                                      3.   Ridderstolpe L, Gill H, Granfeldt H, Ahlfeldt H, Rutberg H. Superficial and deep sternal
     The Figure 8 Sternal Fixation Device was developed to provide a                                  60                                                          34
                                                                                                                                                                                                                                                                               0                                                                                           wound complications: incidence, risk factors and mortality. Eur J Cardiothorac Surg. 2001
sternal closure method that is superior to standard steel wire                                        50                                                          32

                                                                                                                                                                  30
                                                                                                                                                                               30.1
                                                                                                                                                                                                 29.4
                                                                                                                                                                                                                                                                                                  Figure 8 Device                    Steel Wire
                                                                                                                                                                                                                                                                                                                                                                           Dec;20(6):1168-75.
cerclage, while avoiding the flaws of current alternatives. Prior bench-
                                                                                                      40
                                                                                              Years




                                                                                                                                                                  28
                                                                                                                                                                                                                                                                                                                                                                      4.   Losanoff JE, Richman BW, Jones JW. Disruption and infection of median sternotomy: a
                                                                                                                                                            BMI




                                                                                                      30                                                          26
top comparison of the Figure 8 Device to steel wires demonstrated                                                                                                 24                                                                                                                                                                                                       comprehensive review. Eur J Cardiothorac Surg. 2002 May;21(5):831-9.
                                                                                                                                                                                                                  Bypass pump off to surgery end time was used as an
                                                                                                      20

that the device is 70% stronger when placed in a transverse                                           10
                                                                                                                                                                  22

                                                                                                                                                                  20                                                                                                                                                                                                  5.   Voss B, Bauernschmitt R, Will A, et al. Sternal reconstruction with titanium plates in
configuration,  and  40%  stronger  when  constructed  in  the  “X”                                   0
                                                                                                            Figure 8 Cohort   Steel Wire Cohort
                                                                                                                                                                  18
                                                                                                                                                                          Figure 8 Cohort   Steel Wire Cohort
                                                                                                                                                                                                                estimation of sternal closure time. Patients receiving the                                                                                                 complicated sternal dehiscence. Eur J Cardiothorac Surg. 2008 Jul;34(1):139-45.
pattern8. In addition, the device demonstrated significantly reduced                                                                                                                                            Figure 8 Device had significantly increased closure time                                                                                              6.   Casha AR, Gauci M, Yang L, Saleh M, Kay PH, Cooper GJ. Fatigue testing median
cut-through and improved lateral and longitudinal cyclical testing                                                                                                                                                                                                                                                                                                         sternotomy closures. Eur J Cardiothorac Surg. 2001 Mar;19(3):249-53.
                                                                                         Patient age and BMI were statistically equal between the                                                                         (70.82 vs. 53.17 minutes, p=0.0002).
when compared to steel wire cerclage8.                                                                                                                                                                                                                                                                                                                                7.   DiMarco RF, Lee MW, Bekoe S, Grant KJ, Woelfel G, Pellegrini RV. Interlocking figure-of-8
                                                                                                             two cohorts.                                                                                                                                                                                                                                                  closure of the sternum. Ann Thorac Surg. 1989;47:927-929.

                                                                                                                                                                                                                                                                                                                                                                      8.   Wilson R, Ghareeb P, McClellan WT. Biomechanical analysis of the Figure 8 Sternal
                                                                                                                                                                                                                                                                                                                                                                           Fixation Device. In press, Plast Reconstr Surg 2012.

Weitere ähnliche Inhalte

Was ist angesagt?

Decision making in management for large medical equipment
Decision making in management for large medical equipmentDecision making in management for large medical equipment
Decision making in management for large medical equipmentHTAi Bilbao 2012
 
Quality&amp;Info Management Ed Presentation
Quality&amp;Info Management Ed PresentationQuality&amp;Info Management Ed Presentation
Quality&amp;Info Management Ed Presentationcbw0109
 
20120613 7th special interest meeting
20120613 7th special interest meeting20120613 7th special interest meeting
20120613 7th special interest meetingKris Van den Broeck
 
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...10122607 陳俐君(hotel restaurant dining the relationship between perceived value...
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...seashell09
 
BcnInnova presentation- English
BcnInnova presentation- EnglishBcnInnova presentation- English
BcnInnova presentation- EnglishBcnInnova
 
Enhancing SLAP repairs with Fibrin-PRP clots - Poster
Enhancing SLAP repairs with Fibrin-PRP clots - PosterEnhancing SLAP repairs with Fibrin-PRP clots - Poster
Enhancing SLAP repairs with Fibrin-PRP clots - PosterAlan M. Hirahara, M.D., FRCSC
 
Week 5 measurement, scaling, and questionnaire design
Week 5 measurement, scaling, and questionnaire designWeek 5 measurement, scaling, and questionnaire design
Week 5 measurement, scaling, and questionnaire designuniv
 
Rodman renshaw report on onty
Rodman renshaw report on ontyRodman renshaw report on onty
Rodman renshaw report on ontyDailyDoseEquities
 

Was ist angesagt? (8)

Decision making in management for large medical equipment
Decision making in management for large medical equipmentDecision making in management for large medical equipment
Decision making in management for large medical equipment
 
Quality&amp;Info Management Ed Presentation
Quality&amp;Info Management Ed PresentationQuality&amp;Info Management Ed Presentation
Quality&amp;Info Management Ed Presentation
 
20120613 7th special interest meeting
20120613 7th special interest meeting20120613 7th special interest meeting
20120613 7th special interest meeting
 
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...10122607 陳俐君(hotel restaurant dining the relationship between perceived value...
10122607 陳俐君(hotel restaurant dining the relationship between perceived value...
 
BcnInnova presentation- English
BcnInnova presentation- EnglishBcnInnova presentation- English
BcnInnova presentation- English
 
Enhancing SLAP repairs with Fibrin-PRP clots - Poster
Enhancing SLAP repairs with Fibrin-PRP clots - PosterEnhancing SLAP repairs with Fibrin-PRP clots - Poster
Enhancing SLAP repairs with Fibrin-PRP clots - Poster
 
Week 5 measurement, scaling, and questionnaire design
Week 5 measurement, scaling, and questionnaire designWeek 5 measurement, scaling, and questionnaire design
Week 5 measurement, scaling, and questionnaire design
 
Rodman renshaw report on onty
Rodman renshaw report on ontyRodman renshaw report on onty
Rodman renshaw report on onty
 

Andere mochten auch

01 c ce-–-andaime-fachadeiro-–-manual-de-montagem
01 c ce-–-andaime-fachadeiro-–-manual-de-montagem01 c ce-–-andaime-fachadeiro-–-manual-de-montagem
01 c ce-–-andaime-fachadeiro-–-manual-de-montagemJanaina Xavier
 
Ingenieria social borrador
Ingenieria  social   borradorIngenieria  social   borrador
Ingenieria social borradoruearig1
 
BDP Systematic Review Poster Final 2_Gupta
BDP Systematic Review Poster Final 2_GuptaBDP Systematic Review Poster Final 2_Gupta
BDP Systematic Review Poster Final 2_GuptaPriya Gupta
 
Presentación de computadoras
Presentación de computadorasPresentación de computadoras
Presentación de computadoraskenil_palacios
 
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...chivalrouscostu28
 
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...Leonard Davis Institute of Health Economics
 
Fisher House Foundation Celebrates 25 Years
Fisher House Foundation Celebrates 25 YearsFisher House Foundation Celebrates 25 Years
Fisher House Foundation Celebrates 25 YearsUnique Insurance
 

Andere mochten auch (9)

Modelo canvas
Modelo canvasModelo canvas
Modelo canvas
 
01 c ce-–-andaime-fachadeiro-–-manual-de-montagem
01 c ce-–-andaime-fachadeiro-–-manual-de-montagem01 c ce-–-andaime-fachadeiro-–-manual-de-montagem
01 c ce-–-andaime-fachadeiro-–-manual-de-montagem
 
Ingenieria social borrador
Ingenieria  social   borradorIngenieria  social   borrador
Ingenieria social borrador
 
BDP Systematic Review Poster Final 2_Gupta
BDP Systematic Review Poster Final 2_GuptaBDP Systematic Review Poster Final 2_Gupta
BDP Systematic Review Poster Final 2_Gupta
 
Presentación de computadoras
Presentación de computadorasPresentación de computadoras
Presentación de computadoras
 
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...
fou de la baise culbute royalement la chatounette épilée d'une pompeuse chaud...
 
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...
Limited Impact of Reducing Length of Stay on Daily Peak Census at a Children'...
 
Project english pdf
Project english pdfProject english pdf
Project english pdf
 
Fisher House Foundation Celebrates 25 Years
Fisher House Foundation Celebrates 25 YearsFisher House Foundation Celebrates 25 Years
Fisher House Foundation Celebrates 25 Years
 

Mehr von W. Thomas McClellan, MD FACS

IV3000 Presentation ASPS September 2016 Slideshare
IV3000 Presentation ASPS September 2016 SlideshareIV3000 Presentation ASPS September 2016 Slideshare
IV3000 Presentation ASPS September 2016 SlideshareW. Thomas McClellan, MD FACS
 
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...W. Thomas McClellan, MD FACS
 
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...W. Thomas McClellan, MD FACS
 
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System W. Thomas McClellan, MD FACS
 
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...W. Thomas McClellan, MD FACS
 
Breast Augmentation / Breast Implants : An Informed Consent Presentation
Breast Augmentation / Breast Implants : An Informed Consent PresentationBreast Augmentation / Breast Implants : An Informed Consent Presentation
Breast Augmentation / Breast Implants : An Informed Consent PresentationW. Thomas McClellan, MD FACS
 
The Lazy Lateral Incision: An Innovative approach to mastectomy
The Lazy Lateral Incision: An Innovative approach to mastectomyThe Lazy Lateral Incision: An Innovative approach to mastectomy
The Lazy Lateral Incision: An Innovative approach to mastectomyW. Thomas McClellan, MD FACS
 

Mehr von W. Thomas McClellan, MD FACS (20)

McClellan Innovator and BioEntrepreneur
McClellan Innovator and BioEntrepreneurMcClellan Innovator and BioEntrepreneur
McClellan Innovator and BioEntrepreneur
 
IV3000 Presentation ASPS September 2016 Slideshare
IV3000 Presentation ASPS September 2016 SlideshareIV3000 Presentation ASPS September 2016 Slideshare
IV3000 Presentation ASPS September 2016 Slideshare
 
IV3000 Dressing for Fingertip Injuries
IV3000 Dressing for Fingertip InjuriesIV3000 Dressing for Fingertip Injuries
IV3000 Dressing for Fingertip Injuries
 
Top 8 traits of the BioEntrepreneur
Top 8 traits of the BioEntrepreneurTop 8 traits of the BioEntrepreneur
Top 8 traits of the BioEntrepreneur
 
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...
Unilateral Forehead Paralysis Following Operative Repair of Facial Trauma: A ...
 
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...
Acquired Anterior Thoracic Lung Herniation and Repair: A Rare Case and Discus...
 
Metacarpal fractures
Metacarpal fracturesMetacarpal fractures
Metacarpal fractures
 
IV 3000: An Innovative Fingertip Dressing
IV 3000: An Innovative Fingertip DressingIV 3000: An Innovative Fingertip Dressing
IV 3000: An Innovative Fingertip Dressing
 
Radial ArterioVenous Fistula with pictures
Radial ArterioVenous Fistula with picturesRadial ArterioVenous Fistula with pictures
Radial ArterioVenous Fistula with pictures
 
Basal joint arthritis presentation
Basal joint arthritis presentationBasal joint arthritis presentation
Basal joint arthritis presentation
 
Osteoarthritis of the_wrist from Mayo Clinic
Osteoarthritis of the_wrist from Mayo ClinicOsteoarthritis of the_wrist from Mayo Clinic
Osteoarthritis of the_wrist from Mayo Clinic
 
Dupuytrens contracture presentation
Dupuytrens contracture presentationDupuytrens contracture presentation
Dupuytrens contracture presentation
 
WVExecutive Article Tom McClellan, MD FACS
WVExecutive Article Tom McClellan, MD FACSWVExecutive Article Tom McClellan, MD FACS
WVExecutive Article Tom McClellan, MD FACS
 
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System
Prospective Pilot Study: Figure 8 FlatWire Sternal Closure System
 
Eyelid Reconstruction CME Article Dr. McClellan
Eyelid Reconstruction CME Article Dr. McClellanEyelid Reconstruction CME Article Dr. McClellan
Eyelid Reconstruction CME Article Dr. McClellan
 
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...
Reducing Pathogen Transmission in a Hospital Setting. Handshake verses Fistbu...
 
The Inframammary Crease
The Inframammary CreaseThe Inframammary Crease
The Inframammary Crease
 
Breast Augmentation / Breast Implants : An Informed Consent Presentation
Breast Augmentation / Breast Implants : An Informed Consent PresentationBreast Augmentation / Breast Implants : An Informed Consent Presentation
Breast Augmentation / Breast Implants : An Informed Consent Presentation
 
The Lazy Lateral Incision: An Innovative approach to mastectomy
The Lazy Lateral Incision: An Innovative approach to mastectomyThe Lazy Lateral Incision: An Innovative approach to mastectomy
The Lazy Lateral Incision: An Innovative approach to mastectomy
 
Forehead Reconstruction Using an A to T
Forehead Reconstruction Using an A to TForehead Reconstruction Using an A to T
Forehead Reconstruction Using an A to T
 

Kürzlich hochgeladen

Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...narwatsonia7
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableDipal Arora
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 

Kürzlich hochgeladen (20)

Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 

Figure 8 Device Early Clinical Results

  • 1. Prospective, randomized, single blinded trial of a novel sternal fixation device Paul Ghareeb1, Kee Lee2, W. Thomas McClellan1 1. Department of Surgery, Division of Plastic Surgery, West Virginia University School of Medicine, Morgantown WV 2. Department of Surgery, Division of Cardiothoracic Surgery, Monongalia General Hospital, Morgantown WV Abstract Background (Cont) Results (Cont) Results (Cont) Unstable steel wire cerclage following open heart surgery may result in Average Patient-Reported Pain Surgery to Discharge Cost increased pain, risk of sternal cut-through, and dehiscence. These complications 10 90000 lead to longer hospital stays, increased cost, higher morbidity, and patient 9 $82,896 dissatisfaction. The Figure 8 Sternal Device is a novel construct which is simple, 8 80000 intuitive, and inexpensive. Prior bench-top testing has demonstrated superior Pain (Visual Analog 10 Point Scale) Dollars 7 strength and stiffness compared to standard steel wire. We present the initial $73,187 6 clinical data in our prospective, randomized, single blinded study utilizing this 70000 new FDA approved system. 5 4.5 Figure 8 Device Steel Wire 30 patients undergoing elective complete sternotomies at a single 4 institution were randomly assigned to receive either the Figure 8 Sternal Device 3 2.8 60000 Figure 8 Device Steel Wire or standard steel wire closure. All surgeries were performed by a single board 2 certified cardiothoracic surgeon. Data collected included: pump time, pump off 1 1.5 1 1 Cost was reduced by 12% on average in the Figure 8 to end time, design of closure, length of hospital stay, cost from time of surgery The Figure 8  Device  shown  above  in  the  Transverse  (left)  and  “X”  (right)   to discharge, and pain on a visual analog pain scale on the day of discharge, day patterns. 0 Day of Discharge 30 Days 0.22 60 Days Device cohort. 30, and day 60. Patients receiving the Fig. 8 Device had a significant reduction in length of The authors hypothesized that the superior mechanical aspects Conclusions stay compared to patients receiving steel wire (5.67 vs. 7.58 days respectively, of the Figure 8 Sternal Device would translate into significantly Patients receiving the Figure 8 Device reported p<0.03). Additionally those with the Figure 8 Device reported significantly improved clinical outcomes by reducing post-operative pain, leading significantly reduced pain at day of discharge compared The preliminary results presented here suggest that the Figure 8 decreased pain at day of discharge (2.8 vs. 4.5 points on pain scale, p<0.02), with to decreased length of stay and total cost. Sternal Device provides superior in-vivo stability, resulting in equal pain scores at 30 and 60 days. Sternal closure time was increased by 25 to the steel wire cohort (2.8 vs. 4.5, p<0.02). Pain scores percent in patients receiving the device vs. steel wires (70.82 vs. 53.17 minutes, improved clinical outcomes and decreased healthcare costs. p=0.0002). Cost was 12 percent less, on average, for the device cohort. were statistically equal at 30 and 60 days. Prior bench-top testing of the Figure 8 Device demonstrated superior Patients in the device cohort experienced significantly reduced Methods post-operative pain, leading to reduced length of stay. This reduction strength and stiffness to standard steel wire. Early clinical results suggest that the Average Length of Post-Operative Stay Figure 8 Sternal Device provides improved stability, which resulted in 9 in total number of hospital days resulted in decreased cost and diminished post-operative pain at discharge, decreased length of stay, and 30 patients scheduled to undergo elective complete sternotomy at a 8 increased patient satisfaction. 7.58 reduced cost. Further clinical research is warranted to expand upon these initial single institution were randomly divided into two cohorts: one 7 trends and validate long term outcomes. group received the Figure 8 Device, while the other received standard Sternal closure time was slightly increased in the Figure 8 group, steel wire cerclage. The patients were blinded to the type of closure 6 5.66 which may be explained by the fact that application of the device is received. The same board-certified cardiothoracic surgeon performed 5 new to the surgeon. As surgeons become more proficient with the Days Background each procedure, and the configuration of the closure method was 4 device, one could expect closure times to improve. determined  by  surgeon’s  clinical  judgment.    X-rays at day of 3 Prior biomechanical testing demonstrated superior strength and The consequences of sternal separation, cut-through, or discharge and 30 days confirmed alignment and proper placement 2 stiffness when compared to standard steel wire. Our results suggest dehiscence can be profound, with a mortality rate of 10-40%1. The of sternal constructs. 1 that the enhanced mechanical characteristics of the Figure 8 Device risk of complications increase in complex patients and in the elderly2- Pain was measured on a 10 point visual analog pain scale at day of 0 translate into improved clinical outcomes. By improving upon the 3. Over 760,000 procedures requiring sternotomy are performed discharge, day 30, and day 60. Length of post-operative stay was Figure 8 Device Steel Wire current standard of care, the Figure 8 Sternal Fixation Device has the every year, making complications a serious healthcare issue4-5. recorded, and total cost from surgery to discharge was calculated. potential to become the primary method of sternal closure. Length of closure was estimated by using the cardiopulmonary Patients receiving the Figure 8 Device had significantly The median hospital costs for patients developing sternal wound Further clinical research is necessary to expand upon our initial complications following CABG can be up to 2.8 times higher than bypass pump off to surgery end time. Complications were reduced post-operative length of stay compared to those data, and long term outcomes must be validated. for uncomplicated patients (8). The Department of Health and documented. receiving steel wire (5.67 vs. 7.58 days, p<0.03) . Human Resources has identified these complications as hospital- Statistical analyses were performed using SAS software (Version 9.2; acquired events (HAC) for which hospitals should not receive Cary, NC); a student t-test was utilized to compare pain scores and Average Bypass Pump off to Surgery End Time additional payment if the condition was not present upon admission. length of stay between the two groups. 80 70.82 References The most important factor in the prevention of sternal events is 70 60 1. Shih CC, Shih CM, Su YY, Lin SJ. Potential risk of sternal wires. Eur J Cardiothorac Surg. a stable sternal approximation following sternotomy 6-7.There have 50 53.16 2004 May;25(5):812-8. Time (Minutes) been many attempts to improve upon the standard method of steel Results 40 2. Ståhle E, Tammelin A, Bergström R, Hambreus A, Nyström SO, Hansson HE. Sternal wire cerclage, but these systems have failed to gain widespread 30 wound complications--incidence, microbiology and risk factors. Eur J Cardiothorac Surg. 1997 adoption. Average Patient Age Average Patient Body Mass 20 Jun;11(6):1146-53. 70 62.75 Index (BMI) 62.5 10 3. Ridderstolpe L, Gill H, Granfeldt H, Ahlfeldt H, Rutberg H. Superficial and deep sternal The Figure 8 Sternal Fixation Device was developed to provide a 60 34 0 wound complications: incidence, risk factors and mortality. Eur J Cardiothorac Surg. 2001 sternal closure method that is superior to standard steel wire 50 32 30 30.1 29.4 Figure 8 Device Steel Wire Dec;20(6):1168-75. cerclage, while avoiding the flaws of current alternatives. Prior bench- 40 Years 28 4. Losanoff JE, Richman BW, Jones JW. Disruption and infection of median sternotomy: a BMI 30 26 top comparison of the Figure 8 Device to steel wires demonstrated 24 comprehensive review. Eur J Cardiothorac Surg. 2002 May;21(5):831-9. Bypass pump off to surgery end time was used as an 20 that the device is 70% stronger when placed in a transverse 10 22 20 5. Voss B, Bauernschmitt R, Will A, et al. Sternal reconstruction with titanium plates in configuration,  and  40%  stronger  when  constructed  in  the  “X”   0 Figure 8 Cohort Steel Wire Cohort 18 Figure 8 Cohort Steel Wire Cohort estimation of sternal closure time. Patients receiving the complicated sternal dehiscence. Eur J Cardiothorac Surg. 2008 Jul;34(1):139-45. pattern8. In addition, the device demonstrated significantly reduced Figure 8 Device had significantly increased closure time 6. Casha AR, Gauci M, Yang L, Saleh M, Kay PH, Cooper GJ. Fatigue testing median cut-through and improved lateral and longitudinal cyclical testing sternotomy closures. Eur J Cardiothorac Surg. 2001 Mar;19(3):249-53. Patient age and BMI were statistically equal between the (70.82 vs. 53.17 minutes, p=0.0002). when compared to steel wire cerclage8. 7. DiMarco RF, Lee MW, Bekoe S, Grant KJ, Woelfel G, Pellegrini RV. Interlocking figure-of-8 two cohorts. closure of the sternum. Ann Thorac Surg. 1989;47:927-929. 8. Wilson R, Ghareeb P, McClellan WT. Biomechanical analysis of the Figure 8 Sternal Fixation Device. In press, Plast Reconstr Surg 2012.