My presentation on CRO

                                                                                -MEGHA AGRAWAL

                                                                                Ece 2nd year,mnit jaipur.

 Cathode ray oscilloscope:
OBJECTIVE: To learn how to operate a cathode-ray oscilloscope.

APPARATUS: Cathode-ray oscilloscope, multimeter, and oscillator.

INTRODUCTION: The cathode-ray oscilloscope (CRO) is a common laboratory instrument that provides accurate
time and aplitude measurements of voltage signals over a wide range of frequencies. Its reliability, stability, and
ease of operation make it suitable as a general purpose laboratory instrument. The heart of the CRO is a cathode-ray
tube shown schematically in Fig. 1.




  The cathode ray is a beam of electrons which are emitted by the heated cathode (negative electrode) and
accelerated toward the fluorescent screen. The assembly of the cathode, intensity grid, focus grid, and accelerating
anode (positive electrode) is called an electron gun. Its purpose is to generate the electron beam and control its
intensity and focus. Between the electron gun and the fluorescent screen are two pair of metal plates - one oriented
to provide horizontal deflection of the beam and one pair oriented ot give vertical deflection to the beam. These
plates are thus referred to as the horizontal and vertical deflection plates. The combination of these two deflections
allows the beam to reach any portion of the fluorescent screen. Wherever the electron beam hits the screen, the
phosphor is excited and light is emitted from that point. This coversion of electron energy into light allows us to
write with points or lines of light on an otherwise darkened screen.

       In the most common use of the oscilloscope the signal to be studied is first amplified and then applied to the
vertical (deflection) plates to deflect the beam vertically and at the same time a voltage that increases linearly with
time is applied to the horizontal (deflection) plates thus causing the beam to be deflected horizontally at a uniform
(constant> rate. The signal applied to the verical plates is thus displayed on the screen as a function of time. The
horizontal axis serves as a uniform time scale.
The linear deflection or sweep of the beam horizontally is accomplished by use of a sweep generator that is
incorporated in the oscilloscope circuitry. The voltage output of such a generator is that of a sawtooth wave as
shown in Fig. 2. Application of one cycle of this voltage difference, which increases linearly with time, to the
horizontal plates causes the beam to be deflected linearly with time across the tube face. When the voltage suddenly
falls to zero, as at points (a) (b) (c), etc...., the end of each sweep - the beam flies back to its initial position. The
horizontal deflection of the beam is repeated periodically, the frequency of this periodicity is adjustable by external
controls.




       To obtain steady traces on the tube face, an internal number of cycles of the unknown signal that is applied to
the vertical plates must be associated with each cycle of the sweep generator. Thus, with such a matching of
synchronization of the two deflections, the pattern on the tube face repeats itself and hence appears to remain
stationary. The persistance of vision in the human eye and of the glow of the fluorescent screen aids in producing a
stationary pattern. In addition, the electron beam is cut off (blanked) during flyback so that the retrace sweep is not
observed.

CRO Operation: A simplified block diagram of a typical oscilloscope is shown in Fig. 3. In general, the
instrument is operated in the following manner. The signal to be displayed is amplified by the vertical amplifier and
applied to the verical deflection plates of the CRT. A portion of the signal in the vertical amplifier is applied to
the sweep trigger as a triggering signal. The sweep trigger then generates a pulse coincident with a selected point in
the cycle of the triggering signal. This pulse turns on the sweep generator, initiating the sawtooth wave form. The
sawtooth wave is amplified by the horizontal amplifier and applied to the horizontal deflection plates. Usually,
additional provisions signal are made for appliying an external triggering signal or utilizing the 60 Hz line for
triggering. Also the sweep generator may be bypassed and an external signal applied directly to the horizontal
amplifier.

Cro Description:-
Display and general external appearance
The basic oscilloscope, as shown in the illustration, is typically divided into four sections: the display,
vertical controls, horizontal controls and trigger controls.
 The display is usually a CRT or LCD panel which is laid out with both horizontal and vertical reference
lines referred to as the graticule. In addition to the screen, most display sections are equipped with three
basic controls, a focus knob, an intensity knob and a beam finder button.
The vertical section controls the amplitude of the displayed signal. This section carries a Volts-per-
Division (Volts/Div) selector knob, an AC/DC/Ground selector switch and the vertical (primary) input for
the instrument. Additionally, this section is typically equipped with the vertical beam position knob.
The horizontal section controls the time base or “sweep” of the instrument. The primary control is the
Seconds-per-Division (Sec/Div) selector switch. Also included is a horizontal input for plotting dual X-Y
axis signals. The horizontal beam position knob is generally located in this section.
The trigger section controls the start event of the sweep. The trigger can be set to automatically restart
after each sweep or it can be configured to respond to an internal or external event. The principal controls
of this section will be the source and coupling selector switches. An external trigger input (EXT Input) and
level adjustment will also be included.




CRO Controls

      The controls available on most oscilloscopes provide a wide range of operating conditions and thus make the
instrument especially versatile. Since many of these controls are common to most oscilloscopes a brief description
of them follows.




CATHODE-RAY TUBE

Power and Scale Illumination: Turns instrument on and controls illumination of the graticule.

Focus: Focus the spot or trace on the screen.

Intensity: Regulates the brightness of the spot or trace.


VERTICAL AMPLIFIER SECTION

Position: Controls vertical positioning of oscilloscope display.

Sensitivity: Selects the sensitivity of the vertical amplifier in calibrated steps.

Variable Sensitivity: Provides a continuous range of sensitivities between the calibrated steps. Normally the
sensitivity is calibrated only when the variable knob is in the fully clockwise position.
AC-DC-GND: Selects desired coupling (ac or dc) for incoming signal applied to vertical amplifier, or grounds the
amplifier input. Selecting dc couples the input directly to the amplifier; selecting ac send the signal through a
capacitor before going to the amplifier thus blocking any constant component.

HORIZONTAL-SWEEP SECTION

Sweep time/cm: Selects desired sweep rate from calibrated steps or admits external signal to horizontal amplifier.

Sweep time/cm Variable: Provides continuously variable sweep rates. Calibrated position is fully clockwise.

Position: Controls horizontal position of trace on screen.

Horizontal Variable: Controls the attenuation (reduction) of signal applied to horizontal aplifier through Ext. Horiz.
connector.

TRIGGER

The trigger selects the timing of the beginning of the horizontal sweep.

Slope: Selects whether triggering occurs on an increasing (+) or decreasing (-) portion of trigger signal.

Coupling: Selects whether triggering occurs at a specific dc or ac level.

Source: Selects the source of the triggering signal.

      INT - (internal) - from signal on vertical amplifier
      EXT - (external) - from an external signal inserted at the EXT. TRIG. INPUT.
      LINE - 60 cycle triger

Level: Selects the voltage point on the triggering signal at which sweep is triggered. It also allows automatic (auto)
triggering of allows sweep to run free (free run).

CONNECTIONS FOR THE OSCILLOSCOPE

Vertical Input: A pair of jacks for connecting the signal under study to the Y (or vertical) amplifier. The lower jack
is grounded to the case.

Horizontal Input: A pair of jacks for connecting an external signal to the horizontal amplifier. The lower terminal is
graounted to the case of the oscilloscope.

External Tigger Input: Input connector for external trigger signal.

Cal. Out: Provides amplitude calibrated square waves of 25 and 500 millivolts for use in calibrating the gain of the
amplifiers.

      Accuracy of the vertical deflection is + 3%. Sensitivity is variable.

Bandwidth
Bandwidth is a measure of the range of frequencies that can be displayed; it refers primarily to the vertical
amplifier, although the horizontal deflection amplifier has to be fast enough to handle the fastest sweeps.
The bandwidth of the oscilloscope is limited by the vertical amplifiers and the CRT (in analog instruments)
or by the sampling rate of the analog to digital converter in digital instruments. The bandwidth is defined
as the frequency at which the sensitivity is 0.707 of the sensitivity at lower frequency (a drop of 3 dB).
The rise time of the fastest pulse that can be resolved by the scope is related to its bandwidth
approximately:
Bandwidth in Hz x rise time in seconds = 0.35
For example, an oscilloscope intended to resolve pulses with a rise time of 1 nanosecond would have a
bandwidth of 350 MHz.
For a digital oscilloscope, a rule of thumb is that the continuous sampling rate should be ten times the
highest frequency desired to resolve; for example a 20 megasample/second rate would be applicable for
measuring signals up to about 2 megahertz.




      Horizontal sweep should be accurate to within 3%. Range of sweep is variable.

Operating Instructions: Before plugging the oscilloscope into a wall receptacle, set the controls as follows:

      (a) Power switch at off
      (b) Intensity fully counter clockwise
      (c) Vertical centering in the center of range
      (d) Horizontal centering in the center of range
      (e) Vertical at 0.2
      (f) Sweep times 1

Plug line cord into a standard ac wall recepticle (nominally 118 V). Turn power on. Do not advance the Intensity
Control.

Allow the scope to warm up for approximately two minutes, then turn the Intensity Control until the beam is visible
on the screen.




WARNING: Never advance the Intensity Control so far that an excessively bright spot appears. Bright spots imply
burning of the screen. A sharp focused spot of high intensity (great brightness) should never be allowed to remain
fixed in one position on the screen for any length of time as damage to the screen may occur.

Adjust Horizontal and Vertical Centering Controls. Adjust the focus to give a sharp trace. Set trigger to internal,
level to auto.

PROCEDURE:

I. Set the signal generator to a frequency of 1000 cycles per second. Connect the output from the gererator to the
vertical input of the oscilloscope. Establish a steady trace of this input signal on the scope. Adjust (play with)all of
the scope and signal generator controls until you become familiar with the functionof each. The purpose fo such
"playing" is to allow the student to become so familiar with the oscilloscope that it becomes an aid (tool) in making
measurements in other experiments and not as a formidable obstacle. Note: If the vertical gain is set too low, it may
not be possible to obtain a steady trace.
II. Measurement
s of
Voltage: Consid
er the circuit in
Fig. 4(a). The
signal generator
is used to
produce a 1000
hertz sine wave.
The AC
voltmeter and
the leads to the
verticle input of
the oscilloscope
are connected
across the
generator's
output. By
adjusting the
Horizontal Sweep time/cm and trigger, a steady trace of the sine wave may be displayed on the screen. The trace
represents a plot of voltage vs. time, where the vertical deflection of the trace about the line of symmetry CD is
proportional to the magnitude of the voltage at any instant of time.




      To determine the size of the voltage signal appearing at the output of terminals of the signal generator, an AC
(Alternating Current) voltmeter is connected in parallel across these terminals (Fig. 4a). The AC voltmeter is
designed to read the dc "effective value" of the voltage. This effective value is also known as the "Root Mean
Square value" (RMS) value of the voltage.

      The peak or maximum voltage seen on the scope face (Fig. 4b) is V m volts and is represented by the distance
from the symmetry line CD to the maximum deflection. The relationship between the magnitude of the peak voltage
displayed on the scope and the effective or RMS voltage (VRMS) read on the AC voltmeter is

            VRMS = 0.707 Vm (for a sine or cosine wave).

Thus




    Agreement is expected between the voltage reading of the multimeter and that of the oscilloscope. For a
symmetric wave (sine or cosine) the value of V m may be taken as 1/2 the peak to peak signal Vpp

The variable sensitivity control a signal may be used to adjust the display to fill a concenient range of the scope face.
In this position, the trace is no longer calibrated so that you can not just read the size of the signal by counting the
number of divisions and multiplying by the scale factor. However, you can figure out what the new calibration is an
use it as long as the variable control remains unchanged.

Caution: The mathematical prescription given for RMS signals is valid only for sinusoidal signals. The meter will
not indicate the correct voltage when used to measure non-sinusoidal signals.
III. Frequency Measurements: When the horizontal sweep voltage is applied, voltage measurements can still be
taken from the vertical deflection. Moreover, the signal is displayed as a function of time. If the time base (i.e.
sweep) is calibrated, such measurements as pulse duration or signal period can be made. Frequencies can then be
determined as reciprocal of the periods.

      Set the oscillator to 1000 Hz. Display the signal on the CRO and measure the period of the oscillations. Use
the horizontal distance between two points such as C to D in Fig. 4b.

      Set the horizontal gain so that only one complete wave form is displayed.

       Then reset the horizontal until 5 waves are seen. Keep the time base control in a calibrated position. Measure
the distance (and hence time) for 5 complete cycles and calculate the frequency from this measurement. Compare
you result with the value determined above.

      Repeat your measurements for other frequencies of 150 Hz, 5 kHz, 50 kHz as set on the signal generator.

IV. Lissajous Figures: When sine-wave signals of different frequencies are input to the horizontal and vertical
amplifiers a stationary pattern is formed on the CRT when the ratio of the two frequencies is an intergral fraction
such as 1/2, 2/3, 4/3, 1/5, etc. These stationary patterns are known as Lissajous figures and can be used for
comparison measurement of frequencies.

       Use two oscillators to generate some simple Lissajous figures like those shown in Fig. 5. You will find it
difficult to maintain the Lissajous figures in a fixed configuration because the two oscillators are not phase and
frequency locked. Their frequencies and phase drift slowly causing the two different signals to change slightly with
respect to each other.

V. Testing what you have learned: Your instructor will provide you with a small oscillator circuit. Examine the
input to the circuit and output of the circuit using your oscilloscope. Measure such quantities as the voltage and
frequence of the signals. Specify if they are sinusoidal or of some other wave character. If square wave, measure the
frequency of the wave. Also, for square waves, measure the on time (when the voltage is high) and off time (when it
is low).




Examples of use:-
Lissajous figures on an oscilloscope, with 90 degrees phase difference between x and y inputs.

One of the most frequent uses of scopes is troubleshooting malfunctioning electronic equipment. One of the
advantages of a scope is that it can graphically show signals: where a voltmeter may show a totally unexpected
voltage, a scope may reveal that the circuit is oscillating. In other cases the precise shape or timing of a pulse is
important.
In a piece of electronic equipment, for example, the connections between stages (e.g. electronic mixers, electronic
oscillators, amplifiers) may be 'probed' for the expected signal, using the scope as a simple signal tracer. If the
expected signal is absent or incorrect, some preceding stage of the electronics is not operating correctly. Since most
failures occur because of a single faulty component, each measurement can prove that half of the stages of a
complex piece of equipment either work, or probably did not cause the fault.
Once the faulty stage is found, further probing can usually tell a skilled technician exactly which component has
failed. Once the component is replaced, the unit can be restored to service, or at least the next fault can be isolated.
This sort of troubleshooting is typical of radio and TV receivers, as well as audio amplifiers, but can apply to quite-
different devices such as electronic motor drives.
Another use is to check newly designed circuitry. Very often a newly designed circuit will misbehave because of
design errors, bad voltage levels, electrical noise etc. Digital electronics usually operate from a clock, so a dual-trace
scope which shows both the clock signal and a test signal dependent upon the clock is useful. Storage scopes are
helpful for "capturing" rare electronic events that cause defective operation.




                                              Function generator
Definition:- A function generator is usually a piece of electronic test equipment or software used to generate
different types of electrical waveforms over a wide range of frequencies. Some of the most common waveforms
produced by the function generator are the sine, square, triangular and sawtooth shapes. These waveforms can be
either repetitive or single-shot (which requires an internal or external trigger source). Integrated circuits used to
generate waveforms may also be described as function generator ICs.
Feature:- Other important features of the function generator are continuous tuning over wide bands with max-min




frequency ratios of 10:1 or more, a wide range




                                                                        of frequencies from a few Hz to a few
MHz, a flat output amplitude and modulation capabilities like frequency sweeping, frequency modulation and
amplitude modulation.
Although function generators cover both audio and RF frequencies, they are usually not suitable for applications that
need low distortion or stable frequency signals. When those traits are required, other signal generators would be
more appropriate.
Uses:- Function generators are used in the development, test and repair of electronic equipment. For example, they
may be used as a signal source to test amplifiers or to introduce an error signal into a control loop.
Working
Simple function generators usually generate triangular waveform whose frequency can be controlled smoothly as
well as in steps.[4] This triangular wave is used as the basis for all of its other outputs. The triangular wave is
generated by repeatedly charging and discharging a capacitor from a constant current source. This produces
a linearly ascending or descending voltage ramp. As the output voltage reaches upper and lower limits, the charging
and discharging is reversed using a comparator, producing the linear triangle wave. By varying the current and the
size of the capacitor, different frequencies may be obtained. Sawtooth waves can be produced by charging the
capacitor slowly, using a current, but using a diode over the current source to discharge quickly - the polarity of the
diode changes the polarity of the resulting sawtooth, i.e. slow rise and fast fall, or fast rise and slow fall.
A 50% duty cycle square wave is easily obtained by noting whether the capacitor is being charged or discharged,
which is reflected in the current switching comparator output. Other duty cycles (theoretically from 0% to 100%)
can be obtained by using a comparator and the sawtooth or triangle signal. Most function generators also contain a
non-linear diode shaping circuit that can convert the triangle wave into a reasonably accurate sine wave by rounding
off the corners of the triangle wave in a process similar to clipping in audio systems.
A typical function generator can provide frequencies up to 20 MHz. RF generators for higher frequencies are not
function generators in the strict sense since they typically produce pure or modulated sine signals only.
Function generators, like most signal generators, may also contain an attenuator, various means of modulating the
output waveform, and often the ability to automatically and repetitively "sweep" the frequency of the output
waveform (by means of a voltage-controlled oscillator) between two operator-determined limits. This capability
makes it very easy to evaluate the frequency response of a given electronic circuit.
Some function generators can also generate white or pink noise.
More advanced function generators are called arbitrary waveform generators (AWG). They use direct digital
synthesis (DDS) techniques to generate any waveform that can be described by a table of amplitudes.
Specifications
Typical specifications for a general-purpose function generator are:

     Produces sine, square, triangular, sawtooth (ramp), and pulse output. Arbitrary waveform generators can
      produce waves of any shape.
     It can generate a wide range of frequencies. For example, the Tektronix FG 502 (ca 1974) covers 0.1 Hz to
      11 MHz.
     Frequency stability of 0.1 percent per hour for analog generators or 500ppm for a digital generator.
     Maximum sinewave distortion of about 1% (accuracy of diode shaping network) for analog
      generators. Arbitrary waveform generators may have distortion less than -55dB below 50 kHz and less than -
      40dB above 50 kHz.
     Some function generators can be phase locked to an external signal source, which may be a frequency reference
      or another function generator.
     AM or FM modulation may be supported.
     Output amplitude up to 10V peak-to-peak.
     Amplitude can be modified, usually by a calibrated attenuator with decade steps and continuous adjustment
      within each decade.
     Some generators provide a DC offset voltage, e.g. adjustable between -5V to +5V.
     An output impedance of 50 ohms.



    Generator Control Panel -
    What it is and how it’s used
                                   The Control Panel – what is it?
                                   Visually, a control panel is a set of displays that indicate the measurement of
                                   various parameters like voltage, current and frequency, through gauges and
                                   meters. These meters and gauges are set in a metallic body, usually corrosion
                                   proof, to protect from the effect of rain or snow. The panel may be set up on the
                                   body of the generator itself, which is usually the case with small generators. If
                                   they are mounted on the generator, they typically have vibration proof pads that
                                   help isolate the control panel from shocks. Control panels for a larger industrial
                                   generators can be completely separate from the generator and are typically large
    enough to stand upon their own. These units may also be shelf-mounted or wall-mounted next to the generator,
    which is common inside an enclosure or intental application like a data center.
    Control panels are usually fitted with buttons or switches that help to operate the generator such as a switch-off
    button or turn-on key. The switches and gauges are usually grouped on the basis of functionality. This makes the
    panel friendly and safe for use since it minimizes the possibility of an operator accidentally selecting or
    executing the wrong control. Imagine trying to shut down a vibrating generator with a spring loaded lever in the
    middle of the night and you will appreciate why having a simple cut of switch at the control panel makes sense.

    • The front panel is divided into six major control groups:

    1) Frequency Selection Group;
2) Sweep Group;

  3) Amplitude Modulation Group;

  4) DC Offset Group;

  5) Function, or Waveform Group; and

  6) Output Group.




  front panel of function generator

  • The power switch is on the upper left-hand corner of the unit. The green LED will indicate that the unit is on.

  • The three most important groups for this lab are the frequency, function, and output groups. The remaining
  three groups, (sweep, amplitude modulation, and DC offset) will be briefly covered in the lab setup procedures.
  Should you desire more detailed descriptions of these groups, the Leader Function Generator manual is available
  in the lab.

  (1) Frequency Selection Group:- These controls are used to select the operating frequency of the function
      generator. This group consists of the frequency control knob and the eight frequency multiplier selection
      buttons.
  (2) Output Group:- 1. These controls are used to adjust the amplitude of the generator's output signal. The
      group consists of the amplitude-control knob, the three attenuation buttons and the fused 50 ohm BNC
      connector. Although the amplitude knob is not indexed, the amplitude ranges from a few millivolts to
      approximately 20 volts. We will set the amplitude levels by aligning the white line on the amplitude knob
      to the three o'clock position (90 degrees right), the nine o'clock position (90 degrees left), or the twelve
      o'clock position (straight up). Notice that rotating the knob fully to the left does not result in a zero
      amplitude signal.

 • The attenuation buttons are used to attenuate (decrease) the amplitude of the signal by a factor measured in
decibels. The following relationship will assist in working with the attenuation buttons:

  (dB) = -10 * log10 (Pout / Pin) (if power is the unit of measurement)    or

  (dB) = -20 * log10(Vout / Vin) (if voltage is the unit of measurement)

  (3) Function/Waveform Selection Group:- This group is used to select the shape of the generated waveform.
      The group is made up of the six wave-selector buttons. The six waveforms that the function generator can
produce are the sine wave, the square wave, the triangle wave, two sawtooth waves, and the variable-width
      pulse wave.




  How does it work?
  The control panel is becoming an increasingly complex piece of electronics with a microprocessor that can
  manipulate input from sensors to help give feedback to the machine to manage itself. One such feedback could
  be the temperature, indicating overheating, other examples would be over/under speed and low/high oil pressure.
  Typically, a heat sensor inside the generator would sense the build up of heat in the generator body and pass this
  to the microprocessor in the control panel. The microprocessor will then take effective measures to regulate the
  performance of the machine including shutdowns if, for example, the oil pressure is too low or the coolant
  temperature is too high, leading to buildup of heat. In industrial situations, this functionality of control panels is
  becoming increasingly critical. The microprocessor or microcontroller is embedded in the circuitry inside the
  control panel and is programmed to take in the sensor input and react to that with the programmed control rules.
Control panels can be combined with an Automatic Transfer Switch (ATS) to maintain the continuity of electrical
power. The ATS detects an outage of power when your local grid fails. It signals the control panel to start the
generator. Depending on the type of generator being used, the control panel may activate glow plugs (for diesel) for
an adjustable length of time. It will then start the generator using an automatic starter, similar to the one you engage
when you turn the keys in the ignition of your car in the morning. As soon as the engine of the generator reaches an
optimum speed, the starter is disengaged. The ATS then switches to the generator power, and you can go back to
business as usual, without having to frantically scramble to figure out what caused power loss. This aspect of a
control panel makes it extremely useful in homes during bad weather and in industrial situations for ensuring
mission-critical continuity.

Weitere ähnliche Inhalte

Was ist angesagt?

Oscilloscope:
Oscilloscope:Oscilloscope:
Oscilloscope:vipulpop
 
7.instrumentation amplifier
7.instrumentation amplifier7.instrumentation amplifier
7.instrumentation amplifierINDIAN NAVY
 
Cathode ray oscilloscope
Cathode ray oscilloscopeCathode ray oscilloscope
Cathode ray oscilloscopeesmaeel21
 
Electronics 1 : Chapter # 06 : Bipolar Junction Transistor
Electronics 1 : Chapter # 06 : Bipolar Junction TransistorElectronics 1 : Chapter # 06 : Bipolar Junction Transistor
Electronics 1 : Chapter # 06 : Bipolar Junction TransistorSk_Group
 
Basics of amplifier.ppt
Basics of amplifier.pptBasics of amplifier.ppt
Basics of amplifier.pptVijay Togadiya
 
TIME DOMAIN ANALYSIS
TIME DOMAIN ANALYSISTIME DOMAIN ANALYSIS
TIME DOMAIN ANALYSISSyed Saeed
 
digita storage oscilloscope (dso)
digita storage oscilloscope (dso)digita storage oscilloscope (dso)
digita storage oscilloscope (dso)Parind Patel
 
Basics of Sensors & Transducers
Basics of Sensors & TransducersBasics of Sensors & Transducers
Basics of Sensors & TransducersBurdwan University
 
Cro cathode ray oscilloscope working and applications
Cro cathode ray oscilloscope working and applicationsCro cathode ray oscilloscope working and applications
Cro cathode ray oscilloscope working and applicationsEdgefxkits & Solutions
 
Biomedical Instrumentation introduction, Bioamplifiers
Biomedical Instrumentation introduction, BioamplifiersBiomedical Instrumentation introduction, Bioamplifiers
Biomedical Instrumentation introduction, BioamplifiersPoornima D
 
Presentaion on dual trace oscilloscope
Presentaion on dual trace oscilloscopePresentaion on dual trace oscilloscope
Presentaion on dual trace oscilloscopeAliya Sahir
 
Digital frequency meter
Digital frequency meter   Digital frequency meter
Digital frequency meter Soham Gajjar
 

Was ist angesagt? (20)

Oscilloscope:
Oscilloscope:Oscilloscope:
Oscilloscope:
 
7.instrumentation amplifier
7.instrumentation amplifier7.instrumentation amplifier
7.instrumentation amplifier
 
Cathode ray oscilloscope
Cathode ray oscilloscopeCathode ray oscilloscope
Cathode ray oscilloscope
 
Electronics 1 : Chapter # 06 : Bipolar Junction Transistor
Electronics 1 : Chapter # 06 : Bipolar Junction TransistorElectronics 1 : Chapter # 06 : Bipolar Junction Transistor
Electronics 1 : Chapter # 06 : Bipolar Junction Transistor
 
statcom
statcomstatcom
statcom
 
EC6202 ELECTRONIC DEVICES AND CIRCUITS Unit 2
EC6202 ELECTRONIC DEVICES AND CIRCUITS Unit 2EC6202 ELECTRONIC DEVICES AND CIRCUITS Unit 2
EC6202 ELECTRONIC DEVICES AND CIRCUITS Unit 2
 
Basics of amplifier.ppt
Basics of amplifier.pptBasics of amplifier.ppt
Basics of amplifier.ppt
 
Microwave linear beam 31
Microwave linear beam 31Microwave linear beam 31
Microwave linear beam 31
 
Transducers
Transducers Transducers
Transducers
 
TIME DOMAIN ANALYSIS
TIME DOMAIN ANALYSISTIME DOMAIN ANALYSIS
TIME DOMAIN ANALYSIS
 
Pll ppt
Pll pptPll ppt
Pll ppt
 
digita storage oscilloscope (dso)
digita storage oscilloscope (dso)digita storage oscilloscope (dso)
digita storage oscilloscope (dso)
 
Ac servomotor
Ac servomotorAc servomotor
Ac servomotor
 
Forced sensor (Sensor)
 Forced sensor (Sensor) Forced sensor (Sensor)
Forced sensor (Sensor)
 
Basics of Sensors & Transducers
Basics of Sensors & TransducersBasics of Sensors & Transducers
Basics of Sensors & Transducers
 
Cro cathode ray oscilloscope working and applications
Cro cathode ray oscilloscope working and applicationsCro cathode ray oscilloscope working and applications
Cro cathode ray oscilloscope working and applications
 
Biomedical Instrumentation introduction, Bioamplifiers
Biomedical Instrumentation introduction, BioamplifiersBiomedical Instrumentation introduction, Bioamplifiers
Biomedical Instrumentation introduction, Bioamplifiers
 
DUAL TRACE OSCILLOSCOPE
DUAL TRACE OSCILLOSCOPEDUAL TRACE OSCILLOSCOPE
DUAL TRACE OSCILLOSCOPE
 
Presentaion on dual trace oscilloscope
Presentaion on dual trace oscilloscopePresentaion on dual trace oscilloscope
Presentaion on dual trace oscilloscope
 
Digital frequency meter
Digital frequency meter   Digital frequency meter
Digital frequency meter
 

Andere mochten auch

Andere mochten auch (20)

cathode ray oscilloscope
cathode ray oscilloscopecathode ray oscilloscope
cathode ray oscilloscope
 
Cathode ray oscilloscope (cro)
Cathode ray oscilloscope (cro)Cathode ray oscilloscope (cro)
Cathode ray oscilloscope (cro)
 
Unit 1 cro ppt
Unit 1  cro pptUnit 1  cro ppt
Unit 1 cro ppt
 
Function generator
Function generatorFunction generator
Function generator
 
Presentation on cro
Presentation on croPresentation on cro
Presentation on cro
 
Cathode ray oscilloscope
Cathode ray oscilloscopeCathode ray oscilloscope
Cathode ray oscilloscope
 
Oscilloscope
OscilloscopeOscilloscope
Oscilloscope
 
Oscilloscope
OscilloscopeOscilloscope
Oscilloscope
 
(19) electronic instruments 1
(19) electronic instruments 1(19) electronic instruments 1
(19) electronic instruments 1
 
Emi cathode ray oscilloscope
Emi cathode ray oscilloscopeEmi cathode ray oscilloscope
Emi cathode ray oscilloscope
 
Oscilloscope
OscilloscopeOscilloscope
Oscilloscope
 
Cggggg
CgggggCggggg
Cggggg
 
Diode data sheet for alarm type project
Diode data sheet for alarm type projectDiode data sheet for alarm type project
Diode data sheet for alarm type project
 
Cro (emmi) (3)
Cro (emmi) (3)Cro (emmi) (3)
Cro (emmi) (3)
 
Cathode Ray Oscilloscope (CRO)
Cathode Ray Oscilloscope (CRO)Cathode Ray Oscilloscope (CRO)
Cathode Ray Oscilloscope (CRO)
 
Cro (cathod ray oscilloscope)
Cro (cathod ray oscilloscope)Cro (cathod ray oscilloscope)
Cro (cathod ray oscilloscope)
 
The Oscilloscope
The OscilloscopeThe Oscilloscope
The Oscilloscope
 
Lesson revolve
Lesson revolveLesson revolve
Lesson revolve
 
Cathoderaytube final
Cathoderaytube finalCathoderaytube final
Cathoderaytube final
 
Function generator
Function generatorFunction generator
Function generator
 

Ähnlich wie cathode ray oscilloscope &function generator

Cathode Ray Oscilloscope
Cathode Ray OscilloscopeCathode Ray Oscilloscope
Cathode Ray Oscilloscoperajainayat
 
Study of CRO and measusre the voltage and frequency
Study of CRO and measusre the voltage and frequencyStudy of CRO and measusre the voltage and frequency
Study of CRO and measusre the voltage and frequencydreamygyz
 
Analog electronics LAB MANUAL
Analog electronics LAB MANUALAnalog electronics LAB MANUAL
Analog electronics LAB MANUALAJAL A J
 
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKINGCathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKINGAbdul Qayoom Mangrio
 
Handheld oscilloscopes, dso nano v3
Handheld oscilloscopes, dso nano v3Handheld oscilloscopes, dso nano v3
Handheld oscilloscopes, dso nano v3Md. Tosir Uddin
 
157929707-Presentation-on-Cro.pptx
157929707-Presentation-on-Cro.pptx157929707-Presentation-on-Cro.pptx
157929707-Presentation-on-Cro.pptxpranjalmishra68
 
Cro (emmi) (1)
Cro (emmi) (1)Cro (emmi) (1)
Cro (emmi) (1)Ravi Anand
 
Electrical measurement & measuring instruments [emmi (nee-302) -unit-5]
Electrical measurement & measuring instruments [emmi  (nee-302) -unit-5]Electrical measurement & measuring instruments [emmi  (nee-302) -unit-5]
Electrical measurement & measuring instruments [emmi (nee-302) -unit-5]Md Irshad Ahmad
 
Cathode ray oscilloscope and related experiments
Cathode ray oscilloscope and related experimentsCathode ray oscilloscope and related experiments
Cathode ray oscilloscope and related experimentsTrisha Banerjee
 
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...AL- AMIN
 
Emicathoderayoscilloscope 120625075906-phpapp01
Emicathoderayoscilloscope 120625075906-phpapp01Emicathoderayoscilloscope 120625075906-phpapp01
Emicathoderayoscilloscope 120625075906-phpapp01HassanNawazish
 

Ähnlich wie cathode ray oscilloscope &function generator (20)

Cathode Ray Oscilloscope
Cathode Ray OscilloscopeCathode Ray Oscilloscope
Cathode Ray Oscilloscope
 
Emi unit ii ppt
Emi unit ii pptEmi unit ii ppt
Emi unit ii ppt
 
Study of CRO and measusre the voltage and frequency
Study of CRO and measusre the voltage and frequencyStudy of CRO and measusre the voltage and frequency
Study of CRO and measusre the voltage and frequency
 
Cathode Ray Oscilloscope
Cathode Ray OscilloscopeCathode Ray Oscilloscope
Cathode Ray Oscilloscope
 
Oscilloscope tutorial
Oscilloscope tutorialOscilloscope tutorial
Oscilloscope tutorial
 
Analog electronics LAB MANUAL
Analog electronics LAB MANUALAnalog electronics LAB MANUAL
Analog electronics LAB MANUAL
 
Cro ppt
Cro pptCro ppt
Cro ppt
 
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKINGCathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
 
Handheld oscilloscopes, dso nano v3
Handheld oscilloscopes, dso nano v3Handheld oscilloscopes, dso nano v3
Handheld oscilloscopes, dso nano v3
 
157929707-Presentation-on-Cro.pptx
157929707-Presentation-on-Cro.pptx157929707-Presentation-on-Cro.pptx
157929707-Presentation-on-Cro.pptx
 
CRO and FG.ppt
CRO and FG.pptCRO and FG.ppt
CRO and FG.ppt
 
Cro (emmi) (1)
Cro (emmi) (1)Cro (emmi) (1)
Cro (emmi) (1)
 
Electrical measurement & measuring instruments [emmi (nee-302) -unit-5]
Electrical measurement & measuring instruments [emmi  (nee-302) -unit-5]Electrical measurement & measuring instruments [emmi  (nee-302) -unit-5]
Electrical measurement & measuring instruments [emmi (nee-302) -unit-5]
 
Ie practical file
Ie practical fileIe practical file
Ie practical file
 
Cro(nnm)
Cro(nnm)Cro(nnm)
Cro(nnm)
 
Cathode ray oscilloscope and related experiments
Cathode ray oscilloscope and related experimentsCathode ray oscilloscope and related experiments
Cathode ray oscilloscope and related experiments
 
CRO
CROCRO
CRO
 
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
 
Emicathoderayoscilloscope 120625075906-phpapp01
Emicathoderayoscilloscope 120625075906-phpapp01Emicathoderayoscilloscope 120625075906-phpapp01
Emicathoderayoscilloscope 120625075906-phpapp01
 
EMI Unit II
EMI Unit IIEMI Unit II
EMI Unit II
 

cathode ray oscilloscope &function generator

  • 1. My presentation on CRO -MEGHA AGRAWAL Ece 2nd year,mnit jaipur. Cathode ray oscilloscope: OBJECTIVE: To learn how to operate a cathode-ray oscilloscope. APPARATUS: Cathode-ray oscilloscope, multimeter, and oscillator. INTRODUCTION: The cathode-ray oscilloscope (CRO) is a common laboratory instrument that provides accurate time and aplitude measurements of voltage signals over a wide range of frequencies. Its reliability, stability, and ease of operation make it suitable as a general purpose laboratory instrument. The heart of the CRO is a cathode-ray tube shown schematically in Fig. 1. The cathode ray is a beam of electrons which are emitted by the heated cathode (negative electrode) and accelerated toward the fluorescent screen. The assembly of the cathode, intensity grid, focus grid, and accelerating anode (positive electrode) is called an electron gun. Its purpose is to generate the electron beam and control its intensity and focus. Between the electron gun and the fluorescent screen are two pair of metal plates - one oriented to provide horizontal deflection of the beam and one pair oriented ot give vertical deflection to the beam. These plates are thus referred to as the horizontal and vertical deflection plates. The combination of these two deflections allows the beam to reach any portion of the fluorescent screen. Wherever the electron beam hits the screen, the phosphor is excited and light is emitted from that point. This coversion of electron energy into light allows us to write with points or lines of light on an otherwise darkened screen. In the most common use of the oscilloscope the signal to be studied is first amplified and then applied to the vertical (deflection) plates to deflect the beam vertically and at the same time a voltage that increases linearly with time is applied to the horizontal (deflection) plates thus causing the beam to be deflected horizontally at a uniform (constant> rate. The signal applied to the verical plates is thus displayed on the screen as a function of time. The horizontal axis serves as a uniform time scale.
  • 2. The linear deflection or sweep of the beam horizontally is accomplished by use of a sweep generator that is incorporated in the oscilloscope circuitry. The voltage output of such a generator is that of a sawtooth wave as shown in Fig. 2. Application of one cycle of this voltage difference, which increases linearly with time, to the horizontal plates causes the beam to be deflected linearly with time across the tube face. When the voltage suddenly falls to zero, as at points (a) (b) (c), etc...., the end of each sweep - the beam flies back to its initial position. The horizontal deflection of the beam is repeated periodically, the frequency of this periodicity is adjustable by external controls. To obtain steady traces on the tube face, an internal number of cycles of the unknown signal that is applied to the vertical plates must be associated with each cycle of the sweep generator. Thus, with such a matching of synchronization of the two deflections, the pattern on the tube face repeats itself and hence appears to remain stationary. The persistance of vision in the human eye and of the glow of the fluorescent screen aids in producing a stationary pattern. In addition, the electron beam is cut off (blanked) during flyback so that the retrace sweep is not observed. CRO Operation: A simplified block diagram of a typical oscilloscope is shown in Fig. 3. In general, the instrument is operated in the following manner. The signal to be displayed is amplified by the vertical amplifier and applied to the verical deflection plates of the CRT. A portion of the signal in the vertical amplifier is applied to the sweep trigger as a triggering signal. The sweep trigger then generates a pulse coincident with a selected point in the cycle of the triggering signal. This pulse turns on the sweep generator, initiating the sawtooth wave form. The sawtooth wave is amplified by the horizontal amplifier and applied to the horizontal deflection plates. Usually, additional provisions signal are made for appliying an external triggering signal or utilizing the 60 Hz line for triggering. Also the sweep generator may be bypassed and an external signal applied directly to the horizontal amplifier. Cro Description:- Display and general external appearance The basic oscilloscope, as shown in the illustration, is typically divided into four sections: the display, vertical controls, horizontal controls and trigger controls. The display is usually a CRT or LCD panel which is laid out with both horizontal and vertical reference lines referred to as the graticule. In addition to the screen, most display sections are equipped with three basic controls, a focus knob, an intensity knob and a beam finder button. The vertical section controls the amplitude of the displayed signal. This section carries a Volts-per- Division (Volts/Div) selector knob, an AC/DC/Ground selector switch and the vertical (primary) input for the instrument. Additionally, this section is typically equipped with the vertical beam position knob. The horizontal section controls the time base or “sweep” of the instrument. The primary control is the Seconds-per-Division (Sec/Div) selector switch. Also included is a horizontal input for plotting dual X-Y axis signals. The horizontal beam position knob is generally located in this section.
  • 3. The trigger section controls the start event of the sweep. The trigger can be set to automatically restart after each sweep or it can be configured to respond to an internal or external event. The principal controls of this section will be the source and coupling selector switches. An external trigger input (EXT Input) and level adjustment will also be included. CRO Controls The controls available on most oscilloscopes provide a wide range of operating conditions and thus make the instrument especially versatile. Since many of these controls are common to most oscilloscopes a brief description of them follows. CATHODE-RAY TUBE Power and Scale Illumination: Turns instrument on and controls illumination of the graticule. Focus: Focus the spot or trace on the screen. Intensity: Regulates the brightness of the spot or trace. VERTICAL AMPLIFIER SECTION Position: Controls vertical positioning of oscilloscope display. Sensitivity: Selects the sensitivity of the vertical amplifier in calibrated steps. Variable Sensitivity: Provides a continuous range of sensitivities between the calibrated steps. Normally the sensitivity is calibrated only when the variable knob is in the fully clockwise position.
  • 4. AC-DC-GND: Selects desired coupling (ac or dc) for incoming signal applied to vertical amplifier, or grounds the amplifier input. Selecting dc couples the input directly to the amplifier; selecting ac send the signal through a capacitor before going to the amplifier thus blocking any constant component. HORIZONTAL-SWEEP SECTION Sweep time/cm: Selects desired sweep rate from calibrated steps or admits external signal to horizontal amplifier. Sweep time/cm Variable: Provides continuously variable sweep rates. Calibrated position is fully clockwise. Position: Controls horizontal position of trace on screen. Horizontal Variable: Controls the attenuation (reduction) of signal applied to horizontal aplifier through Ext. Horiz. connector. TRIGGER The trigger selects the timing of the beginning of the horizontal sweep. Slope: Selects whether triggering occurs on an increasing (+) or decreasing (-) portion of trigger signal. Coupling: Selects whether triggering occurs at a specific dc or ac level. Source: Selects the source of the triggering signal. INT - (internal) - from signal on vertical amplifier EXT - (external) - from an external signal inserted at the EXT. TRIG. INPUT. LINE - 60 cycle triger Level: Selects the voltage point on the triggering signal at which sweep is triggered. It also allows automatic (auto) triggering of allows sweep to run free (free run). CONNECTIONS FOR THE OSCILLOSCOPE Vertical Input: A pair of jacks for connecting the signal under study to the Y (or vertical) amplifier. The lower jack is grounded to the case. Horizontal Input: A pair of jacks for connecting an external signal to the horizontal amplifier. The lower terminal is graounted to the case of the oscilloscope. External Tigger Input: Input connector for external trigger signal. Cal. Out: Provides amplitude calibrated square waves of 25 and 500 millivolts for use in calibrating the gain of the amplifiers. Accuracy of the vertical deflection is + 3%. Sensitivity is variable. Bandwidth Bandwidth is a measure of the range of frequencies that can be displayed; it refers primarily to the vertical amplifier, although the horizontal deflection amplifier has to be fast enough to handle the fastest sweeps. The bandwidth of the oscilloscope is limited by the vertical amplifiers and the CRT (in analog instruments)
  • 5. or by the sampling rate of the analog to digital converter in digital instruments. The bandwidth is defined as the frequency at which the sensitivity is 0.707 of the sensitivity at lower frequency (a drop of 3 dB). The rise time of the fastest pulse that can be resolved by the scope is related to its bandwidth approximately: Bandwidth in Hz x rise time in seconds = 0.35 For example, an oscilloscope intended to resolve pulses with a rise time of 1 nanosecond would have a bandwidth of 350 MHz. For a digital oscilloscope, a rule of thumb is that the continuous sampling rate should be ten times the highest frequency desired to resolve; for example a 20 megasample/second rate would be applicable for measuring signals up to about 2 megahertz. Horizontal sweep should be accurate to within 3%. Range of sweep is variable. Operating Instructions: Before plugging the oscilloscope into a wall receptacle, set the controls as follows: (a) Power switch at off (b) Intensity fully counter clockwise (c) Vertical centering in the center of range (d) Horizontal centering in the center of range (e) Vertical at 0.2 (f) Sweep times 1 Plug line cord into a standard ac wall recepticle (nominally 118 V). Turn power on. Do not advance the Intensity Control. Allow the scope to warm up for approximately two minutes, then turn the Intensity Control until the beam is visible on the screen. WARNING: Never advance the Intensity Control so far that an excessively bright spot appears. Bright spots imply burning of the screen. A sharp focused spot of high intensity (great brightness) should never be allowed to remain fixed in one position on the screen for any length of time as damage to the screen may occur. Adjust Horizontal and Vertical Centering Controls. Adjust the focus to give a sharp trace. Set trigger to internal, level to auto. PROCEDURE: I. Set the signal generator to a frequency of 1000 cycles per second. Connect the output from the gererator to the vertical input of the oscilloscope. Establish a steady trace of this input signal on the scope. Adjust (play with)all of the scope and signal generator controls until you become familiar with the functionof each. The purpose fo such "playing" is to allow the student to become so familiar with the oscilloscope that it becomes an aid (tool) in making measurements in other experiments and not as a formidable obstacle. Note: If the vertical gain is set too low, it may not be possible to obtain a steady trace.
  • 6. II. Measurement s of Voltage: Consid er the circuit in Fig. 4(a). The signal generator is used to produce a 1000 hertz sine wave. The AC voltmeter and the leads to the verticle input of the oscilloscope are connected across the generator's output. By adjusting the Horizontal Sweep time/cm and trigger, a steady trace of the sine wave may be displayed on the screen. The trace represents a plot of voltage vs. time, where the vertical deflection of the trace about the line of symmetry CD is proportional to the magnitude of the voltage at any instant of time. To determine the size of the voltage signal appearing at the output of terminals of the signal generator, an AC (Alternating Current) voltmeter is connected in parallel across these terminals (Fig. 4a). The AC voltmeter is designed to read the dc "effective value" of the voltage. This effective value is also known as the "Root Mean Square value" (RMS) value of the voltage. The peak or maximum voltage seen on the scope face (Fig. 4b) is V m volts and is represented by the distance from the symmetry line CD to the maximum deflection. The relationship between the magnitude of the peak voltage displayed on the scope and the effective or RMS voltage (VRMS) read on the AC voltmeter is VRMS = 0.707 Vm (for a sine or cosine wave). Thus Agreement is expected between the voltage reading of the multimeter and that of the oscilloscope. For a symmetric wave (sine or cosine) the value of V m may be taken as 1/2 the peak to peak signal Vpp The variable sensitivity control a signal may be used to adjust the display to fill a concenient range of the scope face. In this position, the trace is no longer calibrated so that you can not just read the size of the signal by counting the number of divisions and multiplying by the scale factor. However, you can figure out what the new calibration is an use it as long as the variable control remains unchanged. Caution: The mathematical prescription given for RMS signals is valid only for sinusoidal signals. The meter will not indicate the correct voltage when used to measure non-sinusoidal signals.
  • 7. III. Frequency Measurements: When the horizontal sweep voltage is applied, voltage measurements can still be taken from the vertical deflection. Moreover, the signal is displayed as a function of time. If the time base (i.e. sweep) is calibrated, such measurements as pulse duration or signal period can be made. Frequencies can then be determined as reciprocal of the periods. Set the oscillator to 1000 Hz. Display the signal on the CRO and measure the period of the oscillations. Use the horizontal distance between two points such as C to D in Fig. 4b. Set the horizontal gain so that only one complete wave form is displayed. Then reset the horizontal until 5 waves are seen. Keep the time base control in a calibrated position. Measure the distance (and hence time) for 5 complete cycles and calculate the frequency from this measurement. Compare you result with the value determined above. Repeat your measurements for other frequencies of 150 Hz, 5 kHz, 50 kHz as set on the signal generator. IV. Lissajous Figures: When sine-wave signals of different frequencies are input to the horizontal and vertical amplifiers a stationary pattern is formed on the CRT when the ratio of the two frequencies is an intergral fraction such as 1/2, 2/3, 4/3, 1/5, etc. These stationary patterns are known as Lissajous figures and can be used for comparison measurement of frequencies. Use two oscillators to generate some simple Lissajous figures like those shown in Fig. 5. You will find it difficult to maintain the Lissajous figures in a fixed configuration because the two oscillators are not phase and frequency locked. Their frequencies and phase drift slowly causing the two different signals to change slightly with respect to each other. V. Testing what you have learned: Your instructor will provide you with a small oscillator circuit. Examine the input to the circuit and output of the circuit using your oscilloscope. Measure such quantities as the voltage and frequence of the signals. Specify if they are sinusoidal or of some other wave character. If square wave, measure the frequency of the wave. Also, for square waves, measure the on time (when the voltage is high) and off time (when it is low). Examples of use:-
  • 8. Lissajous figures on an oscilloscope, with 90 degrees phase difference between x and y inputs. One of the most frequent uses of scopes is troubleshooting malfunctioning electronic equipment. One of the advantages of a scope is that it can graphically show signals: where a voltmeter may show a totally unexpected voltage, a scope may reveal that the circuit is oscillating. In other cases the precise shape or timing of a pulse is important. In a piece of electronic equipment, for example, the connections between stages (e.g. electronic mixers, electronic oscillators, amplifiers) may be 'probed' for the expected signal, using the scope as a simple signal tracer. If the expected signal is absent or incorrect, some preceding stage of the electronics is not operating correctly. Since most failures occur because of a single faulty component, each measurement can prove that half of the stages of a complex piece of equipment either work, or probably did not cause the fault. Once the faulty stage is found, further probing can usually tell a skilled technician exactly which component has failed. Once the component is replaced, the unit can be restored to service, or at least the next fault can be isolated. This sort of troubleshooting is typical of radio and TV receivers, as well as audio amplifiers, but can apply to quite- different devices such as electronic motor drives. Another use is to check newly designed circuitry. Very often a newly designed circuit will misbehave because of design errors, bad voltage levels, electrical noise etc. Digital electronics usually operate from a clock, so a dual-trace scope which shows both the clock signal and a test signal dependent upon the clock is useful. Storage scopes are helpful for "capturing" rare electronic events that cause defective operation. Function generator
  • 9. Definition:- A function generator is usually a piece of electronic test equipment or software used to generate different types of electrical waveforms over a wide range of frequencies. Some of the most common waveforms produced by the function generator are the sine, square, triangular and sawtooth shapes. These waveforms can be either repetitive or single-shot (which requires an internal or external trigger source). Integrated circuits used to generate waveforms may also be described as function generator ICs. Feature:- Other important features of the function generator are continuous tuning over wide bands with max-min frequency ratios of 10:1 or more, a wide range of frequencies from a few Hz to a few MHz, a flat output amplitude and modulation capabilities like frequency sweeping, frequency modulation and amplitude modulation. Although function generators cover both audio and RF frequencies, they are usually not suitable for applications that need low distortion or stable frequency signals. When those traits are required, other signal generators would be more appropriate. Uses:- Function generators are used in the development, test and repair of electronic equipment. For example, they may be used as a signal source to test amplifiers or to introduce an error signal into a control loop. Working Simple function generators usually generate triangular waveform whose frequency can be controlled smoothly as well as in steps.[4] This triangular wave is used as the basis for all of its other outputs. The triangular wave is generated by repeatedly charging and discharging a capacitor from a constant current source. This produces a linearly ascending or descending voltage ramp. As the output voltage reaches upper and lower limits, the charging and discharging is reversed using a comparator, producing the linear triangle wave. By varying the current and the size of the capacitor, different frequencies may be obtained. Sawtooth waves can be produced by charging the capacitor slowly, using a current, but using a diode over the current source to discharge quickly - the polarity of the diode changes the polarity of the resulting sawtooth, i.e. slow rise and fast fall, or fast rise and slow fall. A 50% duty cycle square wave is easily obtained by noting whether the capacitor is being charged or discharged, which is reflected in the current switching comparator output. Other duty cycles (theoretically from 0% to 100%) can be obtained by using a comparator and the sawtooth or triangle signal. Most function generators also contain a non-linear diode shaping circuit that can convert the triangle wave into a reasonably accurate sine wave by rounding off the corners of the triangle wave in a process similar to clipping in audio systems. A typical function generator can provide frequencies up to 20 MHz. RF generators for higher frequencies are not function generators in the strict sense since they typically produce pure or modulated sine signals only.
  • 10. Function generators, like most signal generators, may also contain an attenuator, various means of modulating the output waveform, and often the ability to automatically and repetitively "sweep" the frequency of the output waveform (by means of a voltage-controlled oscillator) between two operator-determined limits. This capability makes it very easy to evaluate the frequency response of a given electronic circuit. Some function generators can also generate white or pink noise. More advanced function generators are called arbitrary waveform generators (AWG). They use direct digital synthesis (DDS) techniques to generate any waveform that can be described by a table of amplitudes. Specifications Typical specifications for a general-purpose function generator are:  Produces sine, square, triangular, sawtooth (ramp), and pulse output. Arbitrary waveform generators can produce waves of any shape.  It can generate a wide range of frequencies. For example, the Tektronix FG 502 (ca 1974) covers 0.1 Hz to 11 MHz.  Frequency stability of 0.1 percent per hour for analog generators or 500ppm for a digital generator.  Maximum sinewave distortion of about 1% (accuracy of diode shaping network) for analog generators. Arbitrary waveform generators may have distortion less than -55dB below 50 kHz and less than - 40dB above 50 kHz.  Some function generators can be phase locked to an external signal source, which may be a frequency reference or another function generator.  AM or FM modulation may be supported.  Output amplitude up to 10V peak-to-peak.  Amplitude can be modified, usually by a calibrated attenuator with decade steps and continuous adjustment within each decade.  Some generators provide a DC offset voltage, e.g. adjustable between -5V to +5V.  An output impedance of 50 ohms. Generator Control Panel - What it is and how it’s used The Control Panel – what is it? Visually, a control panel is a set of displays that indicate the measurement of various parameters like voltage, current and frequency, through gauges and meters. These meters and gauges are set in a metallic body, usually corrosion proof, to protect from the effect of rain or snow. The panel may be set up on the body of the generator itself, which is usually the case with small generators. If they are mounted on the generator, they typically have vibration proof pads that help isolate the control panel from shocks. Control panels for a larger industrial generators can be completely separate from the generator and are typically large enough to stand upon their own. These units may also be shelf-mounted or wall-mounted next to the generator, which is common inside an enclosure or intental application like a data center. Control panels are usually fitted with buttons or switches that help to operate the generator such as a switch-off button or turn-on key. The switches and gauges are usually grouped on the basis of functionality. This makes the panel friendly and safe for use since it minimizes the possibility of an operator accidentally selecting or executing the wrong control. Imagine trying to shut down a vibrating generator with a spring loaded lever in the middle of the night and you will appreciate why having a simple cut of switch at the control panel makes sense. • The front panel is divided into six major control groups: 1) Frequency Selection Group;
  • 11. 2) Sweep Group; 3) Amplitude Modulation Group; 4) DC Offset Group; 5) Function, or Waveform Group; and 6) Output Group. front panel of function generator • The power switch is on the upper left-hand corner of the unit. The green LED will indicate that the unit is on. • The three most important groups for this lab are the frequency, function, and output groups. The remaining three groups, (sweep, amplitude modulation, and DC offset) will be briefly covered in the lab setup procedures. Should you desire more detailed descriptions of these groups, the Leader Function Generator manual is available in the lab. (1) Frequency Selection Group:- These controls are used to select the operating frequency of the function generator. This group consists of the frequency control knob and the eight frequency multiplier selection buttons. (2) Output Group:- 1. These controls are used to adjust the amplitude of the generator's output signal. The group consists of the amplitude-control knob, the three attenuation buttons and the fused 50 ohm BNC connector. Although the amplitude knob is not indexed, the amplitude ranges from a few millivolts to approximately 20 volts. We will set the amplitude levels by aligning the white line on the amplitude knob to the three o'clock position (90 degrees right), the nine o'clock position (90 degrees left), or the twelve o'clock position (straight up). Notice that rotating the knob fully to the left does not result in a zero amplitude signal. • The attenuation buttons are used to attenuate (decrease) the amplitude of the signal by a factor measured in decibels. The following relationship will assist in working with the attenuation buttons: (dB) = -10 * log10 (Pout / Pin) (if power is the unit of measurement) or (dB) = -20 * log10(Vout / Vin) (if voltage is the unit of measurement) (3) Function/Waveform Selection Group:- This group is used to select the shape of the generated waveform. The group is made up of the six wave-selector buttons. The six waveforms that the function generator can
  • 12. produce are the sine wave, the square wave, the triangle wave, two sawtooth waves, and the variable-width pulse wave. How does it work? The control panel is becoming an increasingly complex piece of electronics with a microprocessor that can manipulate input from sensors to help give feedback to the machine to manage itself. One such feedback could be the temperature, indicating overheating, other examples would be over/under speed and low/high oil pressure. Typically, a heat sensor inside the generator would sense the build up of heat in the generator body and pass this to the microprocessor in the control panel. The microprocessor will then take effective measures to regulate the performance of the machine including shutdowns if, for example, the oil pressure is too low or the coolant temperature is too high, leading to buildup of heat. In industrial situations, this functionality of control panels is becoming increasingly critical. The microprocessor or microcontroller is embedded in the circuitry inside the control panel and is programmed to take in the sensor input and react to that with the programmed control rules. Control panels can be combined with an Automatic Transfer Switch (ATS) to maintain the continuity of electrical power. The ATS detects an outage of power when your local grid fails. It signals the control panel to start the generator. Depending on the type of generator being used, the control panel may activate glow plugs (for diesel) for an adjustable length of time. It will then start the generator using an automatic starter, similar to the one you engage when you turn the keys in the ignition of your car in the morning. As soon as the engine of the generator reaches an optimum speed, the starter is disengaged. The ATS then switches to the generator power, and you can go back to business as usual, without having to frantically scramble to figure out what caused power loss. This aspect of a control panel makes it extremely useful in homes during bad weather and in industrial situations for ensuring mission-critical continuity.