World Bank & Government of The Netherlands funded
Training module # WQ - 33
Use of Ion Selective Probes
New Delhi, February 2000
CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas,
New Delhi – 11 00 16 India
Tel: 68 61 681 / 84 Fax: (+ 91 11) 68 61 685
E-Mail: dhvdelft@del2.vsnl.net.in
DHV Consultants BV & DELFT HYDRAULICS
with
HALCROW, TAHAL, CES, ORG & JPS
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 1
Table of contents
Page
1 Module context 2
2 Module profile 3
3 Session plan 4
4 Overhead/flipchart master 5
5 Evaluation sheets 21
6 Handout 23
7 Additional handout 28
8 Main text 30
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 2
1. Module context
This module discusses the use of ion selective electrodes recommended for water quality
analysis under the Hydrology Project. Modules in which prior training is required and other
related modules are listed below.
While designing a training course, the relationship between this module and the others,
would be maintained by keeping them close together in the syllabus and place them in a
logical sequence. The actual selection of the topics and the depth of training would, of
course, depend on the training needs of the participants, i.e. their knowledge level and skills
performance upon the start of the course.
No. Module title Code Objectives
1. Basic chemistry concepts WQ - 02 • Convert units from one to another
• Discuss the basic concepts of
quantitative chemistry
• Report analytical results with the
correct number of significant digits.
2. Understanding the hydrogen
ion concentration (pH)
WQ - 06 • Discuss about the concept of pH
• Calculate pH
3. Understanding chemical
oxygen demand test
WQ - 18 • Appreciate significance of COD
measurement
• Understand the chemistry of COD
measurement
4. Basic aquatic chemistry
concepts
WQ - 24 • Understand equilibrium chemistry and
ionisation constants.
• Understand basis of pH and buffers
• Calculate different types of alkalinity.
5. Advanced aquatic chemistry:
solubility equilibria
WQ - 29 • Explain the principles of chemical
equilibrium
• Define solubility product and explain
how this relates to water quality
assessment
• Define the octanol-water partition
coefficient and explain how this
relates to water quality assessment.
6. Advanced aquatic chemistry:
redox equilibria
WQ - 30 • Understand the principles of redox
reactions
• Understand the significance of redox
potential measurement
7. Potentiometric Analysis WQ - 32 • Understand the principles of
potentiometric analysis
• Understand construction of reference
and indicating electrodes
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 3
2. Module profile
Title : Use of Ion Selective Probes
Target group : HIS function(s): Q2, Q3, Q5, Q6
Duration : One session of 45 min
Objectives : After the training the participants will:
• understand precautions required in use of ion selective
electrodes
Key concepts : • ion selective electrodes recommended under HP
• calibration
• precaution
Training methods : Lecture, exercises
Training tools
required
: Board, flipchart
Handouts : As provided in this module
Further reading
and references
: • Chemistry for environmental engineers - C. N. Sawyer, P. L.
McCarty & G. F. Parkin, McGraw - Hill, Inc., 1994
• Standard methods for the examination of water and
wastewaters, AWWA, 19th
edition, 1995
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 4
3. Session plan
No Activities Time Tools
1 Preparations
2 Introduction:
• Discuss various types of indicating electrodes and probes
5 min OHS
3 Performance characteristics
• Describe precautions, limitations and advantanges of
potentiometric measurements using indicating electrodes
and probes
• Discuss calibration requirements and methods
15 min OHS
4 Electrodes and probes recommended under HP
• Discuss applications of the recommended probes and
procedures for their use.
15 min OHS
5 Wrap up
• Discuss experience of participants, if any
• Review precautions and limitations
10 min
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 5
4. Overhead/flipchart master
OHS format guidelines
Type of text Style Setting
Headings: OHS-Title Arial 30-36, with bottom border line (not:
underline)
Text: OHS-lev1
OHS-lev2
Arial 24-26, maximum two levels
Case: Sentence case. Avoid full text in UPPERCASE.
Italics: Use occasionally and in a consistent way
Listings: OHS-lev1
OHS-lev1-Numbered
Big bullets.
Numbers for definite series of steps. Avoid
roman numbers and letters.
Colours: None, as these get lost in photocopying and
some colours do not reproduce at all.
Formulas/Equat
ions
OHS-Equation Use of a table will ease horizontal alignment
over more lines (columns)
Use equation editor for advanced formatting
only
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 6
Ion selective probes (1)
• Electrode potential is linearly related to log of activity of ion
• Glass electrode for H+
measurement is most commonly used.
• Probes for DO, F-
, NH3 and CN-
recommended under HP
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 7
Types and examples
• Glass membrane Na+
, K+
, H+
, NH4
+
• Liquid membrane NO3
-
, Ca2+
, Mg2+
• Solid membrane S2-
, F-
, CN-
, Pb2+
• Diaphragm NH3, O2, HCN
• Metal Ag+
, Cu2+
, Cd2+
, I-
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 8
Performance characteristics (1)
• Temperature effect
- calibration standard and sample at same temperature
- meter in-built correction
• Response time
- steady value, few seconds to minutes
• Potential drift
- recalibrate frequently
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 9
Performance characteristics (2)
• Sensitivity
- working range, depends on properties of membrance
- response 59.15 mV per ten fold conc. change for one electron.
• Selectivity and interference
- other ions may effect reading
- remove interfering ion
- calibrate in presence of the ion.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 10
Performance characteristics (3)
• Activity vs. concentration
- measures activity
- calibration standard of similar ionic strength.
• pH
- may affect ionisation
- formation of complexes
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 11
Performance characteristics (4)
• Presence of complexing agent
- electrodes measure free ions
- remove complexing agents
• Life time
- rate of leaking of liquid ion exchanger
- chemical fouling of solid membranes
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 12
Advantages
• turbid samples
• small volume, sample not destroyed
• do not change existing equilibria
• direct reading of conc. over wide range
• sample does not require pretreatment
• rapid response
• ideal for on-line measurement
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 13
Electrode calibration
• Calibration curve
- treatment of sample and standard for adjustment of pH and ionic
strength
• Standard addition
- when effect of ions in sample matrix can not be duplicated in
standard
- significant liquid junction potential
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 14
Figure 1 Response of Ca electrode as functions of Ca++
ion
concentration and activity
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 15
Standard addition
• Eobs = Eref + Ej - Eind
which can be written as:
• K is the unknown change in potential
• Observe potential before and after adding known standard to
sample and solve for 'a', the analyte concentration
− =
−
log
. /
a
E K
n
obs
0 0591
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 16
Dissolved oxygen probe
• Site and on-line measurements
• Frequent calibration
- air
- water of known DO
• Fouling of membrane
- change when response is sluggish
- do not trap air bubble
• Sample flow across membrane while measuring
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 17
Fluoride electrode
• Evaluate suitability critically
• Al3+
and Fe2+
cause interference
• At pH < 5, HF.HF is formed
• Add complexing agent and pH buffer
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 18
Ammonia probe
• Wide range, 0.03 - 1400 mg/L
• Interference due to amines, Hg & Ag
• Increase pH to convert NH4
+
to NH3
• Stir sample slowly.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 19
Nitrate electrode
• Alternative to UV method and convenient compared to Cd
reduction method
• Adjust pH to 3 eliminate HCO3
-
• Requires complexing agents for other interfering ions
• Check calibration frequently
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 20
Cyanide electrode
• Distillation of sample is required
• Range 0.025 - 10 mg CN-
/L
• Response time at least 5 min, mix well
• Do not work with acidified sample in open, HCN is extemely
toxic gas
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 21
5. Evaluation sheets
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 22
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 23
6. Handout
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 24
Ion selective probes (1)
• Electrode potential is linearly related to log of activity of ion
• Glass electrode for H+
measurement is most commonly used.
• Probes for DO, F-
, NH3 and CN-
recommended under HP
Types and examples
• Glass membrane Na+
, K+
, H+
, NH4
+
• Liquid membrane NO3
-
, Ca2+
, Mg2+
• Solid membrane S2-
, F-
, CN-
, Pb2+
• Diaphragm NH3, O2, HCN
• Metal Ag+
, Cu2+
, Cd2+
, I-
Performance characteristics (1)
• Temperature effect
- calibration standard and sample at same temperature
- meter in-built correction
• Response time
- steady value, few seconds to minutes
• Potential drift
- recalibrate frequently
Performance characteristics (2)
• Sensitivity
- working range, depends on properties of membrance
- response 59.15 mV per ten fold conc. change for one electron
• Selectivity and interference
- other ions may effect reading
- remove interfering ion
- calibrate in presence of the ion.
Performance characteristics (3)
• Activity vs. concentration
- measures activity
- calibration standard of similar ionic strength.
• pH
- may affect ionisation
- formation of complexes
Performance characteristics (4)
• Presence of complexing agent
- electrodes measure free ions
- remove complexing agents
• Life time
- rate of leaking of liquid ion exchanger
- chemical fouling of solid membranes
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 25
Advantages
• turbid samples
• small volume, sample not destroyed
• do not change existing equilibria
• direct reading of conc. over wide range
• sample does not require pretreatment
• rapid response
• ideal for on-line measurement
Electrode calibration
• Calibration curve
- treatment of sample and standard for adjustment of pH and ionic strength
• Standard addition
- when effect of ions in sample matrix can not be duplicated in standard
- significant liquid junction potential
Standard addition
• Eobs = Eref + Ej - Eind
which can be written as:
• K is the unknown change in potential
• Observe potential before and after adding known standard to sample and solve
for 'a', the analyte concentration
Dissolved oxygen probe
• Site and on-line measurements
• Frequent calibration
- air
- water of known DO
• Fouling of membrane
- change when response is sluggish
- do not trap air bubble
• Sample flow across membrane while measuring
Fluoride electrode
• Evaluate suitability critically
• Al3+
and Fe2+
cause interference
• At pH < 5, HF.HF is formed
• Add complexing agent and pH buffer
Ammonia probe
• Wide range, 0.03 - 1400 mg/L
• Interference due to amines, Hg & Ag
• Increase pH to convert NH4
+
to NH3
• Stir sample slowly.
− =
−
log
. /
a
E K
n
obs
0 0591
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 26
Nitrate electrode
• Alternative to UV method and convenient compared to Cd reduction method
• Adjust pH to 3 eliminate HCO3
-
• Requires complexing agents for other interfering ions
• Check calibration frequently
Cyanide electrode
• Distillation of sample is required
• Range 0.025 - 10 mg CN-
/L
• Response time at least 5 min, mix well
• Do not work with acidified sample in open, HCN is extemely toxic gas
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 27
Add copy of Main text in chapter 8, for all participants.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 28
7. Additional handout
These handouts are distributed during delivery and contain test questions, answers to
questions, special worksheets, optional information, and other matters you would not like to
be seen in the regular handouts.
It is a good practice to pre-punch these additional handouts, so the participants can easily
insert them in the main handout folder.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 29
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 30
8. Main text
Contents
1. Introduction 1
2. Performance characteristics of
electrodes 2
3. Electrode calibration 3
4. Recommended probes under HP 5
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 1
Use of Ion Selective Probes
1. Introduction
Ion-selective electrodes belong to a class of electro-chemical sensors which respond to only
the ion of interest. Their use involves a special case of direct potentiometry. When these
electrodes are placed in a conducting solution, they generate a millivolt potential related to
the concentration of the ion. The potential is linearly related to log of concentration of the ion.
To be more accurate, the term ion ‘activity’ should be used instead of concentration. For
simplicity, this text will use the term concentration, which in most cases is a close
approximation of activity.
Historically, the oldest and the best known and most widely used ion-selective electrode is
the glass electrode, which responds to hydrogen ion activity. It remained the only sensing
electrode until mid 1960’s. But the renewed interest in potentiometry has now led to the
commercial availability of a range of ion-selective electrodes.
Basic principles of redox reactions and construction of various types of indicating electrodes
and probes are discussed in module # 30, 'Advanced Aquatic Chemistry Concepts: Redox
Equilibria, and module # 32, 'Potentiometric analysis'. This module describes general
aspects of ion selective probes and use of DO, F-
, NH3
-
and CN-
probes which are
recommended under Hydrology Project.
Table 1 gives an overview of indicator electrodes commonly available
Table 1: Types and examples of ion electrodes
Types of electrode Description Examples of
chemical specie
measured
Glass membrane
electrode
Electrode having glass sensing
membrane
Na+
, K+
, H+
, NH4
+
Liquid membrane
electrode
Electrode having sensing
membrane comprising an ion
exchange liquid held in a rigid
porous membrane
NO3
-
, Ca2+
, Mg+2
,
NH4
+
Solid membrane
electrode
Selective electrode having
sensing membrane made of
poorly soluble metal salt
monocrystal or metal salt
powder
S2-
, F-
, CN-
, Pb2+
,
Ag+
Diaphragm electrode: Electrodes combining gas
transmitting membrane and pH
electrode or ion electrode
NH3, HCN, O2
Metal electrode Electrode comprising the metal
or the metal and a poorly
soluble salt.
Ag+
, Cu2+
, Cd2+
,
Pb2+
, I-
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 2
2. Performance characteristics of electrodes
Temperature effect:
It is necessary to correct for the temperature of the samples because the electrode potential
is temperature dependent. The sample temperature should be measured independently and
a temperature correction should be made with the in-built correction in the meter.
Response time:
The response time of ion-selective electrode is the speed at which it responds (electrode
potential reaches a steady state value), and depends upon the type of electrode and
concentration of ion under measurement. The response time of solid – state electrode is
generally of the order of 5-10 seconds. The response time of some liquid ion-exchanger
electrode is longer, usually many seconds or even few minutes at low concentrations. When
making a reading, enough time must be given for the electrode to reach a stable reading.
Potential drifts:
The potential drifts of ion-selective electrodes is connected with changes in the surface
structure of solid – membrane electrode due to contact with the electrolyte and due to
dissolution of the ion exchanger of liquid-membrane electrodes. All electrodes show drifts
with time and therefore need re-calibration periodically with standardising solutions. This
should be done every few hours if making many analyses.
Sensitivity:
The useful working range of the ion-selective electrodes is knows as sensitivity which is
effectively governed by the properties of the membrane. The working range of electrode is
determined from the concentration (mg/l) to which an electrode responds in a theoretical
fashion (59.15 mV/ten fold change in concentration when one electron is involved in the
reaction).
Selectivity & Interferences:
Although intended to be specific to a single ionic species, most electrodes respond in
varying degrees to the presence of other ions. Where severe interferences are likely to be
encountered, there are two practical courses of action. Either the electrode must be
calibrated in the presence of interfering ion or the interference must be eliminated by some
other means such as buffering, precipitation or by adding a complexing agent.
Activities vs. Concentration:
Electrodes measure activities and not concentrations. The activity varies with ionic strength.
Thus, to use an electrode for concentration measurements, the electrode must be calibrated
in a standard solution of similar ionic strength (see section 3. Electrode calibration).
Effect of pH:
The pH of the sample solution changes the chemistry of the analyte ion. For instance, at low
pH values F-
ions are present as HF, which are not sensed by the fluoride electrode. Thus in
such cases an appropriate pH adjuster should be added to samples to set up a reliable ion-
selective electrode analysis.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 3
Presence of complexing agent:
A number of ions in water samples can be present in more than one form, such as
complexes with other ions or molecules. Electrodes measure only the free ion in solution,
and cannot sense bound or complexed ions.
Life - time:
The life-time of an electrode is determined by the rate at which the liquid ion exchanger
leaks through the membrane in liquid membrane electrodes, or the time required for crystal
surface or the pressed pellet to become chemically fouled in solid membrane electrodes.
Life-time markedly depends upon conditions of use. With good care, electrodes should
provide many months or years of service. If electrode response drops off, rejuvenation of
electrodes is possible by refilling the liquid internals, replacing sensing membrane of liquid
membrane type electrode or mechanical polishing of solid – state membrane tip.
The use of ion-selective electrodes as an analytical techniques in water analysis is
increasing. This is because of the significant advantages such as:
• They are not bothered by turbidity or colour of the sample
• Small volumes of samples can be analysed without loss of sample
• Measure species without upsetting chemical equilibrium
• Measurements are independent of volume of sample
• Concentrations can be read directly on specific ion meter
• Calibration curves can cover even six decades of concentration
• Electrodes are sensitive up to parts per billion
Further application of these electrodes especially for on-line analysis are very much in use
because:
• They do not usually require extensive pre-treatment of the sample
• They have a relatively rapid response-time
• They provide an electrical analogue signal which does not need sophisticated signal
conditioning circuits before a result can be displayed to the instrument user
• Considerable previous experience of on-line instrumentation for pH measurement is
available, which is relevant to other electrodes also.
3. Electrode calibration
Activity vs. concentration:
Electrode response is related to analyte activity rather than to analyte concentration. More
often than not, activity coefficients are not available. The difference between activity and
concentration is illustrated by Figure 1 in which the response of a calcium electrode is
plotted against logarithmic functions of the calcium concentration and calcium activity. The
electrode responds in an ideal fashion, i.e., constant slope, only when concentration values
are changed to activity values.
An obvious way to convert potentiometric measurements from activity to concentration is to
make use of an emperical calibration curve, such as the lower plot in Figure 1. For this
approach to be successful, it is necessary to make the ionic composition of the standards
essentially the same as that for the analyte solution. This is often done by swamping both
sample and standards with a measured excess of an inert electrolyte, thus making the
added effect of electrolyte from the sample matrix negligible.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 4
Figure 1 Response of Ca electrode as functions of Ca++
ion concentration and activity
Standard addition method:
Notwithstanding the advantages of direct potentiometric measurements, there are certain
inherent limitations to the method. Principal among these is the existance of a 'liquid
junction potential' that affects most potentiometric measurements. The junction potential, Ej,
is created at the interface of the electrode and the analyte solution due to unequal mobility of
various cations and anions, which modifies the observed potential, Eobs, according to the
following expression
(1)
where a is the analyte activity, Eind is the actual indicator electrode potential and K is a
constant.
In case the effect of liquid junction potential is to be eliminated and where the effect of the
matrix of the sample cannot be swamped by addition of an inert electrolyte, standard
addition method may be used. It involves determining the potential of the electrode system
before and after a measured volume of a standard has been added to a known volume of
the analyte solution. As in the other method, if required, an excess of an electrolyte is added
to the analyte solution to prevent a major shift in the ionic strength due to addition of the
standard. The following example explains the calculations involved in the method.
E E E E E E E
n a
obs ref j ind ref j= + − = + − −
F
HG I
KJc h 0 0 0591 1.
log
− =
−
log
. /
a
E K
n
obs
0 0591
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 5
Example 1:
A lead ion electrode developed a potential of 0.4706 V (vs. Standard Calomel Electrode)
when immersed in 50.00 mL of a sample. A 5.00 mL addition of standard 0.02000 M lead
solution caused the potential to shift to 0.4490 V. Calculate the molar concentration of lead
in the sample.
Applying Equation 1 and assuming that the activity of Pb2+
is approximately equal to its
molar concentration Cx, we may write
Where E1 is the measured potential (0.4706 V).
After introduction of the standard addition, the potential is now E2 (0.4490 V) and
which can be written as
Subtraction of this equation from the first gives
4. Recommended probes under HP
This section describes the use of probes recommended under HP. As stated earlier, while
the probes have their advantages, their limitations must also be kept in mind when used in
different situations.
For each case the manufacturer's instructions for calibration and operation of the probe and
the ion-meter should be followed.
− =
−
log
.
C
E K
x
1
0 0591
2
−
+
+
=
−
log
. . .
. . .
50 00 5 00 0 02000
50 00 5 00 0 0591
2
2X C X E Kx
− + =
−−
log ( . . )
.
0 9091 1818 10
0 0591
2
3 2
C X
E K
x
−
+
=
−
=
−
=
−
log
. . .
. .
.
.
C
C X
E Ex
x0 9091 1818 10
2
0 0591
2 0 4706 0 4490
0 0591
0 7310
3
1 2b g
b g
C
C X
C X M
x
x
x
0 9091 1818 10
01858
4 06 10
3
4
. .
.
.
+
=
=
−
−
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 6
4.1 Dissolved oxygen probe
Application:
DO probes are useful for on site measurements and when a large number of samples are to
be analysed in the laboratory, for example, when determining BOD. For continuous DO
measurement the potentiometric measurement is the only option.
Construction:
The probe consists of two solid metal electrodes in contact with a supporting electrolyte
separated form the test solution by a selective membrane permeable to molecular oxygen.
Interference:
The probe exhibits a high temperature coefficient. The ion-meter may have in-built
temperature correction or correction factor may be available from the manufacturer.
Prolonged use of the probe in waters containing sulphides lowers the sensitivity of the probe.
Calibration:
The probe may be calibrated by reading against air, a sample containing known DO
concentration (determined by iodometric titration) or a sample containing zero DO (by adding
sulphite to the sample). Periodic cross-check of calibration by different methods is
recommended. Preferably calibrate using the sample water. In case the sample water
contains interfering substances, calibrate with a sample of clean water containing
approximately the same salt content.
Precautions:
Change the probe membrane when the response becomes sluggish or erratic. Take care to
avoid contamination of the sensing element and trapping of minute air bubbles while
changing the membrane. Provide sufficient sample flow across the membrane surface while
making measurements.
4.2 Fluoride electrode
Applications:
Fluoride electrode has been used for determination of fluoride in drinking water supplies.
The analyst should be aware of its limitations as described in the section on interferance
below. Its suitability should be evaluated since even small errors can be critical in terms of
suitability of a water for drinking purpose.
Construction:
The key element in the fluoride electrode is the lanthanum fluoride crystal across which a
potential is established by fluoride solutions of different concentrations, in the sample and
the standard solution in the electrode. The electrode can be used with a standard calomel
reference electrode.
Interference:
Fluoride forms complexes with several cations notably aluminium and iron. At pH values
lower that 5 poorly ionisable HF.HF complex is formed. Cyclohexylene-diamine-tetra acetic
acid and a buffer can be added to the sample to complex the interfering ions. The buffer
maintains a pH of 5. This pH also reduces the effect of the interfering ions. Samples having
TDS in excess of 10,000 mg/L should be distilled.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 7
Calibration:
A calibration curve in the 0.2 to 2 mg/L F-
range may be prepared. The recommended
complexing agent and buffer should be added to the sample as well as the standards. This
would make the ionic strength of the sample similar to that of the standard.
Precautions:
Treat the sample and the standard as described above. Avoid stirring before immersing
electrode so as not to entrap air bubbles. Let electrode remain in solution until the reading is
constant. Recalibrate frequently against atleast one standard.
4.3 Ammonia probe
Applications:
The method can be used for a wide range of ammonia concentration from 0.03 to 1400 mg/L
NH3-N. It is suitable for coloured and turbid water samples.
Construction:
The probe consists of a pH electrode and a reference electrode. It is provided with a
membrane which allows permeation of NH3. The permeated gas reacts with an internal
solution and changes its pH, which is sensed by the pH electrode.
Interference:
The probe has a high temperature coefficient. The temperature correction is generally
available from the manufacturer. Amines cause a positive interference. Hg and Ag form
complexes with ammonia. This can be overcome by adding NaOH/EDTA to the sample.
Calibaration:
Calibrate the instrument using a series of standards in the range of 0.1mg/L - 1000 mg/L
NH3-N prepared from NH4Cl. Sufficient quantity of NaOH is added to convert NH4
+
to NH3.
In case presence of Hg or Ag is suspected in the sample ass NaOH/EDTA solution to the
standard also. Same volume of the reagent should be added to the standards as well as the
sample. A 10 fold increase in NH3 concentration the potential change should be about
59mV.
Precautions:
Calibration standards and samples should be at the same temperature. Do not add NaOH
before immersing the probe to avoid loss of ammonia. Stir slowly and maintain the same
speed while taking readings. For ammonia concentration below 1mg/L, 2 to 3 minutes may
be required to obtain a stable response from the electrode.
4.4 Nitrate electrode
Applications:
Potentiometric analysis for nitrate is an alternative to UV absorption method where presence
of organic matter in the sample interferes with the spectrophotometric method. It is also
convenient and time saving when compared to cadmium reduction method. The use of the
electrode may be limited in case of saline samples (see interferences below). The electrode
responds to NO3
-
-N concentration in 0.14-1400 mg/L range.
Construction:
The electrode has a sensitive liquid ion-exchange membrane held in an inert membrane. It
develops a potentia across the membrane whose one face is kept in contact with the analyte
in the sample and the other in contact with an inner standard solution.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 8
Interferences:
Chloride and bicarbonate ions interfere when their weight raios to NO3
-
-N are in excess of 10
and 5, respectively. Variation in pH gives erratic response. Other ions which may interfere
are NO2
-
, CN-
, S2-
, Br-
, I-
and NO2
-
.
Calibration:
A buffer is added to the standards and the sample to adjust the solutions to similar ionic
strength. The buffer maintains a pH of 3, which eleminates HCO3
-
. The buffer also contains
Ag2(SO4) to remove interference from Cl-
, Br-
, S2-
, and CN-
, and sulfamic acid to remove
NO2
-
.
Precautions:
The standards and the samples should be at same temperature preferably 250
C. The
calibration curve, when mV potential reading is plotted as ordinate versus log NO3
-
-N
concentration as abscissa, should result in a straight line with a slope of +57±3mV/decade.
Check the electrode calibration several times a day using at least one standard.
4.5 Cyanide electrode
Applications:
A preliminary step of distillation of sample and collection of CN-
in alkaline solution is
required. The electrode responds to CN-
concentration in the range of 0.05 to 10 mg/L
Construction:
The electrode uses a solid crystalline membrane sensitive to CN-
.
Interferences:
Since the determination is made on the distillate, ordinarily interference from other ions is not
expected. Lead carbonate may be added to absorbing solution if sulphides are expected in
the distillate.
Calibration:
Use a series of 4 or more standards in the range of 0.025-10mg CN-
/L. Proceed from the
lowest to the highest concentration while taking readings. Mix well using a magnetic stirrer.
Allow at least 5 min for reaching equilibrium.
Precaution:
HCN is an extremely toxic gas. Follow all precautions as detailed in the distillation step in
standard procedures. Do not word with an acidified sample is open. The standards and the
samples should be at the same temperature, preferably as close to 250
C as possible.
Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 9
4.6 Use and care of electrodes
An overview of initial use, storage and recovery for ion electrodes is given in Table 2.
Table 2 : Initiation of use of ion electrodes, storage and recovery
Kind of
electrode
Initiation of use Storage Recovery
Glass
membrane
Immerse in pure water
or diluted standard
liquid for several
hours.
Sufficiently wash in
pure water, immerse in
the diluted standard
liquid. For long term
storage wipe, dry, store
in the air safely.
Wipe the glass
membrane with soft
cloth or sponge
impregnated with
neutral detergent to
clean. If the
electrode is used in
alkali, immerse it in
1N HNO3 for 1 – 2
hours
Crystalline
solid
membrane
Holding the electrode
head, shake gently 1-
2 times, immerse it in
the diluted standard
solution for about 10
min.
Wash in pure water,
immerse in diluted
standard liquid. For
storage, wipe dry, store
in air safely.
Wipe surface with
cloth or absorbent
wadding
impregnated with
ethanol to clean. If
the contamination
persists, gently rub
the surface with soft
wash-leather
Non-
Crystalline
solid
membrane
Wipe it with tissue,
immersed in the
standard liquid for 10
min. If contaminated,
perform the recovery
treatment, immerse
dilute standard liquid
for 10 min.
Wash with pure water,
immerse in diluted
standard liquid. Wipe
thoroughly, dry and
store safely in air.
Drip pure water on
the fine polishing
paper, polish
uniformly and gently.
Wash thoroughly
with pure water, wipe
with tissues.
Liquid
membrane
Holding the electrode
head, shake gently 1-
2 time, immerse in
diluted standard
solution for 10 min.
Wash with pure water,
immerse in diluted
standard liquid. Wipe,
dry and store safely in
the air.
Replenish/replace
the liquid membrane
according to
instruction manual.
Immerse in high
concentration
standard liquid.
Diaphragm Set the internal liquid
and diaphragm,
immerse in pure water
of diluted standard
liquid for 10 min.
Wash with pure water,
immerse in diluted
standard liquid.
Remove the
diaphragm, discharge
the internal liquid wash
with pure water, dry
and store safely the
main body and
diaphragm
Replace the
contaminated /
damaged diaphragm
remove
contamination with
immersing in 1N
HNO3 . Replace the
internal liquid as per
instruction manual.

Weitere ähnliche Inhalte

Was ist angesagt?

Development of an aqueous ammonia-based post-combustion capture technology fo...
Development of an aqueous ammonia-based post-combustion capture technology fo...Development of an aqueous ammonia-based post-combustion capture technology fo...
Development of an aqueous ammonia-based post-combustion capture technology fo...Global CCS Institute
 
Investigation of Solid State Hydrides For Autonomous Fuel Cell Vehicles
Investigation of Solid State Hydrides For Autonomous Fuel Cell VehiclesInvestigation of Solid State Hydrides For Autonomous Fuel Cell Vehicles
Investigation of Solid State Hydrides For Autonomous Fuel Cell Vehicleschrisrobschu
 
Sasha Poster Board Final- NH3
Sasha Poster Board Final- NH3Sasha Poster Board Final- NH3
Sasha Poster Board Final- NH3Sasha Heralall
 
Caro water tech presentation final
Caro water tech presentation finalCaro water tech presentation final
Caro water tech presentation finalPatrick Novak
 
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...Global CCS Institute
 
Decontamination of mercury impacted process equipment
Decontamination of mercury impacted process equipmentDecontamination of mercury impacted process equipment
Decontamination of mercury impacted process equipmentISCT GROUP US LLC
 
cement lock battelle poster final stern
cement lock battelle poster final sterncement lock battelle poster final stern
cement lock battelle poster final sternEric A. Stern
 
Geometric relaxation time of induced polarization fractal dimension for chara...
Geometric relaxation time of induced polarization fractal dimension for chara...Geometric relaxation time of induced polarization fractal dimension for chara...
Geometric relaxation time of induced polarization fractal dimension for chara...Khalid Al-Khidir
 
The South West Hub Project: Developing a project in unconventional geology
The South West Hub Project: Developing a project in unconventional geologyThe South West Hub Project: Developing a project in unconventional geology
The South West Hub Project: Developing a project in unconventional geologyGlobal CCS Institute
 
Environmental forensics for methane source identification
Environmental forensics for methane source identificationEnvironmental forensics for methane source identification
Environmental forensics for methane source identificationChemistry Matters Inc.
 

Was ist angesagt? (12)

Development of an aqueous ammonia-based post-combustion capture technology fo...
Development of an aqueous ammonia-based post-combustion capture technology fo...Development of an aqueous ammonia-based post-combustion capture technology fo...
Development of an aqueous ammonia-based post-combustion capture technology fo...
 
Investigation of Solid State Hydrides For Autonomous Fuel Cell Vehicles
Investigation of Solid State Hydrides For Autonomous Fuel Cell VehiclesInvestigation of Solid State Hydrides For Autonomous Fuel Cell Vehicles
Investigation of Solid State Hydrides For Autonomous Fuel Cell Vehicles
 
Sasha Poster Board Final- NH3
Sasha Poster Board Final- NH3Sasha Poster Board Final- NH3
Sasha Poster Board Final- NH3
 
Caro water tech presentation final
Caro water tech presentation finalCaro water tech presentation final
Caro water tech presentation final
 
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...
Dispersion modelling for CO2 pipelines: Fit for purpose and best practice tec...
 
Decontamination of mercury impacted process equipment
Decontamination of mercury impacted process equipmentDecontamination of mercury impacted process equipment
Decontamination of mercury impacted process equipment
 
cement lock battelle poster final stern
cement lock battelle poster final sterncement lock battelle poster final stern
cement lock battelle poster final stern
 
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
 
Adsorption Materials and Processes for Carbon Capture from Gas-Fired Power Pl...
Adsorption Materials and Processes for Carbon Capture from Gas-Fired Power Pl...Adsorption Materials and Processes for Carbon Capture from Gas-Fired Power Pl...
Adsorption Materials and Processes for Carbon Capture from Gas-Fired Power Pl...
 
Geometric relaxation time of induced polarization fractal dimension for chara...
Geometric relaxation time of induced polarization fractal dimension for chara...Geometric relaxation time of induced polarization fractal dimension for chara...
Geometric relaxation time of induced polarization fractal dimension for chara...
 
The South West Hub Project: Developing a project in unconventional geology
The South West Hub Project: Developing a project in unconventional geologyThe South West Hub Project: Developing a project in unconventional geology
The South West Hub Project: Developing a project in unconventional geology
 
Environmental forensics for methane source identification
Environmental forensics for methane source identificationEnvironmental forensics for methane source identification
Environmental forensics for methane source identification
 

Andere mochten auch

Download-manuals-water quality-wq-training-43howtomeasuretotaliron
 Download-manuals-water quality-wq-training-43howtomeasuretotaliron Download-manuals-water quality-wq-training-43howtomeasuretotaliron
Download-manuals-water quality-wq-training-43howtomeasuretotalironhydrologyproject0
 
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogenhydrologyproject0
 
Download-manuals-water quality-wq-training-36measurementoffluoride
 Download-manuals-water quality-wq-training-36measurementoffluoride Download-manuals-water quality-wq-training-36measurementoffluoride
Download-manuals-water quality-wq-training-36measurementoffluoridehydrologyproject0
 
Download-manuals-water quality-wq-training-42measurementofboron
 Download-manuals-water quality-wq-training-42measurementofboron Download-manuals-water quality-wq-training-42measurementofboron
Download-manuals-water quality-wq-training-42measurementofboronhydrologyproject0
 
Download-manuals-water quality-wq-training-30advancedaquaticchemistry
 Download-manuals-water quality-wq-training-30advancedaquaticchemistry Download-manuals-water quality-wq-training-30advancedaquaticchemistry
Download-manuals-water quality-wq-training-30advancedaquaticchemistryhydrologyproject0
 
Download-manuals-water quality-wq-training-23howtomeasurecoliforms
 Download-manuals-water quality-wq-training-23howtomeasurecoliforms Download-manuals-water quality-wq-training-23howtomeasurecoliforms
Download-manuals-water quality-wq-training-23howtomeasurecoliformshydrologyproject0
 
Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...
 Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil... Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...
Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...hydrologyproject0
 
Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters
 Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters
Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewatershydrologyproject0
 
Download-manuals-water quality-wq-manuals-28majorionsinwater
 Download-manuals-water quality-wq-manuals-28majorionsinwater Download-manuals-water quality-wq-manuals-28majorionsinwater
Download-manuals-water quality-wq-manuals-28majorionsinwaterhydrologyproject0
 
Download-manuals-water quality-wq-training-40measurementofchlorophyll-a
 Download-manuals-water quality-wq-training-40measurementofchlorophyll-a Download-manuals-water quality-wq-training-40measurementofchlorophyll-a
Download-manuals-water quality-wq-training-40measurementofchlorophyll-ahydrologyproject0
 
Download-manuals-water quality-technicalpapers-26basicecologyconcepts
 Download-manuals-water quality-technicalpapers-26basicecologyconcepts Download-manuals-water quality-technicalpapers-26basicecologyconcepts
Download-manuals-water quality-technicalpapers-26basicecologyconceptshydrologyproject0
 
Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...
 Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro... Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...
Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...hydrologyproject0
 
Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts
 Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts
Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconceptshydrologyproject0
 
Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...
 Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua... Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...
Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...hydrologyproject0
 

Andere mochten auch (14)

Download-manuals-water quality-wq-training-43howtomeasuretotaliron
 Download-manuals-water quality-wq-training-43howtomeasuretotaliron Download-manuals-water quality-wq-training-43howtomeasuretotaliron
Download-manuals-water quality-wq-training-43howtomeasuretotaliron
 
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 
Download-manuals-water quality-wq-training-36measurementoffluoride
 Download-manuals-water quality-wq-training-36measurementoffluoride Download-manuals-water quality-wq-training-36measurementoffluoride
Download-manuals-water quality-wq-training-36measurementoffluoride
 
Download-manuals-water quality-wq-training-42measurementofboron
 Download-manuals-water quality-wq-training-42measurementofboron Download-manuals-water quality-wq-training-42measurementofboron
Download-manuals-water quality-wq-training-42measurementofboron
 
Download-manuals-water quality-wq-training-30advancedaquaticchemistry
 Download-manuals-water quality-wq-training-30advancedaquaticchemistry Download-manuals-water quality-wq-training-30advancedaquaticchemistry
Download-manuals-water quality-wq-training-30advancedaquaticchemistry
 
Download-manuals-water quality-wq-training-23howtomeasurecoliforms
 Download-manuals-water quality-wq-training-23howtomeasurecoliforms Download-manuals-water quality-wq-training-23howtomeasurecoliforms
Download-manuals-water quality-wq-training-23howtomeasurecoliforms
 
Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...
 Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil... Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...
Download-manuals-water quality-wq-training-29advancedaquaticchemistrysolubil...
 
Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters
 Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters
Download-manuals-water quality-wq-training-25oxygenbalanceinsurfacewaters
 
Download-manuals-water quality-wq-manuals-28majorionsinwater
 Download-manuals-water quality-wq-manuals-28majorionsinwater Download-manuals-water quality-wq-manuals-28majorionsinwater
Download-manuals-water quality-wq-manuals-28majorionsinwater
 
Download-manuals-water quality-wq-training-40measurementofchlorophyll-a
 Download-manuals-water quality-wq-training-40measurementofchlorophyll-a Download-manuals-water quality-wq-training-40measurementofchlorophyll-a
Download-manuals-water quality-wq-training-40measurementofchlorophyll-a
 
Download-manuals-water quality-technicalpapers-26basicecologyconcepts
 Download-manuals-water quality-technicalpapers-26basicecologyconcepts Download-manuals-water quality-technicalpapers-26basicecologyconcepts
Download-manuals-water quality-technicalpapers-26basicecologyconcepts
 
Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...
 Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro... Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...
Download-manuals-water quality-wq-manuals-31tracecompoundsintheaquaticenviro...
 
Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts
 Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts
Download-manuals-water quality-wq-manuals-27surfacewaterqualityplanningconcepts
 
Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...
 Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua... Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...
Download-manuals-water quality-wq-manuals-32potentiometricanalysisofwaterqua...
 

Ähnlich wie Download-manuals-water quality-wq-manuals-33useofionselectiveprobes

Download-manuals-water quality-wq-training-19howtomeasurecod
 Download-manuals-water quality-wq-training-19howtomeasurecod Download-manuals-water quality-wq-training-19howtomeasurecod
Download-manuals-water quality-wq-training-19howtomeasurecodhydrologyproject0
 
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogenhydrologyproject0
 
Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement
 Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement
Download-manuals-water quality-wq-manuals-11chemistryofdomeasurementhydrologyproject0
 
Download-manuals-water quality-wq-training-18understandingcodtest
 Download-manuals-water quality-wq-training-18understandingcodtest Download-manuals-water quality-wq-training-18understandingcodtest
Download-manuals-water quality-wq-training-18understandingcodtesthydrologyproject0
 
Download-manuals-ground water-training-14howtosamplegroundwater
 Download-manuals-ground water-training-14howtosamplegroundwater Download-manuals-ground water-training-14howtosamplegroundwater
Download-manuals-ground water-training-14howtosamplegroundwaterhydrologyproject0
 
Download-manuals-ground water-training-14howtosamplegroundwater
 Download-manuals-ground water-training-14howtosamplegroundwater Download-manuals-ground water-training-14howtosamplegroundwater
Download-manuals-ground water-training-14howtosamplegroundwaterhydrologyproject001
 
Corrosion Laboratory-slides show
Corrosion Laboratory-slides showCorrosion Laboratory-slides show
Corrosion Laboratory-slides showIgor Kosacki
 
Download-manuals-water quality-wq-training-15understandingbodtest
 Download-manuals-water quality-wq-training-15understandingbodtest Download-manuals-water quality-wq-training-15understandingbodtest
Download-manuals-water quality-wq-training-15understandingbodtesthydrologyproject0
 
Download-manuals-water quality-wq-training-08understandingec
 Download-manuals-water quality-wq-training-08understandingec Download-manuals-water quality-wq-training-08understandingec
Download-manuals-water quality-wq-training-08understandingechydrologyproject0
 
Download-manuals-water quality-wq-manuals-03goodlaboratorypractice
 Download-manuals-water quality-wq-manuals-03goodlaboratorypractice Download-manuals-water quality-wq-manuals-03goodlaboratorypractice
Download-manuals-water quality-wq-manuals-03goodlaboratorypracticehydrologyproject0
 
Download-manuals-water quality-wq-training-09howtomeasureec
 Download-manuals-water quality-wq-training-09howtomeasureec Download-manuals-water quality-wq-training-09howtomeasureec
Download-manuals-water quality-wq-training-09howtomeasureechydrologyproject0
 
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 Download-manuals-water quality-wq-training-13howtosamplesurfacewaters Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
Download-manuals-water quality-wq-training-13howtosamplesurfacewatershydrologyproject0
 
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 Download-manuals-water quality-wq-training-13howtosamplesurfacewaters Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
Download-manuals-water quality-wq-training-13howtosamplesurfacewatershydrologyproject001
 
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...chrisrobschu
 
Download-manuals-water quality-wq-training-17howtomeasurebod
 Download-manuals-water quality-wq-training-17howtomeasurebod Download-manuals-water quality-wq-training-17howtomeasurebod
Download-manuals-water quality-wq-training-17howtomeasurebodhydrologyproject0
 
Download-manuals-water quality-wq-training-02basicchemistryconcepts
 Download-manuals-water quality-wq-training-02basicchemistryconcepts Download-manuals-water quality-wq-training-02basicchemistryconcepts
Download-manuals-water quality-wq-training-02basicchemistryconceptshydrologyproject0
 
Download-manuals-water quality-wq-training-07howtomeasuretheph
 Download-manuals-water quality-wq-training-07howtomeasuretheph Download-manuals-water quality-wq-training-07howtomeasuretheph
Download-manuals-water quality-wq-training-07howtomeasurethephhydrologyproject0
 
24 basic aquatic chemistry concepts
24 basic aquatic chemistry concepts24 basic aquatic chemistry concepts
24 basic aquatic chemistry conceptshydrologyproject0
 

Ähnlich wie Download-manuals-water quality-wq-manuals-33useofionselectiveprobes (20)

Download-manuals-water quality-wq-training-19howtomeasurecod
 Download-manuals-water quality-wq-training-19howtomeasurecod Download-manuals-water quality-wq-training-19howtomeasurecod
Download-manuals-water quality-wq-training-19howtomeasurecod
 
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
Download-manuals-water quality-wq-training-37measurementofoxidisednitrogen
 
Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement
 Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement
Download-manuals-water quality-wq-manuals-11chemistryofdomeasurement
 
Download-manuals-water quality-wq-training-18understandingcodtest
 Download-manuals-water quality-wq-training-18understandingcodtest Download-manuals-water quality-wq-training-18understandingcodtest
Download-manuals-water quality-wq-training-18understandingcodtest
 
Download-manuals-ground water-training-14howtosamplegroundwater
 Download-manuals-ground water-training-14howtosamplegroundwater Download-manuals-ground water-training-14howtosamplegroundwater
Download-manuals-ground water-training-14howtosamplegroundwater
 
Download-manuals-ground water-training-14howtosamplegroundwater
 Download-manuals-ground water-training-14howtosamplegroundwater Download-manuals-ground water-training-14howtosamplegroundwater
Download-manuals-ground water-training-14howtosamplegroundwater
 
Corrosion Laboratory-slides show
Corrosion Laboratory-slides showCorrosion Laboratory-slides show
Corrosion Laboratory-slides show
 
Download-manuals-water quality-wq-training-15understandingbodtest
 Download-manuals-water quality-wq-training-15understandingbodtest Download-manuals-water quality-wq-training-15understandingbodtest
Download-manuals-water quality-wq-training-15understandingbodtest
 
Download-manuals-water quality-wq-training-08understandingec
 Download-manuals-water quality-wq-training-08understandingec Download-manuals-water quality-wq-training-08understandingec
Download-manuals-water quality-wq-training-08understandingec
 
Download-manuals-water quality-wq-manuals-03goodlaboratorypractice
 Download-manuals-water quality-wq-manuals-03goodlaboratorypractice Download-manuals-water quality-wq-manuals-03goodlaboratorypractice
Download-manuals-water quality-wq-manuals-03goodlaboratorypractice
 
Download-manuals-water quality-wq-training-09howtomeasureec
 Download-manuals-water quality-wq-training-09howtomeasureec Download-manuals-water quality-wq-training-09howtomeasureec
Download-manuals-water quality-wq-training-09howtomeasureec
 
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 Download-manuals-water quality-wq-training-13howtosamplesurfacewaters Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 Download-manuals-water quality-wq-training-13howtosamplesurfacewaters Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
Download-manuals-water quality-wq-training-13howtosamplesurfacewaters
 
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...
Investigation of Metal and Chemical Hydrides for Hydrogen Storage in Novel Fu...
 
Download-manuals-water quality-wq-training-17howtomeasurebod
 Download-manuals-water quality-wq-training-17howtomeasurebod Download-manuals-water quality-wq-training-17howtomeasurebod
Download-manuals-water quality-wq-training-17howtomeasurebod
 
Download-manuals-water quality-wq-training-02basicchemistryconcepts
 Download-manuals-water quality-wq-training-02basicchemistryconcepts Download-manuals-water quality-wq-training-02basicchemistryconcepts
Download-manuals-water quality-wq-training-02basicchemistryconcepts
 
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
 
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
Chromatography: Pharmaceutical Analysis of API and Counter Ions in Complex Fo...
 
Download-manuals-water quality-wq-training-07howtomeasuretheph
 Download-manuals-water quality-wq-training-07howtomeasuretheph Download-manuals-water quality-wq-training-07howtomeasuretheph
Download-manuals-water quality-wq-training-07howtomeasuretheph
 
24 basic aquatic chemistry concepts
24 basic aquatic chemistry concepts24 basic aquatic chemistry concepts
24 basic aquatic chemistry concepts
 

Mehr von hydrologyproject0

Mehr von hydrologyproject0 (20)

Specification purewater
Specification purewaterSpecification purewater
Specification purewater
 
Water purifier ion exchange resin column
Water purifier ion exchange resin columnWater purifier ion exchange resin column
Water purifier ion exchange resin column
 
Water purifier, double distillation unit
Water purifier, double distillation unitWater purifier, double distillation unit
Water purifier, double distillation unit
 
Water purifier, distillation unit (automatic)
Water purifier, distillation unit (automatic)Water purifier, distillation unit (automatic)
Water purifier, distillation unit (automatic)
 
Water geyser (heater)
Water geyser (heater)Water geyser (heater)
Water geyser (heater)
 
Water bath, general purpose
Water bath, general purposeWater bath, general purpose
Water bath, general purpose
 
Water bath, bacteriological
Water bath, bacteriologicalWater bath, bacteriological
Water bath, bacteriological
 
Wash bottle
Wash bottleWash bottle
Wash bottle
 
Volumetric flask
Volumetric flaskVolumetric flask
Volumetric flask
 
Visual accumulation tube
Visual accumulation tubeVisual accumulation tube
Visual accumulation tube
 
Vacuum pump
Vacuum pumpVacuum pump
Vacuum pump
 
Turbidity meter (nephelometer)
Turbidity meter (nephelometer)Turbidity meter (nephelometer)
Turbidity meter (nephelometer)
 
Tool kit
Tool kitTool kit
Tool kit
 
Toc analyser, microprocessor controlled
Toc analyser, microprocessor controlledToc analyser, microprocessor controlled
Toc analyser, microprocessor controlled
 
Tissue grinder
Tissue grinderTissue grinder
Tissue grinder
 
Thermometers, set of
Thermometers, set ofThermometers, set of
Thermometers, set of
 
Test tubes
Test tubesTest tubes
Test tubes
 
Test sieves with shaker
Test sieves with shakerTest sieves with shaker
Test sieves with shaker
 
Stirrer, magnetic (with hot plate)
Stirrer, magnetic (with hot plate)Stirrer, magnetic (with hot plate)
Stirrer, magnetic (with hot plate)
 
Sterliser, autoclave
Sterliser, autoclaveSterliser, autoclave
Sterliser, autoclave
 

Kürzlich hochgeladen

From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .Alan Dix
 
What is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfWhat is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfMounikaPolabathina
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Farhan Tariq
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsSergiu Bodiu
 
UiPath Community: Communication Mining from Zero to Hero
UiPath Community: Communication Mining from Zero to HeroUiPath Community: Communication Mining from Zero to Hero
UiPath Community: Communication Mining from Zero to HeroUiPathCommunity
 
Manual 508 Accessibility Compliance Audit
Manual 508 Accessibility Compliance AuditManual 508 Accessibility Compliance Audit
Manual 508 Accessibility Compliance AuditSkynet Technologies
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxLoriGlavin3
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024BookNet Canada
 
Data governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationData governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationKnoldus Inc.
 
Long journey of Ruby standard library at RubyConf AU 2024
Long journey of Ruby standard library at RubyConf AU 2024Long journey of Ruby standard library at RubyConf AU 2024
Long journey of Ruby standard library at RubyConf AU 2024Hiroshi SHIBATA
 
Take control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteTake control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteDianaGray10
 
2024 April Patch Tuesday
2024 April Patch Tuesday2024 April Patch Tuesday
2024 April Patch TuesdayIvanti
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsPixlogix Infotech
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfpanagenda
 
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...Wes McKinney
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesKari Kakkonen
 
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesHow to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesThousandEyes
 
Potential of AI (Generative AI) in Business: Learnings and Insights
Potential of AI (Generative AI) in Business: Learnings and InsightsPotential of AI (Generative AI) in Business: Learnings and Insights
Potential of AI (Generative AI) in Business: Learnings and InsightsRavi Sanghani
 
Time Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsTime Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsNathaniel Shimoni
 

Kürzlich hochgeladen (20)

From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .From Family Reminiscence to Scholarly Archive .
From Family Reminiscence to Scholarly Archive .
 
What is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdfWhat is DBT - The Ultimate Data Build Tool.pdf
What is DBT - The Ultimate Data Build Tool.pdf
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platforms
 
UiPath Community: Communication Mining from Zero to Hero
UiPath Community: Communication Mining from Zero to HeroUiPath Community: Communication Mining from Zero to Hero
UiPath Community: Communication Mining from Zero to Hero
 
Manual 508 Accessibility Compliance Audit
Manual 508 Accessibility Compliance AuditManual 508 Accessibility Compliance Audit
Manual 508 Accessibility Compliance Audit
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
 
Data governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationData governance with Unity Catalog Presentation
Data governance with Unity Catalog Presentation
 
Long journey of Ruby standard library at RubyConf AU 2024
Long journey of Ruby standard library at RubyConf AU 2024Long journey of Ruby standard library at RubyConf AU 2024
Long journey of Ruby standard library at RubyConf AU 2024
 
Take control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test SuiteTake control of your SAP testing with UiPath Test Suite
Take control of your SAP testing with UiPath Test Suite
 
2024 April Patch Tuesday
2024 April Patch Tuesday2024 April Patch Tuesday
2024 April Patch Tuesday
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and Cons
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
 
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
The Future Roadmap for the Composable Data Stack - Wes McKinney - Data Counci...
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examples
 
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesHow to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
 
Potential of AI (Generative AI) in Business: Learnings and Insights
Potential of AI (Generative AI) in Business: Learnings and InsightsPotential of AI (Generative AI) in Business: Learnings and Insights
Potential of AI (Generative AI) in Business: Learnings and Insights
 
Time Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directionsTime Series Foundation Models - current state and future directions
Time Series Foundation Models - current state and future directions
 

Download-manuals-water quality-wq-manuals-33useofionselectiveprobes

  • 1. World Bank & Government of The Netherlands funded Training module # WQ - 33 Use of Ion Selective Probes New Delhi, February 2000 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi – 11 00 16 India Tel: 68 61 681 / 84 Fax: (+ 91 11) 68 61 685 E-Mail: dhvdelft@del2.vsnl.net.in DHV Consultants BV & DELFT HYDRAULICS with HALCROW, TAHAL, CES, ORG & JPS
  • 2. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 1 Table of contents Page 1 Module context 2 2 Module profile 3 3 Session plan 4 4 Overhead/flipchart master 5 5 Evaluation sheets 21 6 Handout 23 7 Additional handout 28 8 Main text 30
  • 3. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 2 1. Module context This module discusses the use of ion selective electrodes recommended for water quality analysis under the Hydrology Project. Modules in which prior training is required and other related modules are listed below. While designing a training course, the relationship between this module and the others, would be maintained by keeping them close together in the syllabus and place them in a logical sequence. The actual selection of the topics and the depth of training would, of course, depend on the training needs of the participants, i.e. their knowledge level and skills performance upon the start of the course. No. Module title Code Objectives 1. Basic chemistry concepts WQ - 02 • Convert units from one to another • Discuss the basic concepts of quantitative chemistry • Report analytical results with the correct number of significant digits. 2. Understanding the hydrogen ion concentration (pH) WQ - 06 • Discuss about the concept of pH • Calculate pH 3. Understanding chemical oxygen demand test WQ - 18 • Appreciate significance of COD measurement • Understand the chemistry of COD measurement 4. Basic aquatic chemistry concepts WQ - 24 • Understand equilibrium chemistry and ionisation constants. • Understand basis of pH and buffers • Calculate different types of alkalinity. 5. Advanced aquatic chemistry: solubility equilibria WQ - 29 • Explain the principles of chemical equilibrium • Define solubility product and explain how this relates to water quality assessment • Define the octanol-water partition coefficient and explain how this relates to water quality assessment. 6. Advanced aquatic chemistry: redox equilibria WQ - 30 • Understand the principles of redox reactions • Understand the significance of redox potential measurement 7. Potentiometric Analysis WQ - 32 • Understand the principles of potentiometric analysis • Understand construction of reference and indicating electrodes
  • 4. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 3 2. Module profile Title : Use of Ion Selective Probes Target group : HIS function(s): Q2, Q3, Q5, Q6 Duration : One session of 45 min Objectives : After the training the participants will: • understand precautions required in use of ion selective electrodes Key concepts : • ion selective electrodes recommended under HP • calibration • precaution Training methods : Lecture, exercises Training tools required : Board, flipchart Handouts : As provided in this module Further reading and references : • Chemistry for environmental engineers - C. N. Sawyer, P. L. McCarty & G. F. Parkin, McGraw - Hill, Inc., 1994 • Standard methods for the examination of water and wastewaters, AWWA, 19th edition, 1995
  • 5. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 4 3. Session plan No Activities Time Tools 1 Preparations 2 Introduction: • Discuss various types of indicating electrodes and probes 5 min OHS 3 Performance characteristics • Describe precautions, limitations and advantanges of potentiometric measurements using indicating electrodes and probes • Discuss calibration requirements and methods 15 min OHS 4 Electrodes and probes recommended under HP • Discuss applications of the recommended probes and procedures for their use. 15 min OHS 5 Wrap up • Discuss experience of participants, if any • Review precautions and limitations 10 min
  • 6. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 5 4. Overhead/flipchart master OHS format guidelines Type of text Style Setting Headings: OHS-Title Arial 30-36, with bottom border line (not: underline) Text: OHS-lev1 OHS-lev2 Arial 24-26, maximum two levels Case: Sentence case. Avoid full text in UPPERCASE. Italics: Use occasionally and in a consistent way Listings: OHS-lev1 OHS-lev1-Numbered Big bullets. Numbers for definite series of steps. Avoid roman numbers and letters. Colours: None, as these get lost in photocopying and some colours do not reproduce at all. Formulas/Equat ions OHS-Equation Use of a table will ease horizontal alignment over more lines (columns) Use equation editor for advanced formatting only
  • 7. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 6 Ion selective probes (1) • Electrode potential is linearly related to log of activity of ion • Glass electrode for H+ measurement is most commonly used. • Probes for DO, F- , NH3 and CN- recommended under HP
  • 8. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 7 Types and examples • Glass membrane Na+ , K+ , H+ , NH4 + • Liquid membrane NO3 - , Ca2+ , Mg2+ • Solid membrane S2- , F- , CN- , Pb2+ • Diaphragm NH3, O2, HCN • Metal Ag+ , Cu2+ , Cd2+ , I-
  • 9. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 8 Performance characteristics (1) • Temperature effect - calibration standard and sample at same temperature - meter in-built correction • Response time - steady value, few seconds to minutes • Potential drift - recalibrate frequently
  • 10. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 9 Performance characteristics (2) • Sensitivity - working range, depends on properties of membrance - response 59.15 mV per ten fold conc. change for one electron. • Selectivity and interference - other ions may effect reading - remove interfering ion - calibrate in presence of the ion.
  • 11. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 10 Performance characteristics (3) • Activity vs. concentration - measures activity - calibration standard of similar ionic strength. • pH - may affect ionisation - formation of complexes
  • 12. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 11 Performance characteristics (4) • Presence of complexing agent - electrodes measure free ions - remove complexing agents • Life time - rate of leaking of liquid ion exchanger - chemical fouling of solid membranes
  • 13. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 12 Advantages • turbid samples • small volume, sample not destroyed • do not change existing equilibria • direct reading of conc. over wide range • sample does not require pretreatment • rapid response • ideal for on-line measurement
  • 14. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 13 Electrode calibration • Calibration curve - treatment of sample and standard for adjustment of pH and ionic strength • Standard addition - when effect of ions in sample matrix can not be duplicated in standard - significant liquid junction potential
  • 15. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 14 Figure 1 Response of Ca electrode as functions of Ca++ ion concentration and activity
  • 16. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 15 Standard addition • Eobs = Eref + Ej - Eind which can be written as: • K is the unknown change in potential • Observe potential before and after adding known standard to sample and solve for 'a', the analyte concentration − = − log . / a E K n obs 0 0591
  • 17. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 16 Dissolved oxygen probe • Site and on-line measurements • Frequent calibration - air - water of known DO • Fouling of membrane - change when response is sluggish - do not trap air bubble • Sample flow across membrane while measuring
  • 18. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 17 Fluoride electrode • Evaluate suitability critically • Al3+ and Fe2+ cause interference • At pH < 5, HF.HF is formed • Add complexing agent and pH buffer
  • 19. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 18 Ammonia probe • Wide range, 0.03 - 1400 mg/L • Interference due to amines, Hg & Ag • Increase pH to convert NH4 + to NH3 • Stir sample slowly.
  • 20. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 19 Nitrate electrode • Alternative to UV method and convenient compared to Cd reduction method • Adjust pH to 3 eliminate HCO3 - • Requires complexing agents for other interfering ions • Check calibration frequently
  • 21. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 20 Cyanide electrode • Distillation of sample is required • Range 0.025 - 10 mg CN- /L • Response time at least 5 min, mix well • Do not work with acidified sample in open, HCN is extemely toxic gas
  • 22. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 21 5. Evaluation sheets
  • 23. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 22
  • 24. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 23 6. Handout
  • 25. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 24 Ion selective probes (1) • Electrode potential is linearly related to log of activity of ion • Glass electrode for H+ measurement is most commonly used. • Probes for DO, F- , NH3 and CN- recommended under HP Types and examples • Glass membrane Na+ , K+ , H+ , NH4 + • Liquid membrane NO3 - , Ca2+ , Mg2+ • Solid membrane S2- , F- , CN- , Pb2+ • Diaphragm NH3, O2, HCN • Metal Ag+ , Cu2+ , Cd2+ , I- Performance characteristics (1) • Temperature effect - calibration standard and sample at same temperature - meter in-built correction • Response time - steady value, few seconds to minutes • Potential drift - recalibrate frequently Performance characteristics (2) • Sensitivity - working range, depends on properties of membrance - response 59.15 mV per ten fold conc. change for one electron • Selectivity and interference - other ions may effect reading - remove interfering ion - calibrate in presence of the ion. Performance characteristics (3) • Activity vs. concentration - measures activity - calibration standard of similar ionic strength. • pH - may affect ionisation - formation of complexes Performance characteristics (4) • Presence of complexing agent - electrodes measure free ions - remove complexing agents • Life time - rate of leaking of liquid ion exchanger - chemical fouling of solid membranes
  • 26. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 25 Advantages • turbid samples • small volume, sample not destroyed • do not change existing equilibria • direct reading of conc. over wide range • sample does not require pretreatment • rapid response • ideal for on-line measurement Electrode calibration • Calibration curve - treatment of sample and standard for adjustment of pH and ionic strength • Standard addition - when effect of ions in sample matrix can not be duplicated in standard - significant liquid junction potential Standard addition • Eobs = Eref + Ej - Eind which can be written as: • K is the unknown change in potential • Observe potential before and after adding known standard to sample and solve for 'a', the analyte concentration Dissolved oxygen probe • Site and on-line measurements • Frequent calibration - air - water of known DO • Fouling of membrane - change when response is sluggish - do not trap air bubble • Sample flow across membrane while measuring Fluoride electrode • Evaluate suitability critically • Al3+ and Fe2+ cause interference • At pH < 5, HF.HF is formed • Add complexing agent and pH buffer Ammonia probe • Wide range, 0.03 - 1400 mg/L • Interference due to amines, Hg & Ag • Increase pH to convert NH4 + to NH3 • Stir sample slowly. − = − log . / a E K n obs 0 0591
  • 27. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 26 Nitrate electrode • Alternative to UV method and convenient compared to Cd reduction method • Adjust pH to 3 eliminate HCO3 - • Requires complexing agents for other interfering ions • Check calibration frequently Cyanide electrode • Distillation of sample is required • Range 0.025 - 10 mg CN- /L • Response time at least 5 min, mix well • Do not work with acidified sample in open, HCN is extemely toxic gas
  • 28. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 27 Add copy of Main text in chapter 8, for all participants.
  • 29. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 28 7. Additional handout These handouts are distributed during delivery and contain test questions, answers to questions, special worksheets, optional information, and other matters you would not like to be seen in the regular handouts. It is a good practice to pre-punch these additional handouts, so the participants can easily insert them in the main handout folder.
  • 30. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 29
  • 31. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 30 8. Main text Contents 1. Introduction 1 2. Performance characteristics of electrodes 2 3. Electrode calibration 3 4. Recommended probes under HP 5
  • 32. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 1 Use of Ion Selective Probes 1. Introduction Ion-selective electrodes belong to a class of electro-chemical sensors which respond to only the ion of interest. Their use involves a special case of direct potentiometry. When these electrodes are placed in a conducting solution, they generate a millivolt potential related to the concentration of the ion. The potential is linearly related to log of concentration of the ion. To be more accurate, the term ion ‘activity’ should be used instead of concentration. For simplicity, this text will use the term concentration, which in most cases is a close approximation of activity. Historically, the oldest and the best known and most widely used ion-selective electrode is the glass electrode, which responds to hydrogen ion activity. It remained the only sensing electrode until mid 1960’s. But the renewed interest in potentiometry has now led to the commercial availability of a range of ion-selective electrodes. Basic principles of redox reactions and construction of various types of indicating electrodes and probes are discussed in module # 30, 'Advanced Aquatic Chemistry Concepts: Redox Equilibria, and module # 32, 'Potentiometric analysis'. This module describes general aspects of ion selective probes and use of DO, F- , NH3 - and CN- probes which are recommended under Hydrology Project. Table 1 gives an overview of indicator electrodes commonly available Table 1: Types and examples of ion electrodes Types of electrode Description Examples of chemical specie measured Glass membrane electrode Electrode having glass sensing membrane Na+ , K+ , H+ , NH4 + Liquid membrane electrode Electrode having sensing membrane comprising an ion exchange liquid held in a rigid porous membrane NO3 - , Ca2+ , Mg+2 , NH4 + Solid membrane electrode Selective electrode having sensing membrane made of poorly soluble metal salt monocrystal or metal salt powder S2- , F- , CN- , Pb2+ , Ag+ Diaphragm electrode: Electrodes combining gas transmitting membrane and pH electrode or ion electrode NH3, HCN, O2 Metal electrode Electrode comprising the metal or the metal and a poorly soluble salt. Ag+ , Cu2+ , Cd2+ , Pb2+ , I-
  • 33. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 2 2. Performance characteristics of electrodes Temperature effect: It is necessary to correct for the temperature of the samples because the electrode potential is temperature dependent. The sample temperature should be measured independently and a temperature correction should be made with the in-built correction in the meter. Response time: The response time of ion-selective electrode is the speed at which it responds (electrode potential reaches a steady state value), and depends upon the type of electrode and concentration of ion under measurement. The response time of solid – state electrode is generally of the order of 5-10 seconds. The response time of some liquid ion-exchanger electrode is longer, usually many seconds or even few minutes at low concentrations. When making a reading, enough time must be given for the electrode to reach a stable reading. Potential drifts: The potential drifts of ion-selective electrodes is connected with changes in the surface structure of solid – membrane electrode due to contact with the electrolyte and due to dissolution of the ion exchanger of liquid-membrane electrodes. All electrodes show drifts with time and therefore need re-calibration periodically with standardising solutions. This should be done every few hours if making many analyses. Sensitivity: The useful working range of the ion-selective electrodes is knows as sensitivity which is effectively governed by the properties of the membrane. The working range of electrode is determined from the concentration (mg/l) to which an electrode responds in a theoretical fashion (59.15 mV/ten fold change in concentration when one electron is involved in the reaction). Selectivity & Interferences: Although intended to be specific to a single ionic species, most electrodes respond in varying degrees to the presence of other ions. Where severe interferences are likely to be encountered, there are two practical courses of action. Either the electrode must be calibrated in the presence of interfering ion or the interference must be eliminated by some other means such as buffering, precipitation or by adding a complexing agent. Activities vs. Concentration: Electrodes measure activities and not concentrations. The activity varies with ionic strength. Thus, to use an electrode for concentration measurements, the electrode must be calibrated in a standard solution of similar ionic strength (see section 3. Electrode calibration). Effect of pH: The pH of the sample solution changes the chemistry of the analyte ion. For instance, at low pH values F- ions are present as HF, which are not sensed by the fluoride electrode. Thus in such cases an appropriate pH adjuster should be added to samples to set up a reliable ion- selective electrode analysis.
  • 34. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 3 Presence of complexing agent: A number of ions in water samples can be present in more than one form, such as complexes with other ions or molecules. Electrodes measure only the free ion in solution, and cannot sense bound or complexed ions. Life - time: The life-time of an electrode is determined by the rate at which the liquid ion exchanger leaks through the membrane in liquid membrane electrodes, or the time required for crystal surface or the pressed pellet to become chemically fouled in solid membrane electrodes. Life-time markedly depends upon conditions of use. With good care, electrodes should provide many months or years of service. If electrode response drops off, rejuvenation of electrodes is possible by refilling the liquid internals, replacing sensing membrane of liquid membrane type electrode or mechanical polishing of solid – state membrane tip. The use of ion-selective electrodes as an analytical techniques in water analysis is increasing. This is because of the significant advantages such as: • They are not bothered by turbidity or colour of the sample • Small volumes of samples can be analysed without loss of sample • Measure species without upsetting chemical equilibrium • Measurements are independent of volume of sample • Concentrations can be read directly on specific ion meter • Calibration curves can cover even six decades of concentration • Electrodes are sensitive up to parts per billion Further application of these electrodes especially for on-line analysis are very much in use because: • They do not usually require extensive pre-treatment of the sample • They have a relatively rapid response-time • They provide an electrical analogue signal which does not need sophisticated signal conditioning circuits before a result can be displayed to the instrument user • Considerable previous experience of on-line instrumentation for pH measurement is available, which is relevant to other electrodes also. 3. Electrode calibration Activity vs. concentration: Electrode response is related to analyte activity rather than to analyte concentration. More often than not, activity coefficients are not available. The difference between activity and concentration is illustrated by Figure 1 in which the response of a calcium electrode is plotted against logarithmic functions of the calcium concentration and calcium activity. The electrode responds in an ideal fashion, i.e., constant slope, only when concentration values are changed to activity values. An obvious way to convert potentiometric measurements from activity to concentration is to make use of an emperical calibration curve, such as the lower plot in Figure 1. For this approach to be successful, it is necessary to make the ionic composition of the standards essentially the same as that for the analyte solution. This is often done by swamping both sample and standards with a measured excess of an inert electrolyte, thus making the added effect of electrolyte from the sample matrix negligible.
  • 35. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 4 Figure 1 Response of Ca electrode as functions of Ca++ ion concentration and activity Standard addition method: Notwithstanding the advantages of direct potentiometric measurements, there are certain inherent limitations to the method. Principal among these is the existance of a 'liquid junction potential' that affects most potentiometric measurements. The junction potential, Ej, is created at the interface of the electrode and the analyte solution due to unequal mobility of various cations and anions, which modifies the observed potential, Eobs, according to the following expression (1) where a is the analyte activity, Eind is the actual indicator electrode potential and K is a constant. In case the effect of liquid junction potential is to be eliminated and where the effect of the matrix of the sample cannot be swamped by addition of an inert electrolyte, standard addition method may be used. It involves determining the potential of the electrode system before and after a measured volume of a standard has been added to a known volume of the analyte solution. As in the other method, if required, an excess of an electrolyte is added to the analyte solution to prevent a major shift in the ionic strength due to addition of the standard. The following example explains the calculations involved in the method. E E E E E E E n a obs ref j ind ref j= + − = + − − F HG I KJc h 0 0 0591 1. log − = − log . / a E K n obs 0 0591
  • 36. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 5 Example 1: A lead ion electrode developed a potential of 0.4706 V (vs. Standard Calomel Electrode) when immersed in 50.00 mL of a sample. A 5.00 mL addition of standard 0.02000 M lead solution caused the potential to shift to 0.4490 V. Calculate the molar concentration of lead in the sample. Applying Equation 1 and assuming that the activity of Pb2+ is approximately equal to its molar concentration Cx, we may write Where E1 is the measured potential (0.4706 V). After introduction of the standard addition, the potential is now E2 (0.4490 V) and which can be written as Subtraction of this equation from the first gives 4. Recommended probes under HP This section describes the use of probes recommended under HP. As stated earlier, while the probes have their advantages, their limitations must also be kept in mind when used in different situations. For each case the manufacturer's instructions for calibration and operation of the probe and the ion-meter should be followed. − = − log . C E K x 1 0 0591 2 − + + = − log . . . . . . 50 00 5 00 0 02000 50 00 5 00 0 0591 2 2X C X E Kx − + = −− log ( . . ) . 0 9091 1818 10 0 0591 2 3 2 C X E K x − + = − = − = − log . . . . . . . C C X E Ex x0 9091 1818 10 2 0 0591 2 0 4706 0 4490 0 0591 0 7310 3 1 2b g b g C C X C X M x x x 0 9091 1818 10 01858 4 06 10 3 4 . . . . + = = − −
  • 37. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 6 4.1 Dissolved oxygen probe Application: DO probes are useful for on site measurements and when a large number of samples are to be analysed in the laboratory, for example, when determining BOD. For continuous DO measurement the potentiometric measurement is the only option. Construction: The probe consists of two solid metal electrodes in contact with a supporting electrolyte separated form the test solution by a selective membrane permeable to molecular oxygen. Interference: The probe exhibits a high temperature coefficient. The ion-meter may have in-built temperature correction or correction factor may be available from the manufacturer. Prolonged use of the probe in waters containing sulphides lowers the sensitivity of the probe. Calibration: The probe may be calibrated by reading against air, a sample containing known DO concentration (determined by iodometric titration) or a sample containing zero DO (by adding sulphite to the sample). Periodic cross-check of calibration by different methods is recommended. Preferably calibrate using the sample water. In case the sample water contains interfering substances, calibrate with a sample of clean water containing approximately the same salt content. Precautions: Change the probe membrane when the response becomes sluggish or erratic. Take care to avoid contamination of the sensing element and trapping of minute air bubbles while changing the membrane. Provide sufficient sample flow across the membrane surface while making measurements. 4.2 Fluoride electrode Applications: Fluoride electrode has been used for determination of fluoride in drinking water supplies. The analyst should be aware of its limitations as described in the section on interferance below. Its suitability should be evaluated since even small errors can be critical in terms of suitability of a water for drinking purpose. Construction: The key element in the fluoride electrode is the lanthanum fluoride crystal across which a potential is established by fluoride solutions of different concentrations, in the sample and the standard solution in the electrode. The electrode can be used with a standard calomel reference electrode. Interference: Fluoride forms complexes with several cations notably aluminium and iron. At pH values lower that 5 poorly ionisable HF.HF complex is formed. Cyclohexylene-diamine-tetra acetic acid and a buffer can be added to the sample to complex the interfering ions. The buffer maintains a pH of 5. This pH also reduces the effect of the interfering ions. Samples having TDS in excess of 10,000 mg/L should be distilled.
  • 38. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 7 Calibration: A calibration curve in the 0.2 to 2 mg/L F- range may be prepared. The recommended complexing agent and buffer should be added to the sample as well as the standards. This would make the ionic strength of the sample similar to that of the standard. Precautions: Treat the sample and the standard as described above. Avoid stirring before immersing electrode so as not to entrap air bubbles. Let electrode remain in solution until the reading is constant. Recalibrate frequently against atleast one standard. 4.3 Ammonia probe Applications: The method can be used for a wide range of ammonia concentration from 0.03 to 1400 mg/L NH3-N. It is suitable for coloured and turbid water samples. Construction: The probe consists of a pH electrode and a reference electrode. It is provided with a membrane which allows permeation of NH3. The permeated gas reacts with an internal solution and changes its pH, which is sensed by the pH electrode. Interference: The probe has a high temperature coefficient. The temperature correction is generally available from the manufacturer. Amines cause a positive interference. Hg and Ag form complexes with ammonia. This can be overcome by adding NaOH/EDTA to the sample. Calibaration: Calibrate the instrument using a series of standards in the range of 0.1mg/L - 1000 mg/L NH3-N prepared from NH4Cl. Sufficient quantity of NaOH is added to convert NH4 + to NH3. In case presence of Hg or Ag is suspected in the sample ass NaOH/EDTA solution to the standard also. Same volume of the reagent should be added to the standards as well as the sample. A 10 fold increase in NH3 concentration the potential change should be about 59mV. Precautions: Calibration standards and samples should be at the same temperature. Do not add NaOH before immersing the probe to avoid loss of ammonia. Stir slowly and maintain the same speed while taking readings. For ammonia concentration below 1mg/L, 2 to 3 minutes may be required to obtain a stable response from the electrode. 4.4 Nitrate electrode Applications: Potentiometric analysis for nitrate is an alternative to UV absorption method where presence of organic matter in the sample interferes with the spectrophotometric method. It is also convenient and time saving when compared to cadmium reduction method. The use of the electrode may be limited in case of saline samples (see interferences below). The electrode responds to NO3 - -N concentration in 0.14-1400 mg/L range. Construction: The electrode has a sensitive liquid ion-exchange membrane held in an inert membrane. It develops a potentia across the membrane whose one face is kept in contact with the analyte in the sample and the other in contact with an inner standard solution.
  • 39. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 8 Interferences: Chloride and bicarbonate ions interfere when their weight raios to NO3 - -N are in excess of 10 and 5, respectively. Variation in pH gives erratic response. Other ions which may interfere are NO2 - , CN- , S2- , Br- , I- and NO2 - . Calibration: A buffer is added to the standards and the sample to adjust the solutions to similar ionic strength. The buffer maintains a pH of 3, which eleminates HCO3 - . The buffer also contains Ag2(SO4) to remove interference from Cl- , Br- , S2- , and CN- , and sulfamic acid to remove NO2 - . Precautions: The standards and the samples should be at same temperature preferably 250 C. The calibration curve, when mV potential reading is plotted as ordinate versus log NO3 - -N concentration as abscissa, should result in a straight line with a slope of +57±3mV/decade. Check the electrode calibration several times a day using at least one standard. 4.5 Cyanide electrode Applications: A preliminary step of distillation of sample and collection of CN- in alkaline solution is required. The electrode responds to CN- concentration in the range of 0.05 to 10 mg/L Construction: The electrode uses a solid crystalline membrane sensitive to CN- . Interferences: Since the determination is made on the distillate, ordinarily interference from other ions is not expected. Lead carbonate may be added to absorbing solution if sulphides are expected in the distillate. Calibration: Use a series of 4 or more standards in the range of 0.025-10mg CN- /L. Proceed from the lowest to the highest concentration while taking readings. Mix well using a magnetic stirrer. Allow at least 5 min for reaching equilibrium. Precaution: HCN is an extremely toxic gas. Follow all precautions as detailed in the distillation step in standard procedures. Do not word with an acidified sample is open. The standards and the samples should be at the same temperature, preferably as close to 250 C as possible.
  • 40. Hydrology Project Training Module File: “ 33 Use of Ion Selective Probes.doc” Version 06/11/02 Page 9 4.6 Use and care of electrodes An overview of initial use, storage and recovery for ion electrodes is given in Table 2. Table 2 : Initiation of use of ion electrodes, storage and recovery Kind of electrode Initiation of use Storage Recovery Glass membrane Immerse in pure water or diluted standard liquid for several hours. Sufficiently wash in pure water, immerse in the diluted standard liquid. For long term storage wipe, dry, store in the air safely. Wipe the glass membrane with soft cloth or sponge impregnated with neutral detergent to clean. If the electrode is used in alkali, immerse it in 1N HNO3 for 1 – 2 hours Crystalline solid membrane Holding the electrode head, shake gently 1- 2 times, immerse it in the diluted standard solution for about 10 min. Wash in pure water, immerse in diluted standard liquid. For storage, wipe dry, store in air safely. Wipe surface with cloth or absorbent wadding impregnated with ethanol to clean. If the contamination persists, gently rub the surface with soft wash-leather Non- Crystalline solid membrane Wipe it with tissue, immersed in the standard liquid for 10 min. If contaminated, perform the recovery treatment, immerse dilute standard liquid for 10 min. Wash with pure water, immerse in diluted standard liquid. Wipe thoroughly, dry and store safely in air. Drip pure water on the fine polishing paper, polish uniformly and gently. Wash thoroughly with pure water, wipe with tissues. Liquid membrane Holding the electrode head, shake gently 1- 2 time, immerse in diluted standard solution for 10 min. Wash with pure water, immerse in diluted standard liquid. Wipe, dry and store safely in the air. Replenish/replace the liquid membrane according to instruction manual. Immerse in high concentration standard liquid. Diaphragm Set the internal liquid and diaphragm, immerse in pure water of diluted standard liquid for 10 min. Wash with pure water, immerse in diluted standard liquid. Remove the diaphragm, discharge the internal liquid wash with pure water, dry and store safely the main body and diaphragm Replace the contaminated / damaged diaphragm remove contamination with immersing in 1N HNO3 . Replace the internal liquid as per instruction manual.