Comparison of the Diaton
Transpalpebral Tonometer
    Versus TonoPen-
       Applination
 Theodore H. Curtis, M.D.1, Douglas L Mackenzie,
 M.D.1, Robert J. Noecker M.D.2, and Malik Y.
 Kahook M.D.1


 1The Rocky Mountain Lions Eye Institute, University of Colorado
 Health Sciences Center, Aurora, CO
 2Eye  and Ear Institute, University of Pittsburgh Medical Center,
 Pittsburgh, PA
Financial Disclosures


•   None of the authors have financial interests
    relevant to the subject discussed.
Purpose

•   To compare intraocular pressure (IOP)
    measurements obtained with Diaton trans-
    palpebral tonometry versus Tonopen
    applanation tonometry in children and adults.
Introduction
• Goldmann applanation is the gold standard for IOP
  measurement
• It has been supplanted by TonoPen applanation in many
  settings because of it’s ease of use, portability,
  convenience, and minimal training requirements.
• The TonoPen requires contact with the corneal surface,
  and has the risks of iatrogenic corneal injury, spread of
  pathogens, and requires topical anesthetics.
Introduction

• The newly-developed Diaton tonometer is a
  handheld device that measures pressure through
  the tarsal plate (Figures 1 & 2).
• It avoids contact with the cornea and the need for
  topical anesthesia.
Figure 1: The Diaton
Transpalpebral Tonometer
Figure 2: Using the
 Diaton Tonometer
Methods
•   We looked at 74 eyes of 38 consecutive patients who received both
    Tonopen and Diaton tonometry
•    TonoPen measurements were taken in the sitting position following
    topical anesthesia with proparicaine.
•   Diaton measurements were performed in the sitting position with the
    patient gazing at a 45° angle, placing the eyelid margin at the
    superior limbus. If necessary, gentle traction was placed on the brow
    to align the lid with the limbus. The device was activated when the
    signaling mechanism indicated the device was vertical.
Results
•   Age range 3-91 years of age (mean 47.5 years).
•   The average IOP with the Diaton was 16.24 (+/-5.11 mm Hg; range
    = 7-32 mmHg).
•   The average IOP with the TonoPen was 16.37 (+/-4.90 mm Hg;
    range = 8-33 mmHg).
•   The mean variation between the two modalities was 1.59 mmHg (+/-
    1.31 mm Hg; range = 0-6 mmHg).
•   Eighty-one percent of all measurements were within 2 mmHg of
    each other (Table 1).
•   There was no statistically significant difference in mean IOP values
    obtained with the two devices (p=0.87).
Comparison of Diaton Tonometer to TonoPen
                                          Tonometer

                 35


                 30


                 25
TonoPen (mmHg)




                 20


                 15


                 10
                                                           y = 0.9955x

                 5


                 0
                      0        5     10     15      20     25      30    35
                                           Diaton (mmHg)
Table
Device     Mean IOP (standard      IOP Range
           deviation)

Tonopen    16.37 (+/-4.90 mm Hg)   8-33 mm Hg

Diaton     16.24 (+/-5.11 mm Hg)   7-32 mm Hg

Variance   1.59 (+/-1.31 mm Hg)    0-6 mm Hg
between
devices
Conclusions
•   The Diaton tonometer pressure measurements correlated well with
    TonoPen measurements in this retrospective review.
•   We did not find problems performing the exam in children, and
    many were reassured by the fact that no drops were needed.
•   There may be a notable benefit in patients after refractive surgery or
    with corneal pathology since the Diaton does not applanate the
    cornea.
•   The Diaton tonometer appears to be a clinically useful device in the
    IOP measurement of both children and adults.
References
• Li J, Herndon LW, Asrani SG, Stinnett S, Allingham RR. Clinical comparison of the
  Proview eye pressure monitor with the goldmann applanation tonometer and the TonoPen.
  Arch Opthalmol 2004;122:1117-21.

• Eisenberg DL, Sherman BG, McKeown CA, Schuman JS. Tonometry in adults and
  children: a manometric evaluation of pneumotonometry, applanation, and TonoPen in vitro
  and in vivo. Ophthalmology 1998;105:1173-81.

• Diaton: digital portable tonometer of intraocular pressure through the eyelid. Operation
  Manual. Ryazan State Instrument Making Enterprise. Ryazan, Russia.

• Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra-
  Pimentel E. Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer:
  comparison with contact and non-contact tonometry. Optom Vis Sci 2005;82:143-50.

• Troost A, Yun SH, Specht K, Krummenauer F, Schwenn. Transpalpebral tonometry:
  reliability and comparison with Goldmann applanation tonometry and palpation in healthy
  volunteers. Br J Ophthalmol 2005;89:280-3.

• Losch A, Scheuerle A, Rupp V, Auffarth G, Becker M. Transpalpebral measurement of
  intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation
  tonometry. Graefes Arch Clin Exp Opthhalmol. 2005;243:313-6.

More Related Content

Similar to Diaton Tonometer Vs Tonopen Correlation

Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
Roman Iospa, CEO, Medical/Ophth Investor
 
Comparison of iop measurements between goldmann applanation tonometry and bal...
Comparison of iop measurements between goldmann applanation tonometry and bal...Comparison of iop measurements between goldmann applanation tonometry and bal...
Comparison of iop measurements between goldmann applanation tonometry and bal...
DevelopAll
 
IOP measurements
IOP measurementsIOP measurements
IOP measurements
Maryam Fida
 

Similar to Diaton Tonometer Vs Tonopen Correlation (20)

Glaucoma Test: Eye Pressure Diaton tonometer vs. Tono pen applanation clinica...
Glaucoma Test: Eye Pressure Diaton tonometer vs. Tono pen applanation clinica...Glaucoma Test: Eye Pressure Diaton tonometer vs. Tono pen applanation clinica...
Glaucoma Test: Eye Pressure Diaton tonometer vs. Tono pen applanation clinica...
 
Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
Diaton Tonometer Clinical Trials Guide 15 trials + Training Videos User Manua...
 
Unique diaton tonometer Glaucoma IOP Eye Test Through the Eyelid Presentation
Unique diaton tonometer Glaucoma IOP Eye Test Through the Eyelid PresentationUnique diaton tonometer Glaucoma IOP Eye Test Through the Eyelid Presentation
Unique diaton tonometer Glaucoma IOP Eye Test Through the Eyelid Presentation
 
Comparison of iop measurements between goldmann applanation tonometry and bal...
Comparison of iop measurements between goldmann applanation tonometry and bal...Comparison of iop measurements between goldmann applanation tonometry and bal...
Comparison of iop measurements between goldmann applanation tonometry and bal...
 
Tonometer Diaton - BiCOM INC - Glaucoma Eye Test Presentation
Tonometer Diaton - BiCOM INC - Glaucoma Eye Test PresentationTonometer Diaton - BiCOM INC - Glaucoma Eye Test Presentation
Tonometer Diaton - BiCOM INC - Glaucoma Eye Test Presentation
 
Comparison Of Diaton Tonometry And Non Contact Tonometry
Comparison Of Diaton Tonometry And Non Contact TonometryComparison Of Diaton Tonometry And Non Contact Tonometry
Comparison Of Diaton Tonometry And Non Contact Tonometry
 
TONOMETRY
TONOMETRYTONOMETRY
TONOMETRY
 
Tonometry
TonometryTonometry
Tonometry
 
tonometry-140624223245-phpapp02-converted.pptx
tonometry-140624223245-phpapp02-converted.pptxtonometry-140624223245-phpapp02-converted.pptx
tonometry-140624223245-phpapp02-converted.pptx
 
Eye Pressure After Lasik Surgery - Study, BiCOM Inc.
Eye Pressure After Lasik Surgery - Study, BiCOM Inc.Eye Pressure After Lasik Surgery - Study, BiCOM Inc.
Eye Pressure After Lasik Surgery - Study, BiCOM Inc.
 
Iop
IopIop
Iop
 
Tonometry
TonometryTonometry
Tonometry
 
Primary care physician and patient experience with non corneal tonometer diat...
Primary care physician and patient experience with non corneal tonometer diat...Primary care physician and patient experience with non corneal tonometer diat...
Primary care physician and patient experience with non corneal tonometer diat...
 
Different types of Tonometry
Different types of TonometryDifferent types of Tonometry
Different types of Tonometry
 
Tonopen for Glaucoma Understanding its Role in Managing the Disease.pdf
Tonopen for Glaucoma Understanding its Role in Managing the Disease.pdfTonopen for Glaucoma Understanding its Role in Managing the Disease.pdf
Tonopen for Glaucoma Understanding its Role in Managing the Disease.pdf
 
IOP measurements
IOP measurementsIOP measurements
IOP measurements
 
Tonometry
 Tonometry Tonometry
Tonometry
 
International Journal of Ophthalmology & Vision Research
International Journal of Ophthalmology & Vision ResearchInternational Journal of Ophthalmology & Vision Research
International Journal of Ophthalmology & Vision Research
 
International Journal of Ophthalmology & Vision Research
International Journal of Ophthalmology & Vision ResearchInternational Journal of Ophthalmology & Vision Research
International Journal of Ophthalmology & Vision Research
 
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTXTONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
 

Recently uploaded

Recently uploaded (20)

Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 

Diaton Tonometer Vs Tonopen Correlation

  • 1. Comparison of the Diaton Transpalpebral Tonometer Versus TonoPen- Applination Theodore H. Curtis, M.D.1, Douglas L Mackenzie, M.D.1, Robert J. Noecker M.D.2, and Malik Y. Kahook M.D.1 1The Rocky Mountain Lions Eye Institute, University of Colorado Health Sciences Center, Aurora, CO 2Eye and Ear Institute, University of Pittsburgh Medical Center, Pittsburgh, PA
  • 2. Financial Disclosures • None of the authors have financial interests relevant to the subject discussed.
  • 3. Purpose • To compare intraocular pressure (IOP) measurements obtained with Diaton trans- palpebral tonometry versus Tonopen applanation tonometry in children and adults.
  • 4. Introduction • Goldmann applanation is the gold standard for IOP measurement • It has been supplanted by TonoPen applanation in many settings because of it’s ease of use, portability, convenience, and minimal training requirements. • The TonoPen requires contact with the corneal surface, and has the risks of iatrogenic corneal injury, spread of pathogens, and requires topical anesthetics.
  • 5. Introduction • The newly-developed Diaton tonometer is a handheld device that measures pressure through the tarsal plate (Figures 1 & 2). • It avoids contact with the cornea and the need for topical anesthesia.
  • 6. Figure 1: The Diaton Transpalpebral Tonometer
  • 7. Figure 2: Using the Diaton Tonometer
  • 8. Methods • We looked at 74 eyes of 38 consecutive patients who received both Tonopen and Diaton tonometry • TonoPen measurements were taken in the sitting position following topical anesthesia with proparicaine. • Diaton measurements were performed in the sitting position with the patient gazing at a 45° angle, placing the eyelid margin at the superior limbus. If necessary, gentle traction was placed on the brow to align the lid with the limbus. The device was activated when the signaling mechanism indicated the device was vertical.
  • 9. Results • Age range 3-91 years of age (mean 47.5 years). • The average IOP with the Diaton was 16.24 (+/-5.11 mm Hg; range = 7-32 mmHg). • The average IOP with the TonoPen was 16.37 (+/-4.90 mm Hg; range = 8-33 mmHg). • The mean variation between the two modalities was 1.59 mmHg (+/- 1.31 mm Hg; range = 0-6 mmHg). • Eighty-one percent of all measurements were within 2 mmHg of each other (Table 1). • There was no statistically significant difference in mean IOP values obtained with the two devices (p=0.87).
  • 10. Comparison of Diaton Tonometer to TonoPen Tonometer 35 30 25 TonoPen (mmHg) 20 15 10 y = 0.9955x 5 0 0 5 10 15 20 25 30 35 Diaton (mmHg)
  • 11. Table Device Mean IOP (standard IOP Range deviation) Tonopen 16.37 (+/-4.90 mm Hg) 8-33 mm Hg Diaton 16.24 (+/-5.11 mm Hg) 7-32 mm Hg Variance 1.59 (+/-1.31 mm Hg) 0-6 mm Hg between devices
  • 12. Conclusions • The Diaton tonometer pressure measurements correlated well with TonoPen measurements in this retrospective review. • We did not find problems performing the exam in children, and many were reassured by the fact that no drops were needed. • There may be a notable benefit in patients after refractive surgery or with corneal pathology since the Diaton does not applanate the cornea. • The Diaton tonometer appears to be a clinically useful device in the IOP measurement of both children and adults.
  • 13. References • Li J, Herndon LW, Asrani SG, Stinnett S, Allingham RR. Clinical comparison of the Proview eye pressure monitor with the goldmann applanation tonometer and the TonoPen. Arch Opthalmol 2004;122:1117-21. • Eisenberg DL, Sherman BG, McKeown CA, Schuman JS. Tonometry in adults and children: a manometric evaluation of pneumotonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 1998;105:1173-81. • Diaton: digital portable tonometer of intraocular pressure through the eyelid. Operation Manual. Ryazan State Instrument Making Enterprise. Ryazan, Russia. • Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra- Pimentel E. Clinical evaluation of the new TGDc-01 “PRA” palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci 2005;82:143-50. • Troost A, Yun SH, Specht K, Krummenauer F, Schwenn. Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers. Br J Ophthalmol 2005;89:280-3. • Losch A, Scheuerle A, Rupp V, Auffarth G, Becker M. Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry. Graefes Arch Clin Exp Opthhalmol. 2005;243:313-6.