1. AMESA-M
Total Mercury Analyser
Monitoring system for fully-automatic and
continuous mercury analysis in smoke gas
(without wet chemistry)
Already MCERTS and TÜV accredited for continuous dioxin and furan
KEYBENEFITS
monitoring, the AMESA - M has been adapted for mercury monitoring.
•
Maintenance-free (6 months) dry
AMESA - M extracts a sample from the flue gas stream under volume
reactor
proportional conditions.
•
•
•
•
Consisting of a sample probe assembly and a control cabinet, the
High operational safety
Easy maintenance
Internal Data Logging Functionality
Low cross sensitivities
The unit can store up to a four week period of sampling data which is
Thermal mass flow meter to control
then available to download via a USB flash drive. Analysis can be achieved
mass flow independent of gas &
with an onsite analyser or a certified laboratory.
pressure
Dry gas meter to determine sample
Measuring Principle
volume
The sample gas is converted into mercury vapour by a combination of
thermal and dry chemical treatment. This is then continuously
measured in a photometer. Probe gas flow is measured after a gas cooler
at 2°C. The concentration is calculated and displayed as dry flue gas.
“
Providing continuous Mercury
monitoring, The AMESA - M is suited
for audit & control purposes
“
•
2. TECHNICALSPECIFICATION
APPLICATIONS
• Waste incinerations
• Sewage sludge incineration
• Contaminated soil burning plants
• Mercury mines and refineries
• Fluorescent light bulb recycling plants
MEASUREMENT PRINCIPLE
• Measurements: total mercury
• Measuring ranges 0–45 to 0–500 μg/Nm³
• Measuring principle: UV-absorption
POWER
• Measuring outputs: 2 x 0 / 4–20 mA/500 Ohm
• Digital outputs: 4 relay outputs, permissible load 250 V,
100 VA
• Digital inputs: 1 potential free outputs
• Supply voltage: 230 VAC, 50 Hz, 1200 VA
sample probe: 650 VA
sample line: 100 VA/m
SPAN & DRIFT
• Accuracy: <1% of measuring range
• Detection limit: <1 μg/Nm³
• Reference point drift: <2% of measuring range/month
• Zero point drift: automatic zero correction
TEMPERATURE & PRESSURE
• Flue gas temperature: 0–250 °C
• Flue gas pressure: -50 up to +50 hPa
• Ambient temperature: +5 bis +30°C
APPROVALS
• MCERTS
• TÜV Certified
PHYSICAL
• Dimensions: cabinet 1600 x 800 x 500 mm (H x W x D)
• Weight: 220 kg
• Duct diameter: >0.5 m
Complete Project Solutions
Our complete project solutions are developed to assist you
in IED compliance. We achieve this through;
•
•
•
•
•
Expert consultancy
Project management
In-house project design & build
Installation & commissioning
After product service & support