CLEANING VALIDATION & 
ANALYTICA METHOD VALIDATION. 
GUIDED BY presented by RENUKA 
Dr. K. KISHORE KUMAR 256213886008
Cleaning Validation 
Principle 
• The objectives of GMP include prevention of possible 
contamination and cross-contamination 
• Contamination by a variety of substances 
– contaminants (e.g. microbes, previous 
products (both API and excipient residues), 
residues of cleaning agents, airborne 
materials (e.g. dust and particulate matter), 
lubricants and ancillary material, such as 
disinfectants 
• Also decomposition residues from product or detergents 
Reference :- [1 ]
Scope 
• Guidelines: General aspects of cleaning validation 
• Excluding specialized cleaning or inactivation 
– e.g. for removal of viral or mycoplasmal 
contaminants in the biological manufacturing 
industry. 
• Normally cleaning validation needed for critical cleaning, e.g. 
– between manufacturing of one product and 
another 
– contact surfaces (products, drug products and 
API). 
Reference :- [1 ]
Cleaning validation protocols (1) 
• Approved by QC or QA and to cover, e.g. 
– disassembly of system; 
– pre-cleaning; 
– cleaning agent, concentration, solution volume, 
water quality; 
– time and temperature; 
– flow rate, pressure and rinsing; 
– complexity and design of the equipment; 
– training of operators; and 
– size of the system. 
Reference :- [2 ]
Cleaning validation protocols (2) 
• The cleaning validation protocol should include: 
– objectives, responsible people; 
– description of the equipment including the make, 
model, serial number or other unique code; 
– time intervals; bioburden; cleaning procedures; 
– equipment used for routine monitoring (e.g. 
conductivity meters, pH meters and total organic 
carbon analysers); 
– number of cleaning cycles; sampling procedures (e.g. 
direct sampling, rinse sampling, in process 
monitoring and sampling locations) and the rationale 
for their use . 
Reference :- [3]
Cleaning validation protocols (3) 
• The cleaning validation protocol should include (2): 
– data on recovery studies (efficiency of the recovery 
of the sampling technique should be established); 
– analytical methods; 
– acceptance criteria (with rationale for setting the 
specific limits) including a margin for error and for 
sampling efficiency; 
– cleaning agent to be used; 
– revalidation requirements. 
Reference :- [3]
Cleaning validation protocols (4) 
• Cleaning agent used, scientifically justified and based on: 
– the solubility of the materials to be removed; 
– the design and construction of the equipment 
and surface materials to be cleaned; 
– the safety of the cleaning agent; 
– the ease of removal and detection; 
– the product attributes; 
– the minimum temperature and volume of 
cleaning agent and rinse solution; and 
– the manufacturer's recommendations 
Reference :- [3]
Cleaning validation protocols (5) 
Bracketing: 
• Very similar cleaning procedures for products and 
processes - no need for individual validation. “Worst case” 
may be acceptable and should be justified. 
• Consider type of products and equipment; allowed only 
where products are similar in nature or property and 
processed on the same equipment; and identical cleaning 
procedures used. 
Reference :- [3 ]
Cleaning validation protocols (6) 
Bracketing: 
• Representative product - most difficult to clean. 
• Equipment - only when it is similar or the same equipment in 
different sizes (e.g. 300 l, 500 l and 1000 l tanks). 
– Alternative approach may be to validate the 
smallest and the largest sizes separately. 
Reference :- [3]
Cleaning validation reports 
• The relevant cleaning records – (signed by the operator, 
checked by production and reviewed by quality assurance) 
– and source data (original results) should be kept. 
• The results of the cleaning validation should be presented 
in cleaning validation reports stating the outcome and 
conclusion. 
Reference :- [4 ]
Equipment 
• If one SOP for cleaning a piece of equipment, review: 
– products being produced, 
– cleaning in a large campaign, 
– cleaning between batches of different 
products. 
• The design of equipment may influence the effectiveness of the 
cleaning process. 
• Consider design, e.g. V-blenders, transfer pumps or filling lines. 
Reference :- [5 ]
Detergents 
• Released by quality control and meet food standards or 
regulations 
• Composition known 
• Easily removed with rinsing - demonstrated - with 
acceptable limits defined 
• If persistent residues (e.g. cationic detergents) - avoided 
• Consider also detergent breakdown 
Reference :- [3 ]
Microbiology 
• Prevent microbial growth and remove contamination 
• Documented evidence 
– routine cleaning 
– storage of equipment 
• The period and conditions 
– storage of unclean equipment before cleaning 
– between cleaning and equipment reuse 
• Equipment stored in a dry condition after cleaning (no stagnant 
water) 
• Control of bioburden important 
Reference :- [5 ]
Sampling 
• Clean as soon as possible after use 
– especially topical products, suspensions and 
bulk drug or 
– where the drying of residues will directly 
affect the efficiency of a cleaning procedure 
• Two methods of sampling: 
– direct surface sampling and 
– rinse samples 
• Combination of the two - most desirable 
Reference :- [4 ]
Sampling 
• Re-sampling: 
– not to be done before or during cleaning 
• Constant re-testing and re-sampling: 
– can show that the cleaning process is not 
validated 
– may indicate presence of unacceptable 
residue and contaminants resulting from an 
ineffective cleaning process. 
Reference :- [4 ]
Direct surface sampling (direct method) 
• Most commonly used method 
• Use “swabs” (inert material) - type of sampling material should not 
interfere with the test 
• Factors to be considered include: 
– supplier of the swab, 
– area swabbed, number of swabs used, 
whether they are wet or dry swabs, 
– swab handling and swabbing technique 
Reference :- [4 ]
Direct surface sampling (direct method) (2) 
• Other factors include: 
– location from which the sample is taken 
(including worst case locations, identified in 
the protocol) 
– composition of the equipment (e.g. glass or 
steel) 
• Critical areas (hardest to clean) 
– e.g. in semi-automatic/fully automatic clean-in- 
place systems 
• Use appropriate sampling medium and solvent 
Reference :- [4 ]
Rinse samples (indirect method) 
• Allows sampling of: 
– a large surface 
– areas that are inaccessible or that cannot be 
routinely disassembled 
• Provides an "overall picture" 
• Useful for checking for residues of cleaning agents 
• In combination with other sampling methods such as surface 
sampling 
Reference :- [4 ]
Rinse samples (indirect method) (2) 
 The manufacturer has to provide evidence that samples are 
accurately recovered 
 What is considered acceptable in terms of recovery? 
Reference :- [4 ]
Levels of cleaning 
• LEVEL USED WHEN CLEANING VALIDATION REQUIRED 
• LEVEL 2 
 • Product changeover of equipment used in final step 
 • Intermediates of one batch to final step of another 
• LEVEL 1 
 • Intermediates or final Step of one product to intermediate of another 
 • Early Step to intermediates in a product sequence 
• progression between level 0 and 2 depending on process and nature of 
contaminant based on scientific rational 
• LEVEL 0- in-campaign, batch to batch changeover 
• no validation required 
• NB: ALL PROCESSES MUST BE EVALUATED INDIVIDUALLY 
Reference :- [5 ]
Cleaning validation process 
step-1 
DETERMINE THE MOST APPROPRIATE CLEANING PROCEDURE 
FOR THE EQUIPMENT - 
• 1. Generate acceptance criteria data for the contaminant. 2. 
The cleaning method will be determined by the process, the 
equipment the cleaning agents and the cleaning techniques 
available. 3. All aspects of the cleaning procedure should be 
clearly defined in SOPs be they manual / CIP or COP 
Reference :- [5 ]
DEVELOP AND VALIDATE THE SAMPLING AND CHOSEN 
ANALYTICAL METHODS FOR THE COMPOUND(S) BEING 
CLEANED 
• 1. Swab 2. Rinse 
• (determine % recovery, limit of detection, limit of 
quantitation, accuracy of method, reproducibility, stability 
over time ...etc.) 
Reference :- [5 ]
EVALUATE EQUIPMENT SURFACES AND DETERMINE 
• 1. Worst case locations to sample (swab sampling) 
• 1. Volume and type of rinse solvent to be employed (rinse 
sampling) 
• 1. Equipment surface area (necessary to calculate carryover 
into subsequent batches) 
Reference :- [5 ]
Step-2 
DEVELOP A CLEANING VALIDATION PROTOCOL FOR THE 
PRODUCT AND THE EQUIPMENT BEING CLEANED 
• That should encompass for example: 1. Introduction 2. Scope 
3. Equipment 4. Cleaning procedure 5. Sampling procedures 6. 
Analytical testing procedure 7. Acceptance/Cleaning limits. 8. 
Acceptance criteria for the validation. 
Reference :- [5 ]
Step-3 
INTERIM REPORT: GENERATE INTERIM CLEANING VALIDATION 
REPORTS ON A CLEAN BY CLEAN BASIS DETAILING THE 
ACCEPTABILITY OF THE CLEANING PROCEDURE FOR THE 
EQUIPMENT AND THE PRODUCT. 
• This is only required where there is a long period of time 
between manufacture of the validation runs (see stage 4 for 
reporting requirements). 
Reference :- [5 ]
Step-4 
FOR THE PRODUCT AND THE EQUIPMENT BEING CLEANED 
• That should encompass for example: 1. Introduction 2. Scope 3. 
Equipment 4. Cleaning procedure 5. Sampling procedures 6. Analytical 
testing procedure 7. Acceptance/Cleaning limits. 8. Acceptance criteria for 
the validation. 
• GENERATE A CLEANING VALIDATION REPORT DETAILING THE 
ACCEPTABILITY OF THE CLEANING PROCEDURE FOR THE EQUIPMENT AND 
THE PRODUCT 
• The report should give a full detailed background and introduction to the 
cleaning Validation study and should evaluate all data generated with 
respect to the acceptance criteria employed for the study. The report 
should also indicate the requirement if any for revalidation (period of time 
/change control etc.) 
Reference :- [5 ]
Validation 
> 80% is considered 
Recovery 
good 
Reference :- [6 ]
Validation 
> 80% is considered 
Recovery 
good 
> 50% is considered 
reasonable 
Reference :- [6 ]
Validation 
Recovery 
< 50% is considered 
questionable 
> 80% is considered 
good 
> 50% is considered 
reasonable 
Reference :- [6 ]
Establishing acceptable limits 
• Limits: Practical, achievable and verifiable 
• Rationale: Logical, based on knowledge of materials 
• Each situation assessed individually 
• Principal reactant and other chemical variations 
• Screening (thin-layer chromatography) in addition to 
chemical analyses where necessary 
Reference :- [6 ]
Establishing acceptable limits (2) 
There should be no residue from: 
• Previous product 
• Reaction by-products and degradants 
• Cleaning process itself (e.g. detergents or solvents) 
Reference :- [6 ]
Establishing acceptable limits (3) 
 The limit-setting approach can: 
– be product-specific 
– group products into families and choose a 
worst case product 
– group products into groups according to risk, 
e.g. very soluble products, products with 
similar potency, highly toxic, or difficult to 
detect products 
– use different safety factors for different 
dosage forms based on physiological 
response (this method is essential for potent 
materials) 
Reference :- [6 ]
Establishing acceptable limits (4) 
 The three most commonly used criteria are: 
– Visually clean No residue visible on equipment 
after cleaning. Spiking studies to determine the 
concentration at which most active ingredients 
are visible. (May not be suitable5 for high 
potency, low-dosage drugs.) 
– No more than 10 ppm of one product will appear 
in another product (basis for heavy metals in 
starting materials). 
– No more than 0.1% of the normal therapeutic 
dose of one product will appear in the maximum 
daily dose of a subsequent product. 
Reference :- [6 ]
Introduction 
• Method validation is the process used to confirm that the 
analytical procedure employed for a specific test is suitable 
for its intended use. Results from method validation can be 
used to judge the quality, reliability and consistency of 
analytical results; it is an integral part of any good analytical 
practice. 
• Analytical methods need to be validated or revalidated 
Reference :- [7 ]
Once the method has been developed and validated, a validation 
report should be prepared that includes the following: 
• Objective and scope of the method (applicability, type). 
• Summary of methodology. 
• Type of compounds and matrix. 
• All chemicals, reagents, reference standards, QC samples with purity, 
grade, their source or detailed instructions on their preparation. 
• Procedures for quality checks of standards and chemicals used. 
• Safety precautions. 
• A plan and procedure for method implementation from the method 
development lab to routine analysis. 
• Method parameters. 
• Critical parameters taken from robustness testing. 
• Listing of equipment and its functional and performance requirements, 
e.g., cell dimensions, baseline noise and column temperature range. 
For complex equipment, a picture or schematic diagram may be useful. 
Reference :- [7 ]
• Detailed conditions on how the experiments were conducted, including 
sample preparation. The report must be detailed enough to ensure that it 
can be reproduced by a competent technician with comparable equipment. 
• Statistical procedures and representative calculations. 
• Procedures for QC in routine analyses, e.g., system suitability tests. 
• Representative plots, e.g., chromatograms, spectra and calibration 
curves. 
• Method acceptance limit performance data. 
•The expected uncertainty of measurement results. 
•Criteria for revalidation. 
•The person(s) who developed and validated the method. 
•References (if any). 
•Summary and conclusions. 
•Approval with names, titles, date and signature of those responsible for 
the review and approval of the analytical test procedure. 
Reference :- [8]
Parameters for analytical methid 
validation 
Reference :- [8 ]
Selectivity $ specificity 
• It is a method that produces a response for a single analyte 
only, while the term selective refers to a method that provides 
responses for a number of chemical entities that may or may 
not be distinguished from each other. If the response is 
distinguished from all other responses, the method is said to 
be selective 
• Selectivity studies should also assess interferences 
that may be caused by the matrix, e.g., urine, blood, 
soil, water or food. Optimized sample preparation can 
eliminate most of the matrix components. The absence 
of matrix interferences for a quantitative method should 
be demonstrated by the analysis of at least five 
independent sources of control matrix. 
Reference :- [8 ]
Precision and Reproducibility 
• The precision of a method (Table 4) is the extent to which 
the individual test results of multiple injections of a 
series of standards agree. The measured standard 
deviation can be subdivided into 3 categories: 
repeatability, intermediate precision and reproducibility 
(4, 5). Repeatability is obtained when the analysis is 
carried out in a laboratory by an operator using a piece 
of equipment over a relatively short time span. At least 6 
determinations of 3 different matrices at 2 or 3 different 
concentrations should be performed, and the RSD 
calculated. 
Reference :- [9 ]
Accuracy and Recovery 
• The accuracy of an analytical method is the extent to which 
test results generated by the method and the true value 
agree. Accuracy can also be described as the closeness of 
agreement between the value that is adopted, either as a 
conventional, true or accepted reference value, and the value 
found. 
Reference :- [10 ]
Linearity and Calibration Curve 
• The linearity of an analytical method is its ability to elicit test results that 
are directly proportional to the concentration of analytes in samples 
within a given range or proportional by means of well-defined 
mathematical transformations. Linearity may be demonstrated directly on 
the test substance (by dilution of a standard stock solution) and/or by 
using separate weighings of synthetic mixtures of the test product 
components, using the proposed procedure. 
• Linearity is determined by a series of 3 to 6 injections of 5 or more 
standards whose concentrations span 80–120 percent of the expected 
concentration range. The response should be directly proportional to the 
concentrations of the analytes or proportional by means of a well-defined 
mathematical calculation. A linear regression equation applied to the 
results should have an intercept not significantly different from 0. If a 
significant nonzero intercept is obtained, it should be demonstrated that 
this has no effect on the accuracy of the method. 
Reference :- [10 ]
Range 
• The range of an analytical method is the interval between the 
upper and lower levels (including these levels) that have been 
demonstrated to be determined with precision, accuracy and 
linearity using the method as written. The range is normally 
expressed in the same units as the test results (e.g., percentage, 
parts per million) obtained by the analytical method. 
• For assay tests, the ICH (5) requires the minimum specified range to 
be 80 to 120 percent of the test concentration, and for the 
determination of an impurity, the range to extend from the limit of 
quantitation, or from 50 percent of the specification of each 
impurity, whichever is greater, to 120 percent of the specification. 
Reference :- [10 ]
Limit of Detection 
• The limit of detection is the point at which a measured value 
is larger than the uncertainty associated with it. It is the 
lowest concentration of analyte in a sample that can be 
detected but not necessarily quantified. The limit of detection 
is frequently confused with the sensitivity of the method. The 
sensitivity of an analytical method is the capability of the 
method to discriminate small differences in concentration or 
mass of the test analyte. In practical terms, sensitivity is the 
slope of the calibration curve that is obtained by plotting the 
response against the analyte concentration or mass. 
• In chromatography, the detection limit is the injected amount 
that results in a peak with a height at least two or three times 
as high as the baseline noise level. Besides this signal/noise 
method, the ICH (4) describes three more methods: 
• Visual inspection: The detection limit is determined by the 
analysis of samples with known concentrations of analyte and 
by establishing the minimum level at which the analyte can be 
reliably detected. 
Reference :- [10 ]
•Standard deviation of the response based on the standard 
deviation of the blank: Measurement of the magnitude of 
analytical background response is performed by analyzing 
an appropriate number of blank samples and calculating the 
standard deviation of these responses. 
•Standard deviation of the response based on the slope of the 
calibration curve: A specific calibration curve is studied using 
samples containing an analyte in the range of the limit of 
detection. The residual standard deviation of a regression line, 
or the standard deviation of y-intercepts of regression lines, may 
be used as the standard deviation. 
Reference :- [10 ]
Ruggedness 
• Ruggedness is a measure of reproducibility of test results 
under normal, expected operational conditions from 
laboratory to laboratory and from analyst to analyst. 
Ruggedness is determined by the analysis of aliquots from 
homogeneous lots in different laboratories. 
Reference :- [10 ]
Robustness 
• Robustness tests examine the effect that operational 
parameters have on the analysis results. For the 
determination of a method’s robustness, a number of 
method parameters, for example, pH, flow rate, column 
temperature, injection volume, detection wavelength or 
mobile phase composition, are varied within a realistic 
range, and the quantitative influence of the variables is 
determined. If the influence of the parameter is within a 
previously specified tolerance, the parameter is said to 
be within the method’s robustness range. 
Reference :- [8 ]
References 
1. ICH Good Manufacturing Practice Guideline for Active 
Pharmaceutical Ingredients. (July 23 1999). 
2. Principles of Qualification and Validation in Pharmaceutical 
Manufacture - Recommendations on Cleaning Validation. (ref. 
Document PR 1/ 99 March 1999. 
3. Zeller, ’Cleaning Validation and residue limits: a contribution 
to current discussions’, pharmaceutical technology Europe. 
(November 1993). 
4. S.W. Harder, ‘The validation of cleaning processes’, 
pharmaceutical technology. (1984) 
5. Mc Arthur, Vasilevsky, ‘Cleaning validation for biological 
products: case study’, pharmaceutical engineering. 
(November / December 1995).
6. James Agalloco, ‘Points to consider in the validation of 
equipment cleaning procedures’, Journal of parenteral 
science and technology. (October 1992). 
7. ISO/IEC 17025.international standad. Eurachem. guidance 
document no.1/WELAC guidance document no.WGD 2. 
8. ICH of technical requirements for the regestration of 
pharmaceuticals for human use. 
9. Green, J.M.A practical guide to analytical method 
validation. 
10. Eurachem. guidance document no.1/WELAC guidance 
document no.WGD 2:accreditation for chemical lads & 
ISO/IEC guide 25 (1993).

Weitere ähnliche Inhalte

Was ist angesagt?

QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERQUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERUshaKhanal3
 
cleaning validation..
cleaning validation..cleaning validation..
cleaning validation..akshara01
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validationVinay Jain
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipmentsPranali Polshettiwar
 
Validation of utility system (water system)
Validation of utility system (water system)Validation of utility system (water system)
Validation of utility system (water system)ShameerAbid
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry personabhishek pandey
 
Qualification of membrane filtration apparatus
Qualification of membrane filtration apparatusQualification of membrane filtration apparatus
Qualification of membrane filtration apparatusPRAVADA
 
Validation of compressed air and nitrogen
Validation of compressed air and nitrogenValidation of compressed air and nitrogen
Validation of compressed air and nitrogenManiKandan1405
 
qualification of lcms
qualification of lcmsqualification of lcms
qualification of lcmsAnkush Sule
 
IPQC of dosage form in pharmaceutical industry
IPQC of dosage form in pharmaceutical industryIPQC of dosage form in pharmaceutical industry
IPQC of dosage form in pharmaceutical industryKay kay shain marma
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamJp Prakash
 
Qualification Of Autoclave
Qualification Of AutoclaveQualification Of Autoclave
Qualification Of Autoclave8669290285
 
STERILE PROCESS OF VALIDATION
STERILE PROCESS OF VALIDATIONSTERILE PROCESS OF VALIDATION
STERILE PROCESS OF VALIDATIONSagar Savale
 

Was ist angesagt? (20)

QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTERQUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
QUALIFICATION OF TAP DENSITY TESTER & DISINTEGRATION TESTER
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
cleaning validation..
cleaning validation..cleaning validation..
cleaning validation..
 
cleaning validation
cleaning validationcleaning validation
cleaning validation
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Qualification Of HPTLC
Qualification Of HPTLCQualification Of HPTLC
Qualification Of HPTLC
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipments
 
Validation of utility system (water system)
Validation of utility system (water system)Validation of utility system (water system)
Validation of utility system (water system)
 
Pharmaceutical validation of water system
Pharmaceutical validation of  water system Pharmaceutical validation of  water system
Pharmaceutical validation of water system
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry person
 
Qualification of membrane filtration apparatus
Qualification of membrane filtration apparatusQualification of membrane filtration apparatus
Qualification of membrane filtration apparatus
 
Validation of compressed air and nitrogen
Validation of compressed air and nitrogenValidation of compressed air and nitrogen
Validation of compressed air and nitrogen
 
qualification of lcms
qualification of lcmsqualification of lcms
qualification of lcms
 
Qualification of DSC
Qualification of DSCQualification of DSC
Qualification of DSC
 
IPQC of dosage form in pharmaceutical industry
IPQC of dosage form in pharmaceutical industryIPQC of dosage form in pharmaceutical industry
IPQC of dosage form in pharmaceutical industry
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steam
 
Qualification Of Autoclave
Qualification Of AutoclaveQualification Of Autoclave
Qualification Of Autoclave
 
STERILE PROCESS OF VALIDATION
STERILE PROCESS OF VALIDATIONSTERILE PROCESS OF VALIDATION
STERILE PROCESS OF VALIDATION
 

Andere mochten auch

Cleaning validation a complete know how
Cleaning validation a complete know howCleaning validation a complete know how
Cleaning validation a complete know howSambhujyoti Das
 
Cleaning validation presentation
Cleaning validation presentationCleaning validation presentation
Cleaning validation presentationVishnu Satpute
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validationSai Vivek Kosaraju
 
Pmtc good cleaning validation practice
Pmtc  good cleaning validation practicePmtc  good cleaning validation practice
Pmtc good cleaning validation practiceRamy Mostafa
 
cleaning validation..
cleaning validation..cleaning validation..
cleaning validation..akshara01
 
Fundamental of cleaning validation
Fundamental of cleaning validationFundamental of cleaning validation
Fundamental of cleaning validationEr. Vishal Katiyar
 
Basic Concepts of Cleaning validation
Basic Concepts of Cleaning validationBasic Concepts of Cleaning validation
Basic Concepts of Cleaning validationAbhishek Morris
 
A practical guide to analytical method validation, including measurement unce...
A practical guide to analytical method validation, including measurement unce...A practical guide to analytical method validation, including measurement unce...
A practical guide to analytical method validation, including measurement unce...Victor Huamaní León
 
Cleaning validation strategy &amp; regulations
Cleaning validation strategy &amp; regulationsCleaning validation strategy &amp; regulations
Cleaning validation strategy &amp; regulationsJahnavi825
 
validation. chemical and biological evaluation, cleaning validation, personal...
validation. chemical and biological evaluation, cleaning validation, personal...validation. chemical and biological evaluation, cleaning validation, personal...
validation. chemical and biological evaluation, cleaning validation, personal...Sharath Hns
 
1 mia q_alimentar_espectrofotometria_2010
1 mia q_alimentar_espectrofotometria_20101 mia q_alimentar_espectrofotometria_2010
1 mia q_alimentar_espectrofotometria_2010Alexander Krogoll
 
High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)Swati Sahani
 
Bioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning ValidationBioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning Validationangelsalaman
 
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...nvasoya
 

Andere mochten auch (20)

Cleaning validation a complete know how
Cleaning validation a complete know howCleaning validation a complete know how
Cleaning validation a complete know how
 
Cleaning validation presentation
Cleaning validation presentationCleaning validation presentation
Cleaning validation presentation
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validation
 
Pmtc good cleaning validation practice
Pmtc  good cleaning validation practicePmtc  good cleaning validation practice
Pmtc good cleaning validation practice
 
Managing risk in cleaning validation
Managing risk in cleaning validationManaging risk in cleaning validation
Managing risk in cleaning validation
 
cleaning validation..
cleaning validation..cleaning validation..
cleaning validation..
 
Cleaning Validation in API Manufacturing Plant.
Cleaning Validation  in  API Manufacturing Plant.Cleaning Validation  in  API Manufacturing Plant.
Cleaning Validation in API Manufacturing Plant.
 
Fundamental of cleaning validation
Fundamental of cleaning validationFundamental of cleaning validation
Fundamental of cleaning validation
 
Basic Concepts of Cleaning validation
Basic Concepts of Cleaning validationBasic Concepts of Cleaning validation
Basic Concepts of Cleaning validation
 
A practical guide to analytical method validation, including measurement unce...
A practical guide to analytical method validation, including measurement unce...A practical guide to analytical method validation, including measurement unce...
A practical guide to analytical method validation, including measurement unce...
 
Analytical validation
Analytical validationAnalytical validation
Analytical validation
 
Cleaning validation strategy &amp; regulations
Cleaning validation strategy &amp; regulationsCleaning validation strategy &amp; regulations
Cleaning validation strategy &amp; regulations
 
Hptlc
HptlcHptlc
Hptlc
 
validation. chemical and biological evaluation, cleaning validation, personal...
validation. chemical and biological evaluation, cleaning validation, personal...validation. chemical and biological evaluation, cleaning validation, personal...
validation. chemical and biological evaluation, cleaning validation, personal...
 
1 mia q_alimentar_espectrofotometria_2010
1 mia q_alimentar_espectrofotometria_20101 mia q_alimentar_espectrofotometria_2010
1 mia q_alimentar_espectrofotometria_2010
 
EMA Qualification & Validation Requirements
EMA Qualification & Validation Requirements EMA Qualification & Validation Requirements
EMA Qualification & Validation Requirements
 
High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)
 
Bioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning ValidationBioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning Validation
 
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...
Lean Clean - Total Organic Carbon Analysis for Cleaning Validation in Pharmac...
 

Ähnlich wie Cleaning and analytical validation

Cleaning validation
Cleaning validationCleaning validation
Cleaning validationRam Kumar
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfLdiaSabino
 
validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfMd. Omar Faruq Khan
 
Good manufacturing practices (schedule m)
Good manufacturing practices (schedule m)Good manufacturing practices (schedule m)
Good manufacturing practices (schedule m)Bindu Kshtriya
 
Single use technology: a regulatory perspective
Single use technology: a regulatory perspectiveSingle use technology: a regulatory perspective
Single use technology: a regulatory perspectiveTGA Australia
 
cleaning validation for SSP.pptx
cleaning validation for SSP.pptxcleaning validation for SSP.pptx
cleaning validation for SSP.pptxSurafelKebede5
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationamol dighe
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validationamol dighe
 
Audit ppt.pptx
Audit ppt.pptxAudit ppt.pptx
Audit ppt.pptxNilam71
 
Cleaning Validation by- Deepak Kumar.
Cleaning Validation by- Deepak Kumar.Cleaning Validation by- Deepak Kumar.
Cleaning Validation by- Deepak Kumar.Deepak Kumar
 
The Fundamentals of Cleaning Validation
The Fundamentals of Cleaning ValidationThe Fundamentals of Cleaning Validation
The Fundamentals of Cleaning ValidationSGS
 
Technology Transfer and Scale-up in Pharmaceutical Industry
Technology Transfer and Scale-up in Pharmaceutical IndustryTechnology Transfer and Scale-up in Pharmaceutical Industry
Technology Transfer and Scale-up in Pharmaceutical IndustryPranjalWagh1
 
Clean Room Pharmaceutical Good Manufacturing Process.pdf
Clean Room Pharmaceutical Good Manufacturing Process.pdfClean Room Pharmaceutical Good Manufacturing Process.pdf
Clean Room Pharmaceutical Good Manufacturing Process.pdfHananZayed4
 
Clean Room Pharmaceutical_ GMP Procedures
Clean Room Pharmaceutical_ GMP ProceduresClean Room Pharmaceutical_ GMP Procedures
Clean Room Pharmaceutical_ GMP ProceduresHananZayed4
 
Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Tim Sandle, Ph.D.
 
Cleaning validation
Cleaning validation Cleaning validation
Cleaning validation Alekya Chinki
 

Ähnlich wie Cleaning and analytical validation (20)

xyz
xyzxyz
xyz
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdf
 
doc_project_part2 (1).ppt
doc_project_part2 (1).pptdoc_project_part2 (1).ppt
doc_project_part2 (1).ppt
 
validation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdfvalidation-of-sterilising-and-virally-retentive-filters.pdf
validation-of-sterilising-and-virally-retentive-filters.pdf
 
Validation Part3
Validation Part3Validation Part3
Validation Part3
 
Good manufacturing practices (schedule m)
Good manufacturing practices (schedule m)Good manufacturing practices (schedule m)
Good manufacturing practices (schedule m)
 
Single use technology: a regulatory perspective
Single use technology: a regulatory perspectiveSingle use technology: a regulatory perspective
Single use technology: a regulatory perspective
 
cleaning validation for SSP.pptx
cleaning validation for SSP.pptxcleaning validation for SSP.pptx
cleaning validation for SSP.pptx
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
Audit ppt.pptx
Audit ppt.pptxAudit ppt.pptx
Audit ppt.pptx
 
Cleaning Validation by- Deepak Kumar.
Cleaning Validation by- Deepak Kumar.Cleaning Validation by- Deepak Kumar.
Cleaning Validation by- Deepak Kumar.
 
The Fundamentals of Cleaning Validation
The Fundamentals of Cleaning ValidationThe Fundamentals of Cleaning Validation
The Fundamentals of Cleaning Validation
 
Technology Transfer and Scale-up in Pharmaceutical Industry
Technology Transfer and Scale-up in Pharmaceutical IndustryTechnology Transfer and Scale-up in Pharmaceutical Industry
Technology Transfer and Scale-up in Pharmaceutical Industry
 
Validation of gc instrument
Validation of gc instrumentValidation of gc instrument
Validation of gc instrument
 
Clean Room Pharmaceutical Good Manufacturing Process.pdf
Clean Room Pharmaceutical Good Manufacturing Process.pdfClean Room Pharmaceutical Good Manufacturing Process.pdf
Clean Room Pharmaceutical Good Manufacturing Process.pdf
 
Clean Room Pharmaceutical_ GMP Procedures
Clean Room Pharmaceutical_ GMP ProceduresClean Room Pharmaceutical_ GMP Procedures
Clean Room Pharmaceutical_ GMP Procedures
 
Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges
 
Cleaning validation
Cleaning validation Cleaning validation
Cleaning validation
 

Mehr von Malla Reddy College of Pharmacy (20)

Rna secondary structure prediction
Rna secondary structure predictionRna secondary structure prediction
Rna secondary structure prediction
 
Proteomics
ProteomicsProteomics
Proteomics
 
Proteins basics
Proteins basicsProteins basics
Proteins basics
 
Protein structure classification
Protein structure classificationProtein structure classification
Protein structure classification
 
Protein identication characterization
Protein identication characterizationProtein identication characterization
Protein identication characterization
 
Protein modeling
Protein modelingProtein modeling
Protein modeling
 
Primerdesign
PrimerdesignPrimerdesign
Primerdesign
 
Phylogenetic studies
Phylogenetic studiesPhylogenetic studies
Phylogenetic studies
 
Multiple sequence alignment
Multiple sequence alignmentMultiple sequence alignment
Multiple sequence alignment
 
Homology modeling tools
Homology modeling toolsHomology modeling tools
Homology modeling tools
 
Homology modeling
Homology modelingHomology modeling
Homology modeling
 
Genome assembly
Genome assemblyGenome assembly
Genome assembly
 
Genome analysis2
Genome analysis2Genome analysis2
Genome analysis2
 
Genome analysis
Genome analysisGenome analysis
Genome analysis
 
Fasta
FastaFasta
Fasta
 
Drug design intro
Drug design introDrug design intro
Drug design intro
 
Drug design
Drug designDrug design
Drug design
 
Data retrieval
Data retrievalData retrieval
Data retrieval
 
Blast
BlastBlast
Blast
 
Biological databases
Biological databasesBiological databases
Biological databases
 

Cleaning and analytical validation

  • 1. CLEANING VALIDATION & ANALYTICA METHOD VALIDATION. GUIDED BY presented by RENUKA Dr. K. KISHORE KUMAR 256213886008
  • 2.
  • 3. Cleaning Validation Principle • The objectives of GMP include prevention of possible contamination and cross-contamination • Contamination by a variety of substances – contaminants (e.g. microbes, previous products (both API and excipient residues), residues of cleaning agents, airborne materials (e.g. dust and particulate matter), lubricants and ancillary material, such as disinfectants • Also decomposition residues from product or detergents Reference :- [1 ]
  • 4. Scope • Guidelines: General aspects of cleaning validation • Excluding specialized cleaning or inactivation – e.g. for removal of viral or mycoplasmal contaminants in the biological manufacturing industry. • Normally cleaning validation needed for critical cleaning, e.g. – between manufacturing of one product and another – contact surfaces (products, drug products and API). Reference :- [1 ]
  • 5. Cleaning validation protocols (1) • Approved by QC or QA and to cover, e.g. – disassembly of system; – pre-cleaning; – cleaning agent, concentration, solution volume, water quality; – time and temperature; – flow rate, pressure and rinsing; – complexity and design of the equipment; – training of operators; and – size of the system. Reference :- [2 ]
  • 6. Cleaning validation protocols (2) • The cleaning validation protocol should include: – objectives, responsible people; – description of the equipment including the make, model, serial number or other unique code; – time intervals; bioburden; cleaning procedures; – equipment used for routine monitoring (e.g. conductivity meters, pH meters and total organic carbon analysers); – number of cleaning cycles; sampling procedures (e.g. direct sampling, rinse sampling, in process monitoring and sampling locations) and the rationale for their use . Reference :- [3]
  • 7. Cleaning validation protocols (3) • The cleaning validation protocol should include (2): – data on recovery studies (efficiency of the recovery of the sampling technique should be established); – analytical methods; – acceptance criteria (with rationale for setting the specific limits) including a margin for error and for sampling efficiency; – cleaning agent to be used; – revalidation requirements. Reference :- [3]
  • 8. Cleaning validation protocols (4) • Cleaning agent used, scientifically justified and based on: – the solubility of the materials to be removed; – the design and construction of the equipment and surface materials to be cleaned; – the safety of the cleaning agent; – the ease of removal and detection; – the product attributes; – the minimum temperature and volume of cleaning agent and rinse solution; and – the manufacturer's recommendations Reference :- [3]
  • 9. Cleaning validation protocols (5) Bracketing: • Very similar cleaning procedures for products and processes - no need for individual validation. “Worst case” may be acceptable and should be justified. • Consider type of products and equipment; allowed only where products are similar in nature or property and processed on the same equipment; and identical cleaning procedures used. Reference :- [3 ]
  • 10. Cleaning validation protocols (6) Bracketing: • Representative product - most difficult to clean. • Equipment - only when it is similar or the same equipment in different sizes (e.g. 300 l, 500 l and 1000 l tanks). – Alternative approach may be to validate the smallest and the largest sizes separately. Reference :- [3]
  • 11. Cleaning validation reports • The relevant cleaning records – (signed by the operator, checked by production and reviewed by quality assurance) – and source data (original results) should be kept. • The results of the cleaning validation should be presented in cleaning validation reports stating the outcome and conclusion. Reference :- [4 ]
  • 12. Equipment • If one SOP for cleaning a piece of equipment, review: – products being produced, – cleaning in a large campaign, – cleaning between batches of different products. • The design of equipment may influence the effectiveness of the cleaning process. • Consider design, e.g. V-blenders, transfer pumps or filling lines. Reference :- [5 ]
  • 13. Detergents • Released by quality control and meet food standards or regulations • Composition known • Easily removed with rinsing - demonstrated - with acceptable limits defined • If persistent residues (e.g. cationic detergents) - avoided • Consider also detergent breakdown Reference :- [3 ]
  • 14. Microbiology • Prevent microbial growth and remove contamination • Documented evidence – routine cleaning – storage of equipment • The period and conditions – storage of unclean equipment before cleaning – between cleaning and equipment reuse • Equipment stored in a dry condition after cleaning (no stagnant water) • Control of bioburden important Reference :- [5 ]
  • 15. Sampling • Clean as soon as possible after use – especially topical products, suspensions and bulk drug or – where the drying of residues will directly affect the efficiency of a cleaning procedure • Two methods of sampling: – direct surface sampling and – rinse samples • Combination of the two - most desirable Reference :- [4 ]
  • 16. Sampling • Re-sampling: – not to be done before or during cleaning • Constant re-testing and re-sampling: – can show that the cleaning process is not validated – may indicate presence of unacceptable residue and contaminants resulting from an ineffective cleaning process. Reference :- [4 ]
  • 17. Direct surface sampling (direct method) • Most commonly used method • Use “swabs” (inert material) - type of sampling material should not interfere with the test • Factors to be considered include: – supplier of the swab, – area swabbed, number of swabs used, whether they are wet or dry swabs, – swab handling and swabbing technique Reference :- [4 ]
  • 18. Direct surface sampling (direct method) (2) • Other factors include: – location from which the sample is taken (including worst case locations, identified in the protocol) – composition of the equipment (e.g. glass or steel) • Critical areas (hardest to clean) – e.g. in semi-automatic/fully automatic clean-in- place systems • Use appropriate sampling medium and solvent Reference :- [4 ]
  • 19. Rinse samples (indirect method) • Allows sampling of: – a large surface – areas that are inaccessible or that cannot be routinely disassembled • Provides an "overall picture" • Useful for checking for residues of cleaning agents • In combination with other sampling methods such as surface sampling Reference :- [4 ]
  • 20. Rinse samples (indirect method) (2)  The manufacturer has to provide evidence that samples are accurately recovered  What is considered acceptable in terms of recovery? Reference :- [4 ]
  • 21. Levels of cleaning • LEVEL USED WHEN CLEANING VALIDATION REQUIRED • LEVEL 2  • Product changeover of equipment used in final step  • Intermediates of one batch to final step of another • LEVEL 1  • Intermediates or final Step of one product to intermediate of another  • Early Step to intermediates in a product sequence • progression between level 0 and 2 depending on process and nature of contaminant based on scientific rational • LEVEL 0- in-campaign, batch to batch changeover • no validation required • NB: ALL PROCESSES MUST BE EVALUATED INDIVIDUALLY Reference :- [5 ]
  • 22. Cleaning validation process step-1 DETERMINE THE MOST APPROPRIATE CLEANING PROCEDURE FOR THE EQUIPMENT - • 1. Generate acceptance criteria data for the contaminant. 2. The cleaning method will be determined by the process, the equipment the cleaning agents and the cleaning techniques available. 3. All aspects of the cleaning procedure should be clearly defined in SOPs be they manual / CIP or COP Reference :- [5 ]
  • 23. DEVELOP AND VALIDATE THE SAMPLING AND CHOSEN ANALYTICAL METHODS FOR THE COMPOUND(S) BEING CLEANED • 1. Swab 2. Rinse • (determine % recovery, limit of detection, limit of quantitation, accuracy of method, reproducibility, stability over time ...etc.) Reference :- [5 ]
  • 24. EVALUATE EQUIPMENT SURFACES AND DETERMINE • 1. Worst case locations to sample (swab sampling) • 1. Volume and type of rinse solvent to be employed (rinse sampling) • 1. Equipment surface area (necessary to calculate carryover into subsequent batches) Reference :- [5 ]
  • 25. Step-2 DEVELOP A CLEANING VALIDATION PROTOCOL FOR THE PRODUCT AND THE EQUIPMENT BEING CLEANED • That should encompass for example: 1. Introduction 2. Scope 3. Equipment 4. Cleaning procedure 5. Sampling procedures 6. Analytical testing procedure 7. Acceptance/Cleaning limits. 8. Acceptance criteria for the validation. Reference :- [5 ]
  • 26. Step-3 INTERIM REPORT: GENERATE INTERIM CLEANING VALIDATION REPORTS ON A CLEAN BY CLEAN BASIS DETAILING THE ACCEPTABILITY OF THE CLEANING PROCEDURE FOR THE EQUIPMENT AND THE PRODUCT. • This is only required where there is a long period of time between manufacture of the validation runs (see stage 4 for reporting requirements). Reference :- [5 ]
  • 27. Step-4 FOR THE PRODUCT AND THE EQUIPMENT BEING CLEANED • That should encompass for example: 1. Introduction 2. Scope 3. Equipment 4. Cleaning procedure 5. Sampling procedures 6. Analytical testing procedure 7. Acceptance/Cleaning limits. 8. Acceptance criteria for the validation. • GENERATE A CLEANING VALIDATION REPORT DETAILING THE ACCEPTABILITY OF THE CLEANING PROCEDURE FOR THE EQUIPMENT AND THE PRODUCT • The report should give a full detailed background and introduction to the cleaning Validation study and should evaluate all data generated with respect to the acceptance criteria employed for the study. The report should also indicate the requirement if any for revalidation (period of time /change control etc.) Reference :- [5 ]
  • 28. Validation > 80% is considered Recovery good Reference :- [6 ]
  • 29. Validation > 80% is considered Recovery good > 50% is considered reasonable Reference :- [6 ]
  • 30. Validation Recovery < 50% is considered questionable > 80% is considered good > 50% is considered reasonable Reference :- [6 ]
  • 31. Establishing acceptable limits • Limits: Practical, achievable and verifiable • Rationale: Logical, based on knowledge of materials • Each situation assessed individually • Principal reactant and other chemical variations • Screening (thin-layer chromatography) in addition to chemical analyses where necessary Reference :- [6 ]
  • 32. Establishing acceptable limits (2) There should be no residue from: • Previous product • Reaction by-products and degradants • Cleaning process itself (e.g. detergents or solvents) Reference :- [6 ]
  • 33. Establishing acceptable limits (3)  The limit-setting approach can: – be product-specific – group products into families and choose a worst case product – group products into groups according to risk, e.g. very soluble products, products with similar potency, highly toxic, or difficult to detect products – use different safety factors for different dosage forms based on physiological response (this method is essential for potent materials) Reference :- [6 ]
  • 34. Establishing acceptable limits (4)  The three most commonly used criteria are: – Visually clean No residue visible on equipment after cleaning. Spiking studies to determine the concentration at which most active ingredients are visible. (May not be suitable5 for high potency, low-dosage drugs.) – No more than 10 ppm of one product will appear in another product (basis for heavy metals in starting materials). – No more than 0.1% of the normal therapeutic dose of one product will appear in the maximum daily dose of a subsequent product. Reference :- [6 ]
  • 35.
  • 36. Introduction • Method validation is the process used to confirm that the analytical procedure employed for a specific test is suitable for its intended use. Results from method validation can be used to judge the quality, reliability and consistency of analytical results; it is an integral part of any good analytical practice. • Analytical methods need to be validated or revalidated Reference :- [7 ]
  • 37. Once the method has been developed and validated, a validation report should be prepared that includes the following: • Objective and scope of the method (applicability, type). • Summary of methodology. • Type of compounds and matrix. • All chemicals, reagents, reference standards, QC samples with purity, grade, their source or detailed instructions on their preparation. • Procedures for quality checks of standards and chemicals used. • Safety precautions. • A plan and procedure for method implementation from the method development lab to routine analysis. • Method parameters. • Critical parameters taken from robustness testing. • Listing of equipment and its functional and performance requirements, e.g., cell dimensions, baseline noise and column temperature range. For complex equipment, a picture or schematic diagram may be useful. Reference :- [7 ]
  • 38. • Detailed conditions on how the experiments were conducted, including sample preparation. The report must be detailed enough to ensure that it can be reproduced by a competent technician with comparable equipment. • Statistical procedures and representative calculations. • Procedures for QC in routine analyses, e.g., system suitability tests. • Representative plots, e.g., chromatograms, spectra and calibration curves. • Method acceptance limit performance data. •The expected uncertainty of measurement results. •Criteria for revalidation. •The person(s) who developed and validated the method. •References (if any). •Summary and conclusions. •Approval with names, titles, date and signature of those responsible for the review and approval of the analytical test procedure. Reference :- [8]
  • 39. Parameters for analytical methid validation Reference :- [8 ]
  • 40. Selectivity $ specificity • It is a method that produces a response for a single analyte only, while the term selective refers to a method that provides responses for a number of chemical entities that may or may not be distinguished from each other. If the response is distinguished from all other responses, the method is said to be selective • Selectivity studies should also assess interferences that may be caused by the matrix, e.g., urine, blood, soil, water or food. Optimized sample preparation can eliminate most of the matrix components. The absence of matrix interferences for a quantitative method should be demonstrated by the analysis of at least five independent sources of control matrix. Reference :- [8 ]
  • 41. Precision and Reproducibility • The precision of a method (Table 4) is the extent to which the individual test results of multiple injections of a series of standards agree. The measured standard deviation can be subdivided into 3 categories: repeatability, intermediate precision and reproducibility (4, 5). Repeatability is obtained when the analysis is carried out in a laboratory by an operator using a piece of equipment over a relatively short time span. At least 6 determinations of 3 different matrices at 2 or 3 different concentrations should be performed, and the RSD calculated. Reference :- [9 ]
  • 42. Accuracy and Recovery • The accuracy of an analytical method is the extent to which test results generated by the method and the true value agree. Accuracy can also be described as the closeness of agreement between the value that is adopted, either as a conventional, true or accepted reference value, and the value found. Reference :- [10 ]
  • 43. Linearity and Calibration Curve • The linearity of an analytical method is its ability to elicit test results that are directly proportional to the concentration of analytes in samples within a given range or proportional by means of well-defined mathematical transformations. Linearity may be demonstrated directly on the test substance (by dilution of a standard stock solution) and/or by using separate weighings of synthetic mixtures of the test product components, using the proposed procedure. • Linearity is determined by a series of 3 to 6 injections of 5 or more standards whose concentrations span 80–120 percent of the expected concentration range. The response should be directly proportional to the concentrations of the analytes or proportional by means of a well-defined mathematical calculation. A linear regression equation applied to the results should have an intercept not significantly different from 0. If a significant nonzero intercept is obtained, it should be demonstrated that this has no effect on the accuracy of the method. Reference :- [10 ]
  • 44. Range • The range of an analytical method is the interval between the upper and lower levels (including these levels) that have been demonstrated to be determined with precision, accuracy and linearity using the method as written. The range is normally expressed in the same units as the test results (e.g., percentage, parts per million) obtained by the analytical method. • For assay tests, the ICH (5) requires the minimum specified range to be 80 to 120 percent of the test concentration, and for the determination of an impurity, the range to extend from the limit of quantitation, or from 50 percent of the specification of each impurity, whichever is greater, to 120 percent of the specification. Reference :- [10 ]
  • 45. Limit of Detection • The limit of detection is the point at which a measured value is larger than the uncertainty associated with it. It is the lowest concentration of analyte in a sample that can be detected but not necessarily quantified. The limit of detection is frequently confused with the sensitivity of the method. The sensitivity of an analytical method is the capability of the method to discriminate small differences in concentration or mass of the test analyte. In practical terms, sensitivity is the slope of the calibration curve that is obtained by plotting the response against the analyte concentration or mass. • In chromatography, the detection limit is the injected amount that results in a peak with a height at least two or three times as high as the baseline noise level. Besides this signal/noise method, the ICH (4) describes three more methods: • Visual inspection: The detection limit is determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be reliably detected. Reference :- [10 ]
  • 46. •Standard deviation of the response based on the standard deviation of the blank: Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. •Standard deviation of the response based on the slope of the calibration curve: A specific calibration curve is studied using samples containing an analyte in the range of the limit of detection. The residual standard deviation of a regression line, or the standard deviation of y-intercepts of regression lines, may be used as the standard deviation. Reference :- [10 ]
  • 47. Ruggedness • Ruggedness is a measure of reproducibility of test results under normal, expected operational conditions from laboratory to laboratory and from analyst to analyst. Ruggedness is determined by the analysis of aliquots from homogeneous lots in different laboratories. Reference :- [10 ]
  • 48. Robustness • Robustness tests examine the effect that operational parameters have on the analysis results. For the determination of a method’s robustness, a number of method parameters, for example, pH, flow rate, column temperature, injection volume, detection wavelength or mobile phase composition, are varied within a realistic range, and the quantitative influence of the variables is determined. If the influence of the parameter is within a previously specified tolerance, the parameter is said to be within the method’s robustness range. Reference :- [8 ]
  • 49. References 1. ICH Good Manufacturing Practice Guideline for Active Pharmaceutical Ingredients. (July 23 1999). 2. Principles of Qualification and Validation in Pharmaceutical Manufacture - Recommendations on Cleaning Validation. (ref. Document PR 1/ 99 March 1999. 3. Zeller, ’Cleaning Validation and residue limits: a contribution to current discussions’, pharmaceutical technology Europe. (November 1993). 4. S.W. Harder, ‘The validation of cleaning processes’, pharmaceutical technology. (1984) 5. Mc Arthur, Vasilevsky, ‘Cleaning validation for biological products: case study’, pharmaceutical engineering. (November / December 1995).
  • 50. 6. James Agalloco, ‘Points to consider in the validation of equipment cleaning procedures’, Journal of parenteral science and technology. (October 1992). 7. ISO/IEC 17025.international standad. Eurachem. guidance document no.1/WELAC guidance document no.WGD 2. 8. ICH of technical requirements for the regestration of pharmaceuticals for human use. 9. Green, J.M.A practical guide to analytical method validation. 10. Eurachem. guidance document no.1/WELAC guidance document no.WGD 2:accreditation for chemical lads & ISO/IEC guide 25 (1993).

Hinweis der Redaktion

  1. 30 September, 2014
  2. 30 September, 2014
  3. 30 September, 2014
  4. 30 September, 2014
  5. 30 September, 2014
  6. 30 September, 2014
  7. 30 September, 2014
  8. 30 September, 2014
  9. 30 September, 2014
  10. 30 September, 2014
  11. 30 September, 2014
  12. 30 September, 2014
  13. 30 September, 2014
  14. 30 September, 2014
  15. 30 September, 2014
  16. 30 September, 2014
  17. 30 September, 2014
  18. 30 September, 2014
  19. 30 September, 2014
  20. 30 September, 2014
  21. 30 September, 2014
  22. 30 September, 2014
  23. 30 September, 2014
  24. 30 September, 2014
  25. 30 September, 2014