Design Controls
FDA Small Business
Regulatory Education for Industry (REdI)
Bethesda, Maryland
September 18, 2014
Stanley Liu
Consumer Safety Officer
Office of Communication and Education
Center for Devices and Radiological Health
U.S. Food and Drug Administration
2
Overview
I. Introduction to Design Controls
II. Requirements for Design Controls
1. Design planning
2. Design input and output
3. Design review
4. Design verification and validation
5. Design changes
6. Design transfer
7. Design history file
III. Resources and FDA contacts
I. Introduction to Design
Controls
3
4
Design Controls – What are they?
• A set of quality practices and procedures
incorporated into the design and development
process
• Control the design process to assure that device
specifications meet:
− User needs
− Intended use
5
Design Controls – Why?
• 44% of voluntary recalls from October 1983 through
September 1989 - may have been prevented by
adequate design controls. Source: ‘‘Device Recalls: A Study of
Quality Problems’’ (see 55 FR 21108, May 22, 1990) and “QS Regulations
(Final Rule)” (see 61 FR 52602, October 07, 1996)
• 90% of all software related device failures from 1983
through 1989 were due to design related error. Source:
“FDA Medical Device Regulation from Premarket Review to Recall”
(FDA/HHS OEI 09-90-0040, February 1991)
• Most frequent causes for recalls related to Design,
Software, and Non-Conforming Materials/Components.
Source: “Medical Recall Report FY2003 to FY2012” (CDRH Office of
Compliance, Division of Analysis and Program Operations)
6
Design Controls- Regulatory
History
• Safe Medical Device Act of 1990 authorized FDA to
add Design Controls to the current Good Manufacturing
Practice (cGMP) requirements for devices.
• The QS Regulation became effective on June 1, 1997,
replacing the 1978 GMP for medical devices.
• Preamble to the QS regulation: extremely important for
understanding Design Controls.
Quality Systems - Key
Definitions
• Design History File (DHF) - compilation of records
which describes the design history of a finished device.
• Device Master Record (DMR) - compilation of records
containing the procedures and specifications for a
finished device.
• Device History Record (DHR) - compilation of records
containing the production history of a finished device.
7
8
Design Controls – Scope
• Design controls apply to:
– All Class II and Class III devices
– The following Class I devices:
1. Devices automated with computer software
2. Tracheobronchial suction catheters
3. Surgeon's gloves
4. Protective restraints
5. Manual radionuclide applicator system
6. Radionuclide teletherapy source
9
Design Controls
• Design control is a Key Subsystem of Quality System
• Quality System (QS) Regulations: 21CFR 820
– Management controls
– Design controls
– Production and process controls
– Corrective and preventive actions
– Documents, Records, & Change Control
– Material Control
– Facility & Equipment Control
10
Design Controls
• Incorporates Quality System of Risk Management and
Analysis:
– Systematic application of management policies, procedures,
practices, insight/judgment, and experience for identifying,
analyzing, controlling, and monitoring risk.
– Risk Management and Analysis are integrated into the Design
Control process.
11
7 Subsystems of the Quality System
Design Control
Material
Control Documents,
Records &
Change Control
Facility &
Equipment
Control
Production &
Process Control
Corrective &
Preventive
Action Control
Management
Control
12
Design Controls (21 CFR 820.30)
a) General requirements
b) Design and development planning
c) Design input
d) Design output
e) Design review
f) Design verification
g) Design validation
h) Design changes
i) Design transfer
j) Design history file
II. Requirements for Design
Controls
13
14
General Requirements
21CFR 820.30(a)
15
Design Controls – General
• Establish procedures to control device design:
– Define
– Document
– Implement
• Maintain procedures to control device design:
– Review
– Approve
– Update
16
Design and Development Planning
21CFR 820.30(b)
17
Design & Development Planning
• Establish, maintain and document Planning:
– Describe or reference design and development
activities.
– Identify, describe, and define interfaces,
responsibilities, and activities impacting device
design.
– Review, document, approve, and update as
developments and changes evolve.
18
Design Input
21CFR 820.30(c)
19
Definition
• Design inputs mean the physical and
performance characteristics of a device
that are used as a bases for device
design.
20
Design Input
• Establish and maintain procedures for
design input:
– Ensure requirements are appropriate by addressing
user needs and intended use in terms that are
measurable
– Address incomplete, ambiguous, or conflicting
requirements
– Document, review, and approve input requirements
21
Sources of Design Input
• Standards
• Focus groups
• Customers
• MDRs
• CAPA
• Service reports
• Complaints
• Marketing surveys
• Sales feedback
• Competitors’ products
Including input from earlier/previous versions of a device
22
Examples of Design Input
• Device functions
• Physical characteristics
• Performance
• Safety
• Reliability
• Standards
• Regulatory
requirements
• Human factors
• Labeling & packaging
• Maintenance
• Sterilization
• Compatibility with other
devices
• Environmental limits
23
Example
Infusion Pump
Issue /Risk:
• Infusion pump stops operating when high-
voltage medical devices are operating in
the vicinity.
24
Example – Questions to
Consider
• What are the user needs?
• In what environment is the infusion pump
supposed to function?
• What other kinds of equipment will it be used
with?
• What kinds of design input are needed to avoid
this problem?
25
Example - Infusion Pump
User Need
Pump must be capable of
functioning in an electromagnetic
environment (e.g., hospital).
Design Input
Pump must function
uninterrupted when used with
other products that generate an
electromagnetic field.
26
Design Output
21CFR 820.30(d)
27
Definition
• Design outputs mean the results of a design
effort at each design phase and at the end of
the total design effort.
28
Design Output
• Establish and maintain procedures for
design output:
– Define and document design output in terms that allow an
adequate evaluation of conformance to design input.
– Reference acceptance criteria (i.e., criteria that are definable,
and measurable).
– Identify design outputs essential for the proper functioning of the
device.
– Review and approve design output before release.
29
In other words...
• Design outputs are the design
specifications which must meet design input
requirements, as confirmed during design
Verification and Validation and ensured
during design Review (more later).
Input Output
30
Design Output
• Design Outputs are included in premarket submissions as
Device Specifications.
• The finished design output is the basis for the Device
Master Record (DMR).
• The total finished design output consists of the device, its
packaging, labeling, and the Device Master Record
(DMR).
31
User Need
Pump must be capable of functioning
in an electromagnetic environment
(e.g., hospital).
Design Input
Pump must function uninterrupted
when used with other products that
generate an electromagnetic field.
Design Output
•Circuit board with filtering or
buffering circuit
•Pump EMI shield
•Software with signal filtering and
error handling/correcting code
Example - Infusion Pump
32
Design Review
21CFR 820.30(e)
33
Design Review
• Design Review means a documented,
comprehensive, systematic examination to:
– Evaluate adequacy of the design requirements.
– Evaluate capability of the design to meet requirements.
– Identify any problems.
34
Design Review
• Establish and maintain procedures.
• Plan and conduct formal documented Design
Reviews of design results at appropriate stages.
• Include at each design review:
– Representatives of all functions concerned and specialists as
needed.
– Individual(s) without direct responsibility for the stage being
reviewed.
35
Design Review
• Document results of design review in Design
History File (DHF), including:
– Identification of design
– Date
– Individuals performing review
36
Design Verification
21CFR 820.30(f)
37
Definition
• Verification means confirmation by
examination and provision of objective
evidence that output meets input
requirements.
Input = Output
38
Design Verification
• Establish and maintain procedures for
Design Verification:
– Confirm through measurable means (e.g., test
reports, etc.).
– Review, approve and document in Design
History File (DHF).
39
Design Verification
• Many test reports associated with Design
Verification are included in premarket
submissions:
– 510(k)s
– de novos
– Premarket Approval Applications (PMAs)
– Investigational Device Exemptions (IDEs)
40
Example - Infusion Pump
User Need
Pump must function in an operating
room environment.
Design Input
Pump must function uninterrupted
when used with other products that
generate an electromagnetic field.
Design Output
•PCB with filtering circuit
•Pump EMI shield
•Software signal filtering code and
error handling code
Design Verification
•Simulated EMI testing on hardware
and software
•Dimensional verification of shield
•Verification of system error
handling due to EMI
Design Review
41
Design Validation
21CFR 820.30(g)
42
Definition
• Design Validation means establishing
by objective evidence that
specifications(specified requirements)
conform with user needs and intended
use(s).
Specifications= user needs
intended use(s)
43
Design Validation
• Establish and maintain procedures for Design
Validation:
– Under defined operating conditions.
– On initial production units, lots, or batches (or their
equivalents).
– Under actual or simulated use conditions.
44
Design Validation
• Perform software validation and risk analysis,
where appropriate.
• Review, approve, and document in Design
History File.
45
Verification vs. Validation
• Design Verification
– Output meets Input
– “Did I make the product correctly?”
• Design Validation
– Specifications meets user needs and intended use(s)
– “Did I make the correct product?”
46
Design Validation
• The results of Design Validation are typically
submitted in premarket submissions.
• Examples:
– Animal study protocols/reports
– Cadaver study protocols/reports
– Clinical study protocols/reports
47
User Need
Pump must function in an operating
room environment.
Design Input
Pump must function uninterrupted
when used with other products that
generate an electromagnetic field.
Design Output
•PCB with filtering circuit
•Pump EMI shield
•Software signal filtering code and
error handling code
Design Verification
•Simulated EMI testing on
hardware and software
•Dimensional verification of shield
•Verification of system error
handling due to EMI
Design Validation
•EMC testing to industry standards
•Simulated EMI testing in high EMI
environment
•Risk Analysis concerning EMI
•Software validation for filtering
code
Design Review
48
Warning Letter Example
Failure to establish and maintain adequate
procedures for Validating the device design and
Risk Analysis, where appropriate, as required
by 21 CFR 820.30(g)
For example: The Design History Files for XXX did not include
documentation that the device had ever been validated before
production and marketing. When requested, the firm was unable to
provide documentation that validation and risk analysis had been
performed.
49
Design Transfer
21CFR 812.30(h)
50
Design Transfer
• Establish and maintain procedures to ensure
correct Design Transfer into production
specifications.
• Is the Design accurately transferred to
Production?
• Although Design Transfer happens
throughout, there frequently is a final stage of
development intended to ensure all outputs
are adequately transferred.
51
Design Changes
21CFR 820.30(i)
52
Design Changes
• Establish and maintain procedures for the
identification, documentation, validation or
where appropriate verification, review, and
approval of Design Changes before their
implementation.
• Is there a system in place to enact future
changes?
• Often overlooked, but of critically importance.
• How can you improve your product if you
have no system for change?
53
Design Changes
• Depending on the scope and impact of the
change, the change may require:
– A new 510(k)
– A new PMA, a PMA supplement, or a PMA 30-Day Notice
– A new IDE or an IDE supplement
• Changes must be communicated to FDA if the
device is under premarket review or IDE review.
54
Design History File
21CFR 820.30(j)
55
Definition
• Design History File (DHF) means a
compilation of records which describes the
design history of a finished device.
• It is a summation record of all Design
actions, from start to transfer, including
changes.
56
Design History File
• Establish and maintain a design history file for
each type of device
• Include in the DHF, or reference records
information necessary to demonstrate that the
design was developed in accordance with the
Design Plan and Quality Systems
requirements
Design Controls - Summary
• Like the Quality Systems regulations
themselves, Design Control should be
viewed and understood as a system – a
subsystem within the main Quality
System.
57
Design Controls - Summary
• Processes –A set of quality practices and procedures
incorporated into the design and development process
• Goal - Control the design process to assure that device
specifications meet:
– User needs
– Intended use
58
59
III. Resources and FDA
Contacts
• CDRH Learn
Training videos focusing on overview, premarket,
registration/listing, and post-market topics
http://www.fda.gov/Training/CDRHLearn/default.htm
• Device Advice
Website accompaniment to CDRH Learn
http://www.fda.gov/MedicalDevices/DeviceRegulatio
nandGuidance/default.htm
60
QS Regulation and Guidance
(General)
• Quality System Regulation and Preamble
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/PostmarketRequi
rements/QualitySystemsRegulations/ucm230127.htm
• Design Control Guidance For Medical Device Manufacturers
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocum
ents/ucm070627.htm
• Do it By Design – An Introduction to Human Factors in Medical Devices
http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guid
anceDocuments/UCM095061.pdf
• Guidance for Industry and FDA Premarket and Design Control Reviewers –
Medical Device Use – Safety: Incorporating Human Factors Engineering into
Risk Management
http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guida
nceDocuments/UCM094461.pdf
• (DRAFT) Applying Human Factors and Usability Engineering to Optimize
Medical Device Design
http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocumen
ts/ucm259748.htm
61
QS Regulation and Guidance
• Guidance for Industry: Part 11, Electronic Records; Electronic Signatures -
Scope and Application
http://www.fda.gov/RegulatoryInformation/Guidances/ucm125067.htm
• General Principles of Software Validation
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu
ments/ucm085281.htm
• Guidance for the Content of Premarket Submissions for Software Contained
in Medical Devices
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu
ments/ucm089543.htm
• Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf
Software Use in Medical Devices
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu
ments/ucm073778.htm
• Guidance: Mobile Medical Applications
http://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/guida
ncedocuments/ucm263366.pdf
62
Need More Information?
Division of Industry and Consumer
Education (DICE)
http://www.fda.gov/MedicalDevices/DeviceRegulationan
dGuidance/ContactUs--
DivisionofIndustryandConsumerEducation/default.htm
• Email: DICE@fda.hhs.gov
• Phone: 301-796-7100 or 800-638-2041

Weitere ähnliche Inhalte

Was ist angesagt?

ISO: 14971 Quality risk management of medical devices
ISO: 14971 Quality risk management  of medical devicesISO: 14971 Quality risk management  of medical devices
ISO: 14971 Quality risk management of medical devicesAtul Bhombe
 
Iso 13485:2016
Iso 13485:2016Iso 13485:2016
Iso 13485:2016Suhas R
 
Presentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidencePresentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidenceTGA Australia
 
Medical Device Regulatory Approval
Medical Device Regulatory ApprovalMedical Device Regulatory Approval
Medical Device Regulatory Approvalruyang89
 
An Overview for Software as a Medical Device (SaMD)
An Overview for Software as a Medical Device (SaMD)An Overview for Software as a Medical Device (SaMD)
An Overview for Software as a Medical Device (SaMD)DePuy Synthes
 
Good design practice for medical devices
Good design practice for medical devicesGood design practice for medical devices
Good design practice for medical devicesJakob Nielsen
 
Fda quality system regulation 21 CFR820_Medical devices_k_trautman
Fda quality system regulation 21 CFR820_Medical devices_k_trautmanFda quality system regulation 21 CFR820_Medical devices_k_trautman
Fda quality system regulation 21 CFR820_Medical devices_k_trautmanLatvian University
 
Presentation: Life cycle of medical devices
Presentation: Life cycle of medical devicesPresentation: Life cycle of medical devices
Presentation: Life cycle of medical devicesTGA Australia
 
21 cfr part 210 and 211
21 cfr part 210 and 21121 cfr part 210 and 211
21 cfr part 210 and 211Bhanu Chava
 
Risk Management for Medical Devices - ISO 14971 Overview
Risk Management for Medical Devices - ISO 14971 Overview Risk Management for Medical Devices - ISO 14971 Overview
Risk Management for Medical Devices - ISO 14971 Overview Greenlight Guru
 
IEC 62304: SDLC Conformance and Management
IEC 62304: SDLC Conformance and Management IEC 62304: SDLC Conformance and Management
IEC 62304: SDLC Conformance and Management MethodSense, Inc.
 
Medical Devices Regulation (MDR) 2017/745 - Clinical investigations
Medical Devices Regulation (MDR)  2017/745 - Clinical investigationsMedical Devices Regulation (MDR)  2017/745 - Clinical investigations
Medical Devices Regulation (MDR) 2017/745 - Clinical investigationsArete-Zoe, LLC
 
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Swapnil Fernandes
 

Was ist angesagt? (20)

ISO Standard 13485
ISO Standard 13485ISO Standard 13485
ISO Standard 13485
 
ISO: 14971 Quality risk management of medical devices
ISO: 14971 Quality risk management  of medical devicesISO: 14971 Quality risk management  of medical devices
ISO: 14971 Quality risk management of medical devices
 
Iso 13485:2016
Iso 13485:2016Iso 13485:2016
Iso 13485:2016
 
Iso13485 ppt
Iso13485 pptIso13485 ppt
Iso13485 ppt
 
Presentation: Conformity Assessment Evidence
Presentation: Conformity Assessment EvidencePresentation: Conformity Assessment Evidence
Presentation: Conformity Assessment Evidence
 
Understanding IEC 62304
Understanding IEC 62304Understanding IEC 62304
Understanding IEC 62304
 
Gamp5 new
Gamp5 newGamp5 new
Gamp5 new
 
Medical Device Regulatory Approval
Medical Device Regulatory ApprovalMedical Device Regulatory Approval
Medical Device Regulatory Approval
 
An Overview for Software as a Medical Device (SaMD)
An Overview for Software as a Medical Device (SaMD)An Overview for Software as a Medical Device (SaMD)
An Overview for Software as a Medical Device (SaMD)
 
ISO 13485.pptx
ISO 13485.pptxISO 13485.pptx
ISO 13485.pptx
 
Good design practice for medical devices
Good design practice for medical devicesGood design practice for medical devices
Good design practice for medical devices
 
Fda quality system regulation 21 CFR820_Medical devices_k_trautman
Fda quality system regulation 21 CFR820_Medical devices_k_trautmanFda quality system regulation 21 CFR820_Medical devices_k_trautman
Fda quality system regulation 21 CFR820_Medical devices_k_trautman
 
Presentation: Life cycle of medical devices
Presentation: Life cycle of medical devicesPresentation: Life cycle of medical devices
Presentation: Life cycle of medical devices
 
21 cfr part 210 and 211
21 cfr part 210 and 21121 cfr part 210 and 211
21 cfr part 210 and 211
 
GHTF study group 3
GHTF study group 3GHTF study group 3
GHTF study group 3
 
Risk Management for Medical Devices - ISO 14971 Overview
Risk Management for Medical Devices - ISO 14971 Overview Risk Management for Medical Devices - ISO 14971 Overview
Risk Management for Medical Devices - ISO 14971 Overview
 
IEC 62304: SDLC Conformance and Management
IEC 62304: SDLC Conformance and Management IEC 62304: SDLC Conformance and Management
IEC 62304: SDLC Conformance and Management
 
GHTF
GHTFGHTF
GHTF
 
Medical Devices Regulation (MDR) 2017/745 - Clinical investigations
Medical Devices Regulation (MDR)  2017/745 - Clinical investigationsMedical Devices Regulation (MDR)  2017/745 - Clinical investigations
Medical Devices Regulation (MDR) 2017/745 - Clinical investigations
 
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
 

Andere mochten auch

FDA Expectations for Traceability in Device & Diagnostic Design
FDA Expectations for Traceability in Device & Diagnostic DesignFDA Expectations for Traceability in Device & Diagnostic Design
FDA Expectations for Traceability in Device & Diagnostic DesignSeapine Software
 
Systems Engineering and Requirements Management in Medical Device Product Dev...
Systems Engineering and Requirements Management in Medical Device Product Dev...Systems Engineering and Requirements Management in Medical Device Product Dev...
Systems Engineering and Requirements Management in Medical Device Product Dev...UBMCanon
 
Medical device development lifecycle
Medical device development lifecycleMedical device development lifecycle
Medical device development lifecycleTim Blair
 
MEDICAL DEVICE DEVELOPMENT_0327
MEDICAL DEVICE DEVELOPMENT_0327MEDICAL DEVICE DEVELOPMENT_0327
MEDICAL DEVICE DEVELOPMENT_0327Robert Stathopulos
 
Common Mistakes in the Medical Device Development Continuum
Common Mistakes in the Medical Device Development ContinuumCommon Mistakes in the Medical Device Development Continuum
Common Mistakes in the Medical Device Development ContinuumNAMSA
 
Medical Product Development cycle
Medical Product Development cycleMedical Product Development cycle
Medical Product Development cyclemax hanafi
 
Life Sciences Product Development
Life Sciences Product DevelopmentLife Sciences Product Development
Life Sciences Product Developmentwraylk
 
Module 01 Introduction To FdA and Quality System Regulation
Module 01 Introduction To FdA and Quality System RegulationModule 01 Introduction To FdA and Quality System Regulation
Module 01 Introduction To FdA and Quality System RegulationQACATX
 
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporation
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow CorporationKey to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporation
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporationmarcus evans Network
 
How to Implement ISO 13485 Updates
How to Implement ISO 13485 UpdatesHow to Implement ISO 13485 Updates
How to Implement ISO 13485 UpdatesApril Bright
 
Fda gmp compliance for the Life Science Industry
Fda gmp compliance for the Life Science IndustryFda gmp compliance for the Life Science Industry
Fda gmp compliance for the Life Science Industrydmanalan
 
BioMedical Strategy Medical Devices Workshop Presentations
BioMedical Strategy Medical Devices Workshop PresentationsBioMedical Strategy Medical Devices Workshop Presentations
BioMedical Strategy Medical Devices Workshop PresentationsBioMedical Strategy (2004) Ltd.
 
Systems Engineering in Medical Devices
Systems Engineering in Medical DevicesSystems Engineering in Medical Devices
Systems Engineering in Medical Deviceschannylaux
 
Elements of a Quality System for Medical Devices
Elements of a Quality System for Medical DevicesElements of a Quality System for Medical Devices
Elements of a Quality System for Medical DevicesRaul Soto
 
Overview of FDA Regulation of Medical Devices
Overview of FDA Regulation of Medical DevicesOverview of FDA Regulation of Medical Devices
Overview of FDA Regulation of Medical DevicesMichael Swit
 
Batch release certificate for medical device
Batch release certificate for medical deviceBatch release certificate for medical device
Batch release certificate for medical deviceMalesh M
 
Software testing ppt
Software testing pptSoftware testing ppt
Software testing pptAjit Waje
 
Mapping Global Value Chains: International Input-Output Table
Mapping Global Value Chains: International Input-Output TableMapping Global Value Chains: International Input-Output Table
Mapping Global Value Chains: International Input-Output Tablecomexcr
 

Andere mochten auch (19)

FDA Expectations for Traceability in Device & Diagnostic Design
FDA Expectations for Traceability in Device & Diagnostic DesignFDA Expectations for Traceability in Device & Diagnostic Design
FDA Expectations for Traceability in Device & Diagnostic Design
 
Systems Engineering and Requirements Management in Medical Device Product Dev...
Systems Engineering and Requirements Management in Medical Device Product Dev...Systems Engineering and Requirements Management in Medical Device Product Dev...
Systems Engineering and Requirements Management in Medical Device Product Dev...
 
Medical device development lifecycle
Medical device development lifecycleMedical device development lifecycle
Medical device development lifecycle
 
MEDICAL DEVICE DEVELOPMENT_0327
MEDICAL DEVICE DEVELOPMENT_0327MEDICAL DEVICE DEVELOPMENT_0327
MEDICAL DEVICE DEVELOPMENT_0327
 
Common Mistakes in the Medical Device Development Continuum
Common Mistakes in the Medical Device Development ContinuumCommon Mistakes in the Medical Device Development Continuum
Common Mistakes in the Medical Device Development Continuum
 
Medical Product Development cycle
Medical Product Development cycleMedical Product Development cycle
Medical Product Development cycle
 
The Medical Device Milestone Map
The Medical Device Milestone MapThe Medical Device Milestone Map
The Medical Device Milestone Map
 
Life Sciences Product Development
Life Sciences Product DevelopmentLife Sciences Product Development
Life Sciences Product Development
 
Module 01 Introduction To FdA and Quality System Regulation
Module 01 Introduction To FdA and Quality System RegulationModule 01 Introduction To FdA and Quality System Regulation
Module 01 Introduction To FdA and Quality System Regulation
 
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporation
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow CorporationKey to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporation
Key to Successful Design to Manufacturing - Siddharth Desai, I-Flow Corporation
 
How to Implement ISO 13485 Updates
How to Implement ISO 13485 UpdatesHow to Implement ISO 13485 Updates
How to Implement ISO 13485 Updates
 
Fda gmp compliance for the Life Science Industry
Fda gmp compliance for the Life Science IndustryFda gmp compliance for the Life Science Industry
Fda gmp compliance for the Life Science Industry
 
BioMedical Strategy Medical Devices Workshop Presentations
BioMedical Strategy Medical Devices Workshop PresentationsBioMedical Strategy Medical Devices Workshop Presentations
BioMedical Strategy Medical Devices Workshop Presentations
 
Systems Engineering in Medical Devices
Systems Engineering in Medical DevicesSystems Engineering in Medical Devices
Systems Engineering in Medical Devices
 
Elements of a Quality System for Medical Devices
Elements of a Quality System for Medical DevicesElements of a Quality System for Medical Devices
Elements of a Quality System for Medical Devices
 
Overview of FDA Regulation of Medical Devices
Overview of FDA Regulation of Medical DevicesOverview of FDA Regulation of Medical Devices
Overview of FDA Regulation of Medical Devices
 
Batch release certificate for medical device
Batch release certificate for medical deviceBatch release certificate for medical device
Batch release certificate for medical device
 
Software testing ppt
Software testing pptSoftware testing ppt
Software testing ppt
 
Mapping Global Value Chains: International Input-Output Table
Mapping Global Value Chains: International Input-Output TableMapping Global Value Chains: International Input-Output Table
Mapping Global Value Chains: International Input-Output Table
 

Ähnlich wie FDA Design Controls

Design & Development.pptx
Design & Development.pptxDesign & Development.pptx
Design & Development.pptxPradeepa J
 
Interconnect Presentation
Interconnect PresentationInterconnect Presentation
Interconnect PresentationEric Deitrick
 
Computer Aided Design - Unit I Introduction
Computer Aided Design - Unit I IntroductionComputer Aided Design - Unit I Introduction
Computer Aided Design - Unit I IntroductionJayavendhanJ
 
Session 25 - Documentation
Session 25 - DocumentationSession 25 - Documentation
Session 25 - DocumentationVidyaIA
 
Introduction to software quality assurance by QuontraSolutions
Introduction to software quality assurance by QuontraSolutionsIntroduction to software quality assurance by QuontraSolutions
Introduction to software quality assurance by QuontraSolutionsQUONTRASOLUTIONS
 
21 CFR 820 and 801 pptx.pdf
21 CFR 820 and 801 pptx.pdf21 CFR 820 and 801 pptx.pdf
21 CFR 820 and 801 pptx.pdfBhavikaAPatel
 
Software Engineering (Software Quality Assurance)
Software Engineering (Software Quality Assurance)Software Engineering (Software Quality Assurance)
Software Engineering (Software Quality Assurance)ShudipPal
 
Risk based conceptual design delivery
Risk based conceptual design deliveryRisk based conceptual design delivery
Risk based conceptual design deliveryNNE
 
vu-re-lecture-06 requirement engineer.ppt
vu-re-lecture-06 requirement engineer.pptvu-re-lecture-06 requirement engineer.ppt
vu-re-lecture-06 requirement engineer.pptubaidullah75790
 
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...Intland Software GmbH
 
Facility Qualification & Consideration of Validation Aspects
Facility Qualification & Consideration of Validation Aspects Facility Qualification & Consideration of Validation Aspects
Facility Qualification & Consideration of Validation Aspects Apoorva Bauskar
 
FMEA.pdf
FMEA.pdfFMEA.pdf
FMEA.pdfinfoqtc
 

Ähnlich wie FDA Design Controls (20)

Design & Development.pptx
Design & Development.pptxDesign & Development.pptx
Design & Development.pptx
 
Presentation1
Presentation1Presentation1
Presentation1
 
Interconnect Presentation
Interconnect PresentationInterconnect Presentation
Interconnect Presentation
 
InHand's Project Developement Process
InHand's Project Developement Process InHand's Project Developement Process
InHand's Project Developement Process
 
Documentation
DocumentationDocumentation
Documentation
 
Computer Aided Design - Unit I Introduction
Computer Aided Design - Unit I IntroductionComputer Aided Design - Unit I Introduction
Computer Aided Design - Unit I Introduction
 
Session 25 - Documentation
Session 25 - DocumentationSession 25 - Documentation
Session 25 - Documentation
 
Introduction to software quality assurance by QuontraSolutions
Introduction to software quality assurance by QuontraSolutionsIntroduction to software quality assurance by QuontraSolutions
Introduction to software quality assurance by QuontraSolutions
 
CAD _ UNIT 1.pptx
CAD _ UNIT 1.pptxCAD _ UNIT 1.pptx
CAD _ UNIT 1.pptx
 
21 CFR 820 and 801 pptx.pdf
21 CFR 820 and 801 pptx.pdf21 CFR 820 and 801 pptx.pdf
21 CFR 820 and 801 pptx.pdf
 
Software Engineering (Software Quality Assurance)
Software Engineering (Software Quality Assurance)Software Engineering (Software Quality Assurance)
Software Engineering (Software Quality Assurance)
 
Maintainability engineering
Maintainability engineeringMaintainability engineering
Maintainability engineering
 
Apqp la
Apqp laApqp la
Apqp la
 
Risk based conceptual design delivery
Risk based conceptual design deliveryRisk based conceptual design delivery
Risk based conceptual design delivery
 
vu-re-lecture-06 requirement engineer.ppt
vu-re-lecture-06 requirement engineer.pptvu-re-lecture-06 requirement engineer.ppt
vu-re-lecture-06 requirement engineer.ppt
 
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...
27 Nov 2019 – Experts Talk: Integrated MedTech Delivery from Requirements thr...
 
Facility Qualification & Consideration of Validation Aspects
Facility Qualification & Consideration of Validation Aspects Facility Qualification & Consideration of Validation Aspects
Facility Qualification & Consideration of Validation Aspects
 
Process Safety Management.pptx
Process Safety Management.pptxProcess Safety Management.pptx
Process Safety Management.pptx
 
Validation
ValidationValidation
Validation
 
FMEA.pdf
FMEA.pdfFMEA.pdf
FMEA.pdf
 

Kürzlich hochgeladen

VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipurparulsinha
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknownarwatsonia7
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 

Kürzlich hochgeladen (20)

VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service JaipurHigh Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
High Profile Call Girls Jaipur Vani 8445551418 Independent Escort Service Jaipur
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 

FDA Design Controls

  • 1. 1 Design Controls FDA Small Business Regulatory Education for Industry (REdI) Bethesda, Maryland September 18, 2014 Stanley Liu Consumer Safety Officer Office of Communication and Education Center for Devices and Radiological Health U.S. Food and Drug Administration
  • 2. 2 Overview I. Introduction to Design Controls II. Requirements for Design Controls 1. Design planning 2. Design input and output 3. Design review 4. Design verification and validation 5. Design changes 6. Design transfer 7. Design history file III. Resources and FDA contacts
  • 3. I. Introduction to Design Controls 3
  • 4. 4 Design Controls – What are they? • A set of quality practices and procedures incorporated into the design and development process • Control the design process to assure that device specifications meet: − User needs − Intended use
  • 5. 5 Design Controls – Why? • 44% of voluntary recalls from October 1983 through September 1989 - may have been prevented by adequate design controls. Source: ‘‘Device Recalls: A Study of Quality Problems’’ (see 55 FR 21108, May 22, 1990) and “QS Regulations (Final Rule)” (see 61 FR 52602, October 07, 1996) • 90% of all software related device failures from 1983 through 1989 were due to design related error. Source: “FDA Medical Device Regulation from Premarket Review to Recall” (FDA/HHS OEI 09-90-0040, February 1991) • Most frequent causes for recalls related to Design, Software, and Non-Conforming Materials/Components. Source: “Medical Recall Report FY2003 to FY2012” (CDRH Office of Compliance, Division of Analysis and Program Operations)
  • 6. 6 Design Controls- Regulatory History • Safe Medical Device Act of 1990 authorized FDA to add Design Controls to the current Good Manufacturing Practice (cGMP) requirements for devices. • The QS Regulation became effective on June 1, 1997, replacing the 1978 GMP for medical devices. • Preamble to the QS regulation: extremely important for understanding Design Controls.
  • 7. Quality Systems - Key Definitions • Design History File (DHF) - compilation of records which describes the design history of a finished device. • Device Master Record (DMR) - compilation of records containing the procedures and specifications for a finished device. • Device History Record (DHR) - compilation of records containing the production history of a finished device. 7
  • 8. 8 Design Controls – Scope • Design controls apply to: – All Class II and Class III devices – The following Class I devices: 1. Devices automated with computer software 2. Tracheobronchial suction catheters 3. Surgeon's gloves 4. Protective restraints 5. Manual radionuclide applicator system 6. Radionuclide teletherapy source
  • 9. 9 Design Controls • Design control is a Key Subsystem of Quality System • Quality System (QS) Regulations: 21CFR 820 – Management controls – Design controls – Production and process controls – Corrective and preventive actions – Documents, Records, & Change Control – Material Control – Facility & Equipment Control
  • 10. 10 Design Controls • Incorporates Quality System of Risk Management and Analysis: – Systematic application of management policies, procedures, practices, insight/judgment, and experience for identifying, analyzing, controlling, and monitoring risk. – Risk Management and Analysis are integrated into the Design Control process.
  • 11. 11 7 Subsystems of the Quality System Design Control Material Control Documents, Records & Change Control Facility & Equipment Control Production & Process Control Corrective & Preventive Action Control Management Control
  • 12. 12 Design Controls (21 CFR 820.30) a) General requirements b) Design and development planning c) Design input d) Design output e) Design review f) Design verification g) Design validation h) Design changes i) Design transfer j) Design history file
  • 13. II. Requirements for Design Controls 13
  • 15. 15 Design Controls – General • Establish procedures to control device design: – Define – Document – Implement • Maintain procedures to control device design: – Review – Approve – Update
  • 16. 16 Design and Development Planning 21CFR 820.30(b)
  • 17. 17 Design & Development Planning • Establish, maintain and document Planning: – Describe or reference design and development activities. – Identify, describe, and define interfaces, responsibilities, and activities impacting device design. – Review, document, approve, and update as developments and changes evolve.
  • 19. 19 Definition • Design inputs mean the physical and performance characteristics of a device that are used as a bases for device design.
  • 20. 20 Design Input • Establish and maintain procedures for design input: – Ensure requirements are appropriate by addressing user needs and intended use in terms that are measurable – Address incomplete, ambiguous, or conflicting requirements – Document, review, and approve input requirements
  • 21. 21 Sources of Design Input • Standards • Focus groups • Customers • MDRs • CAPA • Service reports • Complaints • Marketing surveys • Sales feedback • Competitors’ products Including input from earlier/previous versions of a device
  • 22. 22 Examples of Design Input • Device functions • Physical characteristics • Performance • Safety • Reliability • Standards • Regulatory requirements • Human factors • Labeling & packaging • Maintenance • Sterilization • Compatibility with other devices • Environmental limits
  • 23. 23 Example Infusion Pump Issue /Risk: • Infusion pump stops operating when high- voltage medical devices are operating in the vicinity.
  • 24. 24 Example – Questions to Consider • What are the user needs? • In what environment is the infusion pump supposed to function? • What other kinds of equipment will it be used with? • What kinds of design input are needed to avoid this problem?
  • 25. 25 Example - Infusion Pump User Need Pump must be capable of functioning in an electromagnetic environment (e.g., hospital). Design Input Pump must function uninterrupted when used with other products that generate an electromagnetic field.
  • 27. 27 Definition • Design outputs mean the results of a design effort at each design phase and at the end of the total design effort.
  • 28. 28 Design Output • Establish and maintain procedures for design output: – Define and document design output in terms that allow an adequate evaluation of conformance to design input. – Reference acceptance criteria (i.e., criteria that are definable, and measurable). – Identify design outputs essential for the proper functioning of the device. – Review and approve design output before release.
  • 29. 29 In other words... • Design outputs are the design specifications which must meet design input requirements, as confirmed during design Verification and Validation and ensured during design Review (more later). Input Output
  • 30. 30 Design Output • Design Outputs are included in premarket submissions as Device Specifications. • The finished design output is the basis for the Device Master Record (DMR). • The total finished design output consists of the device, its packaging, labeling, and the Device Master Record (DMR).
  • 31. 31 User Need Pump must be capable of functioning in an electromagnetic environment (e.g., hospital). Design Input Pump must function uninterrupted when used with other products that generate an electromagnetic field. Design Output •Circuit board with filtering or buffering circuit •Pump EMI shield •Software with signal filtering and error handling/correcting code Example - Infusion Pump
  • 33. 33 Design Review • Design Review means a documented, comprehensive, systematic examination to: – Evaluate adequacy of the design requirements. – Evaluate capability of the design to meet requirements. – Identify any problems.
  • 34. 34 Design Review • Establish and maintain procedures. • Plan and conduct formal documented Design Reviews of design results at appropriate stages. • Include at each design review: – Representatives of all functions concerned and specialists as needed. – Individual(s) without direct responsibility for the stage being reviewed.
  • 35. 35 Design Review • Document results of design review in Design History File (DHF), including: – Identification of design – Date – Individuals performing review
  • 37. 37 Definition • Verification means confirmation by examination and provision of objective evidence that output meets input requirements. Input = Output
  • 38. 38 Design Verification • Establish and maintain procedures for Design Verification: – Confirm through measurable means (e.g., test reports, etc.). – Review, approve and document in Design History File (DHF).
  • 39. 39 Design Verification • Many test reports associated with Design Verification are included in premarket submissions: – 510(k)s – de novos – Premarket Approval Applications (PMAs) – Investigational Device Exemptions (IDEs)
  • 40. 40 Example - Infusion Pump User Need Pump must function in an operating room environment. Design Input Pump must function uninterrupted when used with other products that generate an electromagnetic field. Design Output •PCB with filtering circuit •Pump EMI shield •Software signal filtering code and error handling code Design Verification •Simulated EMI testing on hardware and software •Dimensional verification of shield •Verification of system error handling due to EMI Design Review
  • 42. 42 Definition • Design Validation means establishing by objective evidence that specifications(specified requirements) conform with user needs and intended use(s). Specifications= user needs intended use(s)
  • 43. 43 Design Validation • Establish and maintain procedures for Design Validation: – Under defined operating conditions. – On initial production units, lots, or batches (or their equivalents). – Under actual or simulated use conditions.
  • 44. 44 Design Validation • Perform software validation and risk analysis, where appropriate. • Review, approve, and document in Design History File.
  • 45. 45 Verification vs. Validation • Design Verification – Output meets Input – “Did I make the product correctly?” • Design Validation – Specifications meets user needs and intended use(s) – “Did I make the correct product?”
  • 46. 46 Design Validation • The results of Design Validation are typically submitted in premarket submissions. • Examples: – Animal study protocols/reports – Cadaver study protocols/reports – Clinical study protocols/reports
  • 47. 47 User Need Pump must function in an operating room environment. Design Input Pump must function uninterrupted when used with other products that generate an electromagnetic field. Design Output •PCB with filtering circuit •Pump EMI shield •Software signal filtering code and error handling code Design Verification •Simulated EMI testing on hardware and software •Dimensional verification of shield •Verification of system error handling due to EMI Design Validation •EMC testing to industry standards •Simulated EMI testing in high EMI environment •Risk Analysis concerning EMI •Software validation for filtering code Design Review
  • 48. 48 Warning Letter Example Failure to establish and maintain adequate procedures for Validating the device design and Risk Analysis, where appropriate, as required by 21 CFR 820.30(g) For example: The Design History Files for XXX did not include documentation that the device had ever been validated before production and marketing. When requested, the firm was unable to provide documentation that validation and risk analysis had been performed.
  • 50. 50 Design Transfer • Establish and maintain procedures to ensure correct Design Transfer into production specifications. • Is the Design accurately transferred to Production? • Although Design Transfer happens throughout, there frequently is a final stage of development intended to ensure all outputs are adequately transferred.
  • 52. 52 Design Changes • Establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of Design Changes before their implementation. • Is there a system in place to enact future changes? • Often overlooked, but of critically importance. • How can you improve your product if you have no system for change?
  • 53. 53 Design Changes • Depending on the scope and impact of the change, the change may require: – A new 510(k) – A new PMA, a PMA supplement, or a PMA 30-Day Notice – A new IDE or an IDE supplement • Changes must be communicated to FDA if the device is under premarket review or IDE review.
  • 55. 55 Definition • Design History File (DHF) means a compilation of records which describes the design history of a finished device. • It is a summation record of all Design actions, from start to transfer, including changes.
  • 56. 56 Design History File • Establish and maintain a design history file for each type of device • Include in the DHF, or reference records information necessary to demonstrate that the design was developed in accordance with the Design Plan and Quality Systems requirements
  • 57. Design Controls - Summary • Like the Quality Systems regulations themselves, Design Control should be viewed and understood as a system – a subsystem within the main Quality System. 57
  • 58. Design Controls - Summary • Processes –A set of quality practices and procedures incorporated into the design and development process • Goal - Control the design process to assure that device specifications meet: – User needs – Intended use 58
  • 59. 59 III. Resources and FDA Contacts • CDRH Learn Training videos focusing on overview, premarket, registration/listing, and post-market topics http://www.fda.gov/Training/CDRHLearn/default.htm • Device Advice Website accompaniment to CDRH Learn http://www.fda.gov/MedicalDevices/DeviceRegulatio nandGuidance/default.htm
  • 60. 60 QS Regulation and Guidance (General) • Quality System Regulation and Preamble http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/PostmarketRequi rements/QualitySystemsRegulations/ucm230127.htm • Design Control Guidance For Medical Device Manufacturers http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocum ents/ucm070627.htm • Do it By Design – An Introduction to Human Factors in Medical Devices http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guid anceDocuments/UCM095061.pdf • Guidance for Industry and FDA Premarket and Design Control Reviewers – Medical Device Use – Safety: Incorporating Human Factors Engineering into Risk Management http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guida nceDocuments/UCM094461.pdf • (DRAFT) Applying Human Factors and Usability Engineering to Optimize Medical Device Design http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocumen ts/ucm259748.htm
  • 61. 61 QS Regulation and Guidance • Guidance for Industry: Part 11, Electronic Records; Electronic Signatures - Scope and Application http://www.fda.gov/RegulatoryInformation/Guidances/ucm125067.htm • General Principles of Software Validation http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu ments/ucm085281.htm • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu ments/ucm089543.htm • Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu ments/ucm073778.htm • Guidance: Mobile Medical Applications http://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/guida ncedocuments/ucm263366.pdf
  • 62. 62 Need More Information? Division of Industry and Consumer Education (DICE) http://www.fda.gov/MedicalDevices/DeviceRegulationan dGuidance/ContactUs-- DivisionofIndustryandConsumerEducation/default.htm • Email: DICE@fda.hhs.gov • Phone: 301-796-7100 or 800-638-2041