2. Login
• Windows Login
• User - Administrator
• Password – BDIS or bdis
• DiVa Login
• Username - 3 letters first name + 3 letters last name
• Password – as user name
YOUR DISCIPLINE HERE
20. Worksheets
• Global worksheet
• Is not linked to specific data source
• Reloads data on tube selection
• Used when batching
• Different global worksheets can be linked to different
specimens
• Sheet
• Is linked to data
• Does not refresh upon tube selection
• Compensation
• Is linked to compensation tubes
• Must be used to perform acquisition compensation
43. Acquiring data
• Activate tube to collect data
• Hit acquire
• Check stopping gate and numbers
• Hit record
• Hit remove tube
• Remove tube and flick aspirator arm to left
• repeate
47. Cleaning Instrument
1. Run 10 % fresh bleach soln. – 5 mins*
2. Run 4 % extran – 5 mins
3. Run distilled water – 5 mins
4. Shutdown if no other users in the day
* For Biohazardous samples