1. PHOTOSTABILITY
TESTING
PRESENTED BY : HARSHAVARDHAN VIJAY
KONDHARE
1ST YEAR M.PHARM (DRUG REGULATORY AFFAIRS)
GUIDE : DR. R.N. PUROHIT
H.O.D DRUG REGULATORY AFFAIRS
POONA COLLEGE OF PHARMACY
29-12-2015
2. IMPORTANT TERMS
Stability : time period in which the drug product retains the same properties
and characteristics that it possessed at the time of its manufacturing.
Stress testing : is a form of deliberately intense or thorough testing used to
determine the stability of a given drug substance or product.
Photosensitivity : amount to which a drug substance or product reacts
upon receiving photons, especially visible light.
Photostability : stability of a drug substance or product on exposure to light.
2
3. IMPORTANT TERMS
Immediate (primary) pack is that constituent of the packaging that is in
direct contact with the drug substance or drug product, and includes any
appropriate label.
Marketing pack is the combination of immediate pack and other secondary
packaging such as a carton.
Forced degradation testing studies are those undertaken to degrade the
sample deliberately.
3
4. IMPORTANT TERMS
Confirmatory studies are those undertaken to establish photostability
characteristics under standardized conditions.
Parent Guideline : Q1A(R2) Stability Testing of New Drug Substances and
Products.
4
5. INDEX
Why to perform photostability studies ?
When do we do photostability testing?
Factors that influence photostability of drug products
Some typical examples of photostability studies of drugs
Pharmaceutical product sensitivity classification
Photostability testing of new drug substances and products as per ICH
Q1B guideline
Light sources
General procedure
Drug substance: Presentation of samples, Analysis of sample, Judgement of
results
Drug product: Presentation of samples, Analysis of sample, Judgement of
results
Challenges of photostability testing
Solutions to photostability testing challenges
5
6. Why to perform photostability
studies ?
• Instability - may lead to under medication due to lowering of active
drug concentration in dosage form.
• Drug decomposition - leads to formation of toxic products.
• Instability leads to changes in physical appearance.
Quality perspective: To avoid
• Regulatory requirement and scientific necessity during drug
development.
• Mandatory to perform before filing in registration dossier.
Regulatory perspective:
6
7. When do we do photostability
testing?
FDA guidance
states
In phase III of
regulatory
submission process.
The results should
be summarized and
submitted in an
annual report.
Starting early in
preclinical phase or
phase I of clinical
trials is highly
encouraged
7
8. Factors that influence photostability
of drug products
Particle size
Drug content
Tablet geometry
Preparation method
Concentration
pH and Ionization
Ionic strength
Oxidation
8
9. Some typical examples of
photostability studies of drugs
Carbamazepine : polymorphs in
solid dosage form (tablets),
surface discoloured to yellow and
then orange with results
indicating polymorph II to be the
least stable.
Cyanocobalamin : photolysis in
the presence of visible light at
various pH, confirmed that
protonated form was more
susceptible to photolysis.
Furosemide : forms
sulphamoylanthranilic acid after
exposure to sunlight.
9
12. The guideline is
for photostability
information for
Registration for
new molecular
entities.
The guideline
does not cover
the photostability
of drugs after
administration
(i.e. under
conditions of
use).
Alternative
approaches may
be used if they
are scientifically
sound and
justification is
provided.
12
13. The intrinsic
photostability
characteristics of
new drug
substances and
products should be
evaluated to
demonstrate light
exposure does not
result in
unacceptable
change.
Normally, carried
out on a single
batch of material.
Studies should be
repeated if certain
variations and
changes are made
to the product (e.g.,
formulation,
packaging).
13
14. LIGHT SOURCES
The applicant should maintain :
• an appropriate control of temperature to minimize the effect of
localized temperature changes or
• include a dark control in the same environment unless
otherwise justified.
There are two options-
• Option 1
• Option 2
14
15. LIGHT SOURCES
Option 1
Any light source that is designed to
produce an output similar to the D65/ID65*
emission standard such as
an artificial daylight
fluorescent lamp
combining visible
and ultraviolet (UV)
outputs
xenon
metal halide
lamp.
*D65 is the internationally recognized standard for outdoor daylight as
defined in ISO 10977 (1993). ID65 is the equivalent indoor indirect daylight
standard.
15
16. LIGHT SOURCES
Option 2
Similar sample should be exposed
to both the cool white fluorescent
and near ultraviolet lamp.
A cool white fluorescent lamp
designed to produce an output
similar to that specified in ISO
10977(1993) ; and
A near UV fluorescent lamp
having a spectral distribution from
320 nm to 400 nm*.
*a maximum energy emission between 350 nm and 370 nm
16
17. LIGHT SOURCES: ISO 10977
Describes test equipment, test procedure and analytical
methods for
• predicting the long-term dark storage stability of colour
photographic images (based on the Arrhenius method) and
• measuring the colour stability of such products when
subjected to certain illuminants at specified temperatures
and humidities.
Does not specify limits of acceptability for the
stability.
17
18. GENERAL PROCEDURE
For confirmatory studies:
Samples should be exposed to light providing
• illumination of not less than 1.2 million lux hours and
• an integrated near ultraviolet energy of not less than 200 watt
hours/square meter .
Samples may be exposed side-by-side with a validated chemical
actinometric system.
18
19. A systematic
approach to
photostability
testing
Tests on the
drug
substance
Tests on the
exposed drug
product
outside of the
immediate
pack
Tests on the
drug product in
the immediate
pack
Tests on the
drug product
in the
marketing
pack
19
20. DRUG SUBSTANCE
Consist of two
parts:
Forced degradation
testing
Confirmatory
testing.
Forced degradation testing studies: is to evaluate the
overall photosensitivity of the material for method
development purposes and/or degradation pathway
elucidation.
• samples should be in chemically inert and transparent containers.
• variety of exposure conditions may be used
Contd.>
20
21. DRUG SUBSTANCE
Under forcing conditions, decomposition products may be
observed that are unlikely to be formed under the conditions used
for confirmatory studies.
Confirmatory studies: provide the information necessary for
handling, packaging, and labeling.
• If the drug is clearly photostable or photolabile the
photostability characteristics should be confirmed on a single
batch.
• If the results of the confirmatory study are equivocal, testing of
up to two additional batches should be conducted.
21
22. DRUG SUBSTANCE:
PRESENTATION OF SAMPLES
Ensure that the effects of the changes in physical states such
as sublimation, evaporation or melting are minimized.
To provide minimal interference with the exposure of samples
under test.
Possible interactions between the samples and any material
used for containers or for general protection of the sample,
should also be considered and eliminated.
• Solid drug substances spread across the container to give a
thickness of not more than 3 millimeters.
• Liquids should be exposed in chemically inert and transparent
containers.
22
23. DRUG SUBSTANCE: ANALYSIS OF
SAMPLE
the samples should be examined for;
any changes in physical properties (e.g., appearance,
clarity, or color of solution)
for assay and;
degradants by a method suitably validated for products
likely to arise from photochemical degradation
processes.
Analysis of the exposed sample should be performed
concomitantly with protected samples used as dark
23
24. DRUG SUBSTANCE: JUDGEMENT
OF RESULTS
Forced degradation studies:
• It is important to recognize that they form part of the stress
testing and
• are not therefore designed to establish qualitative or
quantitative limits for change.
Confirmatory studies: should identify precautionary
measures needed
• In manufacturing or in formulation of the drug product, and
• If light resistant packaging is needed.
24
25. DRUG PRODUCT
fully exposed
product
in immediate
pack
in marketing
pack.
25
Studies should be carried out in a sequential manner viz.
Testing should progress until the results demonstrate that the drug product
is adequately protected from exposure to light.
The drug product should be exposed to the light conditions described
under the General Procedure.
• If the drug is clearly photostable or photolabile the photostability is
confirmed on a single batch.
• If the results of the confirmatory study are equivocal, testing of up to two
additional batches should be conducted.
27. DRUG PRODUCT: PRESENTATION
OF SAMPLES
The samples should be positioned to provide maximum area of
exposure to the light source. For example, tablets, capsules, etc.,
should be spread in a single layer.
If direct exposure is not practical (e.g., due to oxidation of a
product), the sample should be placed in a suitable protective
inert transparent container (e.g., quartz).
27
28. DRUG PRODUCT: ANALYSIS OF
SAMPLES
the samples should be examined for;
any changes in physical properties (e.g., appearance, clarity, or color of
solution)
for assay and;
degradants by a method suitably validated for products likely to arise from
photochemical degradation processes.
For solid oral dosage form products, testing should be
conducted on an appropriately sized composite , for example,
20 tablets or capsules.
Homogenization or solubilization of the entire sample - creams,
ointments, suspensions, etc.
Analysis of the exposed sample should be performed concomitantly with
protected samples used as dark controls.
28
29. DRUG PRODUCT: JUDGEMENT
OF RESULTS
Depending on the extent of change special
labeling or packaging may be needed.
When evaluating the results of photostability
studies it is important to consider the results
obtained from other formal stability studies.
29
30. DRUG PRODUCT: JUDGEMENT OF
RESULTS
Example showing how confirmatory photostability results
can be used in conjunction with definitive stability results
for the judgement of shelf-life of a drug substance or
product.
30
32. 1. Troublesome Lamp Selection
Option I light sources (fluorescent D65, metal halide or xenon)
cause:
Over exposure: it produces both UVA and visual irradiance in
fixed ration, but the exposure requirement for photopic and UVA
are different, one of the dose gets overexposed.
Production of excessive heat
dark controls are needed to segregate photochemical degradation from
thermal degradation.
Large internal cooling fans are necessary to dissipate this heat and can
pose presentation problems by blowing samples around.
Sample colour changes due to high temperatures cannot easily be
32
33. 1. Troublesome Lamp Selection
Xenon and metal halide lamps have a short life span and need
replaced every 750 to 1500 hours.
Require light filters to eliminate radiation below 320nm.
They also have a relatively small illumination area.
33
34. 2. Irradiance Measurement Difficulty
Chemical actinometers can be used to measure sample dose
ICH describes the use of quinine hydrochloride dehydrate as an
example of a chemical actinometer.
Quinine has a ‘dark reaction’ where the reaction continues after it is
used. Not only is quinine wavelength dependent, it is affected by
temperature and pH variations. Due to these characteristics, quinine
has been shown to be inaccurate with lamps that produce significant
amounts of heat, such as xenon lamps.
Irradiance measurements with instrumental radiometers have
high margins of uncertainty; 10% is not uncommon.
Radiometers need to be calibrated or certified before use.
34
35. 3. Test Completion Control Issues
Chemical actinometers do not provide a mechanism to
automatically turn the lamps off or alert the operator when the
desired exposure level is reached.
Chemical actinometers cannot record irradiance levels
throughout the test.
As lamps age, their intensity decreases → causes irradiance
levels of full-power light sources to fluctuate over time→ light
intensity would terminate prematurely compared to the desired
dose→ This is particularly troublesome for confirmatory studies.
35
36. 4. Distorting the Spectral Power Distribution
Interior chamber materials that reflect light onto samples should
reflect/ absorb radiation uniformly across the UVA and photopic
spectrums.
If not, samples will be subjected to light having a spectral power
distribution different than that specified by ICH.
36
37. 5. Humidity Control Factor
State of hydration affects the photostability of some samples.
This means identical drug substances subjected to
identical irradiance and temperature conditions can have
very different results if exposed to different humidity
levels.
37
39. 1. Prudent Lamp Selection
Option II of ICH guidelines (cool white and near- UV fluorescent lighting)
• Independent control of illuminance and UVA irradiance eliminates overexposure.
• Fluorescent lamps generate minimal heat.
• Eliminate the need for expensive light filters and dark controls.
• Small internal fans can be employed to subtly maintain proper air temperature
without disturbing sample presentation.
• Typically last over ten thousand hours, have low replacement costs, and provide a
large illumination area.
2. Accurate Light Measurements
• Achieved with a built-in radiometer.
• Detectors utilizing a Teflon hemisphere may result in an exceptionally good cosine
response.
• Detectors should be both cosine corrected and calibrated.
• It is best if radiometer displayed units for illuminance and UVA irradiance are
consistent with ICH documentation.
3. Precise Test Completion Controls
• An integrating radiometer combined with chamber controls should be used to
ensure precise dose levels at test completion.
• Advanced systems are capable of running based on exposure level or timed tests,
the radiometer should show irradiance, test time remaining and accumulated dose
levels.
39
40. 4. Preserving the Spectral Power Distribution
• Using specular aluminium on interior reflective surfaces, it reflects light
across both UVA and photopic spectrums
• Available with a 95% total reflection.
• Outshines mirrored stainless steel and white painted surfaces for not only
illuminance reflection but also UVA irradiance.
5. Tight Humidity Control
• Use ultrasonic nebulizers, they vaporize water droplets as small as a 3 micron
mean diameter, this enhances uniform humidity distribution throughout the
chamber without injecting additional unwanted heat.
• Dehumidification is often accomplished through mechanical refrigeration.
6. Value Added Features
Chamber features are available that enhance the end-user experience:
• keep track of accumulated lamp hours
• alert the operator when to replace the lamps
Validation is simple with pre-written IQ/OQ/PQ validation protocols.
Chart recorders assist in demonstrating regulatory compliance by permanently
recording illuminance, UVA irradiance, temperature and humidity testing
conditions.
Sliding shelves, access ports and security lockouts are other features that can40
41. CONCLUSION
Product should be shown to be light stable by photostability
studies.
Showing the dosage form or dosage form in packaging is
photostable.
By light transmission studies showing adequate protection of
the packaging.
For blisters, which are not light protective, store blister in carton.
If product is not light stable; label should state that protect form
light and container should be light protective; light transmission
41
42. CONCLUSION
Photo stability studies are performed to generate primary
degradants of drug substance by exposure to UV or fluorescent
conditions.
Important to help develop determine the degradation pathways
and degradation products of the active ingredients.
The various factors affecting the photostability of dosage form
should be clearly indicated so as to prevent their interference with
the testing and the stability of dosage form.
42
44. REFERENCES
1. ICH Harmonised tripartite guideline, Stability testing: Photostability Testing of New
Drug Substances and Products Q1B Current Step 4 version dated 6 November
1996
2. Jitendra Kumar, Review: Development of forced degradation studies of drugs,
International Journal of Advances in Pharmaceutics 2 (3) 2013; 35,37-38.
3. Abhijjeet Welankiwar, Photostability Testing of Pharmaceutical Products,
International Research Journal of Pharmacy 2013; 11-15.
4. Steven W Baertschi, Commentary: A critical assessment of the ICH guideline on
photostability testing of new drug substances and products (Q1B), Journal of
Pharmaceutical Sciences, vol. 99, No. 7, July 2010; 2934-2939.
5. Robert A. Reed, Implications of Photostability on the Manufacturing, Packaging,
Storage, and Testing of Formulated Pharmaceutical Products; Pharmaceutical
technology March 2005; 70-71.
6. Bob Dotterer, Using photostability chambers to meet the requirements of drug
testing in accordance with ICH, Q1B; CARON; 1-5.
7. PowerPoint Presentation: Gabriel K. Kaddu, Stability Principles and Case Studies:
Active Pharmaceutical Ingredient (API) and Finished Pharmaceutical Product
(FPP); WHO Prequalification of Medicines Programme; Assessment training,
44